Day and Night Closed-loop in Young People With Type 1 Diabetes (DAN05)

February 1, 2021 updated by: Jaeb Center for Health Research

An Open-label, Multi-centre, Randomized, Single-period, Parallel Study to Assess the Efficacy, Safety and Utility of 6 Month Day-and-night Automated Closed-loop Insulin Delivery Under Free Living Conditions Compared to Insulin Pump Therapy in Children and Adolescents With Type 1 Diabetes

The main study objective is to determine whether 24/7 automated closed-loop glucose control combined with low glucose feature will improve glucose control as measured by HbA1c.

This is an open-label, multi-centre, multi-national, single-period, randomised, parallel group design study, involving a 6 month period of home study during which day and night glucose levels will be controlled either by a closed-loop system combined with low glucose feature (intervention group) or by insulin pump therapy alone (control group).

It is expected that a total of up to 150 subjects (aiming for 130 randomised subjects) with type 1 diabetes will be recruited through paediatric outpatient diabetes clinics of the investigation centres. Participants will all be on subcutaneous insulin pump therapy.

Subjects in the intervention group will have proven competencies both in the use of the study insulin pump and the study continuous glucose monitoring (CGM) device, and will receive appropriate training in the safe use of closed-loop insulin delivery system and low glucose feature. All subjects will have regular contact with the study team during the home study phase including 24/7 telephone support. The primary outcome is between group differences in HbA1c levels at 6 months post study arm initiation. Secondary outcomes are the time spent in the glucose target (3.9 to 10.0mmol/l; 70 to 180mg/dl), time spent with glucose levels above and below target, as recorded by CGM, and other CGM-based metrics. Safety evaluation comprises assessment of the frequency of severe hypoglycaemic episodes and diabetic ketoacidosis (DKA).

Study Overview

Detailed Description

  • Purpose of the study: To determine whether 24/7 automated closed-loop glucose control will improve glucose control as measured by glycated haemoglobin and reduce the burden of hypoglycaemia compared to insulin pump therapy alone.
  • Study Objectives:

    1. EFFICACY: The objective is to assess efficacy of day and night automated closed-loop glucose control in improving glucose control as measured by glycated haemoglobin, as compared to insulin pump therapy alone.
    2. SAFETY: The objective is to evaluate the safety of day and night automated closed-loop glucose control, in terms of episodes of severe hypoglycaemia and other adverse events.
    3. UTILITY: The objective is to determine the frequency and duration of the use of the automated closed-loop system.
    4. HUMAN FACTORS: The objective is to assess cognitive, emotional, and behavioural characteristics of participating subjects and family members and their response to the closed-loop system and clinical trial using validated surveys and focus groups.
    5. HEALTH ECONOMICS: The objective is to perform a cost utility analysis to inform reimbursement decision-making.
  • Study Design: An open-label, multi-centre, randomised, single-period parallel study, contrasting day-and-night automated closed-loop glucose control with insulin pump therapy alone.
  • Population: 130 participants randomised (equal proportion of those aged 6 to 12 years and 13 to 18 years, a minimum quota of 25% participants with baseline HbA1c >8.5%)
  • Maximum duration of study for a subject: 8 months
  • Recruitment: The subjects will be recruited through the pediatric outpatient clinics at each center.
  • Consent: Written consent / assent will be obtained from participants and/or guardians according to REC / IRB requirements
  • Screening Assessments: Eligible participants will undergo a screening evaluation where blood samples for full blood count, liver, thyroid function and anti-transglutaminase antibodies (with IgA levels if not done within previous 12 months) will be taken. Non-hypoglycaemia C-peptide, glucose and HbA1c will also be measured, and a urine pregnancy test in females of child-bearing potential will be performed.

Surveys investigating participants' quality of life, psychosocial and cognitive functioning, and response to their current treatment will be distributed.

Participants will be fitted with a blinded continuous glucose monitoring (CGM) device to assess baseline glycaemic control. Instructions on how to safely use, remove and send back the device will be provided.

