Automatisierte Applanationstonometrie - Aktualisiert
Automatisierte Applanations-Tonometrie - Aktualisiert
Goldmann Applanation Tonometry (GAT) is considered the clinical gold standard for eye pressure measurements and yet it is known to be a subjective measurement with limited repeatability and limited portability. Another clinical standard for checking eye pressure is known as the pneumotonometer. This method is more objective but not portable. The purpose of this study is to develop new methods of measuring eye pressure that are more objective, reproducible and portable. In this study, the investigators will be comparing the eye pressure measurements using 2 investigational methods to GAT and pneuumotonometer.
Studienübersicht
Status
Status
Bedingungen
Bedingungen
Intervention / Behandlung
Intervention / Behandlung
Detaillierte Beschreibung
This is a prospective study. Patients presenting for their scheduled eye appointments will be recruited by verbal communication. Following informed consent, intraocular pressure (IOP) measurements will be made by the following methods:
Standard Goldmann Applanation Tonometry (GAT): this is the standard method for IOP measurement in clinical practice. The eye is given topical fluorescein/anesthetic, the GAT prism contacts the eye while the observer looks through the slit lamp machine ocular using blue light illumination to visual the applanation mires. The GAT dial is adjusted until mire alignment is achieved and the IOP measurement is read off the GAT dial.
Pneumotonometer: This is what is done in routine eye care. The participant will receive a topical anesthetic (numbing agent). A probe will touch the front of the eye while the machine measures the eye pressure.
Fixed force GAT: From the patient perspective, this method will feel identical to the standard GAT. The eye is given topical fluorescein/anesthetic. The GAT dial is set at 1.8 or 2.0, a complementary metal oxide semiconductor (C-MOS) camera is connected to one of the oculars of the slit lamp machine and under blue light illumination, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance through the ocular. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter. A similar method using an iPod touch camera has been previously published by this PI.
Upright applanating prototype: With this prototype, an applanating prism (custom manufactured with medical grade acrylic in an ISO-13485 certified facility) is attached to a fixed-force spring that creates a force equivalent to 1.8 or 2.0 on the GAT dial. Blue LED lights on the prototype are used to create the blue illumination similar to the blue light used in clinical practice on the slit lamp or Perkins tonometer. A C-MOS camera is aligned with the GAT prism to image the applanation mires. The eye is given topical fluorescein/anesthetic. Like fixed-force GAT, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter.
The order of the above 4 measurements will be randomized.
Studientyp
Studientyp
Einschreibung (Geschätzt)
Einschreibung
Phase
Phase
- Unzutreffend
Kontakte und Standorte
Studienkontakt
Studienkontakt
- Name: Sarah Jones
- Telefonnummer: 919 681 6584
- E-Mail: sarah.jones1@duke.edu
Studienorte
-
-
North Carolina
-
Durham, North Carolina, Vereinigte Staaten, 27710
- Rekrutierung
- Duke Eye Center
-
Hauptermittler:
- Joanne Wen, MD
-
Kontakt:
- Joanne Wen, MD
- Telefonnummer: +1 919 684 2975
- E-Mail: joanne.wen@duke.edu
-
-
Teilnahmekriterien
Zulassungskriterien
Zulassungskriterien
Studienberechtigtes Alter
Akzeptiert gesunde Freiwillige
Studienberechtigte Geschlechter
Beschreibung
Criteria:
Inclusion Criteria:
- Presenting for a routine eye exam
- ≥ 18 years of age
- Able and willing to give consent
Exclusion Criteria:
- History of corneal scarring
- Active infection of the eye
- Ocular surface trauma or infection
Studienplan
Wie ist die Studie aufgebaut?
Designdetails
- Hauptzweck: Diagnose
- Zuteilung: Zufällig
- Interventionsmodell: Parallele Zuordnung
- Maskierung: Keine (Offenes Etikett)
Anzahl der Arme
Waffen und Interventionen
Teilnehmergruppe / ArmTeilnehmergruppe / Arm |
Intervention / BehandlungIntervention / Behandlung |
|---|---|
|
Aktiver Komparator: Standard GAT
This is the gold-standard method for IOP measurement in clinical practice |
Standard Goldmann Applanation Tonometry (GAT): this is the standard method for IOP measurement in clinical practice |
|
Experimental: Fixed-force GAT
From the patient perspective, this method will feel identical to the standard GAT. The eye is given topical fluorescein/anesthetic. The GAT dial is set at 1.8 or 2.0, a C-MOS camera is connected to one of the oculars of the slit lamp machine and under blue light illumination, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance through the ocular. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter |
An investigational device similar to standard GAT |
|
Experimental: Upright applanating prototype
With this prototype, an applanating prism (custom manufactured with medical grade acrylic in an ISO-13485 certified facility) is attached to a fixed-force spring that creates a force equivalent to 1.8 or 2.0 on the GAT dial. Blue LED lights on the prototype are used to create the blue illumination similar to the blue light used in clinical practice on the slit lamp or Perkins tonometer. A C-MOS camera is aligned with the GAT prism to image the applanation mires. The eye is given topical fluorescein/anesthetic. Like fixed-force GAT, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter |
An investigational device, prism used in standard and fixed-force GAT is attached to a portable device |
|
Aktiver Komparator: Pneumotonometer
This is another standard method for IOP measurement in clinical practice |
This is a standard for checking eye pressure. A topical anesthetic (numbing agent). A probe will touch the front of the eye while the machine measures the eye pressure |
Was misst die Studie?
