Improving Neonatal Hip Screening With Artificial Intelligence
Improving Screening for Developmental Dysplasia of the Hip Using Artificial Intelligence Ultrasound Scans in Neonates: A Pilot Study
The goal of this trial is to pilot a portable ultrasound device that uses artificial intelligence to screen for hip dysplasia. Researchers will gather data to understand the feasibility of performing a larger trial in birthing hospitals. It will also seek to collect initial data on how well the scan compares to the standard hip screening performed soon after birth.
Participants will:
- Have the portable ultrasound performed on their baby before they are discharged from hospital
- Have a diagnostic ultrasound performed on their baby at 6-weeks of age
- Complete a short questionnaire about the experience of having the measurement performed on their baby
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Detailed Description
A recent study by our team showed that at The Royal Women's Hospital, Melbourne, screening at birth using the standard combination of neonatal hip examination and risk-based referral for ultrasound failed to detect 52% (n=100) of cases of Developmental Dysplasia of the Hip (DDH) and that 98.5% (n=2,637) of infants undergoing a screening ultrasound, due to perceived increased risk, do not have DDH. Further, this research team replicated in a regional setting at University Hospital, Geelong in Victoria (n=1,207), that 55.6% of cases of DDH were missed, and of those sent for diagnostic ultrasound scans 92.5% did not have DDH (unpublished data). Together, this means many infants are being scanned, but an unacceptably high proportion of cases are still being missed.
Late detected dysplasia is often resistant to conservative treatment. This form of dysplasia is unpredictable in its presentation and may require surgical intervention to obtain a contained and stable joint. Such patients are at a higher risk of developing degenerative hip disease in early adult life and can suffer considerable disability, often failing to reach their full potential. Thus, many initiatives have been taken to improve our current screening programs, including clinical education programs, streamlined access, and incorporation of hip examinations into child health assessments. However, none of these initiatives has effectively reduced the rate of late detection of dysplasia. Despite selective screening protocols being in place, the incidence of late-diagnosed DDH has increased in South Australia, showing a significant rise from 0.22 per 1000 live births (1988-2003) to 0.77 per 1000 live births (2003-2009).
One part of the solution is optimising screening protocols for DDH in existing care models. A possible solution is utilising artificial intelligence to aid in screening decisions. One such new tool is the Exo Iris, a portable ultrasound device supported by real-time AI-augmented analysis to screen for hip dysplasia. Importantly, use of this device does not require extensive training and could be performed by midwives or paediatricians in standard neonate hip examinations. Initial work has shown that AI could successfully identify the standard plane, make measurements, and classify the hip as normal or abnormal. Scans are simple to conduct, add little time to the overall consultation and are non-invasive without the use of ionising radiation. Importantly, non-experts can easily be trained to use the technology; they are cost-effective and can be used in any clinical environment connected to a standard tablet.
Initial Canadian data suggests that DDH detection rates suggests that artificial intelligence (AI) analysis for hip dysplasia are on par with orthopaedic specialists. Of the infants flagged for follow-up there were 6 subsequently referred to specialist clinics after repeat scan and all were treated for DDH (5 harnessed, 1 surgical intervention). Of these the six infants detected, only two presented with well documented risk factors for increased risk of DDH (female sex, Indigenous, breech, family history), which may not have been detected without initial portable ultrasound screening.
Further to this, Retuve is a new open-source software tool that uses AI-analysis to measures standard indices on hip ultrasound images collected from any manufacturer's ultrasound probe, which can help users make hip screening decisions. This platform generates novel imaging parameters beyond current standards that may also be helpful in further understanding undetected late presentations. However, as this is a relatively new tool there has been little research to fully evaluate its performance and its potential utility as a screening tool.
There is clear scope for this technology to revolutionise screening in Australia by reducing the number of cases of DDH missed and the number of costly conventional ultrasound scans; however, there is limited data assessing the feasibility and accuracy of the new AI analysis strategies for DDH screening in the context of neonatal screening. To date, although the Exo Iris and embedded hip AI software is indicated for use in infants 0-6 months there is a dearth of data examining neonates, and no data exists evaluating it in the context of selective neonatal screening.