  • Study Training: Training sessions on the use of study CGM, insulin pump (and closed loop system for those randomized to be intervention group) will be provided by the research team. Training session on the use of real-time CGM and on how to interpret real-time and retrospective stored data will be provided to all subjects / carers using written material.
  • Run-In Period: During a 1-2 week run-in period, subjects will continue using their own insulin pump. Data obtained from blinded CGM and pump downloads may be utilised for therapy adjustment.
  • Competency Assessment: Competency on the use of study insulin pump and study CGM will be evaluated using a competency assessment tool developed by the research team. Further training may be delivered as required.
  • Randomization: Eligible subjects will be randomised using randomisation software to the use of real-time CGM and low glucose feature combined with day and night closed-loop or to conventional insulin pump therapy alone.

A blood sample will be taken for the measurement of HbA1c and a urine pregnancy test in females of child-bearing potential. A blood sample for centralised analysis of HbA1c will be taken if screening and randomisation are >28 days apart.

  1. Automated day and night closed-loop insulin delivery (intervention arm) combined with low glucose feature (interventional arm) - Participants in the closed-loop group will receive additional training sessions following randomisation covering the use of the study insulin pump and real-time CGM, prior to starting closed-loop insulin delivery.

    Once confident with the use of the study pump and CGM system, participants will receive training required for safe and effective use of the closed-loop system approximately 2-4 weeks after randomisation. During this 2-4 hour session participants will operate the system under the supervision of the clinical team. Competency on the use of closed-loop system will be evaluated.

    Thereafter, participants are expected to use closed-loop for 6 months without direct real-time remote monitoring.

  2. Insulin pump therapy (control arm) - Refresher training on key aspects of insulin pump therapy will be provided.

Subjects will continue using their own insulin pump for 6 months.

  • 3-month and 6 month assessments:

    1. A blood sample will be taken for measurement of HbA1c and a urine pregnancy test in females of child-bearing potential.
    2. Validated surveys evaluating the impact of the devices employed on quality of life, psychosocial and cognitive functioning, diabetes management and treatment satisfaction will be completed.
    3. Participants of both study arms will be fitted with blinded CGM systems at the end of each follow up visit. The sensors will be worn at home for up to 14 days and will be sent back to the research team.

    6 months only: Subjects/guardians will be invited to join follow-up focus groups to gather feedback and reactions to their current treatment (closed-loop or insulin pump), the clinical trial, and quality of life changes.

  • Study Contacts: In between study visits, participants will be contacted by the study team (email or phone) once monthly in order to record any adverse events, device deficiencies, and changes in insulin settings, other medical conditions and/or medication.

In case of any technical device or problems related to diabetes management such as hypo- or hyperglycaemia, subjects will be able to contact a 24-hour telephone helpline to the local research team at any time. The local research team will have access to central 24 hour advice on technical issues.

--Procedures for safety monitoring during trial: Standard operating procedures for monitoring and reporting of all adverse events will be in place, including serious adverse events (SAE), serious adverse device effects (SADE) and specific adverse events (AE) such as severe hypoglycaemia.

Subjects will be asked to test and record blood ketones if their finger prick glucose is > 16.7mmol/l (300mg/dl) upon awakening, >300 for more than 1 hour, or >22.2mmo/l (400mg/dL) at any time as part of the safety assessment for DKA.

A data safety and monitoring board (DSMB) will be informed of all serious adverse events and any unanticipated serious adverse device effects that occur during the study and will review compiled adverse event data at periodic intervals.

--Criteria for withdrawal of patients on safety grounds: A subject, parent, or guardian may terminate participation in the study at any time without necessarily giving a reason and without any personal disadvantage. An investigator can stop the closed-loop intervention after consideration of the benefit/risk ratio. Possible reasons are:

  1. Serious adverse events
  2. Significant protocol violation or non-compliance
  3. Failure to satisfy competency assessment
  4. Decision by the investigator, or the sponsor, that termination is in the subject's best medical interest
  5. Pregnancy, planned pregnancy, or breast feeding
  6. Allergic reaction to insulin

Efforts will be made to retain subjects in follow up for the final primary outcome assessment even if the intervention is discontinued, unless the investigator believes that it will be harmful for the subject to continue in the trial.

