Compare the Accuracy of MindTension Blink Eye Reflex Measuring Device to DANTAC Keypoint Electromyography (EMG) System.
A Clinical Study to Compare the Accuracy and Precision of the MindTension Blink Eye Reflex Measuring (Evoked Response)Device to DANTAC Keypoint EMG System
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Detailed Description
Mindtension device measures the orbicularis muscle response to auditory stimuli, the device is backed up by an algorithm that calculates several parameters of startle reflex and startle related indices. In the present study we aim to test the validity, accuracy and precision of MT1 system compared to an FDA cleared device, the Natus DANTEC Keypoint EMG system (K944547).
The aim of this clinical trial is to demonstrate the accuracy and precision of the MT1 device to the FDA cleared Natus DANTEC Keypoint EMG system that is used in the EMG unit.
Participants (n=38) will be evaluated for eligibility of inclusion. As this is paired-group design study, patients will be randomized in a 1:1 ratio to group 1 (Right facial side with MT1 device), or group 2 (Right facial side Natus keypoint device), each patient will be tested using both devices at the same time. The test is 6 minutes in duration. The results will be compared within and between subject analysis.
Safety will be assessed through collection and analysis of adverse events.
Study Type
Study Type
Enrollment (Actual)
Enrollment
Phase
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
-
Haifa, Israel
- Carmel Medical Center
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
- Adult
Accepts Healthy Volunteers
Description
Inclusion Criteria:
- Subject's informed consent.
- Healthy subjects, with no known chronic disease.
- Both genders, age between 18-45.
Exclusion Criteria:
- Any current co-morbid Chronic illness.
- Hearing loss.
- Pregnancy.
- Substance abuse during the last 3 months.
- Lack of cooperation.
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Screening
- Allocation: N/A
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Experimental: EMG evaluated group
Participants are evaluated in both the MT1 and the Natus Dantec Keypoint systems
|
Auditory evoked potentials in parallel systems
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Average maximal peak levels
Time Frame: Day 1
|
The average peak levels (in millivolts) following auditory stimulus.
|
Day 1
|
|
Average peak index
Time Frame: Day 1
|
The average index (in milliseconds) following auditory stimulus.
|
Day 1
|
|
Average eye blink width
Time Frame: Day 1
|
Average eye blink width (in milliseconds) following auditory stimuli.
|
Day 1
|
|
Average inhibition
Time Frame: Day 1
|
Average inhibition (in percent) following paired auditory stimuli.
|
Day 1
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Average eye blink sum
Time Frame: Day 1
|
Average eye blink sum (in millivolts) following auditory stimuli.
|
Day 1
|
Collaborators and Investigators
Sponsor
Sponsor
Collaborators
Collaborators
Investigators
Investigators
- Study Director: Hadas Shatz Azoulay, PhD, MindTension
Study record dates
Study Major Dates
Study Start (Actual)
Study Start
Primary Completion (Actual)
Primary Completion
Study Completion (Actual)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Estimated)
First Posted
Study Record Updates
Last Update Posted (Estimated)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Other Study ID Numbers
Other Study ID Numbers
- MND-03
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
product manufactured in and exported from the U.S.
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.