Validation of a Miniature SpO2 Sensor in Healthy Subjects and Patients With Long-term Oxygen Therapy. (OxyFrame)
Validation of a Miniature SpO2 Sensor in Healthy Subjects and Patients With Long-term Oxygen Therapy. A Prospective, Method Comparison, Proof of Concept Study.
Panoramica dello studio
Stato
Stato
Condizioni
Condizioni
Intervento / Trattamento
Intervento / Trattamento
Descrizione dettagliata
Objectives:
- To determine accuracy of SpO2 measurements by the novel sensor compared to 2 reference pulse oximeters in healthy subjects undergoing a normobaric hypoxic challenge provoking SpO2 desaturation.
- To determine accuracy of SpO2 measurements by the novel sensor compared to 2 reference pulse oximeters and to arterial blood gas analysis in patients qualifying for LTOT, breathing room air at rest (lying, sitting, standing position) and during exercise (slow walking, 6-MWT, ergometer cycle).
Tipo di studio
Tipo di studio
Iscrizione (Effettivo)
Iscrizione
Fase
Fase
- Non applicabile
Contatti e Sedi
Luoghi di studio
-
-
-
Bern, Svizzera, 3010
- Department of Pulmonary Medicine, University Hospital Bern
-
-
Criteri di partecipazione
Criteri di ammissibilità
Criteri di ammissibilità
Età idonea allo studio
Accetta volontari sani
Sessi ammissibili allo studio
Descrizione
Healthy volunteers
Inclusion Criteria:
- Age 18 - 60 years,
- Non smoker,
- Good general health and physical fitness
- Informed Consent
Exclusion Criteria:
- Current or history of tobacco smoking (> 5 py)
- Current alcohol or drug abuse
- Heart disease or untreated arterial hypertension
- Pulmonary disease
- Epilepsy or other major neurologic disease
- Dark skin colour
- Known anemia
- Severe metabolic disease (e.g. diabetes)
- Pregnancy
Patients
Inclusion Criteria:
- Age ≥ 18 years
- Long term oxygen therapy at rest and/or on physical effort
- Hypoxemia in arterial blood gas analysis at rest PaO2 ≤ 55 mmHg or < 60 mmHg with PHT or polycythemia (men Hb > 174 g/L, women Hb > 140 g/L) or:
Oxygen desaturation on effort (SpO2 < 90%)
- Pulmonary disease including Chronic Obstructive Pulmonary Disease, Interstitial Lung Disease and Sarcoidosis
- Informed Consent
Exclusion Criteria:
- Severe hypoxemia (pO2 <40 mmHg)
- Severe hypercapnia (pCO2 >55 mmHg)
- Acute respiratory infection or systemic infection,
- Severe systolic heart disease (LVEF < 25%)
- Myocardial infarction during the last month before inclusion
- Unstable angina pectoris
- Severe aortic stenosis (mean gradient > 40mmHg, valve area <1 cm2)
- Rhythmogenic heart disease
- Severe or untreated arterial hypertension (blood pressure > 180mmHg systolic, >100mmHg diastolic)
- Anemia (Hb < 120 g/l)
- Known or suspected coagulation disorder (e.g. INR > 4)
- Severe or untreated metabolic disorder
- Neurologic disease or dementia
- Pregnancy
- History of non-compliance to medical treatment
- Current alcohol, drug abuse or current tobacco smoking
- Dark skin color
Piano di studio
Come è strutturato lo studio?
Dettagli di progettazione
- Assegnazione: N / A
- Modello interventistico: Assegnazione di gruppo singolo
- Mascheramento: Nessuno (etichetta aperta)
Numero di armi
Armi e interventi
Gruppo di partecipanti / ArmGruppo di partecipanti / Arm |
Intervento / TrattamentoIntervento / Trattamento |
|---|---|
|
Altro: Healthy Subjects and patients
Validation of SpO2 sensor in healthy subjects as a first step.
Validation of SpO2 sensor in patients as a second step.
|
Validation of a miniaturized pulse oximetry sensor for integration into the frame of oxygen delivery glasses.
|
Cosa sta misurando lo studio?
Misure di risultato primarie
Misure di risultato primarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Assessment of accuracy in patients
Lasso di tempo: At baseline
|
Assessment of the accuracy of the novel pulse oximeter sensor in patients qualifying for LTOT over the entire saturation range by calculating accuracy root mean square (ARMS).
Continuous SpO2 measurement of the novel pulse oximeter will be compared to measurements assessed by 2 standard pulse oximeters and arterial blood gas analysis.
|
At baseline
|
Misure di risultato secondarie
Misure di risultato secondarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Assessment of accuracy in healthy volunteers
Lasso di tempo: At baseline
|
Assessment of accuracy of the novel pulse oximeter sensor in healthy subjects undergoing a normobaric hypoxic challenge provoking SpO2 desaturation.
SpO2 will be measured continuously over the hole procedure and SpO2 range by the novel pulse oximeter sensor and the 2 reference pulse oximeters.
|
At baseline
|
Collaboratori e investigatori
Sponsor
Sponsor
Investigatori
Investigatori
- Investigatore principale: Manuela Funke, MD, Department of Pulmonary Medicine, University Hospital Bern
Studiare le date dei record
Studia le date principali
Inizio studio
Inizio studio
Completamento primario (Effettivo)
Completamento primario
Completamento dello studio (Effettivo)
Completamento dello studio
Date di iscrizione allo studio
Primo inviato
Primo inviato
Primo inviato che soddisfa i criteri di controllo qualità
Primo inviato che soddisfa i criteri di controllo qualità
Primo Inserito (Stima)
Primo Inserito
Aggiornamenti dei record di studio
Ultimo aggiornamento pubblicato (Stima)
Ultimo aggiornamento pubblicato
Ultimo aggiornamento inviato che soddisfa i criteri QC
Ultimo aggiornamento inviato che soddisfa i criteri QC
Ultimo verificato
Ultimo verificato
Maggiori informazioni
Termini relativi a questo studio
Termini MeSH pertinenti aggiuntivi
Altri numeri di identificazione dello studio
Altri numeri di identificazione dello studio
- OxyFrame
Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .