Computational Design, Fabrication, and Evaluation of Optimized Patient-Specific Transtibial Prosthetic Sockets
Study Overview
Status
Conditions
Intervention / Treatment
Detailed Description
Title: Computational Design, Fabrication, and Evaluation of Optimized Patient-Specific Transtibial Prosthetic Sockets
Principle Investigator: Dr. Hugh Herr
Background: The overall goal of this application is to further develop and clinically assess a computational and data-driven design and manufacturing framework for mechanical interfaces that quantitatively produces transtibial prosthetic sockets in a faster and more cost-effective way than conventional processes. Traditionally, prosthetic socket production has been a craft activity, based primarily on the experience of the prosthetist. Even with advances in computer-aided design and computer-aided manufacturing (CAD/CAM), the design process remains manual. The manual nature of the process means it is non-repeatable and currently largely non-data-driven, and quantitative data is either not obtained or insufficiently employed. Furthermore, discomfort, skin problems and pressure ulcer formation remain prevalent. Through the proposed computational modeling framework, a repeatable, data-driven and patient-specific design process is made available which is based on scientific rationale.
Objective/hypothesis: The main hypothesis of this proposal is that a socket, designed using the novel computational design framework, is equivalent to, or better than, a conventional socket (designed by a prosthetist) in terms of: 1) gait symmetry, 2) skin contact pressures, 3) skin surface temperature, and 4) comfort as evaluated from a questionnaire. Our hypothesis is supported by the presented pilot data which shows reduced or equivalent skin contact pressures and subject reported comfort levels for several critical anatomical regions.
Specific Aims: 1) Subject-specific biomechanical modeling for N=18 subjects, 2) Computational design and fabrication of sockets for N=18 subjects, and 3) Clinical evaluation of novel sockets for N=18 subjects.
Study Design: A cohort of 18 subjects will be recruited for this study. MRI or CT data will be recorded for all subjects. Through image segmentation geometrically accurate 3D finite element analysis (FEA) models will be constructed. The resulting predictive FEA models will then be used in a novel, data-driven, and automated computational design framework for prosthetic sockets, to design prosthetic sockets for all subjects. The framework optimizes the socket designs, as assessed by skin contact pressures and internal tissue strain, through iterative adjustment of the virtual tests sockets. Final designs are subsequently 3D printed. To evaluate the prosthetic sockets with each of the subjects each subject will do a standing and walking exercise using their conventional sockets or the novel sockets. Meanwhile skin contact pressures and temperatures are recorded, gait symmetry data is collecetd, and socket comfort is assessed using a questionnaire. Together this data provides a quantitative and qualitative evaluation and comparison of the novel and conventional sockets.
Study Type
Enrollment (Actual)
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
Jalisco
-
Nuevo Mexico, Jalisco, Mexico, 45201
- Tecnologico de Monterrey, Biomechatronics Lab
-
-
-
-
Massachusetts
-
Cambridge, Massachusetts, United States, 02142
- Massachusetts Institute of Technology, Media Lab
-
-
Participation Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Description
Inclusion Criteria:
- Age: 18-64 years old
- Amputation type: Transtibial amputation (bilateral or unilateral) which took place >1 year prior to study
- Activity or K-level: At least K3
- Socket quality: The subject's conventional socket(s) should be deemed of high quality and comfortable
- MRI safety: Subjects should be able to undergo MRI
Exclusion Criteria:
-
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Basic Science
- Allocation: N/A
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Arms and Interventions
Participant Group / Arm |
Intervention / Treatment |
|---|---|
|
Experimental: Prosthetic socket evaluation
Intervention: Prosthetic socket for transtibial amputee.
A subject's conventional prosthetic socket is compared to a novel prosthetic socket designed as part of the research.
For standing and walking exercises the following will be assessed.
1) local skin contact pressures, 2) metabolic power, 3) gait parameters (symmetry indices for joint angles, positions, torques, and also ground reaction forces), and 4) the socket evaluation questionnaire.
|
To wear a prosthetic socket for evaluation of comfort and biomechanical metrics such as skin loading conditions.
|
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Local skin pressure while wearing a socket
Time Frame: Assessed once during (clinical) socket comfort evaluation phase. Measurements are for standing and walking exercises for the conventional and novel socket.
|
Local skin contact pressures (in Newton per square meter) are assessed using pressure sensors positioned on the skin surface.
Local pressures relate to comfort which is the topic of investigation.
These measurements are performed during the (clinical) socket comfort evaluation phase which is typically 3 months after the initial data acquisition phase for each subject.
Measurements take place once and in a single session.
As such only a single time point is investigated (the time point of the experiment is not a relevant or investigated parameter).
The data is compared across subjects and socket types.
|
Assessed once during (clinical) socket comfort evaluation phase. Measurements are for standing and walking exercises for the conventional and novel socket.
|
|
Skin surface temperature difference
Time Frame: Assessed once during (clinical) socket comfort evaluation phase. Measurements are for standing and walking exercises for the conventional and novel socket.
|
Skin surface temperatures (in Celsius) are assessed using a Flir thermal camera, with images taken immediately post-kinematic walking trial.
From these measurements the average residuum skin surface temperature is assessed for 4 different view directions: anterior, posterior, medial, and lateral.
Measurements take place once and in a single session.
As such only a single time point is investigated (the time point of the experiment is not a relevant or investigated parameter).
