Automatisierte Applanationstonometrie - Aktualisiert

2. November 2026 aktualisiert von: Duke University

Automatisierte Applanations-Tonometrie - Aktualisiert

Goldmann Applanation Tonometry (GAT) is considered the clinical gold standard for eye pressure measurements and yet it is known to be a subjective measurement with limited repeatability and limited portability. Another clinical standard for checking eye pressure is known as the pneumotonometer. This method is more objective but not portable. The purpose of this study is to develop new methods of measuring eye pressure that are more objective, reproducible and portable. In this study, the investigators will be comparing the eye pressure measurements using 2 investigational methods to GAT and pneuumotonometer.

Studienübersicht

Status

Rekrutierung

Bedingungen

Intervention / Behandlung

Detaillierte Beschreibung

This is a prospective study. Patients presenting for their scheduled eye appointments will be recruited by verbal communication. Following informed consent, intraocular pressure (IOP) measurements will be made by the following methods:

Standard Goldmann Applanation Tonometry (GAT): this is the standard method for IOP measurement in clinical practice. The eye is given topical fluorescein/anesthetic, the GAT prism contacts the eye while the observer looks through the slit lamp machine ocular using blue light illumination to visual the applanation mires. The GAT dial is adjusted until mire alignment is achieved and the IOP measurement is read off the GAT dial.

Pneumotonometer: This is what is done in routine eye care. The participant will receive a topical anesthetic (numbing agent). A probe will touch the front of the eye while the machine measures the eye pressure.

Fixed force GAT: From the patient perspective, this method will feel identical to the standard GAT. The eye is given topical fluorescein/anesthetic. The GAT dial is set at 1.8 or 2.0, a complementary metal oxide semiconductor (C-MOS) camera is connected to one of the oculars of the slit lamp machine and under blue light illumination, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance through the ocular. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter. A similar method using an iPod touch camera has been previously published by this PI.

Upright applanating prototype: With this prototype, an applanating prism (custom manufactured with medical grade acrylic in an ISO-13485 certified facility) is attached to a fixed-force spring that creates a force equivalent to 1.8 or 2.0 on the GAT dial. Blue LED lights on the prototype are used to create the blue illumination similar to the blue light used in clinical practice on the slit lamp or Perkins tonometer. A C-MOS camera is aligned with the GAT prism to image the applanation mires. The eye is given topical fluorescein/anesthetic. Like fixed-force GAT, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter.

The order of the above 4 measurements will be randomized.

Studientyp

Interventionell

Einschreibung (Geschätzt)

200

Phase

  • Unzutreffend

Kontakte und Standorte

Dieser Abschnitt enthält die Kontaktdaten derjenigen, die die Studie durchführen, und Informationen darüber, wo diese Studie durchgeführt wird.

Studienkontakt

Studienorte

    • North Carolina
      • Durham, North Carolina, Vereinigte Staaten, 27710
        • Rekrutierung
        • Duke Eye Center
        • Hauptermittler:
          • Joanne Wen, MD
        • Kontakt:

Teilnahmekriterien

Forscher suchen nach Personen, die einer bestimmten Beschreibung entsprechen, die als Auswahlkriterien bezeichnet werden. Einige Beispiele für diese Kriterien sind der allgemeine Gesundheitszustand einer Person oder frühere Behandlungen.

Zulassungskriterien

Studienberechtigtes Alter

18 Jahre und älter (Erwachsene, Älterer Erwachsener)

Akzeptiert gesunde Freiwillige

Nein

Studienberechtigte Geschlechter

Alle

Beschreibung

Criteria:

Inclusion Criteria:

  • Presenting for a routine eye exam
  • ≥ 18 years of age
  • Able and willing to give consent

Exclusion Criteria:

  • History of corneal scarring
  • Active infection of the eye
  • Ocular surface trauma or infection

Studienplan

Dieser Abschnitt enthält Einzelheiten zum Studienplan, einschließlich des Studiendesigns und der Messung der Studieninhalte.

Wie ist die Studie aufgebaut?

Designdetails

  • Hauptzweck: Diagnose
  • Zuteilung: Zufällig
  • Interventionsmodell: Parallele Zuordnung
  • Maskierung: Keine (Offenes Etikett)

Waffen und Interventionen

Teilnehmergruppe / Arm
Intervention / Behandlung
Aktiver Komparator: Standard GAT

This is the gold-standard method for IOP measurement in clinical practice

Standard Goldmann Applanation Tonometry (GAT): this is the standard method for IOP measurement in clinical practice

Experimental: Fixed-force GAT

From the patient perspective, this method will feel identical to the standard GAT. The eye is given topical fluorescein/anesthetic. The GAT dial is set at 1.8 or 2.0, a C-MOS camera is connected to one of the oculars of the slit lamp machine and under blue light illumination, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance through the ocular. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter

An investigational device similar to standard GAT

Experimental: Upright applanating prototype

With this prototype, an applanating prism (custom manufactured with medical grade acrylic in an ISO-13485 certified facility) is attached to a fixed-force spring that creates a force equivalent to 1.8 or 2.0 on the GAT dial. Blue LED lights on the prototype are used to create the blue illumination similar to the blue light used in clinical practice on the slit lamp or Perkins tonometer. A C-MOS camera is aligned with the GAT prism to image the applanation mires. The eye is given topical fluorescein/anesthetic. Like fixed-force GAT, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter

An investigational device, prism used in standard and fixed-force GAT is attached to a portable device

Aktiver Komparator: Pneumotonometer

This is another standard method for IOP measurement in clinical practice

This is a standard for checking eye pressure. A topical anesthetic (numbing agent). A probe will touch the front of the eye while the machine measures the eye pressure

Was misst die Studie?

