Calibre vs. COSMED
Correlation of Cardiopulmonary Exercise Testing (CPET) VO2 Max Parameters Between the Calibre Device and COSMED Metabolic Cart
Studienübersicht
Status
Status
Bedingungen
Bedingungen
Intervention / Behandlung
Intervention / Behandlung
Studientyp
Studientyp
Einschreibung (Geschätzt)
Einschreibung
Phase
Phase
- Unzutreffend
Teilnahmekriterien
Zulassungskriterien
Zulassungskriterien
Studienberechtigtes Alter
- Erwachsene
Akzeptiert gesunde Freiwillige
Beschreibung
Inclusion Criteria:
- Age 18-55
- Able to speak English
- Able to provide informed consent
Exclusion Criteria:
- Acute myocardial infarction
- Unstable angina
- Syncope
- Uncontrolled asthma
- Arterial desaturation at rest on room air <85%
- Severe untreated arterial hypertension at rest (>200 mm Hg systolic, >120 mm Hg diastolic)
- Significant pulmonary hypertension
- Active thrombosis of the lower extremity, unless treated for at least 2 weeks prior to participation.
- Within 2 weeks of acute symptomatic pulmonary embolus
- Considered inappropriate to participate by Principal Investigator
- Previously experienced a heart attack
- Pregnancy
Studienplan
Wie ist die Studie aufgebaut?
Designdetails
- Hauptzweck: Sonstiges
- Zuteilung: Zufällig
- Interventionsmodell: Crossover-Aufgabe
- Maskierung: Keine (Offenes Etikett)
Anzahl der Arme
Waffen und Interventionen
Teilnehmergruppe / ArmTeilnehmergruppe / Arm |
Intervention / BehandlungIntervention / Behandlung |
|---|---|
|
Experimental: Calibre First Sequence
Participants undergo cardiopulmonary exercise testing using the investigational Calibre device during Visit 1, followed by testing with the COSMED Quark CPET reference system during Visit 2 (at least 2 days later).
|
Investigational wearable facemask device with integrated sensors for breath-by-breath measurement of respiratory gas exchange and ventilatory parameters during cardiopulmonary exercise testing, evaluated against the COSMED Quark CPET reference system.
|
|
Experimental: COSMED First Sequence
Participants undergo cardiopulmonary exercise testing using the COSMED Quark CPET reference system during Visit 1, followed by testing with the investigational Calibre device during Visit 2 (at least 2 days later).
|
FDA-cleared cardiopulmonary exercise testing (CPET) metabolic cart used as the reference comparator for measurement of respiratory gas exchange and ventilatory parameters during ramp cycle ergometry.
Participants wear a facemask connected to the system for breath-by-breath measurement of oxygen consumption (VO2), carbon dioxide production (VCO2), minute ventilation (VE), anaerobic threshold, and ventilatory efficiency during maximal exercise testing.
|
Was misst die Studie?
Primäre Ergebnismessungen
Primäre Ergebnismessungen
Ergebnis Maßnahme |
Maßnahmenbeschreibung |
Zeitfenster |
|---|---|---|
|
VO₂ peak correlation, as measured by Pearson or Spearman correlation coefficient between Calibre and COSMED
Zeitfenster: During CPET testing across Visit 1 and Visit 2, up to 7 days.
|
VO₂ peak measured during CPET will be compared between devices using Pearson or Spearman correlation coefficient, depending on data distribution.
|
During CPET testing across Visit 1 and Visit 2, up to 7 days.
|
|
VO₂ peak agreement, as measured by Bland-Altman bias and limits of agreement between Calibre and COSMED
Zeitfenster: During CPET testing across Visit 1 and Visit 2, up to 7 days.
|
VO₂ peak measured during CPET will be compared between devices using Bland-Altman analysis to estimate average bias and limits of agreement.
|
During CPET testing across Visit 1 and Visit 2, up to 7 days.
|
|
Anaerobic threshold correlation, as measured by Pearson or Spearman correlation coefficient between Calibre and COSMED
Zeitfenster: During CPET testing across Visit 1 and Visit 2, up to 7 days
|
Anaerobic threshold measured during CPET will be compared between devices using Pearson or Spearman correlation coefficient, depending on data distribution.
|
During CPET testing across Visit 1 and Visit 2, up to 7 days
|
|
Anaerobic threshold agreement, as measured by Bland-Altman bias and limits of agreement between Calibre and COSMED
Zeitfenster: During CPET testing across Visit 1 and Visit 2, up to 7 days
|
Anaerobic threshold measured during CPET will be compared between devices using Bland-Altman analysis to estimate average bias and limits of agreement.
|
During CPET testing across Visit 1 and Visit 2, up to 7 days
|
|
VE/VCO₂ correlation, as measured by Pearson or Spearman correlation coefficient between Calibre and COSMED
Zeitfenster: During CPET testing across Visit 1 and Visit 2, up to 7 days
|
Ventilatory equivalent for carbon dioxide (VE/VCO₂) measured during CPET will be compared between devices using Pearson or Spearman correlation coefficient, depending on data distribution.
|
During CPET testing across Visit 1 and Visit 2, up to 7 days
|
|
VE/VCO₂ agreement, as measured by Bland-Altman bias and limits of agreement between Calibre and COSMED
Zeitfenster: During CPET testing across Visit 1 and Visit 2, up to 7 days
|
Ventilatory equivalent for carbon dioxide (VE/VCO₂) measured during CPET will be compared between devices using Bland-Altman analysis to estimate average bias and limits of agreement.
|
During CPET testing across Visit 1 and Visit 2, up to 7 days
|
Mitarbeiter und Ermittler
Sponsor
Sponsor
Ermittler
Ermittler
- Hauptermittler: David MacLeod, M.B.B.S., Duke University
Studienaufzeichnungsdaten
Haupttermine studieren
Studienbeginn (Geschätzt)
Studienbeginn
Primärer Abschluss (Geschätzt)
Primärer Abschluss
Studienabschluss (Geschätzt)
Studienabschluss
Studienanmeldedaten
Zuerst eingereicht
Zuerst eingereicht
Zuerst eingereicht, das die QC-Kriterien erfüllt hat
Zuerst eingereicht, das die QC-Kriterien erfüllt hat
Zuerst gepostet (Tatsächlich)
Zuerst gepostet
Studienaufzeichnungsaktualisierungen
Letztes Update gepostet (Tatsächlich)
Letztes Update gepostet
Letztes eingereichtes Update, das die QC-Kriterien erfüllt
Letztes eingereichtes Update, das die QC-Kriterien erfüllt
Zuletzt verifiziert
Zuletzt verifiziert
Mehr Informationen
Begriffe im Zusammenhang mit dieser Studie
Schlüsselwörter
Andere Studien-ID-Nummern
Andere Studien-ID-Nummern
- Pro00118791
Plan für individuelle Teilnehmerdaten (IPD)
Planen Sie, individuelle Teilnehmerdaten (IPD) zu teilen?
Arzneimittel- und Geräteinformationen, Studienunterlagen
Studiert ein von der US-amerikanischen FDA reguliertes Arzneimittelprodukt
Studiert ein von der US-amerikanischen FDA reguliertes Geräteprodukt
Produkt, das in den USA hergestellt und aus den USA exportiert wird
Diese Informationen wurden ohne Änderungen direkt von der Website clinicaltrials.gov abgerufen. Wenn Sie Ihre Studiendaten ändern, entfernen oder aktualisieren möchten, wenden Sie sich bitte an register@clinicaltrials.gov. Sobald eine Änderung auf clinicaltrials.gov implementiert wird, wird diese automatisch auch auf unserer Website aktualisiert .