Study Type

Interventional

Enrollment (Actual)

131

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

      • Liverpool, United Kingdom
        • Alder Hey Children's NHS Foundation Trust
      • Oxford, United Kingdom
        • Oxford Children's Hospital
    • Cambridgeshire County
      • Cambridge, Cambridgeshire County, United Kingdom, CB2 0QQ
        • University of Cambridge
    • England
      • Nottingham, England, United Kingdom, NG5 1PB
        • Nottingham Children's Hospital
      • Southampton, England, United Kingdom, SO16 6YD
        • Southampton Children's Hospital
    • West Yorkshire
      • Leeds, West Yorkshire, United Kingdom, LS9 7TF
        • The Leeds Teaching Hospitals Nhs Trust
    • California
      • Palo Alto, California, United States, 95032
        • Stanford University
    • Colorado
      • Aurora, Colorado, United States, 80045
        • University of Colorado Denver School of Medicine Barbara Davis Center
    • Connecticut
      • Hartford, Connecticut, United States, 06520
        • Yale University
    • Florida
      • Jacksonville, Florida, United States, 32207
        • Nemours Children's Health System
    • Indiana
      • Indianapolis, Indiana, United States, 43202
        • Indiana University

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

2 years to 14 years (Child, Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

  1. Age ≥6 and <19 years
  2. Type 1 diabetes as defined by WHO (51) for at least 1 year [WHO definition: 'The aetiological type named type 1 encompasses the majority of cases with are primarily due to beta-cell destruction, and are prone to ketoacidosis. Type 1 includes those cases attributable to an autoimmune process, as well as those with beta-cell destruction for which neither an aetiology nor a pathogenesis is known (idiopathic). It does not include those forms of beta-cell destruction or failure to which specific causes can be assigned (e.g. cystic fibrosis, mitochondrial defects, etc.).']
  3. Use of an insulin pump for at least 3 months, with good knowledge of insulin self-adjustment by subject or caregiver as judged by the investigator
  4. Using U-100 rapid acting insulin analogues insulin Aspart or Lispro only
  5. Willing to perform regular finger-prick blood glucose monitoring, with at least 4 blood glucose measurements per day day
  6. Screening HbA1c ≥ 7.0% (53 mmol/mol) and ≤10 % (86mmol/mol) based on analysis from local laboratory
  7. Literate in English
  8. Willing to wear glucose sensor
  9. Willing to wear closed-loop system at home
  10. Willing to follow study specific instructions
  11. Willing to upload pump and CGM data at regular intervals
  12. Access to WiFi.
  13. Lives with someone who is trained to administer intramuscular glucagon and is able to seek emergency assistance

Exclusion Criteria:

  1. Living alone
  2. Current use of any closed-loop system
  3. Any other physical or psychological disease likely to interfere with the normal conduct of the study and interpretation of the study results as judged by the investigator
  4. Untreated coeliac disease, adrenal insufficiency, or untreated thyroid disease
  5. Current treatment with drugs known to interfere with glucose metabolism, e.g. systemic corticosteroids, non-selective beta-blockers and MAO inhibitors etc.
  6. Known or suspected allergy to insulin
  7. Clinically significant nephropathy (eGFR < 45ml/min) or on dialysis, neuropathy or active retinopathy (defined as presence of maculopathy or proliferative changes) as judged by the investigator
  8. Recurrent incidents of severe hypoglycaemia (>1 episode) during the previous 6 months (adolescents: severe hypoglycaemia is defined as an event requiring assistance of another person to actively administer carbohydrates, glucagon, or take other corrective actions including episodes of hypoglycaemia severe enough to cause unconsciousness, seizures or attendance at hospital; children: severe hypoglycaemia is defined as an event associated with a seizure or loss of consciousness)
  9. Recurrent incidents of diabetic ketoacidosis (>1 episode) during previous 6 months
  10. Unwilling to avoid regular use of acetaminophen
  11. Lack of reliable telephone facility for contact
  12. Total daily insulin dose ≥ 2 IU/kg/day
  13. Total daily insulin dose < 15 IU/day
  14. Pregnancy, planned pregnancy, or breast feeding
  15. Severe visual impairment
  16. Severe hearing impairment
  17. Seizure disorder
  18. Medically documented allergy towards the adhesive (glue) of plasters or unable to tolerate tape adhesive in the area of sensor placement
  19. Serious skin diseases (e.g. psoriasis vulgaris, bacterial skin diseases) located at places of the body, which potentially are possible to be used for localisation of the glucose sensor)
  20. Illicit drugs abuse
  21. Subject is currently abusing prescription drugs
  22. Alcohol abuse
  23. Use of pramlintide (Symlin), or other non-insulin glucose lowering agents including sulphonylureas, biguanides, DPP4-Inhibitors, , GLP-1 analogues, SGLT-1/ 2 inhibitors at time of screening
  24. Shift work with working hours between 10pm and 8am
  25. Sickle cell disease, haemoglobinopathy, or has received red blood cell transfusion or erythropoietin within 3 months prior to time of screening
  26. Eating disorder such as anorexia or bulimia
  27. Employed by Medtronic Diabetes or with immediate family members employed by Medtronic Diabetes

