AI-Based Analysis of Port Catheter Tip Position
Determination of Port Catheter Tip Position Using Artificial Intelligence-Based Automated Image Analysis
Przegląd badań
Status
Status
Warunki
Warunki
Interwencja / Leczenie
Interwencja / Leczenie
Szczegółowy opis
Before prospective enrollment, a two-dimensional nnU-Net-based multitask model will be developed and internally tested using retrospective chest radiographs from the same center. The model parameters, ensemble, preprocessing steps, and technical output rules will be locked before prospective enrollment. The retrospective development and internal test datasets are not part of the enrollment reported for this record.
The prospective component will include one observational cohort of adults aged 18 years or older undergoing right internal jugular vein port implantation during routine clinical care. Implantation, chest radiography, computed tomography, and all clinical management decisions will remain under the responsibility of the treating clinical team and will not be assigned by the research protocol. The automated output will not be made available for procedural guidance or patient management.
For each evaluable participant, the locked automated method will generate an estimate based on chest radiographic image analysis and a patient-specific Delta adjustment. A prespecified manual A+B measurement method will be applied as a comparator in the same participants. Reference measurements will be derived from the first suitable post-implantation noncontrast chest computed tomography examination obtained for a routine clinical indication, provided that both the catheter tip and the cavoatrial junction are visible and no catheter manipulation occurred beforehand. No additional radiographic or computed tomography examination will be requested for the study.
The primary objective is to quantify the error of the automated estimate using mean absolute error in the prospective cohort. Prespecified secondary analyses will evaluate additional error, agreement, pragmatic accuracy, technical assessability, and repeatability measures. The ±10 mm criterion will be treated as a pragmatic accuracy threshold rather than a clinical safety boundary. The study is a same-center prospective performance evaluation; it is not an external validation and is not designed to assess clinical benefit or to establish that the system is ready for clinical use.
Typ studiów
Typ studiów
Zapisy (Szacowany)
Zapisy
Kontakty i lokalizacje
Kontakt w sprawie studiów
Kontakt w sprawie studiów
- Nazwa: Balamir Utku Seker
- Numer telefonu: +90 539 433 79 62
- E-mail: b.utku.seker@gmail.com
Lokalizacje studiów
-
-
Istanbul
-
Bağcılar, Istanbul, Turcja (Türkiye), 34200
- Bagcilar Training and Research Hospital
-
Kontakt:
- Serdar Demirgan
- Numer telefonu: +90 212 440 40 00
- E-mail: istanbuleah30@saglik.gov.tr
-
-
Kryteria uczestnictwa
Kryteria kwalifikacji
Kryteria kwalifikacji
Wiek uprawniający do nauki
- Dorosły
- Starszy dorosły
Akceptuje zdrowych ochotników
Metoda próbkowania
Badana populacja
Opis
Inclusion Criteria:
- Age 18 years or older.
- Planned placement of a totally implantable venous access port through the right internal jugular vein as part of routine clinical care.
- Availability of a suitable pre-implantation chest radiograph obtained as part of routine clinical care.
- Availability of a clinically indicated post-implantation noncontrast chest computed tomography examination.
- Absence of documented catheter manipulation or revision before the selected post-implantation computed tomography examination.
- Provision of written informed consent.
Exclusion Criteria:
- Age younger than 18 years.
- Cancellation of the routine clinical port implantation plan.
- Planned or performed port placement through a venous access route other than the right internal jugular vein.
- No suitable pre-implantation chest radiograph available for study assessment.
- Advanced thoracic deformity or mediastinal distortion precluding the required study measurements.
- Inability or unwillingness to provide written informed consent.
Plan studiów
Jak projektuje się badanie?
Szczegóły projektu
Liczba grup / kohort
Kohorty i interwencje
Grupa / KohortaGrupa / Kohorta |
Interwencja / LeczenieInterwencja / Leczenie |
|---|---|
|
Prospective Port Catheter Cohort
Adults aged 18 years or older undergoing totally implantable venous access port implantation through the right internal jugular vein as part of routine clinical care will be enrolled in a single prospective observational cohort.
