Validation of a Miniature SpO2 Sensor in Healthy Subjects and Patients With Long-term Oxygen Therapy. (OxyFrame)
Validation of a Miniature SpO2 Sensor in Healthy Subjects and Patients With Long-term Oxygen Therapy. A Prospective, Method Comparison, Proof of Concept Study.
Studienübersicht
Status
Status
Bedingungen
Bedingungen
Intervention / Behandlung
Intervention / Behandlung
Detaillierte Beschreibung
Objectives:
- To determine accuracy of SpO2 measurements by the novel sensor compared to 2 reference pulse oximeters in healthy subjects undergoing a normobaric hypoxic challenge provoking SpO2 desaturation.
- To determine accuracy of SpO2 measurements by the novel sensor compared to 2 reference pulse oximeters and to arterial blood gas analysis in patients qualifying for LTOT, breathing room air at rest (lying, sitting, standing position) and during exercise (slow walking, 6-MWT, ergometer cycle).
Studientyp
Studientyp
Einschreibung (Tatsächlich)
Einschreibung
Phase
Phase
- Unzutreffend
Kontakte und Standorte
Studienorte
-
-
-
Bern, Schweiz, 3010
- Department of Pulmonary Medicine, University Hospital Bern
-
-
Teilnahmekriterien
Zulassungskriterien
Zulassungskriterien
Studienberechtigtes Alter
Akzeptiert gesunde Freiwillige
Studienberechtigte Geschlechter
Beschreibung
Healthy volunteers
Inclusion Criteria:
- Age 18 - 60 years,
- Non smoker,
- Good general health and physical fitness
- Informed Consent
Exclusion Criteria:
- Current or history of tobacco smoking (> 5 py)
- Current alcohol or drug abuse
- Heart disease or untreated arterial hypertension
- Pulmonary disease
- Epilepsy or other major neurologic disease
- Dark skin colour
- Known anemia
- Severe metabolic disease (e.g. diabetes)
- Pregnancy
Patients
Inclusion Criteria:
- Age ≥ 18 years
- Long term oxygen therapy at rest and/or on physical effort
- Hypoxemia in arterial blood gas analysis at rest PaO2 ≤ 55 mmHg or < 60 mmHg with PHT or polycythemia (men Hb > 174 g/L, women Hb > 140 g/L) or:
Oxygen desaturation on effort (SpO2 < 90%)
- Pulmonary disease including Chronic Obstructive Pulmonary Disease, Interstitial Lung Disease and Sarcoidosis
- Informed Consent
Exclusion Criteria:
- Severe hypoxemia (pO2 <40 mmHg)
- Severe hypercapnia (pCO2 >55 mmHg)
- Acute respiratory infection or systemic infection,
- Severe systolic heart disease (LVEF < 25%)
- Myocardial infarction during the last month before inclusion
- Unstable angina pectoris
- Severe aortic stenosis (mean gradient > 40mmHg, valve area <1 cm2)
- Rhythmogenic heart disease
- Severe or untreated arterial hypertension (blood pressure > 180mmHg systolic, >100mmHg diastolic)
- Anemia (Hb < 120 g/l)
- Known or suspected coagulation disorder (e.g. INR > 4)
- Severe or untreated metabolic disorder
- Neurologic disease or dementia
- Pregnancy
- History of non-compliance to medical treatment
- Current alcohol, drug abuse or current tobacco smoking
- Dark skin color
Studienplan
Wie ist die Studie aufgebaut?
Designdetails
- Zuteilung: N / A
- Interventionsmodell: Einzelgruppenzuweisung
- Maskierung: Keine (Offenes Etikett)
Anzahl der Arme
Waffen und Interventionen
Teilnehmergruppe / ArmTeilnehmergruppe / Arm |
Intervention / BehandlungIntervention / Behandlung |
|---|---|
|
Sonstiges: Healthy Subjects and patients
Validation of SpO2 sensor in healthy subjects as a first step.
Validation of SpO2 sensor in patients as a second step.
|
Validation of a miniaturized pulse oximetry sensor for integration into the frame of oxygen delivery glasses.
|
Was misst die Studie?
Primäre Ergebnismessungen
Primäre Ergebnismessungen
Ergebnis Maßnahme |
Maßnahmenbeschreibung |
Zeitfenster |
|---|---|---|
|
Assessment of accuracy in patients
Zeitfenster: At baseline
|
Assessment of the accuracy of the novel pulse oximeter sensor in patients qualifying for LTOT over the entire saturation range by calculating accuracy root mean square (ARMS).
Continuous SpO2 measurement of the novel pulse oximeter will be compared to measurements assessed by 2 standard pulse oximeters and arterial blood gas analysis.
|
At baseline
|
Sekundäre Ergebnismessungen
Sekundäre Ergebnismessungen
Ergebnis Maßnahme |
Maßnahmenbeschreibung |
Zeitfenster |
|---|---|---|
|
Assessment of accuracy in healthy volunteers
Zeitfenster: At baseline
|
Assessment of accuracy of the novel pulse oximeter sensor in healthy subjects undergoing a normobaric hypoxic challenge provoking SpO2 desaturation.
SpO2 will be measured continuously over the hole procedure and SpO2 range by the novel pulse oximeter sensor and the 2 reference pulse oximeters.
|
At baseline
|
Mitarbeiter und Ermittler
Sponsor
Sponsor
Ermittler
Ermittler
- Hauptermittler: Manuela Funke, MD, Department of Pulmonary Medicine, University Hospital Bern
Studienaufzeichnungsdaten
Haupttermine studieren
Studienbeginn
Studienbeginn
Primärer Abschluss (Tatsächlich)
Primärer Abschluss
Studienabschluss (Tatsächlich)
Studienabschluss
Studienanmeldedaten
Zuerst eingereicht
Zuerst eingereicht
Zuerst eingereicht, das die QC-Kriterien erfüllt hat
Zuerst eingereicht, das die QC-Kriterien erfüllt hat
Zuerst gepostet (Schätzen)
Zuerst gepostet
Studienaufzeichnungsaktualisierungen
Letztes Update gepostet (Schätzen)
Letztes Update gepostet
Letztes eingereichtes Update, das die QC-Kriterien erfüllt
Letztes eingereichtes Update, das die QC-Kriterien erfüllt
Zuletzt verifiziert
Zuletzt verifiziert
Mehr Informationen
Begriffe im Zusammenhang mit dieser Studie
Zusätzliche relevante MeSH-Bedingungen
Andere Studien-ID-Nummern
Andere Studien-ID-Nummern
- OxyFrame
Diese Informationen wurden ohne Änderungen direkt von der Website clinicaltrials.gov abgerufen. Wenn Sie Ihre Studiendaten ändern, entfernen oder aktualisieren möchten, wenden Sie sich bitte an register@clinicaltrials.gov. Sobald eine Änderung auf clinicaltrials.gov implementiert wird, wird diese automatisch auch auf unserer Website aktualisiert .