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Treatment
  • Allocation: Randomized
  • Interventional Model: Parallel Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: 24/7 closed loop insulin delivery
The study system includes (1) a CGM that measures glucose levels, (2) a computer program on a smartphone that determines how much insulin is needed, and (3) an insulin pump that delivers the insulin. The name of this closed-loop system used in the US is FlorenceM (Medtronic 640G pump and Guardian3 sensor). The name of this closed-loop system in the UK is FlorenceX (DANA pump and Dexcom sensor). Half of the individuals taking part in the study will use the closed-loop study system for 6 months.

The automated closed loop system (FlorenceM in US) will consist of:

Next generation sensor augmented Medtronic insulin pump 640G (Medtronic Minimed, CA, USA) incorporating the Medtronic Guardian3 CGM and glucose suspend feature.

The automated closed loop system (FlorenceX in UK) will consist of:

The DANA Diabecare R insulin pump (Sooil Development, Korea) incorporating the Dexcom G6 CGM.

An Android smartphone containing the Cambridge model predictive algorithm and communicating wirelessly with the insulin pump using a proprietary translator device.

Active Comparator: Insulin pump therapy
Half of the Subjects will continue using their own insulin pump for 6 months.
Subjects will continue using their own insulin pump for 6 months.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
The primary outcome is the centralised measurement of glycated haemoglobin (HbA1c) at 6 months.
Time Frame: HbA1c will be taken at baseline, 3 and 6 months
The objective is to assess efficacy of day and night automated closed-loop glucose control combined with low glucose feature in improving HbA1c, as compared with insulin pump therapy alone.
HbA1c will be taken at baseline, 3 and 6 months

