- ICH GCP
- US Clinical Trials Registry
- Clinical Trial NCT05817968
Solid State vs. Balloon Esophageal Catheter for Estimation of Pleural Pressure
Comparison of a Solid State Versus Balloon Esophageal Catheter for Estimation of Pleural Pressure in Surgical ICU Patients
Measurements of esophageal pressure (Pes) as surrogate for pleural pressure are routinely performed in selected ICU patients to facilitate lung-protective ventilation and assess breathing effort. Pes is clinically measured via a nasogastric esophageal catheter. Current techniques involve balloon catheters but have some important disadvantages as they could deflate over time and require a very precise positioning and filling volume. A solid-state sensor does not have disadvantages associated with balloon catheters and may therefore be a useful alternative in clinical practice.
This method-comparison study in adult mechanically ventilated ICU patients evaluates the accuracy of Pes measured using an esophageal catheter with a solid-state sensor as compared to a balloon catheter as reference standard.
Study Overview
Status
Conditions
Intervention / Treatment
Detailed Description
Measurements of esophageal pressure (Pes) as surrogate for pleural pressure are routinely performed in selected ICU patients to facilitate lung-protective ventilation and assess breathing effort. Pes is clinically measured via a nasogastric esophageal catheter. Current techniques involve balloon catheters but have some important disadvantages as they could deflate over time and require a very precise positioning and filling volume. A solid state sensor does not have disadvantages associated with balloon catheters and may therefore be a useful alternative in clinical practice.
This method-comparison study in adult mechanically ventilated ICU patients evaluates the accuracy of Pes measured using an esophageal catheter with a solid-state sensor as compared to a balloon catheter as reference standard.
Subjects will receive a standard balloon esophageal catheter and a solid-state pressure catheter for simultaneous measurements of Pes. Study population will be post-surgical ICU patients. Measurements will be performed during clinical mechanical ventilation settings which include a first phase of passive controlled ventilation (patient is still sedated after surgery) as well a phase of partially-assisted ventilation where the patient's breathing effort has resumed and is assisted by the ventilator. During both phases, 10-15 minutes of tidal breathing will be recorded.
Study Type
Enrollment (Actual)
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
Zuid-Holland
-
Rotterdam, Zuid-Holland, Netherlands, 3015GD
- Erasmus MC
-
-
Participation Criteria
Eligibility Criteria
Ages Eligible for Study
- Adult
- Older Adult
Accepts Healthy Volunteers
Description
Inclusion Criteria:
- Signed informed consent
- Patient requiring mechanical ventilation at the ICU following cardiothoracic or abdominal surgery
- Age ≥ 18 year
Exclusion Criteria:
- Pregnancy
- Upper airway/esophageal/mouth or face pathology (i.e. recent surgery, esophageal varices, diaphragmatic hernia)
- Nasal bleeding within the last <2 weeks
- Presence of pneumothorax
- Use of anticoagulants that increase the risk of catheter insertion
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Other
- Allocation: N/A
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Arms and Interventions
Participant Group / Arm |
Intervention / Treatment |
|---|---|
|
Experimental: Esophageal manometry using a solid state sensor vs. balloon catheter
Placement of both a solid state and balloon esophageal pressure (Pes) catheter.
Pes recordings of these catheters will be acquired simultaneously during both controlled mechanical ventilation and assisted mechanical ventilation, for 10-15 minutes per phase.
Ventilator settings/protocol will be as per standard-of-care.
|
Placement of the iEPC nasogastric catheter with solid state sensor for esophageal manometry. Comparator: Esophageal balloon catheter (NutriVent).
Other Names:
|
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Difference in absolute Pes-derived parameters obtained by solid state catheter vs. balloon catheter.
Time Frame: 30 minutes
|
Absolute Pes value (in cmH2O) at end-expiration and at peak inspiration (based on flow recordings) will be recorded.
|
30 minutes
|
|
Difference in relative Pes-derived parameters obtained by solid state catheter vs. balloon catheter.
Time Frame: 30 minutes
|
Relative Pes value (cmH2O) will be measured as the inspiratory amplitude in Pes.
|
30 minutes
|
Collaborators and Investigators
Sponsor
Collaborators
Investigators
- Principal Investigator: Annemijn Jonkman, PhD, Erasmus Medical Center
Study record dates
Study Major Dates
Study Start (Actual)
Primary Completion (Actual)
Study Completion (Actual)
Study Registration Dates
First Submitted
First Submitted That Met QC Criteria
First Posted (Actual)
Study Record Updates
Last Update Posted (Actual)
Last Update Submitted That Met QC Criteria
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
- MEC-2023-0119
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.