In each evaluable participant, the locked AI-based automated A+B+Delta method and the prespecified manual A+B measurement method will be assessed using reference measurements from clinically indicated post-implantation noncontrast chest computed tomography.
No research-specific imaging will be performed, and the automated output will not guide port implantation or alter clinical decisions.
|
The locked automated method will use chest radiographic image analysis and a patient-specific Delta adjustment to estimate port catheter tip position.
The method will be evaluated offline using reference measurements derived from clinically indicated post-implantation noncontrast chest computed tomography.
Its output will not be provided to the clinical team and will not guide port implantation or alter clinical decisions.
The prespecified manual A+B method will estimate port catheter tip position using manual measurements on the pre-implantation chest radiograph.
It will be applied as a comparator in the same participants.
The manual research measurement will not guide port implantation or alter clinical care.
|
Co mierzy badanie?
Podstawowe miary wyniku
Podstawowe miary wyniku
Miara wyniku |
Opis środka |
Ramy czasowe |
|---|---|---|
|
Mean Absolute Error of Catheter Advancement Length Estimated by the AI-Based Automated A+B+Delta Method
Ramy czasowe: At the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care, before any documented catheter manipulation or revision
|
Mean absolute error (MAE), in millimeters, will be calculated as the arithmetic mean of the absolute differences between the catheter advancement length estimated by the locked AI-based automated A+B+Delta method and the reference catheter advancement length derived from the first eligible post-implantation noncontrast chest computed tomography examination.
The analysis population will include participants with both a valid automated output and an evaluable reference measurement.
Lower values indicate smaller estimation error.
A 95% confidence interval will be estimated using 10,000 participant-level bias-corrected and accelerated bootstrap resamples.
|
At the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care, before any documented catheter manipulation or revision
|
Miary wyników drugorzędnych
Miary wyników drugorzędnych
Miara wyniku |
Opis środka |
Ramy czasowe |
|---|---|---|
|
Median Absolute Error of Catheter Advancement Length Estimated by the AI-Based Automated A+B+Delta Method
Ramy czasowe: At the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care, before any documented catheter manipulation or revision
|
Median absolute error (MedAE), in millimeters, will be calculated as the median of the absolute differences between the catheter advancement length estimated by the locked AI-based automated A+B+Delta method and the reference catheter advancement length derived from the first eligible post-implantation noncontrast chest computed tomography examination.
The analysis population will include participants with both a valid automated output and an evaluable reference measurement.
Lower values indicate smaller estimation error.
A 95% confidence interval will be estimated using participant-level bootstrap resampling.
|
At the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care, before any documented catheter manipulation or revision
|
|
Root Mean Squared Error of Catheter Advancement Length Estimated by the AI-Based Automated A+B+Delta Method
Ramy czasowe: At the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care, before any documented catheter manipulation or revision
|
Root mean squared error (RMSE), in millimeters, will be calculated as the square root of the arithmetic mean of the squared differences between the catheter advancement length estimated by the locked AI-based automated A+B+Delta method and the reference catheter advancement length derived from the first eligible post-implantation noncontrast chest computed tomography examination.
The analysis population will include participants with both a valid automated output and an evaluable reference measurement.
Lower values indicate smaller overall estimation error, and larger individual errors have greater influence because differences are squared before averaging.
A 95% confidence interval will be estimated using participant-level bootstrap resampling.
|
At the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care, before any documented catheter manipulation or revision
|
|
Mean Signed Error of Catheter Advancement Length Estimated by the AI-Based Automated A+B+Delta Method
Ramy czasowe: At the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care, before any documented catheter manipulation or revision
|
Mean signed error, in millimeters, will be calculated as the arithmetic mean of the catheter advancement length estimated by the locked AI-based automated A+B+Delta method minus the reference catheter advancement length derived from the first eligible post-implantation noncontrast chest computed tomography examination.
