자원봉사자를 대상으로 한 양압기(PAP) 시스템 평가
자원자를 대상으로 한 새로운 양압 기도 시스템 평가
This study will test the control and measurements of a new system for airway support in healthy volunteers. The goals of the study are to evaluate how accurately the system can stabilize airway pressure to a desired level, evaluate the accuracy of the system's reported volume of each breath (tidal volume), evaluate the accuracy of the system's reported respiratory rate, evaluate the accuracy of exhaled carbon dioxide monitoring with a standard carbon dioxide monitor when the novel PAP system is used, and evaluate general comfort of the volunteers while using the system.
연구 개요
상태
상태
정황
정황
개입 / 치료
개입 / 치료
상세 설명
Sedation procedures involve administering drugs to patients that reduce awareness and pain without complete loss of consciousness. A common side effect of these drugs is partial or complete obstruction of the airway by the airway tissue, inhibiting the flow of oxygen to the lungs. Research has shown that this risk can be effectively reduced by administering PAP therapy during sedation, which involves directing air flow into a tight-fitting mask over the patient's nostrils and/or mouth. However, non-invasive ventilators used for administering PAP therapy are expensive and large, and often not available in areas of the hospital where sedation procedures take place.
The investigators have developed a small and low-cost novel PAP system for use in sedation procedures that controls airway pressure to desired levels (a feature that is generally only available in larger and higher-cost devices), provides estimates for tidal volumes (generally only available with a non-invasive ventilator), provides respiratory rate estimates, and allows for more accurate monitoring of exhaled carbon dioxide with a standard carbon dioxide monitor using a novel approach that to the investigator's knowledge has not been published. The goal of investigator's study is to demonstrate the accuracy of these features in human subjects. The system is much smaller and less expensive than a non-invasive ventilator, provides enhanced respiratory monitoring, and has the potential to be a practical alternative that would make PAP therapy during sedation much more feasible.
The investigator's evaluation will primarily include evaluating the investigator's system's performance against an FDA cleared commercial respiratory monitor to verify accuracy. The investigators anticipate that the investigator's system will automatically adjust oxygen flow to maintain a desired airway pressure even in the presence of variables such as mask leak and patient inhalation flow. The investigators also anticipate that the investigator's system will provide tidal volume estimates within 15% of the respiratory monitor reference measurement, as well as respiratory rate estimates that closely match the respiratory monitor. The investigator's system employs a method that detects when the patient exhales and removes mask pressure during exhalation, and the investigators anticipate that this will result in improved comfort for the patient, as well as improved carbon dioxide monitoring accuracy, since removing flow during exhalation reduces dilution of the expired gas sample that is used for measuring exhaled carbon dioxide levels.
연구 유형
연구 유형
등록 (실제)
등록
단계
단계
- 해당 없음
연락처 및 위치
연구 장소
-
-
Utah
-
Salt Lake City, Utah, 미국, 84108
- University of Utah Pain Management Center
-
-
참여기준
자격 기준
자격 기준
공부할 수 있는 나이
건강한 자원 봉사자를 받아들입니다
연구 대상 성별
설명
Inclusion Criteria:
- Healthy volunteers
Exclusion Criteria:
- Have any respiratory or cardiovascular conditions
- i.e., cold, allergies, congestion, pneumonia, lung cancer, COPD, acute respiratory distress syndrome, etc.
공부 계획
연구는 어떻게 설계됩니까?
디자인 세부사항
- 주 목적: 장치 타당성
- 할당: 해당 없음
- 중재 모델: 단일 그룹 할당
- 마스킹: 없음(오픈 라벨)
팔의 수
무기와 개입
참가자 그룹 / 팔참가자 그룹 / 팔 |
개입 / 치료개입 / 치료 |
|---|---|
|
실험적: CPAP
Enrolled healthy volunteers to examine airway pressure support through the investigational device. |
Airway pressure support through the investigational device |
연구는 무엇을 측정합니까?
