Automatizovaná aplanační tonometrie - Aktualizováno
Goldmann Applanation Tonometry (GAT) is considered the clinical gold standard for eye pressure measurements and yet it is known to be a subjective measurement with limited repeatability and limited portability. Another clinical standard for checking eye pressure is known as the pneumotonometer. This method is more objective but not portable. The purpose of this study is to develop new methods of measuring eye pressure that are more objective, reproducible and portable. In this study, the investigators will be comparing the eye pressure measurements using 2 investigational methods to GAT and pneuumotonometer.
Přehled studie
Postavení
Postavení
Podmínky
Podmínky
Intervence / Léčba
Intervence / Léčba
Detailní popis
This is a prospective study. Patients presenting for their scheduled eye appointments will be recruited by verbal communication. Following informed consent, intraocular pressure (IOP) measurements will be made by the following methods:
Standard Goldmann Applanation Tonometry (GAT): this is the standard method for IOP measurement in clinical practice. The eye is given topical fluorescein/anesthetic, the GAT prism contacts the eye while the observer looks through the slit lamp machine ocular using blue light illumination to visual the applanation mires. The GAT dial is adjusted until mire alignment is achieved and the IOP measurement is read off the GAT dial.
Pneumotonometer: This is what is done in routine eye care. The participant will receive a topical anesthetic (numbing agent). A probe will touch the front of the eye while the machine measures the eye pressure.
Fixed force GAT: From the patient perspective, this method will feel identical to the standard GAT. The eye is given topical fluorescein/anesthetic. The GAT dial is set at 1.8 or 2.0, a complementary metal oxide semiconductor (C-MOS) camera is connected to one of the oculars of the slit lamp machine and under blue light illumination, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance through the ocular. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter. A similar method using an iPod touch camera has been previously published by this PI.
Upright applanating prototype: With this prototype, an applanating prism (custom manufactured with medical grade acrylic in an ISO-13485 certified facility) is attached to a fixed-force spring that creates a force equivalent to 1.8 or 2.0 on the GAT dial. Blue LED lights on the prototype are used to create the blue illumination similar to the blue light used in clinical practice on the slit lamp or Perkins tonometer. A C-MOS camera is aligned with the GAT prism to image the applanation mires. The eye is given topical fluorescein/anesthetic. Like fixed-force GAT, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter.
The order of the above 4 measurements will be randomized.
Typ studie
Typ studie
Zápis (Odhadovaný)
Zápis
Fáze
Fáze
- Nelze použít
Kontakty a umístění
Studijní kontakt
Studijní kontakt
- Jméno: Sarah Jones
- Telefonní číslo: 919 681 6584
- E-mail: sarah.jones1@duke.edu
Studijní místa
-
-
North Carolina
-
Durham, North Carolina, Spojené státy, 27710
- Nábor
- Duke Eye Center
-
Vrchní vyšetřovatel:
- Joanne Wen, MD
-
Kontakt:
- Joanne Wen, MD
- Telefonní číslo: +1 919 684 2975
- E-mail: joanne.wen@duke.edu
-
-
Kritéria účasti
Kritéria způsobilosti
Kritéria způsobilosti
Věk způsobilý ke studiu
Přijímá zdravé dobrovolníky
Pohlaví způsobilá ke studiu
Popis
Criteria:
Inclusion Criteria:
- Presenting for a routine eye exam
- ≥ 18 years of age
- Able and willing to give consent
Exclusion Criteria:
- History of corneal scarring
- Active infection of the eye
- Ocular surface trauma or infection
Studijní plán
Jak je studie koncipována?
Detaily designu
- Primární účel: Diagnostický
- Přidělení: Randomizované
- Intervenční model: Paralelní přiřazení
- Maskování: Žádné (otevřený štítek)
Počet zbraní
Zbraně a zásahy
Skupina účastníků / ArmSkupina účastníků / Arm |
Intervence / LéčbaIntervence / Léčba |
|---|---|
|
Aktivní komparátor: Standard GAT
This is the gold-standard method for IOP measurement in clinical practice |
Standard Goldmann Applanation Tonometry (GAT): this is the standard method for IOP measurement in clinical practice |
|
Experimentální: Fixed-force GAT
From the patient perspective, this method will feel identical to the standard GAT. The eye is given topical fluorescein/anesthetic. The GAT dial is set at 1.8 or 2.0, a C-MOS camera is connected to one of the oculars of the slit lamp machine and under blue light illumination, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance through the ocular. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter |
An investigational device similar to standard GAT |
|
Experimentální: Upright applanating prototype
With this prototype, an applanating prism (custom manufactured with medical grade acrylic in an ISO-13485 certified facility) is attached to a fixed-force spring that creates a force equivalent to 1.8 or 2.0 on the GAT dial. Blue LED lights on the prototype are used to create the blue illumination similar to the blue light used in clinical practice on the slit lamp or Perkins tonometer. A C-MOS camera is aligned with the GAT prism to image the applanation mires. The eye is given topical fluorescein/anesthetic. Like fixed-force GAT, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter |
An investigational device, prism used in standard and fixed-force GAT is attached to a portable device |
|
Aktivní komparátor: Pneumotonometer
This is another standard method for IOP measurement in clinical practice |
This is a standard for checking eye pressure. A topical anesthetic (numbing agent). A probe will touch the front of the eye while the machine measures the eye pressure |
Co je měření studie?