The analysis population will include participants with both a valid automated output and an evaluable reference measurement.
Positive values indicate that the automated method estimates a longer advancement length than the reference, negative values indicate a shorter estimate, and zero indicates no average directional error.
A 95% confidence interval will be estimated using participant-level bootstrap resampling.
|
At the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care, before any documented catheter manipulation or revision
|
|
Mean Paired Difference in Absolute Error Between the AI-Based Automated A+B+Delta and Manual A+B Methods
Ramy czasowe: At the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care, before any documented catheter manipulation or revision
|
For each participant, absolute error will be calculated for both the locked AI-based automated A+B+Delta method and the manual A+B method relative to the reference catheter advancement length derived from the first eligible post-implantation noncontrast chest computed tomography examination.
The paired difference, in millimeters, will be calculated as automated absolute error minus manual absolute error, and its arithmetic mean will be reported.
The analysis population will include participants with an evaluable reference measurement and valid estimates from both methods.
Negative values indicate a smaller absolute error for the automated method, positive values indicate a smaller absolute error for the manual method, and zero indicates no average paired difference.
A 95% confidence interval will be estimated using participant-level bootstrap resampling.
|
At the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care, before any documented catheter manipulation or revision
|
|
Proportion of AI-Based Automated A+B+Delta Estimates With Absolute Error Within 10 mm
Ramy czasowe: At the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care, before any documented catheter manipulation or revision
|
The proportion of evaluable participants whose absolute error is 10 mm or less will be calculated for the locked AI-based automated A+B+Delta method.
Absolute error will be defined as the absolute difference between the automated estimate and the reference catheter advancement length derived from the first eligible post-implantation noncontrast chest computed tomography examination.
The denominator will include participants with both a valid automated output and an evaluable reference measurement, and the result will be reported as a percentage with a 95% Wilson score confidence interval.
A higher percentage indicates that a larger proportion of estimates meet this prespecified pragmatic accuracy criterion.
The 10 mm threshold is an operational technical accuracy criterion and not a clinical safety boundary.
|
At the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care, before any documented catheter manipulation or revision
|
|
Paired Difference in Proportions of Estimates With Absolute Error Within 10 mm Between the AI-Based Automated A+B+Delta and Manual A+B Methods
Ramy czasowe: At the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care, before any documented catheter manipulation or revision
|
For each participant, the locked AI-based automated A+B+Delta estimate and the manual A+B estimate will each be classified by whether the absolute error from the reference catheter advancement length is 10 mm or less.
The paired proportion difference, in percentage points, will be calculated as the automated qualifying proportion minus the manual qualifying proportion.
The analysis population will include participants with an evaluable reference measurement and valid estimates from both methods.
Positive values indicate a higher qualifying proportion for the automated method, and negative values indicate a higher qualifying proportion for the manual method.
A 95% confidence interval will be calculated using Newcombe's method 10; an exact two-sided McNemar test will be used as an auxiliary analysis.
The 10 mm threshold is a prespecified pragmatic accuracy criterion and not a clinical safety boundary.
|
At the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care, before any documented catheter manipulation or revision
|
|
Technical Success Rate of the AI-Based Automated A+B+Delta Method
Ramy czasowe: At baseline, during offline analysis of the pre-implantation chest radiograph obtained before port implantation
|
Technical success will be defined as the generation of a valid numerical output by the AI-based automated A+B+Delta method after the method has been locked for prospective evaluation.
The numerator will be the number of participants enrolled in the prospective cohort for whom the method generates a valid automated output, and the denominator will be all participants enrolled in the prospective cohort.
A valid output requires the prespecified image-scale, anatomical-landmark, projection, image-quality, and uncertainty criteria to be satisfied.
The result will be reported as a percentage with a 95% Wilson score confidence interval.
Technical success will be reported separately from measurement accuracy among participants with evaluable reference data.
