Sensor Validation Study - Quality Assurance
A Quality Assurance Study of Respiratory Rate Measurements on Obese Patients With a Novel Monitoring Technology
The respiratory rate is an important parameter in clinical medicine. It is defined as the number of breaths per minute. Currently this is measured at the bedside in clinical practice by counting the breaths, however the gold standard for measuring this vital sign is the capnograph. For the patient, this involves wearing a tube in their nose and around their ears while trying to minimise their movement and talking so that the measurements can be taken.
A new respiratory rate monitor, RespiraSense, is non-invasive and measures the respiratory rate by measuring the displacement between the ribs and abdomen. This research study is intended to validate that this technology is effective and accurate on people with a bigger body mass.
Subjects with a BMI > 35 will be invited to participate should they meet all of the eligibility criteria. If patients agree to participate, following informed consent, subjects will be monitored for one hour with both the capnograph and RespiraSense measuring at the same time so their results can be compared
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Study Type
Study Type
Enrollment (Actual)
Enrollment
Contacts and Locations
Study Locations
-
-
-
Portsmouth, United Kingdom, PO6 3LY
- Queen Alexandra Hospital
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Sampling Method
Study Population
Description
Inclusion Criteria:
- ≥ 18 Years
Have a BMI ≥35
- For the sub-group analysis at least 5 subjects with a BMI 35-49 and at least 5 BMI > 50 will be needed
- Are willing to voluntarily sign a statement of informed consent to participate in this investigation
Exclusion Criteria:
- Allergic to medical grade skin adhesive
- Continuous oral long term steroid use. To be free from oral steroid use this is defined as not using oral steroids in the 4 weeks previous to enrolment
- Patients under the influence of substance abuse (drug or alcohol) that may interfere with their ability to cooperate and comply with the investigation procedures
- Any disorder, including cognitive dysfunction, which would affect the ability to accurately complete questionnaires and freely give full informed consent
Study Plan
How is the study designed?
Design Details
- Observational Models: Cohort
- Time Perspectives: Prospective
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Comparison of PMD RespiraSense against capnography in obese patients.
Time Frame: 1 hour
|
Analysed using a Bland Altman (BA) analysis to measure limits of agreement between both methods.
The BA analysis will correct for repeated measures within each subject if required.
In addition, a Deming Regression will be performed on the collected data.
|
1 hour
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Comparison of PMD RespiraSense in obese patients with BMI below 50 and those with BMI of 50 or above.
Time Frame: 1 hour
|
Analysed using a Bland Altman (BA) analysis to measure limits of agreement between both methods.
The BA analysis will correct for repeated measures within each subject if required.
In addition, a Deming Regression will be performed on the collected data.
|
1 hour
|
|
Identification of any accuracy spread patterns as BMI increases.
Time Frame: 1 hour
|
Analysed using a Bland Altman (BA) analysis to measure limits of agreement between both methods.
The BA analysis will correct for repeated measures within each subject if required.
In addition, a Deming Regression will be performed on the collected data.
|
1 hour
|
|
Collection of feedback on device comfort when worn by patients with high BMI.
Time Frame: 1 hour
|
Will not be analysed formally.
|
1 hour
|
Other Outcome Measures
Other Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Exploratory: To measure RespiraSense device alarm generation and assess the rationale behind these to determine if false alarms and therefore alarm fatigue is occurring
Time Frame: 1 hour
|
Will not be analysed formally.
|
1 hour
|
Collaborators and Investigators
Sponsor
Sponsor
Investigators
Investigators
- Principal Investigator: Lorraine Albon, Portsmouth Hospitals NHS Trust
Study record dates
Study Major Dates
Study Start (Actual)
Study Start
Primary Completion (Actual)
Primary Completion
Study Completion (Actual)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Actual)
First Posted
Study Record Updates
Last Update Posted (Actual)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Other Study ID Numbers
Other Study ID Numbers
- PMD-CS-011
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Respiratory Rate
-
NCT04179279Completed
-
NCT03754231CompletedHeart Rate | Respiratory Rate | Vital Sign
-
NCT03832205CompletedExercise | Heart Rate | Respiratory Rate
-
NCT03750318Active, not recruitingHeart Rate | Respiratory Rate | Vital Sign
-
NCT07217327RecruitingDetection of Heart Rate and Respiratory Rate
-
NCT07477834CompletedSleep | Heart Rate | Respiratory Rate | Air Pollution Exposure | SpO2 | HRV | Resting Heart Rate
-
NCT05889832CompletedHeart Rate Variability | Heart Rate | Respiratory Rate
-
NCT06911541CompletedRespiratory Rate | Capnography | Wireless Vital Signs Monitoring System | Respiratory Rate Accuracy Test
Clinical Trials on RespiraSense Respiratory Rate Monitor
-
NCT04003415CompletedAsthma | Copd | Respiratory Effort | Contactless Vital Sign Monitoring
-
NCT06911541CompletedRespiratory Rate | Capnography | Wireless Vital Signs Monitoring System | Respiratory Rate Accuracy Test
-
NCT01176539Completed
-
NCT04215887Completed
-
NCT03993886CompletedHeart Failure | Chronic Kidney Diseases
-
NCT03067558Terminated
-
NCT04712240CompletedValidation of Respiratory Rate Monitor Versus Capnography
-
NCT02530021UnknownAngina Pectoris | Ischemic Heart Disease