주요 결과 측정
주요 결과 측정
결과 측정 |
측정값 설명 |
기간 |
|---|---|---|
|
Tidal Volume Accuracy
기간: During the intervention, up to 20 minutes
|
The mean and standard deviation of the percent difference in tidal volume measurement between our PAP system and the respiratory monitor |
During the intervention, up to 20 minutes
|
2차 결과 측정
2차 결과 측정
결과 측정 |
측정값 설명 |
기간 |
|---|---|---|
|
Respiratory Rate
기간: During the intervention, up to 20 minutes
|
The average respiratory rate reported by the respiratory monitor and rate reported by the PAP system for each volunteer will be compared. The average difference for each volunteer will be calculated, and a 95% confidence interval will be constructed, with the ideal mean difference being zero with low variance. |
During the intervention, up to 20 minutes
|
|
EtCO2
기간: During the intervention, up to 20 minutes
|
End-tidal carbon dioxide measurements at 10 cmH2O of mask pressure will be compared between when the system turned off flow during exhalation and when delivering flow throughout the entire breath. We anticipate that end-tidal CO2 levels will read lower when flow is delivered continuously during exhalation because of dilution of the expired gas, and this will be demonstrated by calculating the average difference in end-tidal CO2 between these two approaches. |
During the intervention, up to 20 minutes
|
|
Mask Pressure Control Accuracy
기간: During the intervention, up to 20 minutes
|
The measured pressure in the mask compared to the set pressure in the 50 milliseconds immediately preceding inhalation will be assessed for evaluating pressure control. Average error will be calculated and a 95% confidence interval will be constructed. We anticipate the error to be less than 1 cmH2O. |
During the intervention, up to 20 minutes
|
공동 작업자 및 조사자
협력자
협력자
수사관
수사관
- 수석 연구원: Kai Kuck, PhD, University of Utah
간행물 및 유용한 링크
일반 간행물
- Burk KM, Sakata DJ, Kuck K, Orr JA. Comparing Nasal End-Tidal Carbon Dioxide Measurement Variation and Agreement While Delivering Pulsed and Continuous Flow Oxygen in Volunteers and Patients. Anesth Analg. 2020 Mar;130(3):715-724. doi: 10.1213/ANE.0000000000004004.
- Fogarty M, Orr JA, Sakata D, Brewer L, Johnson K, Fang JC, Kuck K. A comparison of ventilation with a non-invasive ventilator versus standard O2 with a nasal cannula for colonoscopy with moderate sedation using propofol. J Clin Monit Comput. 2020 Dec;34(6):1215-1221. doi: 10.1007/s10877-019-00426-5. Epub 2019 Nov 23.
연구 기록 날짜
연구 주요 날짜
연구 시작 (실제)
연구 시작
기본 완료 (실제)
기본 완료
연구 완료 (실제)
연구 완료
연구 등록 날짜
최초 제출
최초 제출
QC 기준을 충족하는 최초 제출
QC 기준을 충족하는 최초 제출
처음 게시됨 (실제)
처음 게시됨
연구 기록 업데이트
마지막 업데이트 게시됨 (실제)
마지막 업데이트 게시됨
QC 기준을 충족하는 마지막 업데이트 제출
QC 기준을 충족하는 마지막 업데이트 제출
마지막으로 확인됨
마지막으로 확인됨
추가 정보
이 연구와 관련된 용어
키워드
기타 연구 ID 번호
기타 연구 ID 번호
- IRB_00190165
개별 참가자 데이터(IPD) 계획
개별 참가자 데이터(IPD)를 공유할 계획입니까?
약물 및 장치 정보, 연구 문서
미국 FDA 규제 의약품 연구
미국 FDA 규제 기기 제품 연구
미국에서 제조되어 미국에서 수출되는 제품
이 정보는 변경 없이 clinicaltrials.gov 웹사이트에서 직접 가져온 것입니다. 귀하의 연구 세부 정보를 변경, 제거 또는 업데이트하도록 요청하는 경우 register@clinicaltrials.gov. 문의하십시오. 변경 사항이 clinicaltrials.gov에 구현되는 즉시 저희 웹사이트에도 자동으로 업데이트됩니다. .