A higher percentage indicates greater technical output availability and does not represent clinical effectiveness or safety.
|
At baseline, during offline analysis of the pre-implantation chest radiograph obtained before port implantation
|
|
Intraobserver Agreement of the CT-Derived Catheter Tip-to-Cavoatrial Junction Distance Assessed by ICC(A,1)
Ramy czasowe: At repeat assessment of the CT-derived reference measurement after a 4-week washout period
|
Intraobserver agreement of the computed tomography (CT)-derived catheter tip-to-cavoatrial junction distance (dBT) will be assessed using a two-way random-effects, absolute-agreement, single-measure intraclass correlation coefficient [ICC(A,1)].
The same radiologist will repeat the dBT measurement in a prespecified random subset of 60 cases after a 4-week washout period, using newly assigned identifiers and reshuffled images while blinded to the initial measurement.
Only cases evaluable at both readings will be analyzed.
Higher ICC values indicate greater relative agreement.
The estimate will be reported with a 95% confidence interval.
|
At repeat assessment of the CT-derived reference measurement after a 4-week washout period
|
|
Mean Intraobserver Difference in the CT-Derived Catheter Tip-to-Cavoatrial Junction Distance
Ramy czasowe: At repeat assessment of the CT-derived reference measurement after a 4-week washout period
|
The mean signed difference (second reading minus first reading), in millimeters, will be calculated for the computed tomography-derived catheter tip-to-cavoatrial junction distance in the prespecified random subset of 60 cases measured twice by the same radiologist after a 4-week washout period.
The radiologist will use newly assigned identifiers and reshuffled images and will be blinded to the initial measurement.
Only cases evaluable at both readings will be analyzed.
Positive values indicate a larger measurement at the second reading, negative values indicate a smaller measurement, and values closer to zero indicate less average systematic difference.
A 95% confidence interval will be reported.
|
At repeat assessment of the CT-derived reference measurement after a 4-week washout period
|
|
Within-Subject Standard Deviation of the CT-Derived Catheter Tip-to-Cavoatrial Junction Distance
Ramy czasowe: At repeat assessment of the CT-derived reference measurement after a 4-week washout period
|
Within-subject standard deviation (wSD), in millimeters, will quantify within-observer variation in the computed tomography-derived catheter tip-to-cavoatrial junction distance.
In the prespecified random subset of 60 cases, the same radiologist will repeat the measurement after a 4-week washout period using newly assigned identifiers and reshuffled images while blinded to the initial measurement.
Only cases evaluable at both readings will be analyzed.
For these cases, wSD will be calculated as the standard deviation of the paired differences (second reading minus first reading) divided by the square root of 2. Lower values indicate less within-observer measurement variation.
A chi-square-based 95% confidence interval will be reported.
|
At repeat assessment of the CT-derived reference measurement after a 4-week washout period
|
|
Repeatability Coefficient of the CT-Derived Catheter Tip-to-Cavoatrial Junction Distance
Ramy czasowe: At repeat assessment of the CT-derived reference measurement after a 4-week washout period
|
The repeatability coefficient (RC), in millimeters, will quantify within-observer repeatability of the computed tomography-derived catheter tip-to-cavoatrial junction distance in the prespecified random subset of 60 cases measured twice by the same radiologist after a 4-week washout period.
Paired differences will be defined as the second reading minus the first reading.
The RC will be calculated as 1.96 times the standard deviation of the paired differences.
Only cases evaluable at both readings will be analyzed.
The radiologist will use newly assigned identifiers and reshuffled images and will be blinded to the initial measurement.
Lower RC values indicate less within-observer measurement variation.
A chi-square-based 95% confidence interval will be reported.
|
At repeat assessment of the CT-derived reference measurement after a 4-week washout period
|
|
Mean Absolute Error of Catheter Advancement Length Estimated by the Manual A+B Method
Ramy czasowe: At the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care, before any documented catheter manipulation or revision
|
Mean absolute error (MAE), in millimeters, will be calculated as the arithmetic mean of the absolute differences between the catheter advancement length estimated by the prespecified manual A+B measurement method and the CT-derived reference catheter advancement length.
