Shock Efficacy Study Comparing Fixed Tilt and Fixed Pulse Width in a General Implantable-Cardioverter Population (EFICAZ)
Shock Efficacy, a Randomized Study Comparing Fixed Tilt and Fixed Pulse Width in a General Implantable-Cardioverter Population
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Detailed Description
Study Type
Study Type
Enrollment (Actual)
Enrollment
Phase
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
-
Madrid, Spain
- Hospital Puerta de Hierro
-
-
Andalucia
-
Malaga, Andalucia, Spain
- Hospital Virgen de la Victoria
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Description
Inclusion Criteria:
Subjects who meet the current ESC indications for ICD or CRT-D implant. This includes upgrades indications from pacemaker to ICD or to CRT-D or replacements of ICDs or CRT-Ds provided that there is no previous record of ineffective shocks
- Subjects who are willing to be registered on the Merlin.net remote follow-up program
- Subject ≥18 years
- Subjects with life expectancy of at least 1 year
- Subjects who agree to comply with the follow-up program included in this protocol
Exclusion Criteria:
• Subjects on the waiting list for a heart transplant or under consideration for a transplant over the next 12 months
- Subjects with Myocardial infarction or unstable angina within 40 days prior to enrollment
- Subjects who have undergone a recent cardiac revascularization (PTCA, Stent or CABG) procedure in the 4 weeks prior to enrollment
- Functional NYHA Class IV
- Subjects who are participating in another ICD or CRT-D study
- Subjects who are pregnant at the time of enrollment or women of childbearing potential that do not take contraceptives
- Any disability or limitations to correctly understand or complete the study (physical, intellectual, logistical)
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Treatment
- Allocation: Randomized
- Interventional Model: Parallel Assignment
- Masking: None (Open Label)
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Active Comparator: Fixed Tilt (65%)
This is the control group, so device programming for shock energy is the default setting
|
This is the standard group, so device programming for shock energy is the default setting
|
|
Active Comparator: Fixed Pulse Width
This is the Study group.
|
The device has to be programmed with fixed pulse width for each phase of the biphasic waveform
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Efficacy of the first appropriate shock to revert a tachyarrhythmia to sinus rhythm
Time Frame: 5 years
|
Rate of rhythm conversion from Ventricular Tachyarrhythmia to sinus rhythm by the first appropriate shock programmed to 25 Joules (or equivalent voltage) programming the shock waveform in Fixed Pulse Width (Study Group) versus the Fixed Tilt with value 65% (Control Group).
|
5 years
|
Collaborators and Investigators
Sponsor
Sponsor
Investigators
Investigators
- Principal Investigator: Ignacio Fernandez, MD
- Principal Investigator: Javier Alzueta, MD
Study record dates
Study Major Dates
Study Start (Actual)
Study Start
Primary Completion (Actual)
Primary Completion
Study Completion (Actual)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Actual)
First Posted
Study Record Updates
Last Update Posted (Actual)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
Other Study ID Numbers
- CRD_914
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
product manufactured in and exported from the U.S.
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.