Belun SpO2 Accuracy Comparison to Arterial Blood CO-Oximetry

April 30, 2021 updated by: Belun Technology Company Limited
This study was a comparative, single-center, non-randomized study conducted to evaluate the SpO2 accuracy per standards and guidelines identified above for SpO2 accuracy for pulse oximetry equipment over the range of 70-100% SaO2 under non-motion conditions. Arterial blood sampling measured by functional SaO2 CO-Oximetry, was used as the basis for comparison. Testing was conducted under normal office environment conditions.

Study Overview

Status

Completed

Conditions

Intervention / Treatment

Detailed Description

Pulse oximetry monitoring is considered a standard physiological measurement and is used by clinicians in everyday situations to estimate arterial oxygen saturation. A pulse oximeter is a device that measures the oxygen saturation of arterial blood non-invasively. In general, pulse oximeters use two wavelength absorption spectrophotometry to measure oxygen saturation based on the amount of reflected or scattered radiation. The wavelengths are selected to provide the best separation of absorbencies of oxy-hemoglobin (O2Hb) and deoxy-hemoglobin (HHb) states. The ratio of the two absorbencies is used to calculate the oxygen saturation (SpO2) value. Because an arterial sample of blood is not required to make the measurement, the pulse oximeter can provide non-invasive real time information. The clinical use of pulse oximeters has reduced the frequency and necessity of invasive arterial blood sampling, and has helped to improve patient safety by providing information to clinicians about patients' oxygenation status.

The purpose of this study was to evaluate the SpO2 accuracy performance of the Belun Ring Oximeter placed on the index fingers during non-motion conditions over the range of 70-100% SaO2 to arterial blood samples assessed by CO-Oximetry for an SpO2 validation.

The primary goal, was to show the SpO2 accuracy performance for the investigational device below. As a secondary goal, the pulse rate accuracy performance for the Belun ring was calculated.

It was expected that the Accuracy Root Mean Square (ARMS) performance would meet the required specification of ARMS 3.5% or lower in non-motion conditions for the range of 70 - 100% SaO2 thereby demonstrating an acceptable SpO2 accuracy performance specification. For the pulse rate accuracy, the desired goal was to have an accuracy performance of an ARMS of 3.

No risks or adverse device effects were expected. There were no contraindications for use in the proposed study / study population.

The study was conducted in accordance to the code of federal regulations for non-significant medical device studies and applicable ISO 14155 (2nd edition 2011-02-01), applicable sections of ISO 80601-2-61, second edition 2017-12, Corrected version 2018-02, and Pulse Oximeters - Premarket Notifications Submissions [510(k)s] Guidance For Industry and Food and Drug Administration Staff (issued: March 4, 2013).

Study Type

Observational

Enrollment (Actual)

10

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • Colorado
      • Louisville, Colorado, United States, 80027
        • Clinimark, LLC

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years to 50 years (Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Sampling Method

Non-Probability Sample

Study Population

The study population will include 10 to 15 healthy non-smoking (or has refrained from smoking for 2 days) competent adults, ages 18 to 50 years. The subject selection will be an equitable distribution of males and females of any race with varying skin tones including at least 2 darkly pigmented subjects or 15% of the subject pool, whichever is larger.

Description

Inclusion Criteria:

  • 10-15 Adults with a minimum of 3 males and a minimum of 3 females, with the balance made up of either
  • Subject must have the ability to understand and provide written informed consent
  • Subject is 18 to 50 years of age
  • Subject must be willing and able to comply with study procedures and duration
  • Subject is a non-smoker or who has not smoked within 2 days prior to the study

Exclusion Criteria:

  • Subject is considered as being morbidly obese (defined as BMI >39.5)
  • Compromised circulation, injury, or physical malformation of fingers, wrist, hands, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study. (Note: Certain malformations may still allow subjects to participate if the condition is noted and would not affect the particular sites utilized.)
  • Female subjects that are actively trying to get pregnant or are pregnant (confirmed by positive urine pregnancy test unless the subject is known to be not of child-bearing potential).
  • Smoker Subjects who have refrained will be screened for COHb levels >3% as assessed with a Masimo Radical 7 (Rainbow)
  • Subjects with known respiratory conditions such as: (self-reported)