The analysis population will include participants with both a valid manual A+B estimate and an evaluable CT-derived reference measurement.
Lower values indicate smaller estimation error.
A 95% confidence interval will be estimated using participant-level bootstrap resampling.
|
At the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care, before any documented catheter manipulation or revision
|
|
Median Absolute Error of Catheter Advancement Length Estimated by the Manual A+B Method
Ramy czasowe: At the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care, before any documented catheter manipulation or revision
|
Median absolute error (MedAE), in millimeters, will be calculated as the median of the absolute differences between the catheter advancement length estimated by the prespecified manual A+B measurement method and the CT-derived reference catheter advancement length.
The analysis population will include participants with both a valid manual A+B estimate and an evaluable CT-derived reference measurement.
Lower values indicate smaller estimation error.
A 95% confidence interval will be estimated using participant-level bootstrap resampling.
|
At the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care, before any documented catheter manipulation or revision
|
|
Root Mean Squared Error of Catheter Advancement Length Estimated by the Manual A+B Method
Ramy czasowe: At the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care, before any documented catheter manipulation or revision
|
Root mean squared error (RMSE), in millimeters, will be calculated as the square root of the arithmetic mean of the squared differences between the catheter advancement length estimated by the prespecified manual A+B measurement method and the CT-derived reference catheter advancement length.
The analysis population will include participants with both a valid manual A+B estimate and an evaluable CT-derived reference measurement.
Lower values indicate smaller overall estimation error, and larger individual errors have greater influence because differences are squared before averaging.
A 95% confidence interval will be estimated using participant-level bootstrap resampling.
|
At the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care, before any documented catheter manipulation or revision
|
|
Mean Signed Error of Catheter Advancement Length Estimated by the Manual A+B Method
Ramy czasowe: At the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care, before any documented catheter manipulation or revision
|
Mean signed error, in millimeters, will be calculated as the arithmetic mean of the catheter advancement length estimated by the prespecified manual A+B measurement method minus the CT-derived reference catheter advancement length.
The analysis population will include participants with both a valid manual A+B estimate and an evaluable CT-derived reference measurement.
Positive values indicate that the manual method estimates a longer advancement length than the reference, negative values indicate a shorter estimate, and zero indicates no average directional error.
A 95% confidence interval will be estimated using participant-level bootstrap resampling.
|
At the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care, before any documented catheter manipulation or revision
|
|
Mean Absolute Error of the AI-Based Automated A+B+Delta Method Within 24 Hours After Port Implantation and Before Catheter Manipulation
Ramy czasowe: Within 24 hours after port implantation and before any documented catheter manipulation, at the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care
|
In the prespecified sensitivity subset of participants whose first eligible post-implantation noncontrast chest computed tomography examination is obtained within 24 hours after port implantation and before any documented catheter manipulation, mean absolute error (MAE), in millimeters, will be calculated as the arithmetic mean of the absolute differences between the catheter advancement length estimated by the locked AI-based automated A+B+Delta method and the CT-derived reference catheter advancement length.
Only participants with a valid automated output and an evaluable CT-derived reference measurement will be included.
Lower values indicate smaller estimation error.
A 95% confidence interval will be estimated using participant-level bootstrap resampling.
|
Within 24 hours after port implantation and before any documented catheter manipulation, at the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care
|
|
Proportion of AI-Based Automated A+B+Delta Estimates With Absolute Error Within 10 mm Within 24 Hours After Port Implantation and Before Catheter Manipulation
Ramy czasowe: Within 24 hours after port implantation and before any documented catheter manipulation, at the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care
|
In the prespecified sensitivity subset of participants whose first eligible post-implantation noncontrast chest computed tomography examination is obtained within 24 hours after port implantation and before any documented catheter manipulation, the proportion of evaluable participants whose absolute error is 10 mm or less will be calculated for the locked AI-based automated A+B+Delta method.
