Effect of Prone Position on Continuous Noninvasive Hemoglobin Measurement
Does the Prone Position Affect Continuous Noninvasive Hemoglobin Measurement?
This prospective observational study aims to evaluate whether the prone position affects the agreement between noninvasive hemoglobin (SpHb) measurements and laboratory hemoglobin values in patients undergoing lumbar stabilization surgery under general anesthesia.
SpHb provides continuous, noninvasive monitoring of hemoglobin using multi-wavelength pulse CO-oximetry. However, its accuracy may be affected by changes in peripheral perfusion, hemodynamic status, and patient position. The prone position used during spine surgery may alter cardiovascular function and peripheral perfusion and may therefore influence SpHb measurements.
In this study, SpHb values obtained in the supine and prone positions will be compared with simultaneously obtained laboratory hemoglobin values. The study will also evaluate the ability of SpHb to follow changes in hemoglobin during surgery. The results may help clarify the clinical usefulness of continuous noninvasive hemoglobin monitoring during surgery performed in the prone position.
Przegląd badań
Status
Status
Warunki
Warunki
Interwencja / Leczenie
Interwencja / Leczenie
Szczegółowy opis
This is a prospective, single-center observational study involving adult patients undergoing lumbar stabilization surgery under general anesthesia. Eligible patients will be older than 18 years and classified as American Society of Anesthesiologists (ASA) physical status I-III.
During surgery, noninvasive hemoglobin concentration (SpHb) will be monitored using a pulse CO-oximetry sensor placed on the finger as part of intraoperative monitoring. SpHb measurements will be compared with laboratory hemoglobin values obtained from blood gas analysis. Measurements obtained during supine and prone positioning will be evaluated to determine whether prone positioning affects the agreement between noninvasive and laboratory hemoglobin measurements.
Demographic and clinical characteristics, intraoperative hemodynamic variables, laboratory and blood gas parameters, estimated intraoperative blood loss, amount of transfused blood products, postoperative intensive care unit stay, and total hospital length of stay will also be recorded. The study will additionally assess the potential of SpHb for following intraoperative hemoglobin trends.
Typ studiów
Typ studiów
Zapisy (Szacowany)
Zapisy
Kontakty i lokalizacje
Kontakt w sprawie studiów
Kontakt w sprawie studiów
- Nazwa: Sule Altuncu, MD
- Numer telefonu: 02166063300
- E-mail: sancaktepeetikkurul@gmail.com
Lokalizacje studiów
-
-
Istanbul
-
Istanbul, Istanbul, Turcja (Türkiye), 34000
- Rekrutacyjny
- Sancaktepe Sehit Prof. Dr. Ilhan Varank Training and Research Hospital
-
Kontakt:
- Sule Altuncu, MD
- Numer telefonu: 02166063300
- E-mail: sancaktepeetikkurul@gmail.com
-
-
Kryteria uczestnictwa
Kryteria kwalifikacji
Kryteria kwalifikacji
Wiek uprawniający do nauki
- Dorosły
- Starszy dorosły
Akceptuje zdrowych ochotników
Metoda próbkowania
Badana populacja
Opis
Inclusion Criteria:
- Age 18 years or older
- American Society of Anesthesiologists (ASA) physical status I-III
- Scheduled to undergo lumbar stabilization surgery under general anesthesia
- Provision of written informed consent for participation in the study
Exclusion Criteria:
- Refusal to participate in the study
- ASA physical status IV-V
- Hemoglobin level <10 g/dL
- Presence of peripheral arterial disease
Plan studiów
Jak projektuje się badanie?
Szczegóły projektu
Liczba grup / kohort
Kohorty i interwencje
Grupa / KohortaGrupa / Kohorta |
Interwencja / LeczenieInterwencja / Leczenie |
|---|---|
|
Patients Undergoing Lumbar Stabilization Surgery
Adult patients undergoing lumbar stabilization surgery under general anesthesia.
Noninvasive hemoglobin (SpHb) and laboratory hemoglobin measurements will be obtained and compared during supine and prone positioning.
|
Prone positioning is performed as part of standard surgical care for lumbar stabilization surgery and is not assigned by the investigators.
The study will evaluate the association between prone positioning and the agreement between noninvasive hemoglobin (SpHb) measurements and simultaneously obtained laboratory hemoglobin values.
|
Co mierzy badanie?
Podstawowe miary wyniku
Podstawowe miary wyniku
Miara wyniku |
Opis środka |
Ramy czasowe |
|---|---|---|
|
Agreement Between Noninvasive Hemoglobin (SpHb) and Laboratory Hemoglobin Measurements
Ramy czasowe: Intraoperative period, during supine and prone positioning on the day of surgery
|
Paired noninvasive hemoglobin (SpHb) and laboratory hemoglobin values will be obtained at corresponding intraoperative time points in the supine and prone positions.
