Solid State vs. Balloon Esophageal Catheter for Estimation of Pleural Pressure

May 1, 2024 updated by: Annemijn Jonkman, Erasmus Medical Center

Comparison of a Solid State Versus Balloon Esophageal Catheter for Estimation of Pleural Pressure in Surgical ICU Patients

Measurements of esophageal pressure (Pes) as surrogate for pleural pressure are routinely performed in selected ICU patients to facilitate lung-protective ventilation and assess breathing effort. Pes is clinically measured via a nasogastric esophageal catheter. Current techniques involve balloon catheters but have some important disadvantages as they could deflate over time and require a very precise positioning and filling volume. A solid-state sensor does not have disadvantages associated with balloon catheters and may therefore be a useful alternative in clinical practice.

This method-comparison study in adult mechanically ventilated ICU patients evaluates the accuracy of Pes measured using an esophageal catheter with a solid-state sensor as compared to a balloon catheter as reference standard.

Study Overview

Detailed Description

Measurements of esophageal pressure (Pes) as surrogate for pleural pressure are routinely performed in selected ICU patients to facilitate lung-protective ventilation and assess breathing effort. Pes is clinically measured via a nasogastric esophageal catheter. Current techniques involve balloon catheters but have some important disadvantages as they could deflate over time and require a very precise positioning and filling volume. A solid state sensor does not have disadvantages associated with balloon catheters and may therefore be a useful alternative in clinical practice.

This method-comparison study in adult mechanically ventilated ICU patients evaluates the accuracy of Pes measured using an esophageal catheter with a solid-state sensor as compared to a balloon catheter as reference standard.

Subjects will receive a standard balloon esophageal catheter and a solid-state pressure catheter for simultaneous measurements of Pes. Study population will be post-surgical ICU patients. Measurements will be performed during clinical mechanical ventilation settings which include a first phase of passive controlled ventilation (patient is still sedated after surgery) as well a phase of partially-assisted ventilation where the patient's breathing effort has resumed and is assisted by the ventilator. During both phases, 10-15 minutes of tidal breathing will be recorded.

Study Type

Interventional

Enrollment (Actual)

16

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • Zuid-Holland
      • Rotterdam, Zuid-Holland, Netherlands, 3015GD
        • Erasmus MC

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

  • Adult
  • Older Adult

Accepts Healthy Volunteers

No

Description

Inclusion Criteria:

  • Signed informed consent
  • Patient requiring mechanical ventilation at the ICU following cardiothoracic or abdominal surgery
  • Age ≥ 18 year

Exclusion Criteria:

  • Pregnancy
  • Upper airway/esophageal/mouth or face pathology (i.e. recent surgery, esophageal varices, diaphragmatic hernia)
  • Nasal bleeding within the last <2 weeks
  • Presence of pneumothorax
  • Use of anticoagulants that increase the risk of catheter insertion

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Other
  • Allocation: N/A
  • Interventional Model: Single Group Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: Esophageal manometry using a solid state sensor vs. balloon catheter
Placement of both a solid state and balloon esophageal pressure (Pes) catheter. Pes recordings of these catheters will be acquired simultaneously during both controlled mechanical ventilation and assisted mechanical ventilation, for 10-15 minutes per phase. Ventilator settings/protocol will be as per standard-of-care.

Placement of the iEPC nasogastric catheter with solid state sensor for esophageal manometry.

Comparator: Esophageal balloon catheter (NutriVent).

Other Names:
  • Solid state esophageal pressure catheter, PulmoTech B.V.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Difference in absolute Pes-derived parameters obtained by solid state catheter vs. balloon catheter.
Time Frame: 30 minutes
Absolute Pes value (in cmH2O) at end-expiration and at peak inspiration (based on flow recordings) will be recorded.
30 minutes
Difference in relative Pes-derived parameters obtained by solid state catheter vs. balloon catheter.
Time Frame: 30 minutes
Relative Pes value (cmH2O) will be measured as the inspiratory amplitude in Pes.
30 minutes

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Collaborators

Investigators

  • Principal Investigator: Annemijn Jonkman, PhD, Erasmus Medical Center

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

October 20, 2023

Primary Completion (Actual)

March 21, 2024

Study Completion (Actual)

March 21, 2024

Study Registration Dates

First Submitted

April 3, 2023

First Submitted That Met QC Criteria

April 17, 2023

First Posted (Actual)

April 18, 2023

Study Record Updates

Last Update Posted (Actual)

May 2, 2024

Last Update Submitted That Met QC Criteria

May 1, 2024

Last Verified

May 1, 2024

More Information

Terms related to this study

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

NO

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.