Evaluation of the Argos System to Estimate Cardiac Output

20 de julio de 2026 actualizado por: Christopher Lai, Bicetre Hospital

Evaluation of the Argos System to Estimate Cardiac Output in Critically Ill Patients

The goal of this observational study is to evaluate the reliability of the Argos system to track changes in cardiac output in critically ill patients. The main question it aims to answer is:

Does cardiac output changes induced by fluid administration or changes in norepinephrine dosage can be tracked by the Argos system, compared to transpulmonary thermodilution device.

Descripción general del estudio

Estado

Reclutamiento

Condiciones

Intervención / Tratamiento

Descripción detallada

In patients with acute circulatory failure, hemodynamic monitoring including cardiac index (CI) assessment is recommended for most severe patients. Transpulmonary thermodilution (TPTD) is one of the methods that are recommended for this purpose, in particular because it is considered as reliable in this type of patients. The PiCCO2 system (Pulsion Medical Systems, Getinge, Feldkirchen, Germany) relies on TPTD method and associate two methods to measure CI. On the one hand, TPTD is considered as one of the gold standards for CI evaluation (CITPTD). On the other hand, pulse contour analysis calculates stroke volume from the waveform of the systolic part of the arterial pressure curve (CIpwa). The way this estimation is performed and it is regularly re-evaluated differs among the devices that are available on the market. This estimation of CIpwa is calibrated when TPTD is performed. Thus, it allows a real-time pulse contour derived CI assessment, which is very precise. However, TPTD system are considered invasive, expensive and are not available in all settings.

The Argos monitor (Retia Medical, Valhalla, USA) uses the so-called multi-beat analysis (MBA) to estimate CI after analyzing the arterial pressure waveform over periods of multiple heartbeats (up to 20) and scaling CI estimations to biometric data. The algorithm estimates the theoretical central arterial blood pressure waveform using MBA, minimizing and avoiding the cofounding effect of the wave reflection phenomena. Based on this central arterial blood pressure waveform and biometric patient data, the system estimates cardiac output. The Argos monitor has a second advantage. It simply connects to the analogue output of the blood pressure signal from a compatible multi-parameter monitor and needs no additional specific pressure sensor. During cardiac surgery, the Argos monitor showed reasonable agreement and trending ability compared to the pulmonary artery catheter. However, in patients after cardiac surgery, including those with low cardiac output and arrhythmia, CI estimated by the device was not interchangeable with the continuous thermodilution method via a PAC.

The primary objective of the study is to evaluate the reliability of the Argos system to track changes of CI induced by changes in norepinephrine or fluid administration, compared to CITPTD in critically ill patients.

Tipo de estudio

De observación

Inscripción (Estimado)

120

Contactos y Ubicaciones

Esta sección proporciona los datos de contacto de quienes realizan el estudio e información sobre dónde se lleva a cabo este estudio.

Estudio Contacto

Ubicaciones de estudio

      • Le Kremlin-Bicêtre, Francia
        • Reclutamiento
        • Bicetre Hospital
        • Contacto:

Criterios de participación

Los investigadores buscan personas que se ajusten a una determinada descripción, denominada criterio de elegibilidad. Algunos ejemplos de estos criterios son el estado de salud general de una persona o tratamientos previos.

Criterio de elegibilidad

Edades elegibles para estudiar

  • Adulto
  • Adulto Mayor

Acepta Voluntarios Saludables

No

Método de muestreo

Muestra no probabilística

Población de estudio

Patients with shock defined by one sign of hypoperfusion:

  • systolic arterial pressure <90mmHg
  • mean arterial pressure <65mmHg
  • prolonged capillary refill time >3 sec
  • mottling score > 2
  • urine output < 0.1 mL/kg/h

Descripción

Inclusion Criteria:

  • Age ≥18 years old
  • Hospitalisation in intensive care
  • Haemodynamic monitoring with transpulmonary thermodilution already in place
  • Hemodynamic intervention deemed necessary by the clinician in charge of the patient

    • Administration of a fluid bolus
    • Change in norepinephrine dosage or performance of a vasopressor test

Exclusion Criteria:

  • Pregnancy
  • Other therapeutic modification during volume expansion or the vasopressor test
  • Refusal to participate by relatives of the patient or the patient himself or herself
  • Non availability of investigators

Plan de estudios

Esta sección proporciona detalles del plan de estudio, incluido cómo está diseñado el estudio y qué mide el estudio.

¿Cómo está diseñado el estudio?

Detalles de diseño

Cohortes e Intervenciones

Grupo / Cohorte
Intervención / Tratamiento
patients with shock

Patients with shock defined by one sign of hypoperfusion:

  • systolic arterial pressure <90mmHg
  • mean arterial pressure <65mmHg
  • prolonged capillary refill time >3 sec
  • mottling score > 2
  • urine output < 0.1 mL/kg/h
Giving a fluid bolus of 500mL of normal saline (NaCl 0.9%)
brief, transient increase in norepinephrine to achieve a 15mmHg increase in mean arterial pressure and evaluates if tissue perfusion improves

¿Qué mide el estudio?

Medidas de resultado primarias

Medida de resultado
Periodo de tiempo
To compare the changes in cardiac output provided by the Argos system induced by changes in norepinephrine or fluid bolus, compared to cardiac output provided by the transpulmonary thermodilution device in critically ill patients.
Periodo de tiempo: 1 hour (before and after intervention)
1 hour (before and after intervention)

Medidas de resultado secundarias

Medida de resultado
Medida Descripción
Periodo de tiempo
Accuracy of the Argos system
Periodo de tiempo: 1 hour (before and after intervention)
To evaluate the accuracy of the Argos system to measure absolute values of cardiac output compared to transpulmonary thermodilution
1 hour (before and after intervention)
Precision of the Argos system
Periodo de tiempo: 1 hour (before and after intervention)
To evaluate the precision and least significant change of cardiac output provided by the Argos system
1 hour (before and after intervention)

Colaboradores e Investigadores

Aquí es donde encontrará personas y organizaciones involucradas en este estudio.

Patrocinador

Fechas de registro del estudio

Estas fechas rastrean el progreso del registro del estudio y los envíos de resultados resumidos a ClinicalTrials.gov. Los registros del estudio y los resultados informados son revisados ​​por la Biblioteca Nacional de Medicina (NLM) para asegurarse de que cumplan con los estándares de control de calidad específicos antes de publicarlos en el sitio web público.

Fechas importantes del estudio

Inicio del estudio (Actual)

1 de febrero de 2026

Finalización primaria (Estimado)

31 de diciembre de 2026

Finalización del estudio (Estimado)

31 de enero de 2027

Fechas de registro del estudio

Enviado por primera vez

20 de julio de 2026

Primero enviado que cumplió con los criterios de control de calidad

20 de julio de 2026

Publicado por primera vez (Actual)

24 de julio de 2026

Actualizaciones de registros de estudio

Última actualización publicada (Actual)

24 de julio de 2026

Última actualización enviada que cumplió con los criterios de control de calidad

20 de julio de 2026

Última verificación

1 de julio de 2026

Más información

Términos relacionados con este estudio

Otros números de identificación del estudio

  • Argos study
  • CE-SRLF 24-100 (Otro identificador: SRLF)

Plan de datos de participantes individuales (IPD)

¿Planea compartir datos de participantes individuales (IPD)?

INDECISO

Información sobre medicamentos y dispositivos, documentos del estudio

Estudia un producto farmacéutico regulado por la FDA de EE. UU.

No

Estudia un producto de dispositivo regulado por la FDA de EE. UU.

No

producto fabricado y exportado desde los EE. UU.

No

Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .