Real-World Validation of Huawei Wearable Technologies for Hypertension Risk Screening in Malaysia (Huawei-SunU1)
Background:
Hypertension (high blood pressure) is a major risk factor for heart disease and stroke, yet many people do not know they have it. In Malaysia, nearly half of all adults with elevated blood pressure are unaware of their condition. Standard clinic blood pressure checks can miss cases of high blood pressure or falsely flag temporary stress-related spikes. While 24-hour Ambulatory Blood Pressure Monitoring (ABPM) is the standard method for an accurate diagnosis, it is not practical for large-scale screening.
Purpose:
The purpose of this study is to evaluate whether a novel Huawei smartwatch, which uses photoplethysmography (PPG) technology to estimate blood pressure, can accurately identify individuals at high risk for hypertension in a real-world setting.
Study Design & Procedures:
Participants: Approximately 500 adults will be enrolled over a 24-month study period.
Screening: All participants will use the Huawei smartwatch screening tool to monitor their blood pressure estimates.
Validation: A subgroup of approximately 50 participants identified as high-risk by the smartwatch will undergo 24-hour ABPM (the clinical gold standard) to confirm whether they have high blood pressure.
Goal:
Researchers aim to validate the accuracy of the smartwatch screening tool against 24-hour ABPM to determine if wearable technology can support early detection and timely treatment of undiagnosed hypertension in the community.
Panoramica dello studio
Stato
Stato
Condizioni
Condizioni
Tipo di studio
Tipo di studio
Iscrizione (Stimato)
Iscrizione
Contatti e Sedi
Contatto studio
Contatto studio
- Nome: Yook Chin Chia
- Numero di telefono: +60374918622
- Email: ycchia@sunway.edu.my
Backup dei contatti dello studio
- Nome: Yee How Say
- Numero di telefono: +60374918622
- Email: yeehows@sunway.edu.my
Luoghi di studio
-
-
Selangor
-
Subang Jaya, Selangor, Malaysia, 47500
- Faculty of Medical and Life Sciences, Sunway University
-
Contatto:
- Yook Chin Chia
- Numero di telefono: +60374918622
- Email: ycchia@sunway.edu.my
-
Contatto:
- Yee How Say
- Numero di telefono: +60374918622
- Email: yeehows@sunway.edu.my
-
Investigatore principale:
- Yook Chin Chia
-
Investigatore principale:
- Yee How Say
-
-
Criteri di partecipazione
Criteri di ammissibilità
Criteri di ammissibilità
Età idonea allo studio
- Adulto
- Adulto più anziano
Accetta volontari sani
Metodo di campionamento
Popolazione di studio
Descrizione
Inclusion Criteria:
- Possesses the specific Huawei wearable required for the study
- Aged ≥18 years
- Malaysian citizen/permanent resident
- Understands English, Malay, or Chinese
- Able to download the required application on a compatible smartphone
- Willing to wear devices continuously
- Able to provide informed consent.
Exclusion Criteria:
- Current pregnancy
- Implantable cardiac devices (e.g., permanent pacemakers, implantable cardioverter-defibrillators [ICDs])
- Uncontrolled arrhythmias (e.g., active atrial fibrillation, frequent premature ventricular contractions [PVCs])
- Movement disorders or active tremors that interfere with signal acquisition (e.g., severe Parkinson's disease, essential tremor)
- Acute medical conditions, including:
- Acute coronary syndromes (e.g., recent myocardial infarction or unstable angina within the past 3 months)
- Acute cerebrovascular events (e.g., stroke or transient ischemic attack [TIA] within the past 3 months)
- Acute decompensated heart failure or severe respiratory distress
- Active, severe acute infections or systemic inflammatory illness (e.g., sepsis, acute pneumonia)
- Hypertensive emergencies or urgency
- Currently on antihypertensive medications
- Resting BP ≥180/110 mmHg.
Piano di studio
Come è strutturato lo studio?
Dettagli di progettazione
Numero di gruppi/coorti
Coorti e interventi
Gruppo / CoorteGruppo / Coorte |
|---|
|
Screening Population
Adults enrolled in the community screening phase who undergo non-invasive blood pressure estimation using the Huawei PPG smartwatch over the 24-month study period.
|
|
High-Risk Validation Cohort
A targeted subset of participants (n = 50) identified as high-risk for hypertension by the Huawei smartwatch screening tool, who undergo 24-hour Ambulatory Blood Pressure Monitoring (ABPM) to confirm their diagnostic status.
|
Cosa sta misurando lo studio?
Misure di risultato primarie
Misure di risultato primarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Diagnostic Accuracy (AUC-ROC) of the Huawei PPG Smartwatch Hypertension Risk Algorithm Compared to 24-Hour Ambulatory Blood Pressure Monitoring (ABPM)
Lasso di tempo: Month 11 through Month 18.
|
Measure Description: Area Under the Receiver Operating Characteristic Curve (AUC-ROC) evaluated by comparing the binary high-risk classification flags generated by the Huawei PPG smartwatch algorithm directly against diagnostic confirmation from clinical 24-hour ABPM. Unit of Measure: Score on a scale from 0.0 to 1.0 (where 0.5 represents chance performance and 1.0 represents perfect discrimination) |
Month 11 through Month 18.
|
|
Diagnostic Sensitivity of the Huawei PPG Smartwatch Hypertension Risk Algorithm
Lasso di tempo: Month 11 through Month 18.
|
Measure Description: The proportion of participants with true ABPM-confirmed hypertension who are correctly identified as high-risk by the Huawei PPG smartwatch algorithm. Unit of Measure: Percentage of true positive cases |
Month 11 through Month 18.
|
|
Diagnostic Specificity of the Huawei PPG Smartwatch Hypertension Risk Algorithm
Lasso di tempo: Month 11 through Month 18.
|
Measure Description: The proportion of participants without ABPM-confirmed hypertension who are correctly identified as low-risk by the Huawei PPG smartwatch algorithm. Unit of Measure: Percentage of true negative cases |
Month 11 through Month 18.
|
|
Positive Predictive Value (PPV) of the Huawei PPG Smartwatch Hypertension Risk Algorithm
Lasso di tempo: Month 11 through Month 18
|
Measure Description: The probability that a participant flagged as high-risk by the Huawei PPG smartwatch algorithm actually has hypertension confirmed by 24-hour ABPM. Unit of Measure: Percentage |
Month 11 through Month 18
|
|
Negative Predictive Value (NPV) of the Huawei PPG Smartwatch Hypertension Risk Algorithm
Lasso di tempo: Month 11 through Month 18
|
Measure Description: The probability that a participant flagged as low-risk by the Huawei PPG smartwatch algorithm actually does not have hypertension confirmed by 24-hour ABPM. Unit of Measure: Percentage |
Month 11 through Month 18
|
Misure di risultato secondarie
Misure di risultato secondarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Participant Adherence to Huawei Wearable Monitoring Protocols
Lasso di tempo: Month 1 through Month 12 (up to 12 months per participant)
|
Measure Description: The proportion of enrolled participants who achieve minimum complete wearable datasets, defined as continuous daily wear of the Huawei smartwatch to track PPG, heart rate, physical activity, and sleep metrics. Unit of Measure: Percentage of enrolled participants |
Month 1 through Month 12 (up to 12 months per participant)
|
|
Participant Usability and Acceptance Score Measured via Post-Monitoring Questionnaire
Lasso di tempo: At the completion of the wearable monitoring phase (Month 12).
|
Measure Description: Total score evaluating participant satisfaction, device comfort, and acceptance of the Huawei wearable for hypertension screening, measured using a standardized post-monitoring user experience questionnaire.
Unit of Measure: Mean score on a 1 to 5 Likert scale (where 1 = Strongly Disagree/Poor and 5 = Strongly Agree/Excellent)
|
At the completion of the wearable monitoring phase (Month 12).
|
Altre misure di risultato
Altre misure di risultato
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Prevalence of Wearable-Identified High-Risk Flags in the Study Population
Lasso di tempo: Month 3 through Month 12 (during community-based wearable data collection).
|
Measure Description: The percentage of participants flagged as high-risk for hypertension by the Huawei wearable platform within the total recruited screening cohort. Unit of Measure: Percentage of total participants |
Month 3 through Month 12 (during community-based wearable data collection).
|
|
Association Between Wearable Digital Biomarkers and ABPM-Confirmed Hypertension Risk
Lasso di tempo: Month 17 through Month 20 (during data cleaning and analytical modelling).
|
Measure Description: Adjusted Odds Ratios derived from logistic regression models assessing the relationship between wearable physiological parameters (heart rate, heart rate variability, sleep duration/quality, physical activity, and PPG blood pressure estimates) and ABPM-confirmed hypertension status. Unit of Measure: Adjusted Odds Ratio (OR) with 95% Confidence Intervals |
Month 17 through Month 20 (during data cleaning and analytical modelling).
|
Collaboratori e investigatori
Sponsor
Sponsor
Collaboratori
Collaboratori
Investigatori
Investigatori
- Investigatore principale: Yook Chin Chia, Sunway University
Pubblicazioni e link utili
Pubblicazioni generali
- Islam SMS, Chow CK, Daryabeygikhotbehsara R, Subedi N, Rawstorn J, Tegegne T, Karmakar C, Siddiqui MU, Lambert G, Maddison R. Wearable cuffless blood pressure monitoring devices: a systematic review and meta-analysis. Eur Heart J Digit Health. 2022 May 2;3(2):323-337. doi: 10.1093/ehjdh/ztac021. eCollection 2022 Jun.