Primäre Ergebnismessungen
Primäre Ergebnismessungen
Ergebnis Maßnahme |
Maßnahmenbeschreibung |
Zeitfenster |
|---|---|---|
|
Intraokularer Druck, gemessen mit dem liegenden Pneumotonometer
Zeitfenster: Ausgangswert
|
Ausgangswert
|
|
|
Intraocular pressure (IOP) as measured by standard GAT
Zeitfenster: Baseline
|
this is the standard method for IOP measurement in clinical practice. The eye is given topical fluorescein/anesthetic, the GAT prism contacts the eye while the observer looks through the slit lamp machine ocular using blue light illumination to visual the applanation mires. The GAT dial is adjusted until mire alignment is achieved and the IOP measurement is read off the GAT dial. |
Baseline
|
|
Intraocular pressure (IOP) as measured by fixed force GAT
Zeitfenster: Baseline
|
From the patient perspective, this method will feel identical to the standard GAT. The eye is given topical fluorescein/anesthetic. The GAT dial is set at 1.8 or 2.0, a C-MOS camera is connected to one of the oculars of the slit lamp machine and under blue light illumination, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance through the ocular. |
Baseline
|
|
Intraocular pressure as measured by upright applanating
Zeitfenster: Baseline
|
With this prototype, an applanating prism (custom manufactured with medical grade acrylic in an ISO-13485 certified facility) is attached to a fixed-force spring that creates a force equivalent to 1.8 or 2.0 on the GAT dial. Blue LED lights on the prototype are used to create the blue illumination similar to the blue light used in clinical practice on the slit lamp or Perkins tonometer. A C-MOS camera is aligned with the GAT prism to image the applanation mires. The eye is given topical fluorescein/anesthetic. Like fixed-force GAT, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter. |
Baseline
|
Mitarbeiter und Ermittler
Sponsor
Sponsor
Ermittler
Ermittler
- Hauptermittler: Joanne Wen, MD, Duke Eye Center
Studienaufzeichnungsdaten
Haupttermine studieren
Studienbeginn (Tatsächlich)
Studienbeginn
Primärer Abschluss (Geschätzt)
Primärer Abschluss
Studienabschluss (Geschätzt)
Studienabschluss
Studienanmeldedaten
Zuerst eingereicht
Zuerst eingereicht
Zuerst eingereicht, das die QC-Kriterien erfüllt hat
Zuerst eingereicht, das die QC-Kriterien erfüllt hat
Zuerst gepostet (Tatsächlich)
Zuerst gepostet
Studienaufzeichnungsaktualisierungen
Letztes Update gepostet (Tatsächlich)
Letztes Update gepostet
Letztes eingereichtes Update, das die QC-Kriterien erfüllt
Letztes eingereichtes Update, das die QC-Kriterien erfüllt
Zuletzt verifiziert
Zuletzt verifiziert
Mehr Informationen
Begriffe im Zusammenhang mit dieser Studie
Zusätzliche relevante MeSH-Bedingungen
Andere Studien-ID-Nummern
Andere Studien-ID-Nummern
- Pro00106111_1
Plan für individuelle Teilnehmerdaten (IPD)
Planen Sie, individuelle Teilnehmerdaten (IPD) zu teilen?
Beschreibung des IPD-Plans
No Currently, there is no plan to share data with other researchers.
Arzneimittel- und Geräteinformationen, Studienunterlagen
Studiert ein von der US-amerikanischen FDA reguliertes Arzneimittelprodukt
Studiert ein von der US-amerikanischen FDA reguliertes Geräteprodukt
Produkt, das in den USA hergestellt und aus den USA exportiert wird
Diese Informationen wurden ohne Änderungen direkt von der Website clinicaltrials.gov abgerufen. Wenn Sie Ihre Studiendaten ändern, entfernen oder aktualisieren möchten, wenden Sie sich bitte an register@clinicaltrials.gov. Sobald eine Änderung auf clinicaltrials.gov implementiert wird, wird diese automatisch auch auf unserer Website aktualisiert .