Technolas 217z Laser With Zyoptix Aspheric Algorithm Used for the LASIK Treatment of Myopia and Myopic Astigmatism.
A Prospective Trial to Evaluate the Safety and Effectiveness of the Bausch & Lomb Technolas® 217z Laser With Zyoptix® Aspheric Algorithm When Used for the LASIK Treatment of Myopia and Myopic Astigmatism
Studieoversigt
Status
Status
Betingelser
Betingelser
Intervention / Behandling
Intervention / Behandling
Undersøgelsestype
Undersøgelsestype
Tilmelding (Faktiske)
Tilmelding
Fase
Fase
- Ikke anvendelig
Kontakter og lokationer
Studiesteder
-
-
Ontario
-
Toronto, Ontario, Canada, M5X1C8
- Lasik MD, 130 King Street
-
-
Deltagelseskriterier
Berettigelseskriterier
Berettigelseskriterier
Aldre berettiget til at studere
Tager imod sunde frivillige
Køn, der er berettiget til at studere
Beskrivelse
Inclusion Criteria:
- Subjects who were of legal age (at least 18) and had full legal capacity to volunteer.
- Subjects must be willing to have both eyes treated with the laser.
- Subjects must have up to -12.00D of absolute spherical myopia (not spherical equivalent) with up to -6.00D of refractive astigmatism, with a spherical equivalent up to -15.00D as measured by manifest subjective refraction.
- Subjects who had less than 0.75 D of difference in the preoperative manifest spherical equivalent and cycloplegic spherical equivalent subjective refractions.
(Manifest subjective refractions adjusted for optical infinity were used.)
- High contrast, manifest best spectacle-corrected distance logMAR visual acuity correctable to at least 0.3 in both eyes, and must not differ between eyes by more than 1 line (5 letters).
- Contact lens wearers must have 2 Orbscan measurements and 2 manifest subjective refractions taken preoperatively at least 1 week apart.
- Contact lens wearers must have gas permeable lenses discontinued for at least 3 weeks and soft lenses discontinued for at least 2 weeks prior to the preoperative evaluation.
Exclusion
- Subjects with uncorrected visual acuity better than 20/40 (Snellen) or 0.3 logMAR
- Subjects who were immuno-compromised or had a diagnosis of connective tissue disease, clinically significant atopic disease, diabetes, autoimmune diseases and other acute or chronic illnesses that increased the risk to the subject or could have confounded the outcomes of this study.
- Subjects with uncontrolled glaucoma or glaucoma who were under treatment in the study eye or with ocular hypertension.
- Subjects who demonstrated evidence of retinal vascular disease.
- Subjects who have an ocular muscle disorder including strabismus or nystagmus, or other disorders that affected fixation.
- Subjects who were at risk for angle closure or who had a potentially occludable angle, as evidenced by a shallow anterior chamber on slit lamp exam, confirmed by gonioscopy as deemed necessary by the study Investigator.
- Subjects who had undergone previous intraocular or corneal surgery of any kind, including any type of excimer laser surgery for either refractive or therapeutic purposes.
- Subjects who were taking systemic medications likely to affect wound healing such as corticosteroids or anti-metabolites or any other medications that may increase the risk to the subject or confound the outcomes of this study.
- Subjects for whom the combination of their preoperative corneal thickness and the planned operative parameters for the LASIK procedure could have resulted in less than 250 microns of remaining posterior corneal thickness below the flap postoperatively.
Studieplan
Hvordan er undersøgelsen tilrettelagt?
Design detaljer
- Primært formål: Behandling
- Tildeling: N/A
- Interventionel model: Enkelt gruppeopgave
- Maskning: Ingen (Åben etiket)
Antal våben
Våben og indgreb
Deltagergruppe / ArmDeltagergruppe / Arm |
Intervention / BehandlingIntervention / Behandling |
|---|---|
|
Eksperimentel: Technolas 217z Excimer Laser
Bausch & Lomb Zyoptix Aspheric Algorithm for LASIK correction of myopia and myopic astigmatism.
|
LASIK correction of myopia and myopic astigmatism
|
Hvad måler undersøgelsen?
Primære resultatmål
Primære resultatmål
Resultatmål |
Foranstaltningsbeskrivelse |
Tidsramme |
|---|---|---|
|
Percentage of Eyes With Uncorrected High Contrast Distance Visual Acuity of 20/40 or Better
Tidsramme: 3 months
|
3 months
|
|
|
Percentage of Eyes With Manifest Subjective Refraction (MRSE) Within ± 1.00 Diopter (D) and 0.50 D of Emmetropia
Tidsramme: 3 months
|
Manifest refractions (eyeglass prescriptions) have three components: sphere, cylinder, and axis.