    • uncontrolled / severe asthma,
    • flu,
    • pneumonia / bronchitis,
    • shortness of breath / respiratory distress,
    • unresolved respiratory or lung surgery with continued indications of health issues,
    • emphysema, COPD, lung disease
  • Subjects with known heart or cardiovascular conditions such as: (self-reported, except for blood pressure and ECG review)

    • hypertension: systolic >140mmHg, Diastolic >90mmHg on 3 consecutive readings (reviewed during health screen).
    • have had cardiovascular surgery
    • Chest pain (angina)
    • heart rhythms other than a normal sinus rhythm or with respiratory sinus arrhythmia (reviewed during health screen)
    • previous heart attack
    • blocked artery
    • unexplained shortness of breath
    • congestive heart failure (CHF)
    • history of stroke
    • transient ischemic attack
    • carotid artery disease
    • myocardial ischemia
    • myocardial infarction
    • cardiomyopathy
  • Self-reported health conditions as identified in the Health Assessment Form (self-reported)

    • diabetes,
    • uncontrolled thyroid disease,
    • kidney disease / chronic renal impairment,
    • history of seizures (except childhood febrile seizures),
    • epilepsy,
    • history of unexplained syncope,
    • recent history of frequent migraine headaches,
    • recent symptomatic head injury (within the last 2 months)
    • cancer / chemotherapy
  • Subjects with known clotting disorders (self-reported)

    • history of bleeding disorders or personal history of prolonged bleeding from injury
    • history of blood clots
    • hemophilia
    • current use of blood thinner: prescription or daily use of aspirin
  • Subjects with severe contact allergies to standard adhesives, latex or other materials found in pulse oximetry sensors, ECG electrodes, respiration monitor electrodes or other medical sensors (self-reported)
  • Subjects with severe allergies to iodine (only applicable if iodine is used)
  • Subjects with severe allergies to lidocaine (or similar pharmacological agents, e.g. Novocain)
  • Failure of the Perfusion Index Ulnar/Ulnar+Radial Ratio test (Ratio < 0.4)
  • Unwillingness or inability to remove colored nail polish from test digits.
  • Other known health condition, should be considered upon disclosure in health assessment form

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Observational Models: Case-Only
  • Time Perspectives: Prospective

Cohorts and Interventions

Group / Cohort
Intervention / Treatment
healthy adult volunteer
Subject is 18 to 50 years of age. Subject is a non-smoker or who has not smoked within 2 days prior to the study.
The investigational device will be placed on the finger of the left or right hand per the instructions for use. Simultaneous data collection will be set up for the system under test and control oximeter.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Collect SpO2 data for accuracy statistical analysis by 10 subjects
Time Frame: 3 days
Subjects were given medical grade mixtures of oxygen and nitrogen to induce stable plateaus across the range of 100% to 70%. The goal was to have an equal distribution of data by decade. The stable plateaus allowed data collection in the following SaO2 ranges 95-100, 90-95, 85-90, 80-85, 75-80, 70-75. In general, 4 to 8 discrete points were collected at each of the levels.
3 days
Accuracy data analysis
Time Frame: 3 days
Data analysis follows ISO80601-2-61:2017, Annex EE and Pulse Oximeters - Premarket Notifications Submissions [510(k)s] Guidance For Industry and Food and Drug Administration Staff (issued: March 4, 2013).
3 days

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Collaborators

Investigators

  • Principal Investigator: Arthur Cabrera, M.D., Clinimark, LLC

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

August 25, 2020

Primary Completion (Actual)

August 28, 2020

Study Completion (Actual)

August 28, 2020

Study Registration Dates

First Submitted

April 30, 2021

First Submitted That Met QC Criteria

April 30, 2021

First Posted (Actual)

May 5, 2021

Study Record Updates

Last Update Posted (Actual)

May 5, 2021

Last Update Submitted That Met QC Criteria

April 30, 2021

Last Verified

April 1, 2021

More Information

Terms related to this study

Additional Relevant MeSH Terms

Other Study ID Numbers

  • PR2020-376

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

No

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

Yes

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Hypoxia

Clinical Trials on Pulse Oximeter