Primäre Ergebnismessungen

Ergebnis Maßnahme
Maßnahmenbeschreibung
Zeitfenster
Intraokularer Druck, gemessen mit dem liegenden Pneumotonometer
Zeitfenster: Ausgangswert
Ausgangswert
Intraocular pressure (IOP) as measured by standard GAT
Zeitfenster: Baseline

this is the standard method for IOP measurement in clinical practice. The eye is given topical fluorescein/anesthetic, the GAT prism contacts the eye while the observer looks through the slit lamp machine ocular using blue light illumination to visual the applanation mires. The GAT dial is adjusted until mire alignment is achieved and the IOP measurement is read off the GAT dial.

Baseline
Intraocular pressure (IOP) as measured by fixed force GAT
Zeitfenster: Baseline

From the patient perspective, this method will feel identical to the standard GAT. The eye is given topical fluorescein/anesthetic. The GAT dial is set at 1.8 or 2.0, a C-MOS camera is connected to one of the oculars of the slit lamp machine and under blue light illumination, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance through the ocular.

Baseline
Intraocular pressure as measured by upright applanating
Zeitfenster: Baseline

With this prototype, an applanating prism (custom manufactured with medical grade acrylic in an ISO-13485 certified facility) is attached to a fixed-force spring that creates a force equivalent to 1.8 or 2.0 on the GAT dial. Blue LED lights on the prototype are used to create the blue illumination similar to the blue light used in clinical practice on the slit lamp or Perkins tonometer. A C-MOS camera is aligned with the GAT prism to image the applanation mires. The eye is given topical fluorescein/anesthetic. Like fixed-force GAT, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter.

Baseline

Mitarbeiter und Ermittler

Hier finden Sie Personen und Organisationen, die an dieser Studie beteiligt sind.

Sponsor

Ermittler

  • Hauptermittler: Joanne Wen, MD, Duke Eye Center

Studienaufzeichnungsdaten

Diese Daten verfolgen den Fortschritt der Übermittlung von Studienaufzeichnungen und zusammenfassenden Ergebnissen an ClinicalTrials.gov. Studienaufzeichnungen und gemeldete Ergebnisse werden von der National Library of Medicine (NLM) überprüft, um sicherzustellen, dass sie bestimmten Qualitätskontrollstandards entsprechen, bevor sie auf der öffentlichen Website veröffentlicht werden.

Haupttermine studieren

Studienbeginn (Tatsächlich)

2. März 2026

Primärer Abschluss (Geschätzt)

29. Dezember 2026

Studienabschluss (Geschätzt)

29. Dezember 2026

Studienanmeldedaten

Zuerst eingereicht

19. Dezember 2025

Zuerst eingereicht, das die QC-Kriterien erfüllt hat

19. Dezember 2025

Zuerst gepostet (Tatsächlich)

23. Dezember 2025

Studienaufzeichnungsaktualisierungen

Letztes Update gepostet (Tatsächlich)

3. November 2026

Letztes eingereichtes Update, das die QC-Kriterien erfüllt

2. November 2026

Zuletzt verifiziert

1. November 2026

Mehr Informationen

Begriffe im Zusammenhang mit dieser Studie

Zusätzliche relevante MeSH-Bedingungen

Andere Studien-ID-Nummern

  • Pro00106111_1

Plan für individuelle Teilnehmerdaten (IPD)

Planen Sie, individuelle Teilnehmerdaten (IPD) zu teilen?

NEIN

Beschreibung des IPD-Plans

No Currently, there is no plan to share data with other researchers.

Arzneimittel- und Geräteinformationen, Studienunterlagen

Studiert ein von der US-amerikanischen FDA reguliertes Arzneimittelprodukt

Nein

Studiert ein von der US-amerikanischen FDA reguliertes Geräteprodukt

Ja

Produkt, das in den USA hergestellt und aus den USA exportiert wird

Nein

Diese Informationen wurden ohne Änderungen direkt von der Website clinicaltrials.gov abgerufen. Wenn Sie Ihre Studiendaten ändern, entfernen oder aktualisieren möchten, wenden Sie sich bitte an register@clinicaltrials.gov. Sobald eine Änderung auf clinicaltrials.gov implementiert wird, wird diese automatisch auch auf unserer Website aktualisiert .