- McEvoy JW, McCarthy CP, Bruno RM, Brouwers S, Canavan MD, Ceconi C, Christodorescu RM, Daskalopoulou SS, Ferro CJ, Gerdts E, Hanssen H, Harris J, Lauder L, McManus RJ, Molloy GJ, Rahimi K, Regitz-Zagrosek V, Rossi GP, Sandset EC, Scheenaerts B, Staessen JA, Uchmanowicz I, Volterrani M, Touyz RM; ESC Scientific Document Group. 2024 ESC Guidelines for the management of elevated blood pressure and hypertension. Eur Heart J. 2024 Oct 7;45(38):3912-4018. doi: 10.1093/eurheartj/ehae178. No abstract available.
- Wang Y, Zhou SS, Liu YQ, Li DD, Hu SY, Wang X, Yi L, Yu YN, Chen YD. Accessibility and usage patterns of wearable devices among Chinese adults: the Huawei Blood Pressure Health Study. Eur Heart J Digit Health. 2025 Aug 5;6(6):1264-1272. doi: 10.1093/ehjdh/ztaf088. eCollection 2025 Nov.
- Stergiou GS, Menti A, Mariglis D, Kollias A. The quest for accurate wearable blood pressure monitors. Hypertens Res. 2026 Mar;49(3):1025-1029. doi: 10.1038/s41440-025-02410-w. Epub 2025 Nov 5.
- Odeh VA, Chen Y, Wang W, Ding X. Recent Advances in the Wearable Devices for Monitoring and Management of Heart Failure. Rev Cardiovasc Med. 2024 Oct 28;25(10):386. doi: 10.31083/j.rcm2510386. eCollection 2024 Oct.
- Chia J, Bhatia KS, Mihailidou AS, Kanagaratnam LB. The Role of Ambulatory Blood Pressure Monitoring in Current Clinical Practice. Heart Lung Circ. 2022 Oct;31(10):1333-1340. doi: 10.1016/j.hlc.2022.06.670. Epub 2022 Aug 4.
Studiare le date dei record
Studia le date principali
Inizio studio (Stimato)
Inizio studio
Completamento primario (Stimato)
Completamento primario
Completamento dello studio (Stimato)
Completamento dello studio
Date di iscrizione allo studio
Primo inviato
Primo inviato
Primo inviato che soddisfa i criteri di controllo qualità
Primo inviato che soddisfa i criteri di controllo qualità
Primo Inserito (Effettivo)
Primo Inserito
Aggiornamenti dei record di studio
Ultimo aggiornamento pubblicato (Effettivo)
Ultimo aggiornamento pubblicato
Ultimo aggiornamento inviato che soddisfa i criteri QC
Ultimo aggiornamento inviato che soddisfa i criteri QC
Ultimo verificato
Ultimo verificato
Maggiori informazioni
Termini relativi a questo studio
Parole chiave
- Ipertensione
- Ipertensione mascherata
- Alta pressione sanguigna
- Fotopletismografia (PPG)
- Rischio di malattie cardiovascolari
- Screening e diagnosi precoce
- Monitoraggio Ambulatoriale della Pressione Arteriosa (ABPM)
- 24-Hour Ambulatory Blood Pressure
- Wearable Health Devices
- Smartwatch Blood Pressure Screening
- White-Coat Hypertension
Termini MeSH pertinenti aggiuntivi
Altri numeri di identificazione dello studio
Altri numeri di identificazione dello studio
- Huawei-SU-HTN-01
Piano per i dati dei singoli partecipanti (IPD)
Hai intenzione di condividere i dati dei singoli partecipanti (IPD)?
Descrizione del piano IPD
Periodo di condivisione IPD
Criteri di accesso alla condivisione IPD
Tipo di informazioni di supporto alla condivisione IPD
- STUDIO_PROTOCOLLO
- LINFA
- ICF
Informazioni su farmaci e dispositivi, documenti di studio
Studia un prodotto farmaceutico regolamentato dalla FDA degli Stati Uniti
Studia un dispositivo regolamentato dalla FDA degli Stati Uniti
Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .