mGFR vs tGFR in AKI
Comparison of Agreement Between mGFR and tGFR in ICU Patients With Acute Kidney Injury
The goal of this study is to learn if the TGFR device (which uses telmapirazin to continuously measure real-time GFR) works to provide accurate and dynamic kidney function assessment in critically ill adults with acute kidney injury (AKI). It will also learn how tGFR compares with the gold standard measured GFR (mGFR) and other commonly used estimate methods. The main questions it aims to answer are:
- Does tGFR show good agreement with mGFR over the 0-12 hour period in AKI patients?
- How do tGFR, mGFR, 4-hour creatinine clearance, kinetic GFR (keGFR), and CKD-EPI eGFR compare across different AKI stages?
- What are the bias and variability of each method over time?
Researchers will compare tGFR to mGFR (iohexol clearance) to see if tGFR is a reliable alternative for bedside, continuous GFR monitoring. They will also compare tGFR against 4-hour creatinine clearance, keGFR, and eGFR to assess consistency across AKI stages.
Participants will:
- Wear a single-use tGFR sensor patch on the chest for continuous GFR monitoring
- Receive iohexol injection for mGFR measurement (gold standard)
- Provide blood and urine samples for 4-hour creatinine clearance and other tests
- Undergo repeated GFR assessments over the 12-hour study period
Descripción general del estudio
Estado
Estado
Condiciones
Condiciones
Intervención / Tratamiento
Intervención / Tratamiento
Tipo de estudio
Tipo de estudio
Inscripción (Estimado)
Inscripción
Fase
Fase
- Fase 4
Contactos y Ubicaciones
Estudio Contacto
Estudio Contacto
- Nombre: Bin Du
- Número de teléfono: 86+69155643
- Correo electrónico: dubin98@gmail.com
Criterios de participación
Criterio de elegibilidad
Criterio de elegibilidad
Edades elegibles para estudiar
- Adulto
- Adulto Mayor
Acepta Voluntarios Saludables
Descripción
Inclusion Criteria:
- CU stay ≤ 72 hours;
- meets KDIGO AKI criteria;
- age 18-85 years;
- able to complete blood/urine collection and monitoring
Exclusion Criteria:
- prior history of CKD stage 4 or higher;
- anuria;
- recent contrast agent exposure or kidney transplant history;
- pregnancy or breastfeeding; expected survival < 24 hours.
Plan de estudios
¿Cómo está diseñado el estudio?
Detalles de diseño
- Propósito principal: Diagnóstico
- Asignación: N / A
- Modelo Intervencionista: Asignación de un solo grupo
- Enmascaramiento: Ninguno (etiqueta abierta)
Número de brazos
Armas e Intervenciones
Grupo de participantes/brazoGrupo de participantes/brazo |
Intervención / TratamientoIntervención / Tratamiento |
|---|---|
|
Experimental: grupo de intervención
|
The tGFR detection system comprised 3 distinct components as follows: (i) the fluorescent tracer agent, Lumitrace, (ii) a single small module containing both the light source and photo detector attached to the upper chest area using a biocompatible adhesive, and (iii) a display monitor that digitizes the acquired fluorescent intensity data and contains the algorithms to convert these data to GFR. Background endogenous fluorescence was acquired after attachment of the module to the chest, followed by a 7 ml i.v.
bolus dose of Lumitrace.
The monitor shows the fluorescence intensity as it is being acquired and then displays the determined tGFR after data acquisition is completed.
|
¿Qué mide el estudio?
Medidas de resultado primarias
Medidas de resultado primarias
Medida de resultado |
Periodo de tiempo |
|---|---|
|
To validate the agreement of TGFR with mGFR in AKI patients at both the segmented and overall 0-12 hour levels.
Periodo de tiempo: 12 hours
|
12 hours
|
Colaboradores e Investigadores
Patrocinador
Patrocinador
Publicaciones y enlaces útiles
Publicaciones Generales
- Dorshow RB, Debreczeny MP, Goldstein SL. GFR Measurement Using Transdermal Detection Methodology. J Am Soc Nephrol. 2025 Feb 7;36(8):1592-1602. doi: 10.1681/ASN.0000000639.
- Dorshow RB, Debreczeny MP, Goldstein SL, Shieh JJ. Clinical validation of the novel fluorescent glomerular filtration rate tracer agent relmapirazin (MB-102). Kidney Int. 2024 Oct;106(4):679-687. doi: 10.1016/j.kint.2024.06.012. Epub 2024 Jul 2.
- Dixon JJ, Lane K, Dalton RN, Turner C, MacPhee IAM, Chis Ster I, Philips BJ. Continuous Infusion of Low-Dose Iohexol Measures Changing Glomerular Filtration Rate in Critically Ill Patients. Crit Care Med. 2018 Mar;46(3):e190-e197. doi: 10.1097/CCM.0000000000002870.
- Erley CM, Bader BD, Berger ED, Vochazer A, Jorzik JJ, Dietz K, Risler T. Plasma clearance of iodine contrast media as a measure of glomerular filtration rate in critically ill patients. Crit Care Med. 2001 Aug;29(8):1544-50. doi: 10.1097/00003246-200108000-00008.
- Molitoris BA. Measuring glomerular filtration rate in the intensive care unit: no substitutes please. Crit Care. 2013 Sep 4;17(5):181. doi: 10.1186/cc12876.
- Inker LA, Titan S. Measurement and Estimation of GFR for Use in Clinical Practice: Core Curriculum 2021. Am J Kidney Dis. 2021 Nov;78(5):736-749. doi: 10.1053/j.ajkd.2021.04.016. Epub 2021 Sep 11.
- Dorshow RB, Friedemann J, Ross-Jones J, Johnson JR, Vitale L, Riley IR, Debreczeny MP, Harr J, Trelford LP, Goldstein SL. Accuracy of Point-of-Care Transdermal GFR Methodology. Kidney Int Rep. 2026 Jan 22;11(4):103797. doi: 10.1016/j.ekir.2026.103797. eCollection 2026 Apr.
Fechas de registro del estudio
Fechas importantes del estudio
Inicio del estudio (Estimado)
Inicio del estudio
Finalización primaria (Estimado)
Finalización primaria
Finalización del estudio (Estimado)
Finalización del estudio
Fechas de registro del estudio
Enviado por primera vez
Enviado por primera vez
Primero enviado que cumplió con los criterios de control de calidad
Primero enviado que cumplió con los criterios de control de calidad
Publicado por primera vez (Actual)
Publicado por primera vez
Actualizaciones de registros de estudio
Última actualización publicada (Actual)
Última actualización publicada
Última actualización enviada que cumplió con los criterios de control de calidad
Última actualización enviada que cumplió con los criterios de control de calidad
Última verificación
Última verificación
Más información
Términos relacionados con este estudio
Palabras clave
Términos MeSH relevantes adicionales
Otros números de identificación del estudio
Otros números de identificación del estudio
- k9467
Plan de datos de participantes individuales (IPD)
¿Planea compartir datos de participantes individuales (IPD)?
Información sobre medicamentos y dispositivos, documentos del estudio
Estudia un producto farmacéutico regulado por la FDA de EE. UU.
Estudia un producto de dispositivo regulado por la FDA de EE. UU.
Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .