Real-World Validation of Huawei Wearable Technologies for Hypertension Risk Screening in Malaysia (Huawei-SunU1)
Background:
Hypertension (high blood pressure) is a major risk factor for heart disease and stroke, yet many people do not know they have it. In Malaysia, nearly half of all adults with elevated blood pressure are unaware of their condition. Standard clinic blood pressure checks can miss cases of high blood pressure or falsely flag temporary stress-related spikes. While 24-hour Ambulatory Blood Pressure Monitoring (ABPM) is the standard method for an accurate diagnosis, it is not practical for large-scale screening.
Purpose:
The purpose of this study is to evaluate whether a novel Huawei smartwatch, which uses photoplethysmography (PPG) technology to estimate blood pressure, can accurately identify individuals at high risk for hypertension in a real-world setting.
Study Design & Procedures:
Participants: Approximately 500 adults will be enrolled over a 24-month study period.
Screening: All participants will use the Huawei smartwatch screening tool to monitor their blood pressure estimates.
Validation: A subgroup of approximately 50 participants identified as high-risk by the smartwatch will undergo 24-hour ABPM (the clinical gold standard) to confirm whether they have high blood pressure.
Goal:
Researchers aim to validate the accuracy of the smartwatch screening tool against 24-hour ABPM to determine if wearable technology can support early detection and timely treatment of undiagnosed hypertension in the community.
Přehled studie
Postavení
Postavení
Podmínky
Podmínky
Typ studie
Typ studie
Zápis (Odhadovaný)
Zápis
Kontakty a umístění
Studijní kontakt
Studijní kontakt
- Jméno: Yook Chin Chia
- Telefonní číslo: +60374918622
- E-mail: ycchia@sunway.edu.my
Studijní záloha kontaktů
- Jméno: Yee How Say
- Telefonní číslo: +60374918622
- E-mail: yeehows@sunway.edu.my
Studijní místa
-
-
Selangor
-
Subang Jaya, Selangor, Malajsie, 47500
- Faculty of Medical and Life Sciences, Sunway University
-
Kontakt:
- Yook Chin Chia
- Telefonní číslo: +60374918622
- E-mail: ycchia@sunway.edu.my
-
Kontakt:
- Yee How Say
- Telefonní číslo: +60374918622
- E-mail: yeehows@sunway.edu.my
-
Vrchní vyšetřovatel:
- Yook Chin Chia
-
Vrchní vyšetřovatel:
- Yee How Say
-
-
Kritéria účasti
Kritéria způsobilosti
Kritéria způsobilosti
Věk způsobilý ke studiu
- Dospělý
- Starší dospělý
Přijímá zdravé dobrovolníky
Metoda odběru vzorků
Studijní populace
Popis
Inclusion Criteria:
- Possesses the specific Huawei wearable required for the study
- Aged ≥18 years
- Malaysian citizen/permanent resident
- Understands English, Malay, or Chinese
- Able to download the required application on a compatible smartphone
- Willing to wear devices continuously
- Able to provide informed consent.
Exclusion Criteria:
- Current pregnancy
- Implantable cardiac devices (e.g., permanent pacemakers, implantable cardioverter-defibrillators [ICDs])
- Uncontrolled arrhythmias (e.g., active atrial fibrillation, frequent premature ventricular contractions [PVCs])
- Movement disorders or active tremors that interfere with signal acquisition (e.g., severe Parkinson's disease, essential tremor)
- Acute medical conditions, including:
- Acute coronary syndromes (e.g., recent myocardial infarction or unstable angina within the past 3 months)
- Acute cerebrovascular events (e.g., stroke or transient ischemic attack [TIA] within the past 3 months)
- Acute decompensated heart failure or severe respiratory distress
- Active, severe acute infections or systemic inflammatory illness (e.g., sepsis, acute pneumonia)
- Hypertensive emergencies or urgency
- Currently on antihypertensive medications
- Resting BP ≥180/110 mmHg.
Studijní plán
Jak je studie koncipována?
Detaily designu
Počet skupin / kohort
Kohorty a intervence
Skupina / kohortaSkupina / kohorta |
|---|
|
Screening Population
Adults enrolled in the community screening phase who undergo non-invasive blood pressure estimation using the Huawei PPG smartwatch over the 24-month study period.
|
|
High-Risk Validation Cohort
A targeted subset of participants (n = 50) identified as high-risk for hypertension by the Huawei smartwatch screening tool, who undergo 24-hour Ambulatory Blood Pressure Monitoring (ABPM) to confirm their diagnostic status.
|
Co je měření studie?
Primární výstupní opatření
Primární výstupní opatření
Měření výsledku |
Popis opatření |
Časové okno |
|---|---|---|
|
Diagnostic Accuracy (AUC-ROC) of the Huawei PPG Smartwatch Hypertension Risk Algorithm Compared to 24-Hour Ambulatory Blood Pressure Monitoring (ABPM)
Časové okno: Month 11 through Month 18.
|
Measure Description: Area Under the Receiver Operating Characteristic Curve (AUC-ROC) evaluated by comparing the binary high-risk classification flags generated by the Huawei PPG smartwatch algorithm directly against diagnostic confirmation from clinical 24-hour ABPM. Unit of Measure: Score on a scale from 0.0 to 1.0 (where 0.5 represents chance performance and 1.0 represents perfect discrimination) |
Month 11 through Month 18.
|
|
Diagnostic Sensitivity of the Huawei PPG Smartwatch Hypertension Risk Algorithm
Časové okno: Month 11 through Month 18.
|
Measure Description: The proportion of participants with true ABPM-confirmed hypertension who are correctly identified as high-risk by the Huawei PPG smartwatch algorithm. Unit of Measure: Percentage of true positive cases |
Month 11 through Month 18.
|
|
Diagnostic Specificity of the Huawei PPG Smartwatch Hypertension Risk Algorithm
Časové okno: Month 11 through Month 18.
|
Measure Description: The proportion of participants without ABPM-confirmed hypertension who are correctly identified as low-risk by the Huawei PPG smartwatch algorithm. Unit of Measure: Percentage of true negative cases |
Month 11 through Month 18.
|
|
Positive Predictive Value (PPV) of the Huawei PPG Smartwatch Hypertension Risk Algorithm
Časové okno: Month 11 through Month 18
|
Measure Description: The probability that a participant flagged as high-risk by the Huawei PPG smartwatch algorithm actually has hypertension confirmed by 24-hour ABPM. Unit of Measure: Percentage |
Month 11 through Month 18
|
|
Negative Predictive Value (NPV) of the Huawei PPG Smartwatch Hypertension Risk Algorithm
Časové okno: Month 11 through Month 18
|
Measure Description: The probability that a participant flagged as low-risk by the Huawei PPG smartwatch algorithm actually does not have hypertension confirmed by 24-hour ABPM. Unit of Measure: Percentage |
Month 11 through Month 18
|
Sekundární výstupní opatření
Sekundární výstupní opatření
Měření výsledku |
Popis opatření |
Časové okno |
|---|---|---|
|
Participant Adherence to Huawei Wearable Monitoring Protocols
Časové okno: Month 1 through Month 12 (up to 12 months per participant)
|
Measure Description: The proportion of enrolled participants who achieve minimum complete wearable datasets, defined as continuous daily wear of the Huawei smartwatch to track PPG, heart rate, physical activity, and sleep metrics. Unit of Measure: Percentage of enrolled participants |
Month 1 through Month 12 (up to 12 months per participant)
|
|
Participant Usability and Acceptance Score Measured via Post-Monitoring Questionnaire
Časové okno: At the completion of the wearable monitoring phase (Month 12).
|
Measure Description: Total score evaluating participant satisfaction, device comfort, and acceptance of the Huawei wearable for hypertension screening, measured using a standardized post-monitoring user experience questionnaire.
Unit of Measure: Mean score on a 1 to 5 Likert scale (where 1 = Strongly Disagree/Poor and 5 = Strongly Agree/Excellent)
|
At the completion of the wearable monitoring phase (Month 12).
|
Další výstupní opatření
Další výstupní opatření
Měření výsledku |
Popis opatření |
Časové okno |
|---|---|---|
|
Prevalence of Wearable-Identified High-Risk Flags in the Study Population
Časové okno: Month 3 through Month 12 (during community-based wearable data collection).
|
Measure Description: The percentage of participants flagged as high-risk for hypertension by the Huawei wearable platform within the total recruited screening cohort. Unit of Measure: Percentage of total participants |
Month 3 through Month 12 (during community-based wearable data collection).
|
|
Association Between Wearable Digital Biomarkers and ABPM-Confirmed Hypertension Risk
Časové okno: Month 17 through Month 20 (during data cleaning and analytical modelling).
|
Measure Description: Adjusted Odds Ratios derived from logistic regression models assessing the relationship between wearable physiological parameters (heart rate, heart rate variability, sleep duration/quality, physical activity, and PPG blood pressure estimates) and ABPM-confirmed hypertension status. Unit of Measure: Adjusted Odds Ratio (OR) with 95% Confidence Intervals |
Month 17 through Month 20 (during data cleaning and analytical modelling).
|
Spolupracovníci a vyšetřovatelé
Sponzor
Sponzor
Spolupracovníci
Spolupracovníci
Vyšetřovatelé
Vyšetřovatelé
- Vrchní vyšetřovatel: Yook Chin Chia, Sunway University
Publikace a užitečné odkazy
Obecné publikace
- Islam SMS, Chow CK, Daryabeygikhotbehsara R, Subedi N, Rawstorn J, Tegegne T, Karmakar C, Siddiqui MU, Lambert G, Maddison R. Wearable cuffless blood pressure monitoring devices: a systematic review and meta-analysis. Eur Heart J Digit Health. 2022 May 2;3(2):323-337. doi: 10.1093/ehjdh/ztac021. eCollection 2022 Jun.
- McEvoy JW, McCarthy CP, Bruno RM, Brouwers S, Canavan MD, Ceconi C, Christodorescu RM, Daskalopoulou SS, Ferro CJ, Gerdts E, Hanssen H, Harris J, Lauder L, McManus RJ, Molloy GJ, Rahimi K, Regitz-Zagrosek V, Rossi GP, Sandset EC, Scheenaerts B, Staessen JA, Uchmanowicz I, Volterrani M, Touyz RM; ESC Scientific Document Group. 2024 ESC Guidelines for the management of elevated blood pressure and hypertension. Eur Heart J. 2024 Oct 7;45(38):3912-4018. doi: 10.1093/eurheartj/ehae178. No abstract available.
- Wang Y, Zhou SS, Liu YQ, Li DD, Hu SY, Wang X, Yi L, Yu YN, Chen YD. Accessibility and usage patterns of wearable devices among Chinese adults: the Huawei Blood Pressure Health Study. Eur Heart J Digit Health. 2025 Aug 5;6(6):1264-1272. doi: 10.1093/ehjdh/ztaf088. eCollection 2025 Nov.
- Stergiou GS, Menti A, Mariglis D, Kollias A. The quest for accurate wearable blood pressure monitors. Hypertens Res. 2026 Mar;49(3):1025-1029. doi: 10.1038/s41440-025-02410-w. Epub 2025 Nov 5.
- Odeh VA, Chen Y, Wang W, Ding X. Recent Advances in the Wearable Devices for Monitoring and Management of Heart Failure. Rev Cardiovasc Med. 2024 Oct 28;25(10):386. doi: 10.31083/j.rcm2510386. eCollection 2024 Oct.
- Chia J, Bhatia KS, Mihailidou AS, Kanagaratnam LB. The Role of Ambulatory Blood Pressure Monitoring in Current Clinical Practice. Heart Lung Circ. 2022 Oct;31(10):1333-1340. doi: 10.1016/j.hlc.2022.06.670. Epub 2022 Aug 4.
Termíny studijních záznamů
Hlavní termíny studia
Začátek studia (Odhadovaný)
Začátek studia
Primární dokončení (Odhadovaný)
Primární dokončení
Dokončení studie (Odhadovaný)
Dokončení studie
Termíny zápisu do studia
První předloženo
První předloženo
První předloženo, které splnilo kritéria kontroly kvality
První předloženo, které splnilo kritéria kontroly kvality
První zveřejněno (Aktuální)
První zveřejněno
Aktualizace studijních záznamů
Poslední zveřejněná aktualizace (Aktuální)
Poslední zveřejněná aktualizace
Odeslaná poslední aktualizace, která splnila kritéria kontroly kvality
Odeslaná poslední aktualizace, která splnila kritéria kontroly kvality
Naposledy ověřeno
Naposledy ověřeno
Více informací
Termíny související s touto studií
Klíčová slova
- Hypertenze
- Maskovaná hypertenze
- Vysoký krevní tlak
- Fotopletysmografie (PPG)
- Riziko kardiovaskulárních onemocnění
- Screening a včasná detekce
- Ambulatorní monitorování krevního tlaku (ABPM)
- 24-Hour Ambulatory Blood Pressure
- Wearable Health Devices
- Smartwatch Blood Pressure Screening
- White-Coat Hypertension
Další relevantní podmínky MeSH
Další identifikační čísla studie
Další identifikační čísla studie
- Huawei-SU-HTN-01
Plán pro data jednotlivých účastníků (IPD)
Plánujete sdílet data jednotlivých účastníků (IPD)?
Popis plánu IPD
Časový rámec sdílení IPD
Kritéria přístupu pro sdílení IPD
Typ podpůrných informací pro sdílení IPD
- PROTOKOL STUDY
- MÍZA
- ICF
Informace o lécích a zařízeních, studijní dokumenty
Studuje lékový produkt regulovaný americkým FDA
Studuje produkt zařízení regulovaný americkým úřadem FDA
Tyto informace byly beze změn načteny přímo z webu clinicaltrials.gov. Máte-li jakékoli požadavky na změnu, odstranění nebo aktualizaci podrobností studie, kontaktujte prosím register@clinicaltrials.gov. Jakmile bude změna implementována na clinicaltrials.gov, bude automaticky aktualizována i na našem webu .