Primární výstupní opatření
Primární výstupní opatření
Měření výsledku |
Popis opatření |
Časové okno |
|---|---|---|
|
Intraokulární tlak měřený vleže pomocí pneumotonometru
Časové okno: Výchozí stav
|
Výchozí stav
|
|
|
Intraocular pressure (IOP) as measured by standard GAT
Časové okno: Baseline
|
this is the standard method for IOP measurement in clinical practice. The eye is given topical fluorescein/anesthetic, the GAT prism contacts the eye while the observer looks through the slit lamp machine ocular using blue light illumination to visual the applanation mires. The GAT dial is adjusted until mire alignment is achieved and the IOP measurement is read off the GAT dial. |
Baseline
|
|
Intraocular pressure (IOP) as measured by fixed force GAT
Časové okno: Baseline
|
From the patient perspective, this method will feel identical to the standard GAT. The eye is given topical fluorescein/anesthetic. The GAT dial is set at 1.8 or 2.0, a C-MOS camera is connected to one of the oculars of the slit lamp machine and under blue light illumination, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance through the ocular. |
Baseline
|
|
Intraocular pressure as measured by upright applanating
Časové okno: Baseline
|
With this prototype, an applanating prism (custom manufactured with medical grade acrylic in an ISO-13485 certified facility) is attached to a fixed-force spring that creates a force equivalent to 1.8 or 2.0 on the GAT dial. Blue LED lights on the prototype are used to create the blue illumination similar to the blue light used in clinical practice on the slit lamp or Perkins tonometer. A C-MOS camera is aligned with the GAT prism to image the applanation mires. The eye is given topical fluorescein/anesthetic. Like fixed-force GAT, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter. |
Baseline
|
Spolupracovníci a vyšetřovatelé
Sponzor
Sponzor
Vyšetřovatelé
Vyšetřovatelé
- Vrchní vyšetřovatel: Joanne Wen, MD, Duke Eye Center
Termíny studijních záznamů
Hlavní termíny studia
Začátek studia (Aktuální)
Začátek studia
Primární dokončení (Odhadovaný)
Primární dokončení
Dokončení studie (Odhadovaný)
Dokončení studie
Termíny zápisu do studia
První předloženo
První předloženo
První předloženo, které splnilo kritéria kontroly kvality
První předloženo, které splnilo kritéria kontroly kvality
První zveřejněno (Aktuální)
První zveřejněno
Aktualizace studijních záznamů
Poslední zveřejněná aktualizace (Aktuální)
Poslední zveřejněná aktualizace
Odeslaná poslední aktualizace, která splnila kritéria kontroly kvality
Odeslaná poslední aktualizace, která splnila kritéria kontroly kvality
Naposledy ověřeno
Naposledy ověřeno
Více informací
Termíny související s touto studií
Další relevantní podmínky MeSH
Další identifikační čísla studie
Další identifikační čísla studie
- Pro00106111_1
Plán pro data jednotlivých účastníků (IPD)
Plánujete sdílet data jednotlivých účastníků (IPD)?
Popis plánu IPD
No Currently, there is no plan to share data with other researchers.
Informace o lécích a zařízeních, studijní dokumenty
Studuje lékový produkt regulovaný americkým FDA
Studuje produkt zařízení regulovaný americkým úřadem FDA
produkt vyrobený a vyvážený z USA
Tyto informace byly beze změn načteny přímo z webu clinicaltrials.gov. Máte-li jakékoli požadavky na změnu, odstranění nebo aktualizaci podrobností studie, kontaktujte prosím register@clinicaltrials.gov. Jakmile bude změna implementována na clinicaltrials.gov, bude automaticky aktualizována i na našem webu .