The primary outcome is the agreement between SpHb and laboratory hemoglobin measurements and whether this agreement changes after positioning the patient prone.
|
Intraoperative period, during supine and prone positioning on the day of surgery
|
Miary wyników drugorzędnych
Miary wyników drugorzędnych
Miara wyniku |
Opis środka |
Ramy czasowe |
|---|---|---|
|
Ability of SpHb to Track Intraoperative Hemoglobin Changes
Ramy czasowe: Intraoperative period on the day of surgery
|
Changes in noninvasive hemoglobin (SpHb) measurements over the intraoperative period will be compared with corresponding changes in laboratory hemoglobin values to evaluate the ability of SpHb to track the direction and magnitude of hemoglobin changes during surgery.
|
Intraoperative period on the day of surgery
|
Współpracownicy i badacze
Sponsor
Sponsor
Śledczy
Śledczy
- Główny śledczy: Altuncu, Şehit Prof. Dr. İlhan Varank Sancaktepe Training and Research Hospital
Publikacje i pomocne linki
Publikacje ogólne
- Tang B, Yu X, Xu L, Zhu A, Zhang Y, Huang Y. Continuous noninvasive hemoglobin monitoring estimates timing for detecting anemia better than clinicians: a randomized controlled trial. BMC Anesthesiol. 2019 May 17;19(1):80. doi: 10.1186/s12871-019-0755-1.
- Berkow L, Rotolo S, Mirski E. Continuous noninvasive hemoglobin monitoring during complex spine surgery. Anesth Analg. 2011 Dec;113(6):1396-402. doi: 10.1213/ANE.0b013e318230b425. Epub 2011 Sep 29.
- Park SK, Hur C, Kim YW, Yoo S, Lim YJ, Kim JT. Noninvasive hemoglobin monitoring for maintaining hemoglobin concentration within the target range during major noncardiac surgery: A randomized controlled trial. J Clin Anesth. 2024 May;93:111326. doi: 10.1016/j.jclinane.2023.111326. Epub 2023 Nov 20.
- Adel A, Awada W, Abdelhamid B, Omar H, Abd El Dayem O, Hasanin A, Rady A. Accuracy and trending of non-invasive hemoglobin measurement during different volume and perfusion statuses. J Clin Monit Comput. 2018 Dec;32(6):1025-1031. doi: 10.1007/s10877-018-0101-z. Epub 2018 Jan 15.
- Erdogan Kayhan G, Colak YZ, Sanli M, Ucar M, Toprak HI. Accuracy of non-invasive hemoglobin monitoring by pulse CO-oximeter during liver transplantation. Minerva Anestesiol. 2017 May;83(5):485-492. doi: 10.23736/S0375-9393.17.11652-4. Epub 2017 Jan 20.
Daty zapisu na studia
Główne daty studiów
Rozpoczęcie studiów (Rzeczywisty)
Rozpoczęcie studiów
Zakończenie podstawowe (Szacowany)
Zakończenie podstawowe
Ukończenie studiów (Szacowany)
Ukończenie studiów
Daty rejestracji na studia
Pierwszy przesłany
Pierwszy przesłany
Pierwszy przesłany, który spełnia kryteria kontroli jakości
Pierwszy przesłany, który spełnia kryteria kontroli jakości
Pierwszy wysłany (Rzeczywisty)
Pierwszy wysłany
Aktualizacje rekordów badań
Ostatnia wysłana aktualizacja (Rzeczywisty)
Ostatnia wysłana aktualizacja
Ostatnia przesłana aktualizacja, która spełniała kryteria kontroli jakości
Ostatnia przesłana aktualizacja, która spełniała kryteria kontroli jakości
Ostatnia weryfikacja
Ostatnia weryfikacja
Więcej informacji
Terminy związane z tym badaniem
Słowa kluczowe
Dodatkowe istotne warunki MeSH
Inne numery identyfikacyjne badania
Inne numery identyfikacyjne badania
- PRONE-NIHb-2026-486
Plan dla danych uczestnika indywidualnego (IPD)
Planujesz udostępniać dane poszczególnych uczestników (IPD)?
Informacje o lekach i urządzeniach, dokumenty badawcze
Bada produkt leczniczy regulowany przez amerykańską FDA
Bada produkt urządzenia regulowany przez amerykańską FDA
Te informacje zostały pobrane bezpośrednio ze strony internetowej clinicaltrials.gov bez żadnych zmian. Jeśli chcesz zmienić, usunąć lub zaktualizować dane swojego badania, skontaktuj się z register@clinicaltrials.gov. Gdy tylko zmiana zostanie wprowadzona na stronie clinicaltrials.gov, zostanie ona automatycznie zaktualizowana również na naszej stronie internetowej .