Can the Electronic Nose Smell COVID-19 Antibodies?

July 16, 2020 updated by: Nicole Bouvy, Maastricht University Medical Center

Can the Electronic Nose Smell COVID-19 Antibodies? A Proof-of-principle Study

Corona Virus Disease (COVID-19), spread worldwide and has become an emergency of major international concern. In March 2020, the WHO declared the COVID-19 outbreak a global pandemic. Accurate and fast diagnosis is crucial in managing the pandemic. Current diagnostic approaches raise several difficulties: they are time-consuming, expensive, invasive, and most important lacking high sensitivity. The gold standard diagnostic test for COVID-19, reverse transcription polymerase chain reaction (RT-PCR), is highly dependent on adequate deep sampling of the swab in the naso- and oropharynx. A new diagnostic test that can correctly and rapidly identify infected patients and asymptomatic carriers is urgently required to prevent further virus transmission and thus reduce mortality rates.

Aim: This proof-of-principle study aims to investigate if an electronic nose (Aeonose) can distinguish individuals with antibodies from individuals without antibodies against COVID-19 based on analysis of volatile organic compounds (VOCs).

Methods: between April and July 2020, persons undergoing RT-PCR and a serology test for COVID-19 were recruited at Maastricht UMC+ for breath analysis. All participants had to breathe through the Aeonose for five consecutive minutes. The VOC pattern in their exhaled breath was then linked to the matching RT-PCR and serological test results.

Study Overview

Status

Completed

Conditions

Intervention / Treatment

Study Type

Interventional

Enrollment (Actual)

222

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

      • Maastricht, Netherlands, 6229 HX
        • Maastricht University Medical Center

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years and older (Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

  • Participants of whom an oropharyngeal and nasopharyngeal swab was collected for RT-PCR and on whom serology testing for the detection of antibodies was performed.

Exclusion Criteria:

  • Participants who where experiencing dyspnea or needed supplemental oxygen.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Diagnostic
  • Allocation: N/A
  • Interventional Model: Single Group Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Other: COVID-19 suspected
Participants where included if an oropharyngeal and nasopharyngeal swab was collected for RT-PCR and serology testing had been performed, or if participants have had a confirmed COVID-19 diagnosis in the previous days or weeks with an indication for re-testing via PCR and serology testing at the moment of inclusion
All participants breathed through the Aeonose for five minutes. This device contains metal-oxide sensors that change in conductivity upon reaction with VOCs in exhaled breath. These conductivity changes are input data for machine-learning and used for pattern recognition. A nose clip was placed on the nose of each participant to avoid entry of non-filtered air in the device. Before measuring, the Aeonose was flushed with room air, guided through a carbon filter as well. During each measurement, a video was displayed to distract the participant and to reduce the chance of hyperventilation. Failed breath tests were excluded from analysis; the reason for failure was documented. Four similar Aeonose devices were used for breath analysis. A full-measurement procedure required sixteen minutes.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
COVID-19 antibodies vs COVID-19 negative
Time Frame: 3 months
Ability of the electronic nose (Aeonose) to distinguish individuals with antibodies from individuals without antibodies against COVID-19 based on analysis of volatile organic compounds (VOCs).
3 months

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Sponsor

Publications and helpful links

The person responsible for entering information about the study voluntarily provides these publications. These may be about anything related to the study.

General Publications

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

April 6, 2020

Primary Completion (Actual)

June 30, 2020

Study Completion (Actual)

July 1, 2020

Study Registration Dates

First Submitted

July 15, 2020

First Submitted That Met QC Criteria

July 16, 2020

First Posted (Actual)

July 17, 2020

Study Record Updates

Last Update Posted (Actual)

July 17, 2020

Last Update Submitted That Met QC Criteria

July 16, 2020

Last Verified

July 1, 2020

More Information

Terms related to this study

Other Study ID Numbers

  • eNoseCOVIDantibodies

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

YES

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Covid19

Clinical Trials on Aeonose

Search Similar Trials