The data is compared across subjects and socket types.
|
Assessed once during (clinical) socket comfort evaluation phase. Measurements are for standing and walking exercises for the conventional and novel socket.
|
|
Gait parameter asymmetry
Time Frame: Assessed once during (clinical) socket comfort evaluation phase. Measurements are for standing and walking exercises for the conventional and novel socket.
|
Socket fit (and discomfort) influences musculoskeletal behavior and may alter gait patterns.
Subjects are equipped with markers tracked (units of mm) using a motion capture system (T40s, Vicon Motion Systems Ltd, Oxford, UK).
Ground reaction forces (units Newton) are also recorded.
From these measurements gait symmetry indices for: joint angles, positions, torques, and ground reaction forces, are computed.
These measurements are performed during the (clinical) socket comfort evaluation phase which is typically 3 months after the initial data acquisition phase for each subject.
Measurements take place once and in a single session.
As such only a single time point is investigated (the time point of the experiment is not a relevant or investigated parameter).
The data is compared across subjects and socket types.
|
Assessed once during (clinical) socket comfort evaluation phase. Measurements are for standing and walking exercises for the conventional and novel socket.
|
|
Socket evaluation questionnaire
Time Frame: Assessed once during (clinical) socket comfort evaluation phase. Measurements are for sitting, standing and walking exercises for the conventional and novel socket.
|
Socket fit and discomfort is assessed using a questionnaire created to document patient experiences and preferences while using prosthetic interfaces (liners, sockets) both over a longitudinal timeline and on an immediate basis.
After establishing subject baselines for socket use, patients complete a section that instructs them to evaluate the novel prosthetic socket provided against their conventional prosthetic socket.
Subjects complete the questionnaire via a HIPAA compliant surveying instrument on Redcap, and questions are tailored to minimize bias through adoption of verified metrics (such as those from the TAPES-R instrument) or by the use of sliding scales that eliminate the need for constrained integer number input.
The data is analyzed by subject, across subjects as a whole, and socket types.
|
Assessed once during (clinical) socket comfort evaluation phase. Measurements are for sitting, standing and walking exercises for the conventional and novel socket.
|
Collaborators and Investigators
Investigators
- Principal Investigator: Hugh M Herr, PhD, Massachusetts Institute of Technology, Media Lab
Study record dates
Study Major Dates
Study Start (Actual)
Primary Completion (Actual)
Study Completion (Actual)
Study Registration Dates
First Submitted
First Submitted That Met QC Criteria
First Posted (Actual)
Study Record Updates
Last Update Posted (Actual)
Last Update Submitted That Met QC Criteria
Last Verified
More Information
Terms related to this study
Keywords
Other Study ID Numbers
- 1101004280
- R01EB024531-01 (U.S. NIH Grant/Contract)
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Amputation Stump
-
Brooke Army Medical CenterUniversity of WashingtonCompleted
-
Istituto Nazionale Assicurazione contro gli Infortuni...Scuola Superiore Sant'Anna di PisaCompleted
-
Helsinki University Central HospitalCompletedAmputation StumpFinland
-
Gaziler Physical Medicine and Rehabilitation Education...CompletedLower Extremity Amputation | Residual Limb Pain | Stump NeuromaTurkey (Türkiye)
-
Jason HighsmithFlorida High Tech Corridor Council; Otto Bock HealthcareCompletedOther and Unspecified Complications of Amputation StumpUnited States
-
London Health Sciences CentreUnknownInfection | Amputation Wound | Amputation Stump Complication | Amputation Stump; Infection
-
Liberating Technologies, Inc.Vivonics, Inc.CompletedLower Limb Amputation Below Knee (Injury) | Amputation | Prosthesis User | Amputation; Traumatic, Leg, Lower | Limb; Absence, Congenital, Lower | Amputation StumpUnited States
-
Baskent UniversityAnkara Yildirim Beyazıt University; Gulhane School of MedicineCompletedAmputation | Amputation Stump | Amputees | Amputation of Knee | VascularisationTurkey
-
Liberating Technologies, Inc.Vivonics, Inc.TerminatedAmputation | Prosthesis | Prosthesis User | Amputation Stump | Amputees | Lower Limb DeformityUnited States
-
Centre hospitalier de l'Université de Montréal...Not yet recruitingAmputation Stump | Amputation Neuroma | Amputation, Traumatic/Surgery | Amputation, Surgical
Clinical Trials on Prosthetic socket for transtibial amputee
-
Ohio Willow WoodUnited States Department of DefenseCompleted
-
Eunice Kennedy Shriver National Institute of Child...Unknown
-
University of Wisconsin, MilwaukeeUnknown
-
Ohio Willow WoodOhio State UniversityUnknownLower Limb AmputationUnited States
-
University of SouthamptonRadii Devices Ltd; Opcare LtdCompletedTranstibial Amputation | ProstheticUnited Kingdom
-
VA Office of Research and DevelopmentRecruitingTranstibial Amputation | Transfemoral AmputationUnited States
-
VA Office of Research and DevelopmentUniversity of TexasCompletedDiabetes Complications | AmputationUnited States
-
VA Office of Research and DevelopmentNot yet recruiting
-
Ohio Willow WoodOhio State UniversityUnknown
-
University of South FloridaSouthern Bone & Joint Specialists, P.A.CompletedSocket Designs of Transfemoral Amputees.United States