Validation Study of an Advanced Blood Pressure Monitor (TALISMAN)
Study of Validation of the Advanced Blood Pressure Monitor Withings WPM04 for the Detection of Anomalies of Arterial Pressure, Arrhythmia and Valvular Heart Diseases
The proposed clinical study aims to validate the effectiveness of the WPM04 blood pressure monitor developed by Withings as a diagnostic tool for cardiovascular disorders.
The detection of cardiovascular disorders is oriented along three axes, namely the measurement of blood pressure, the identification of rhythm disorders such as atrial fibrillation and the identification of heart murmur indicating valvular heart disease.
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Study Type
Study Type
Enrollment (Actual)
Enrollment
Phase
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
-
Aix-en-Provence, France, 13090
- Clinique AXIUM Centre de cardiologie
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Description
Inclusion Criteria:
- Male or female over 18 years old,
- Subject who signed the consent form,
- Subject affiliated to a social security scheme or entitled.
Exclusion Criteria:
- Underage subject;
- Subject who refused to participate in the study;
- Subject participating in other research or clinical studies;
- Subject with body mass index greater than 35 kg/m2;
- Subject in linguistic or psychic incapacity to sign an informed consent;
- Subject with congenital cardiomyopathy;
- Subject with aorto-venous fistula;
- Subject with pulmonary pathology that may degrade the quality of the recording;
- Subject presenting a serious pathology compromising the realization of the examinations envisaged in the test;
- Subject with an arm injury incompatible with the use of an armband sphygmomanometer;
- Subject receiving an intravenous perfusion on the left arm;
- Subject with electrical stimulation by pacemaker, implantable cardioverter-defibrillator;
- Subject with prosthetic heart valve;
- Subject under kidney dialysis;
- Hemodynamically unstable subject (acute heart failure or cardiac assistance).
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Diagnostic
- Allocation: Non-Randomized
- Interventional Model: Parallel Assignment
- Masking: None (Open Label)
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Other: Blood pressure (BP)
Group for the validation of blood pressure measurement by the device
|
sequential BP measurement with Withings WPM04 and sphygmomanometer with dual observer auscultation
phonocardiogram with digital stethoscope of Withings WPM04 and Doppler echocardiography
|
|
Other: Atrial fibrillation (AF)
Group for the validation of detection of AF by the device
|
phonocardiogram with digital stethoscope of Withings WPM04 and Doppler echocardiography
single lead ECG with Withings WPM04 and 12 lead ECG
|
|
Other: Valvular heart disease (VHD)
Group for the validation of detection of VHD by the device
|
sequential BP measurement with Withings WPM04 and sphygmomanometer with dual observer auscultation
phonocardiogram with digital stethoscope of Withings WPM04 and Doppler echocardiography
single lead ECG with Withings WPM04 and 12 lead ECG
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
BP precision and accuracy
Time Frame: 1 visit (45 minutes)
|
mean error and standard deviation on measurements, and standard deviation on patients (criteria 1 and 2 of ANSI/AAMI/ISO 81060-2:2013) against sphygmomanometer dual observer auscultation
|
1 visit (45 minutes)
|
|
AF and SR classification accuracy
Time Frame: 1 visit (30 minutes)
|
sensitivity and specificity of detection of AF and SR from 1-lead ECG against 12-lead ECG
|
1 visit (30 minutes)
|
|
VHD detection accuracy
Time Frame: 1 visit (30 minutes)
|
sensitivity and specificity of detection of VHD from phonocardiogram against Doppler echocardiogram
|
1 visit (30 minutes)
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
safety evaluation
Time Frame: 1 visit (1 hour)
|
frequency of adverse events
|
1 visit (1 hour)
|
Collaborators and Investigators
Sponsor
Sponsor
Collaborators
Collaborators
Study record dates
Study Major Dates
Study Start (Actual)
Study Start
Primary Completion (Actual)
Primary Completion
Study Completion (Actual)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Actual)
First Posted
Study Record Updates
Last Update Posted (Actual)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Keywords
Additional Relevant MeSH Terms
Other Study ID Numbers
Other Study ID Numbers
- 2018-A02847-48
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Hypertension
-
NCT04863508RecruitingHypertension,Essential | Hypertension, Masked
-
NCT05395403CompletedHypertension | Hypertension, Resistant to Conventional Therapy | Uncontrolled Hypertension | Hypertension, White Coat
-
NCT07487441Not yet recruitingPulmonary Hypertension | Pulmonary Arterial Hypertension (PAH)
-
NCT07577973Not yet recruiting
-
NCT07399912Enrolling by invitationHypertension,Essential
-
NCT07363447CompletedHypertension | Arterial Hypertension | Systemic Hypertension
-
NCT07598760Not yet recruitingHypertension, Systolic | Hypertension (HTN) | Heart Failure With Preserved Ejection Fraction (HFpEF
-
NCT07566650Not yet recruiting
-
NCT07486453Not yet recruitingPrimary Hypertension
-
NCT07147595CompletedHypertension (HTN) | Hypertension Arterial
Clinical Trials on BP measurement
-
NCT02321618Completed
-
NCT02198144Completed
-
NCT06491433CompletedHypertension | Prehypertension | Normal Blood Pressure | Low Blood Pressure
-
NCT06248463Not yet recruitingHypertension | Pregnancy Induced Hypertension
-
NCT01835249Completed