- ICH GCP
- Register voor klinische proeven in de VS.
- Klinische proef NCT01955018
Comparison of the Ultra-low-dose Veo Algorithm With the Gold Standard Filtered Back Projection for Detecting Pulmonary Asbestos-related Conditions
Asbestos fibers were intensively used throughout the 20th century and remain prevalent in developing countries. However, asbestos exposure induces a variety of benign and malignant pleural and lung diseases. The most common asbestos-induced neoplasm is lung cancer. Moreover, thin-section computed tomography (CT) is more sensitive than a chest x-ray for detecting early asbestos-related conditions. Increased exposure to radiation underpins the consequences of cancer induction. However, reducing CT doses increases image noise from the filtered back projection (FBP) reconstruction. Strategies to reduce radiation exposure include the use of iterative reconstruction algorithms. A new algorithm called VeoTM (General Electric Healthcare, Milwaukee, MI, USA) decreases the image noise up to 70% compared with the gold standard FBP model. Moreover, Veo improves spatial resolution with excellent detection of low and high contrast objects from a CT Dose Index (CTDIvol) equal to 0.3 mGy.
The objective of the present study is to compare Veo with the gold standard FBP for detecting pulmonary asbestos-related conditions among workers previously exposed to asbestos. Comparisons included radiation delivered and image quality.
Studie Overzicht
Toestand
Conditie
Interventie / Behandeling
Gedetailleerde beschrijving
Asbestos-exposed workers will be recruited following referral to our radiology department for the evaluation of asbestos-related disease. CT examinations will be performed with a 64-slice CT system (Discovery CT 750HD; GE Healthcare, Milwaukee, WI) and will consist of two successive acquisitions. Each examination, the normal-dose (FBP acquisition) and ultra-low-dose (Veo acquisition) spiral CT, will be performed in supine position, on the entire thorax, at full inspiration and without contrast agent injection.
In order to perform the two acquisitions without increasing radiation, standard acquisition will be performed with the same kV with mA equal to patient's body weight minus 10. To conserve image quality, 60 mA will be the inferior limit. The other CT parameters will be rotation time 0.5 s and pitch 1.375. Image data will be reconstructed with FBP algorithm.
The Veo acquisition will be performed with constant CT parameters including: a tube voltage of 100 kV, a tube current of 20 mA, pitch of 0.984 and rotation time 0.4 s. Image data will be reconstructed with the Veo algorithm.
Each CT acquisition will be viewed independently by two experienced radiologists (2 to 7 years of experience). The low-dose images with Veo reconstruction will be interpreted before the standard CT and on separate weeks to minimize recall bias. Then, the more experienced radiologist will evaluate the detection and characterization of pleuroparenchymal abnormalities by a second and simultaneous reading of the Veo and FBP acquisitions. Because FBP images are benchmark practice, when a lesion will be found only on Veo images, it will be regarded as a false positive.
The following asbestos-related pleural and parenchymal abnormalities will be recorded as present or absent. Pleural abnormalities considered will be: pleural plaques, diffuse pleural thickening and pleural effusion.
CT features of asbestosis will include subpleural dots and branching opacities, curvilinear subpleural lines, areas of ground glass opacities, septal lines, reticulations and honeycombing.
Presence of nodules will also be recorded. We will note for each abnormality: localization (side, table position) and nature (non-solid, part-solid, solid or calcified).
Studietype
Inschrijving (Werkelijk)
Fase
- Fase 4
Contacten en locaties
Studie Locaties
-
-
-
Clermont-Ferrand, Frankrijk, 63003
- CHU Clermont-Ferrand
-
-
Deelname Criteria
Geschiktheidscriteria
Leeftijden die in aanmerking komen voor studie
Accepteert gezonde vrijwilligers
Geslachten die in aanmerking komen voor studie
Beschrijving
Inclusion Criteria:
- asbestos-exposed workers
- ability to give a written informed consent
Exclusion Criteria:
- - previous history of cancer
- previous history of thoracic surgery
- other interstitial pathology known
Studie plan
Hoe is de studie opgezet?
Ontwerpdetails
- Primair doel: Diagnostisch
- Toewijzing: Gerandomiseerd
- Interventioneel model: Opdracht voor een enkele groep
- Masker: Enkel
Wapens en interventies
Deelnemersgroep / Arm |
Interventie / Behandeling |
---|---|
Experimenteel: VeoTM
A new algorithm called VeoTM (General Electric Healthcare, Milwaukee, MI, USA) decreases the image noise up to 70% compared with the gold standard FBP model.
Moreover, Veo improves spatial resolution with excellent detection of low and high contrast objects from a CT Dose Index (CTDIvol) equal to 0.3 mGy
|
The objective of the present study is to compare Veo with the gold standard FBP for detecting pulmonary asbestos-related conditions among workers previously exposed to asbestos.
Comparisons included radiation delivered and image quality.
|
Ander: gold standard FBP model
The objective of the present study is to compare Veo with the gold standard FBP for detecting pulmonary asbestos-related conditions among workers previously exposed to asbestos.
Comparisons included radiation delivered and image quality
|
The objective of the present study is to compare Veo with the gold standard FBP for detecting pulmonary asbestos-related conditions among workers previously exposed to asbestos.
Comparisons included radiation delivered and image quality.
|
Wat meet het onderzoek?
Primaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
---|---|---|
Pleural and parenchymal abnormalities
Tijdsspanne: at day 1
|
The kappa coefficient will be used to measure agreement for categorical parameters and Pearson's correlation coefficient and Lin concordance correlation coefficient for quantitative data.
|
at day 1
|
Secundaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
---|---|---|
Radiation dose
Tijdsspanne: at day 1
|
The dose length product (DLP) will be recorded.
|
at day 1
|
Quality images assessment
Tijdsspanne: at day 1
|
Respiratory artifacts will be graded on a three-point scale (1 = negligible, 2 = moderate, 3 = salient). Images noise will be studied in the axial and coronal planes. A similar scale will be used for subjective image quality in the mediastinum and parenchyma windows. Objective image noise (Standard Deviation) and average CT numbers (in Hounsfield's units) will be measured with circular regions of interest (ROI) on different anatomical levels, 10 mm in diameter. The signal to noise ratio (SNR) will also be calculated using the equation SNR = CT numbers / noise. |
at day 1
|
Medewerkers en onderzoekers
Publicaties en nuttige links
Studie record data
Bestudeer belangrijke data
Studie start
Primaire voltooiing (Werkelijk)
Studie voltooiing (Werkelijk)
Studieregistratiedata
Eerst ingediend
Eerst ingediend dat voldeed aan de QC-criteria
Eerst geplaatst (Schatting)
Updates van studierecords
Laatste update geplaatst (Schatting)
Laatste update ingediend die voldeed aan QC-criteria
Laatst geverifieerd
Meer informatie
Termen gerelateerd aan deze studie
Trefwoorden
Andere studie-ID-nummers
- CHU-0166
- 2012-A00618-35
Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .