Oxygen Saturation: Analytical Comparison for COVID (SHOCOVID)
Oxygen Hemoglobin Saturation in COVID Patients: Comparison of Arterial Gasometry and Pulse Oximetry
In many patients, respiratory Sars-Cov2 infection causes arterial hypoxemia, which remains without signs of verbalized respiratory distress, up to a point. This phenomenon, called "happy" or "silent" hypoxemia, has a plural pathophysiological basis. Hypoxemia has been shown to be predictively associated with admission to the ICU. Therefore, the question of constant monitoring of oxygenation, practiced on a large scale, at home, in asymptomatic patients or contact cases, arises. A large number of portable pulse oximeter are currently freely available on the market; however, their clinical validation remains sometimes doubtful, or even absent from FDA standards.
The objective of this study is to evaluate the accuracy of SpO2 values provided by portable pulse oximeter in COVID patients, in comparison with the reference method. The study will be conducted on a population of adult patients with COVID, hospitalized in the ICU, for whom gasometry sampling is already scheduled in the usual management.
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Detailed Description
The constant monitoring of saturation by the pulse oximeter, the result of three centuries of technological progress, has introduced a true revolution in medical management that now relies on this vital parameter. In many patients, Sars-Cov2 respiratory infection causes arterial hypoxemia, which remains without signs of verbalized respiratory distress, up to a certain point. This phenomenon, called "happy" or "silent" hypoxemia, has a plural pathophysiological basis. Hypoxemia has been shown to be predictively associated with admission to the ICU. Therefore, the question of constant monitoring of oxygenation, practiced on a large scale, at home, in asymptomatic patients or contact cases, arises.
Despite its many advantages (non-invasive, kinetic), the pulse oximeter can be misused in several pathophysiological situations. The reference method for measuring hemoglobin oxygen saturation remains arterial gasometry.
However, a large number of portable pulse oximeter are currently freely available on the market; however, their clinical validation is sometimes questionable or even absent from FDA standards.
The objective of this study is to evaluate the accuracy of SpO2 values provided by portable pulse oximeter in COVID patients, in comparison with the reference method (arterial gasometry). The study will be conducted on a population of adult patients with COVID, hospitalized in the ICU, for whom gasometry sampling is already scheduled in the usual management.
Daily at 6am, during the systematic monitoring, arterial gasometry will be taken (as part of routine care) by the nurse or medical staff on duty. Synchronous measurement of Sp02 given by three portable pulse oximeter (1 measurement point per device, therefore 3 measurements/patient/day for a minimum of 3 days).
Study Type
Study Type
Phase
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
-
Paris, France, 75013
- Service de Pneumologie, Médecine Intensive et Réanimation (SPMIR-R3S) Hôpital Pitié-Salpêtrière
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Description
Inclusion Criteria:
- Patients hospitalized in the intensive care unit for COVID
- Age ≥ 18 years
- Patients or family/trusted person informed and not objecting to participating in the research
Exclusion Criteria:
- Carbon monoxide poisoning
- Homozygous sickle cell disease
- Severe anemia (< 7 g/L) with associated hypoxemia
- Methemoglobinemia
- Sulfhemoglobinemia
- Sepsis, vasoconstrictive drugs
- Patients under guardianship / curators
- Patient under state medical aid
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Other
- Allocation: N/A
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Experimental: Major patient admitted to the ICU for COVID
|
Measurement of oxygen saturation by pulse oximetry once a day, at 6 a.m., for a minimum of 3 days, thanks to three pulse oximeters (1 measurement point per device/day), in parallel with a synchronous sampling, programmed as part of the usual care, of arterial gasometry.
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
To evaluate the accuracy of SpO2 value provided by portable pulse oximeter in COVID patients compared with the reference method.
Time Frame: once a day
|
Measurement of oxygen saturation by pulse oximetry once a day, at 6 a.m., for a minimum of 3 days, using three pulse oximeters (1 measurement point per device/day), in parallel with a synchronous, programmed arterial gasometry sampling Evaluation according to the standards of the SFBC (French Society of Clinical Biology) for the comparison of methods: minimum 40 points per apparatus, evenly distributed over the pathophysiological range of the saturations explored (9, 10): 60-100%.
|
once a day
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Identification of pathophysiological or analytical biases specific to COVID patients.
Time Frame: through study completion, an average of 7 months
|
Data analysis will allow to identify potential biases in the measurement of oxygen saturation in COVID patients
|
through study completion, an average of 7 months
|
Collaborators and Investigators
Sponsor
Sponsor
Study record dates
Study Major Dates
Study Start (Actual)
Study Start
Primary Completion (Actual)
Primary Completion
Study Completion (Actual)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Actual)
First Posted
Study Record Updates
Last Update Posted (Actual)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
Other Study ID Numbers
- APHP210142
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.