Absolute error will be defined as the absolute difference between the automated estimate and the CT-derived reference catheter advancement length.
The denominator will include participants with both a valid automated output and an evaluable CT-derived reference measurement.
The result will be reported as a percentage with a 95% Wilson score confidence interval.
The 10 mm threshold is a prespecified pragmatic technical accuracy criterion and not a clinical safety boundary.
|
Within 24 hours after port implantation and before any documented catheter manipulation, at the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care
|
|
Bland-Altman Limits of Agreement for the CT-Derived Catheter Tip-to-Cavoatrial Junction Distance
Ramy czasowe: At repeat assessment of the CT-derived reference measurement after a 4-week washout period
|
Bland-Altman limits of agreement (LoA), in millimeters, will be calculated for the CT-derived catheter tip-to-cavoatrial junction distance in a prespecified random subset of 60 cases measured twice by the same radiologist after a 4-week washout.
Paired differences will be the second reading minus the first.
The mean difference and LoA (mean difference +/- 1.96 times the standard deviation of paired differences) will be reported.
Only cases evaluable at both readings will be analyzed.
The radiologist will use new identifiers and reshuffled images while blinded to the initial measurement.
The mean difference will have a t-based 95% confidence interval.
Separate two-sided 95% confidence intervals for each LoA will be calculated using the exact noncentral t method, assuming normally distributed paired differences.
Narrower limits indicate less within-observer measurement variation.
|
At repeat assessment of the CT-derived reference measurement after a 4-week washout period
|
|
Bland-Altman Limits of Agreement Between the AI-Based Automated A+B+Delta Estimate and the CT-Derived Reference
Ramy czasowe: At the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care, before any documented catheter manipulation or revision
|
For participants with a valid automated output and an evaluable CT-derived reference measurement, paired differences, in millimeters, will be defined as the catheter advancement length estimated by the locked AI-based automated A+B+Delta method minus the CT-derived reference catheter advancement length.
The mean difference and lower and upper 95% limits of agreement (LoA), calculated as the mean difference +/- 1.96 times the standard deviation of paired differences, will be reported.
The mean difference will have a t-based 95% confidence interval.
Separate two-sided 95% confidence intervals for each LoA will be calculated using the exact noncentral t method, assuming normally distributed paired differences.
Positive differences indicate longer automated estimates than the reference, and negative differences indicate shorter estimates.
This analysis describes agreement and does not establish clinical interchangeability or safety.
|
At the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care, before any documented catheter manipulation or revision
|
|
Bland-Altman Limits of Agreement Between the Manual A+B Estimate and the CT-Derived Reference
Ramy czasowe: At the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care, before any documented catheter manipulation or revision
|
For participants with a valid manual A+B estimate and an evaluable CT-derived reference measurement, paired differences, in millimeters, will be defined as the catheter advancement length estimated by the prespecified manual A+B method minus the CT-derived reference catheter advancement length.
The mean difference and lower and upper 95% limits of agreement (LoA), calculated as the mean difference +/- 1.96 times the standard deviation of paired differences, will be reported.
The mean difference will have a t-based 95% confidence interval.
Separate two-sided 95% confidence intervals for each LoA will be calculated using the exact noncentral t method, assuming normally distributed paired differences.
Positive differences indicate longer manual estimates than the reference, and negative differences indicate shorter estimates.
This analysis describes agreement and does not establish clinical interchangeability or safety.
|
At the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care, before any documented catheter manipulation or revision
|
Inne miary wyników
Inne miary wyników
Miara wyniku |
Opis środka |
Ramy czasowe |
|---|---|---|
|
Huber Regression Slope for the Association Between the Port Implantation-to-CT Interval and Signed Error
Ramy czasowe: At the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care, before any documented catheter manipulation or revision
|
In this prespecified exploratory analysis, the association between the interval from port implantation to the first eligible post-implantation noncontrast chest computed tomography examination and the signed error of the locked AI-based automated A+B+Delta method will be assessed among participants with a valid automated output and an evaluable CT-derived reference measurement.