This study will provide foundational pilot evidence on the accuracy and feasibility of using the Exo Iris probe and an alternate AI software platform as part of standard neonatal screening. In addition to this, long-term follow-up as part of the VicHip parent study will aid in understanding potential parameters associated with missed presentations. To do this, 100 infants will be recruited from Monash Medical Centre, Victoria, Australia and will undergo both the AI-ultrasound scan that will be analysed both by the embedded software and Retuve and results from the AI analyses will be compared to a 6-week full diagnostic ultrasound.
Study Type
Study Type
Enrollment (Estimated)
Enrollment
Phase
Phase
- Not Applicable
Contacts and Locations
Study Contact
Study Contact
- Name: Brian Loh
- Phone Number: +61383416200
- Email: brian.loh@rch.org.au
Study Contact Backup
- Name: Natalie Hyde
- Phone Number: +61383416200
- Email: natalie.hyde@mcri.edu.au
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
- Child
- Adult
- Older Adult
Accepts Healthy Volunteers
Description
Inclusion Criteria:
- Enrolled in the Victorian Hip Dysplasia Registry (VicHip) study
- Infant born at term (≥37 weeks gestation)
- Infant and caregiver admitted to the post-natal ward
- Caregivers indicate they are willing to attend a 6-week ultrasound
- Caregivers can provide a signed and dated informed consent form and is a legally acceptable representative capable of understanding the informed consent document and providing consent on the infant's behalf.
Exclusion Criteria:
- Any known congenital anomalies in the infant precluding examination of the hips
- Any known congenital neuromuscular conditions in the infant
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Screening
- Allocation: N/A
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Experimental: All active participants
All infants will undergo the AI-augmented ultrasound measure
|
The hip ultrasound is performed using a handheld device (Exo Iris) that is a pocket-sized ultrasound probe and is run through an application on an IoS (Apple mobile) operation system.
A real-time algorithm detects and records the anatomical landmarks.
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Feasibility of the AI-ultrasound method as determined by a study-specific questionnaire administered to care givers at Day 1
Time Frame: Day 1
|
Caregiver perspectives will be captured via a study-specific questionnaire administered at Day 1 and this will enable determination of the feasibility of the AI-ultrasound method.
|
Day 1
|
|
Number of infants unable to be scanned with the AI ultrasound
Time Frame: Day 1
|
The proportion of infants unable to be successfully scanned with the AI-ultrasound will be calculated.
|
Day 1
|
|
Reasons for failure to obtain AI ultrasound scan as determined by the performing research assistant
Time Frame: Day 1
|
The reasons why infants were unable to be scanned as determined by the research assistant performing the AI scan will be documented as follows: Unsettled baby, technical failure, body habitus or other.
|
Day 1
|
|
Proportion of infants lost to follow-up between the AI-ultrasound and 6-week (corrected) diagnostic scan
Time Frame: Week 6
|
The proportion of infants that had an initial AI scan at Day 1 but did not return for a scan at Week 6 will be calculated.
|
Week 6
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Specificity of AI-ultrasound device as determined by comparison of the geometric measures (femoral head coverage and alpha angle) and expert reviews between the Day 1 and Week 6 scans
Time Frame: Day 1, Week 6
|
AI-ultrasound results will be compared to 6-week diagnostic ultrasound imaging to calculate specificity ([number of true negative cases detected/(number of false positive cases detected + number of true negatives cases detected] X 100).
Hips will be classified according to the AI-recommendations, "unremarkable" i.e. no follow-up required or "follow-up recommended" to define negative and positive cases; those that receive an "unremarkable" classification will be defined as negative cases, and those that receive a "follow-up recommended" scan will be considered as positive cases for the reliability calculations.
Diagnosis will be defined from the 6-week ultrasound scan and will be defined by both expert conclusion and geometric measures (femoral head coverage and alpha angle).
Calculations will be performed at both the hip and individual level.
|
Day 1, Week 6
|
|
Sensitivity of AI-ultrasound measure determined by comparison of the geometric measures (femoral head coverage and alpha angle) and expert reviews between the Day 1 and Week 6 scans
Time Frame: Day 1, Week 6
|
AI-ultrasound results will be compared to 6-week diagnostic ultrasound imaging to calculate sensitivity ([number of true positive cases detected/(number of true positive cases detected + number of false negative cases detected)] X 100).