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Time spent in the target glucose range (3.9 to 10mmol/l) (70 to 180mg/dl)
Time Frame: 6 months
Secondary endpoints regarding glucose levels will be based on sensor glucose data.
6 months
Time spent below target glucose (3.9mmol/l)(70mg/dl)
Time Frame: 6 months
Secondary endpoints regarding glucose levels will be based on sensor glucose data.
6 months
Time spent above target glucose (10.0 mmol/l) (180 mg/dl)
Time Frame: 6 months
Secondary endpoints regarding glucose levels will be based on sensor glucose data.
6 months
Mean and standard deviation or percentiles sensor glucose
Time Frame: 6 months
Secondary endpoints regarding glucose levels will be based on sensor glucose data.
6 months
Coefficient of variation of glucose levels
Time Frame: 6 months
Secondary endpoints regarding glucose levels will be based on sensor glucose data.
6 months
Time with glucose levels < 3.5 mmol/l (63 mg/dl)
Time Frame: 6 months
Secondary endpoints regarding glucose levels will be based on sensor glucose data.
6 months
Time with glucose levels <3.0 mmol/l (54 mg/dl)
Time Frame: 6 months
Secondary endpoints regarding glucose levels will be based on sensor glucose data.
6 months
Time with glucose levels in significant hyperglycaemia (glucose levels > 16.7 mmol/l) (300mg/dl)
Time Frame: 6 months
Secondary endpoints regarding glucose levels will be based on sensor glucose data.
6 months
Changes in total basal and bolus insulin dose
Time Frame: 6 months
Secondary endpoints regarding glucose levels will be based on CRF and insulin pump data.
6 months
AUC of glucose below 3.5mmol/l (63mg/dl)
Time Frame: 6 months
Secondary endpoints regarding glucose levels will be based on sensor glucose data.
6 months
AUC glucose above 10.0mmol/L (180mg/dL)
Time Frame: 6 months
Secondary endpoints regarding glucose levels will be based on sensor glucose data.
6 months
HbA1c <7.0%, HbA1c <7.5%, Relative reduction ≥10% from baseline. o Absolute reduction ≥0.5% from baseline o Absolute reduction ≥1% from baseline o Absolute reduction ≥1% from baseline or HbA1c <7.0%
Time Frame: 6 months
Binary metrics for HbA1c
6 months

Other Outcome Measures

Outcome Measure
Measure Description
Time Frame
Safety Evaluation
Time Frame: 6 months
Frequency of severe hypoglycaemic episodes as defined by American Diabetes Association (adolescents), and International Society for Pediatric and Adolescent Diabetes (children), frequency of diabetic ketoacidosis (DKA), frequency of severe hyperglycaemia (>16.7 mmol/l)(>300mg/dl) with significant ketosis (plasma ketones >0.6mmol/l) and nature and severity of other adverse events
6 months
Utility evaluation
Time Frame: 6 months
Assessment of the frequency and duration of use of the closed-loop system
6 months
Human Factors Assessment
Time Frame: 6 months
Cognitive, emotional, and behavioural characteristics of participating subjects and family members and their response to the closed-loop system and clinical trial will be assessed gathering both quantitative (validated surveys) and qualitative data (focus groups)
6 months
Health Economic Evaluation
Time Frame: 6 months
Cost utility analysis on the benefits of closed loop insulin delivery to inform reimbursement decision-making
6 months

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Principal Investigator: Nelly Mauras, MD, Nemours Children's Health System
  • Study Chair: Roman Hovorka, PhD, University of Cambridge
  • Principal Investigator: Ajay Thankamony, MD, University of Cambridge
  • Principal Investigator: Fiona Campbell, MD, The Leeds Teaching Hospitals Nhs Trust
  • Principal Investigator: Stuart Weinzimer, MD, Yale University
  • Principal Investigator: Linda DiMeglio, MD, Indiana University
  • Principal Investigator: Paul Wadwa, MD, University of Colorado, Denver
  • Principal Investigator: Korey Hood, PhD, Stanford University
  • Principal Investigator: Dana Goldman, PhD, University of Southern California
  • Principal Investigator: Nikki C Davis, MD, Southampton Children's Hospital
  • Principal Investigator: Louise Denvir, MD, Nottingham Children's Hospital
  • Principal Investigator: Rachel Besser, MD, Oxford Children's Hospital
  • Principal Investigator: Atrayee Ghatak, MD, Alder Hey Children's NHS Foundation Trust

Publications and helpful links

The person responsible for entering information about the study voluntarily provides these publications. These may be about anything related to the study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

May 12, 2017

Primary Completion (Actual)

August 27, 2020

Study Completion (Actual)

August 27, 2020

Study Registration Dates

First Submitted

October 4, 2016

First Submitted That Met QC Criteria

October 4, 2016

First Posted (Estimate)

October 5, 2016

Study Record Updates

Last Update Posted (Actual)

February 2, 2021

Last Update Submitted That Met QC Criteria

February 1, 2021

Last Verified

February 1, 2021

More Information

Terms related to this study

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

No

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

Yes

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Glucose Metabolism Disorders

Clinical Trials on FlorenceM (US) and FlorenceX (UK)

Subscribe