Signed error, in millimeters, will be defined as the automated catheter advancement length estimate minus the CT-derived reference length.
The interval, recorded in hours, will be transformed as log2(1 + hours).
A robust Huber regression slope with a two-sided 95% confidence interval will be reported; Spearman's rank correlation coefficient with a two-sided 95% confidence interval will be an auxiliary analysis.
Raw and Holm-adjusted Spearman p-values will be reported.
The slope is expressed in millimeters per log2(1 + hours); positive values indicate increasingly positive signed error with longer intervals.
|
At the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care, before any documented catheter manipulation or revision
|
|
Huber Regression Slope for the Association Between the Port Implantation-to-CT Interval and Absolute Error
Ramy czasowe: At the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care, before any documented catheter manipulation or revision
|
In this prespecified exploratory analysis, the association between the interval from port implantation to the first eligible post-implantation noncontrast chest computed tomography examination and the absolute error of the locked AI-based automated A+B+Delta method will be assessed among participants with a valid automated output and an evaluable CT-derived reference measurement.
Absolute error, in millimeters, will be defined as the absolute difference between the automated catheter advancement length estimate and the CT-derived reference length.
The interval, recorded in hours, will be transformed as log2(1 + hours).
A robust Huber regression slope with a two-sided 95% confidence interval will be reported; Spearman's rank correlation coefficient with a two-sided 95% confidence interval will be an auxiliary analysis.
Raw and Holm-adjusted p-values will be reported for the Spearman test.
Slope is in millimeters per log2(1 + hours); positive values indicate increasing absolute error.
|
At the first eligible post-implantation noncontrast chest computed tomography examination obtained during routine care, before any documented catheter manipulation or revision
|
Współpracownicy i badacze
Sponsor
Sponsor
Daty zapisu na studia
Główne daty studiów
Rozpoczęcie studiów (Szacowany)
Rozpoczęcie studiów
Zakończenie podstawowe (Szacowany)
Zakończenie podstawowe
Ukończenie studiów (Szacowany)
Ukończenie studiów
Daty rejestracji na studia
Pierwszy przesłany
Pierwszy przesłany
Pierwszy przesłany, który spełnia kryteria kontroli jakości
Pierwszy przesłany, który spełnia kryteria kontroli jakości
Pierwszy wysłany (Rzeczywisty)
Pierwszy wysłany
Aktualizacje rekordów badań
Ostatnia wysłana aktualizacja (Rzeczywisty)
Ostatnia wysłana aktualizacja
Ostatnia przesłana aktualizacja, która spełniała kryteria kontroli jakości
Ostatnia przesłana aktualizacja, która spełniała kryteria kontroli jakości
Ostatnia weryfikacja
Ostatnia weryfikacja
Więcej informacji
Terminy związane z tym badaniem
Słowa kluczowe
Inne numery identyfikacyjne badania
Inne numery identyfikacyjne badania
- E-10840098-50.04-1975
Plan dla danych uczestnika indywidualnego (IPD)
Planujesz udostępniać dane poszczególnych uczestników (IPD)?
Opis planu IPD
Informacje o lekach i urządzeniach, dokumenty badawcze
Bada produkt leczniczy regulowany przez amerykańską FDA
Bada produkt urządzenia regulowany przez amerykańską FDA
Te informacje zostały pobrane bezpośrednio ze strony internetowej clinicaltrials.gov bez żadnych zmian. Jeśli chcesz zmienić, usunąć lub zaktualizować dane swojego badania, skontaktuj się z register@clinicaltrials.gov. Gdy tylko zmiana zostanie wprowadzona na stronie clinicaltrials.gov, zostanie ona automatycznie zaktualizowana również na naszej stronie internetowej .