Hips will be classified according to the AI-recommendations, "unremarkable" i.e. no follow-up required or "follow-up recommended" to define negative and positive cases; those that receive an "unremarkable" classification will be defined as negative cases, and those that receive a "follow-up recommended" scan will be considered as positive cases for the reliability calculations.
Diagnosis will be defined from the 6-week ultrasound scan and will be defined by both expert conclusion and geometric measures (femoral head coverage and alpha angle).
Calculations will be performed at both the hip and individual level.
|
Day 1, Week 6
|
|
The correlation between the alpha angle degree as reported by the AI-ultrasound analysis and expert analysis of the 6-week diagnostic ultrasound scan
Time Frame: Day 1, Week 6
|
Dependent on the distribution of the data Pearsons or Spearman's Rho correlation coefficients will be reported.
|
Day 1, Week 6
|
|
The correlation between the percentage femoral head coverage as reported by the AI-ultrasound analysis and expert analysis of the 6-week diagnostic ultrasound scan
Time Frame: Day 1, Week 6
|
Dependent on the distribution of the data Pearsons or Spearman's Rho correlation coefficients will be reported.
|
Day 1, Week 6
|
Other Outcome Measures
Other Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Difference in specificity of standard neonatal screening and the AI-ultrasound scan to discriminate for DDH
Time Frame: Day 1
|
The specificity of standard neonatal screening and the AI-ultrasound screening will be compared.
|
Day 1
|
|
Difference in specificity of Retuve and the AI-ultrasound scan to discriminate for DDH
Time Frame: Day 1
|
The specificity of an alternate AI-analysis platform will be compared to the AI-ultrasound scan results.
|
Day 1
|
Collaborators and Investigators
Sponsor
Sponsor
Collaborators
Collaborators
Investigators
Investigators
- Principal Investigator: Brian Loh, MBBS BBiomedSc FRACS FAOrthoA, Murdoch Children's Research Institute & Monash Health
Publications and helpful links
General Publications
- McArthur A, Wichuk S, Burnside S, et al. Retuve: Automated multi-modality analysis of hip dysplasia with open source AI. Software Impacts. 2025/10/01/ 2025;26:100791. doi:https://doi.org/10.1016/j.simpa.2025.100791
- Jaremko JL, Hareendranathan A, Bolouri SES, Frey RF, Dulai S, Bailey AL. AI aided workflow for hip dysplasia screening using ultrasound in primary care clinics. Sci Rep. 2023 Jun 7;13(1):9224. doi: 10.1038/s41598-023-35603-9.
- from 3D ultrasound using artificial intelligence: A two-center multi-year study. Informatics in Medicine Unlocked. 09/01 2022;33:101082. doi:10.1016/j.imu.2022.101082
- Studer K, Williams N, Antoniou G, Gibson C, Scott H, Scheil WK, Foster BK, Cundy PJ. Increase in late diagnosed developmental dysplasia of the hip in South Australia: risk factors, proposed solutions. Med J Aust. 2016 Apr 4;204(6):240. doi: 10.5694/mja15.01082.
- Johnson MD, Kuschel C, Donnan L. Neonatal clinical examination and selective ultrasound screening are not reliable for the early diagnosis of hip dysplasia: A retrospective cohort study. J Paediatr Child Health. 2023 Oct;59(10):1146-1151. doi: 10.1111/jpc.16472. Epub 2023 Aug 7.
Study record dates
Study Major Dates
Study Start (Estimated)
Study Start
Primary Completion (Estimated)
Primary Completion
Study Completion (Estimated)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Actual)
First Posted
Study Record Updates
Last Update Posted (Actual)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
Other Study ID Numbers
- 124710
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
IPD Plan Description
IPD Sharing Time Frame
IPD Sharing Access Criteria
IPD Sharing Supporting Information Type
- STUDY_PROTOCOL
- ICF
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
product manufactured in and exported from the U.S.
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.