Sphere and cylinder are both expressed in diopters, which is a continuous measure of the focusing power needed to make the image sharp on the retina.
Manifest refraction spherical equivalent (MRSE) is equal to sphere + ½ x cylinder.
It is expressed in diopters (D).
|
3 months
|
|
Preservation of Manifest Best Corrected High Contrast Distance Visual Acuity
Tidsramme: Baseline, 3 months
|
Mean change in best-corrected high contrast distance visual acuity.
Corrected means spectacle corrected.
Visual acuity (VA) was reported as letters read correctly, and the letters read were converted to logarithm of the minimum angle of resolution (logMAR) acuity.
The participant read rows of letters on a standard LogMAR chart until the participant misses 3 letters on a row with 5 letters.
The total number of correct letters were recorded.
A decrease in the number of correct letters indicates a decrease in visual acuity.
Acuity was averaged across visits.
|
Baseline, 3 months
|
|
Percentage of Eyes With Manifest Best-corrected Distance Photopic Visual Acuity Decrease of 2 or More Lines
Tidsramme: 3 months
|
3 months
|
|
|
Percentage of Eyes With > 2.00 Diopter of Induced Postoperative Refractive Cylinder
Tidsramme: 3 months
|
Manifest refractions (eyeglass prescriptions) have three components: sphere, cylinder, and axis.
Sphere and cylinder are both expressed in diopters, which is a continuous measure of the focusing power needed to make the image sharp on the retina.
Manifest refraction spherical equivalent (MRSE) is equal to sphere + ½ x cylinder.
It is expressed in diopters (D).
|
3 months
|
|
Percentage of Eyes That Had High Contrast Best Spectacle Corrected Visual Acuity (HCBSCVA) of 20/20 or Better at Baseline Had a Visual Outcome Worse Than 20/40 at 3 Months
Tidsramme: 3 months
|
Corrected means spectacle corrected.
Visual acuity (VA) was reported as letters read correctly, and the letters read were converted to logarithm of the minimum angle of resolution (logMAR) acuity.
The participant read rows of letters on a standard LogMAR chart until the participant misses 3 letters on a row with 5 letters.
The total number of correct letters were recorded.
A decrease in the number of correct letters indicates a decrease in visual acuity.
Acuity was averaged across visits.
|
3 months
|
Sekundære resultatmål
Sekundære resultatmål
Resultatmål |
Foranstaltningsbeskrivelse |
Tidsramme |
|---|---|---|
|
Percentage Change From Baseline in Higher Order Aberrations.
Tidsramme: Baseline, 3 months
|
The aberrations in the eye's optical system were measured over the entire pupil by the Bausch & Lomb Zywave II WaveFront Aberrometer, a Hartmann-Shack type device.
The pupil should have been at least 6 mm in diameter using 2.5% phenylephrine as a dilating agent to maximize the pupil diameter and hence the amount of wavefront information collected by the Zywave II.
The wavefront data was then fitted with Zernike polynomials representing the 2nd, 3rd, 4th and 5th order Zernike terms.
These terms represent the wavefront aberration of the eye.
|
Baseline, 3 months
|
Samarbejdspartnere og efterforskere
Sponsor
Sponsor
Efterforskere
Efterforskere
- Ledende efterforsker: Mark Cohen, MD, Lasik MD
Datoer for undersøgelser
Studer store datoer
Studiestart
Studiestart
Primær færdiggørelse (Faktiske)
Primær færdiggørelse
Studieafslutning (Faktiske)
Studieafslutning
Datoer for studieregistrering
Først indsendt
Først indsendt
Først indsendt, der opfyldte QC-kriterier
Først indsendt, der opfyldte QC-kriterier
Først opslået (Skøn)
Først opslået
Opdateringer af undersøgelsesjournaler
Sidste opdatering sendt (Faktiske)
Sidste opdatering sendt
Sidste opdatering indsendt, der opfyldte kvalitetskontrolkriterier
Sidste opdatering indsendt, der opfyldte kvalitetskontrolkriterier
Sidst verificeret
Sidst verificeret
Mere information
Begreber relateret til denne undersøgelse
Yderligere relevante MeSH-vilkår
Andre undersøgelses-id-numre
Andre undersøgelses-id-numre
- 564
Disse oplysninger blev hentet direkte fra webstedet clinicaltrials.gov uden ændringer. Hvis du har nogen anmodninger om at ændre, fjerne eller opdatere dine undersøgelsesoplysninger, bedes du kontakte register@clinicaltrials.gov. Så snart en ændring er implementeret på clinicaltrials.gov, vil denne også blive opdateret automatisk på vores hjemmeside .