SPIKE FLOW Endocardial Lead Safety and Performance Study (SPIKE FLOW)
A Prospective Multi-center Single Arm Study Evaluating the Safety and Performance of the SPIKE FLOW Endocardial Lead
The objective of this study is to assess the safety and performance of the SPIKE FLOW endocardial lead.
Primary objectives are to assess:
- safety in terms of freedom from Serious Adverse Events(SADEs);
- performance in terms of Mean Pacing Capture Threshold.
Secondary objectives are to assess:
- safety in terms of freedom from Adverse Events (ADEs);
- performance in terms of Mean Pacing Impedance;
- performance in terms of Sensed R-wave Amplitude. The study is a prospective, non-randomized, single-arm, multi-center investigation in 131 adult subjects with 3 investigational sites in Italy.
The study is expected to take approximately 1 year to enroll. Enrolled subjects will have visit timepoints at: baseline, lead insertion, lead removal, discharge and 30 days post lead removal.
Descripción general del estudio
Estado
Estado
Condiciones
Condiciones
Intervención / Tratamiento
Intervención / Tratamiento
Descripción detallada
SPIKE FLOW temporary leads are intended for temporary intravenous cardiac pacing and recording of intravenous cardiac electrical signals. The products are designed to be used for a continuous period of not more than 7 days. The indications for temporary pacing typically are the following:
- Acute myocardial infarction;
- Asystole;
- Bradycardia;
- AV block;
- Ventricular tachyarrhythmias;
- Overdrive suppression of tachyarrhythmias;
- High frequency pacing during percutaneous procedures;
- Cardiac surgery requiring backup pacing, surveillance of cardiac activity. The leads may be introduced, connected, and put into operation only by experienced and trained medical staff in heart pacing and percutaneous access.
The physicians entitled to place a temporary pacemaker should be familiar with the insertion equipment (puncture needle and sheath), electro-catheters and connectors, and the external pulse generator available in their own hospital settings.
Endocardial SPIKE FLOW family leads are single-use sterile devices, class III according to MDR (EU) 2017/745, used for temporary cardiac pacing, for short-term therapy of diseases which require the stimulation of the heart by means of electrical energy supplied by external pulse generators.
The rationale of the study is to evaluate, through clinical investigation according to its intended conditions of use, the effectiveness of SPIKE FLOW device's performance and to confirm the safety of the medical device.
Specifically, to assess safety and effectiveness of the SPIKE FLOW when using the balloon to facilitate lead insertion. Furthermore, we want to identify any risks (previously known or previously unidentified) associated with adverse events and/or complications that could occur during use in order to confirm or, if necessary, revise the assessments made during the risk management process.
The sample size requirement for this study is 131 adult subjects that can be enrolled in 3 different investigational sites in Italy. To ensure balanced enrollment and shorten the study's timeframe, each center will be required to enroll a minimum of 30 patients, while the remaining sample will be competitively recruited among centers until a total of 131 patients is reached. The enrollment is anticipated to take approximately 1 year.
Potential subjects must meet ALL of the following criteria to be eligible for enrollment into the study:
- Subject undergoing a surgical or an interventional procedure, requiring temporary cardiac pacing per investigator's medical judgment or subjects with bradycardia who are under evaluation to determine if the cause is transient or permanent;
- Subject is at least 18 years of age;
- Subject provides written informed consent to the study.
Potential subjects will be excluded if ANY of the following criteria apply:
- Subject has a co-existing temporary pacing or an implanted active permanent pacemaker or a cardiac defibrillator;
- Subject in cardiogenic shock;
- Subject is pregnant or breastfeeding;
- Subject had a stroke within 6 months prior to enrollment;
- Subject has renal insufficiency with creatinine >2 mg/dl;
- Subject with known bleeding diathesis;
- Subject with a history of deep vein thrombosis or pulmonary embolization within 6 months;
- Subject with known tricuspid valve disease that may impede catheter advancement to and withdrawal from the right ventricle;
- Subject with known intracardiac thrombus or vegetation on echocardiography;
- Subject with significant neck abnormalities such that placement of a right and/or left internal jugular vein line would be difficult;
- Subject has an active systemic infection or local infection at or around the insertion site;
- Subject is involved in another clinical study that could influence the safety or outcome measures of this study.
Potentially eligible subjects who meet all inclusion criteria, and no exclusion criteria and who have consented to participate in the trial will be enrolled in the study. Sites will maintain a subject enrollment log. Subjects can withdraw from the study at any time.
Study success will be defined as success on the primary endpoints. Primary Study Endpoint Safety: Freedom from Serious ADEs [Time Frame: from insertion up to 30 days post-lead removal].
Performance: Mean Pacing Capture Threshold (>2V) [Time Frame: from insertion to Lead removal (Up to 7 days)].
Secondary Study Endpoint Safety: Freedom from ADEs [Time Frame: from insertion up to 30 days post-lead removal].
Performance: Mean Pacing Impedance (Ohms) [Time Frame: from insertion to Lead removal (Up to 7 days)].
Impedance measured while pacing the heart (in multiple postures). Performance: Sensed R-wave Amplitude [Time Frame: from insertion to Lead removal (Up to 7 days)].
Summary statistics for sensed R-wave amplitude for Lead from insertion through removal.
Tipo de estudio
Tipo de estudio
Inscripción (Estimado)
Inscripción
Fase
Fase
- No aplica
Contactos y Ubicaciones
Ubicaciones de estudio
-
-
Bari
-
Acquaviva delle Fonti, Bari, Italia, 70021
- Aún no reclutando
- Ospedale Generale Regionale "F. Miulli"
-
Contacto:
- Dr. Federico Quadrini
- Número de teléfono: +39 0803054111
- Correo electrónico: federico.quadrini@tin.it
-
-
Massa-Carrara
-
Massa, Massa-Carrara, Italia, 54100
- Aún no reclutando
- Ospedale del Cuore 'G. Pasquinucci' Fondazione Toscana 'Gabriele Monasterio'
-
Contacto:
- Dr. Marco Ciardetti
- Número de teléfono: +39 0503153711
- Correo electrónico: mciard@ftgm.it
-
-
Pisa
-
Pisa, Pisa, Italia, 56126
- Reclutamiento
- Azienda Ospedaliero Universitaria Pisana
-
Contacto:
- Prof. Marco De Carlo
- Número de teléfono: +39 050992111
- Correo electrónico: marco.decarlo@unipi.it
-
-
Criterios de participación
Criterio de elegibilidad
Criterio de elegibilidad
Edades elegibles para estudiar
- Adulto
- Adulto Mayor
Acepta Voluntarios Saludables
Descripción
Inclusion Criteria:
- Subject undergoing a surgical or an interventional procedure, requiring temporary cardiac pacing per investigator's medical judgment or subjects with bradycardia who are under evaluation to determine if the cause is transient or permanent;
- Subject is at least 18 years of age;
- Subject provides written informed consent to the study.
Exclusion Criteria:
- Subject has a co-existing temporary pacing or an implanted active permanent pacemaker or a cardiac defibrillator;
- Subject in cardiogenic shock;
- Subject is currently in atrial fibrillation;
- Subject is pregnant or breastfeeding;
- Subject had a stroke within 6 months prior to enrollment;
- Subject has renal insufficiency with creatinine >2 mg/dl;
- Subject with known bleeding diathesis;
- Subject with a history of deep vein thrombosis or pulmonary embolism within 6 months;
- Subject with known tricuspid valve disease that may impede catheter advancement to and withdrawal from the right ventricle;
- Subject with known intracardiac thrombus or vegetation on echocardiography;
- Subject with significant neck abnormalities such that placement of a right and/or left internal jugular vein line would be difficult;
- Subject has an active systemic infection or local infection at or around the insertion site;
- Subject is involved in another clinical study that could influence the safety or outcome measures of this study.
Plan de estudios
¿Cómo está diseñado el estudio?
Detalles de diseño
- Propósito principal: Otro
- Asignación: N / A
- Modelo Intervencionista: Asignación de un solo grupo
- Enmascaramiento: Ninguno (etiqueta abierta)
Número de brazos
Armas e Intervenciones
Grupo de participantes/brazoGrupo de participantes/brazo |
Intervención / TratamientoIntervención / Tratamiento |
|---|---|
|
Experimental: experimental single-arm
|
Implantation of the device should be performed as per the manufacturer's Instructions for Use (IFU). Procedural details will be captured as appropriate on the procedural worksheets. Any Adverse Events or Device Deficiencies observed shall also be recorded. The following tests and examinations must be performed immediately after lead insertion:
Removal of the device should be performed as per the manufacturer's IFU. Procedural details will be captured as appropriate on the procedural worksheets. Any adverse events or device deficiencies observed shall also be recorded. The following tests and examinations must be performed immediately after lead removal:
|
¿Qué mide el estudio?
Medidas de resultado primarias
Medidas de resultado primarias
Medida de resultado |
Periodo de tiempo |
|---|---|
|
Safety: Freedom from Severe ADEs
Periodo de tiempo: from insertion up to 30 days post-lead removal
|
from insertion up to 30 days post-lead removal
|
|
Performance: Mean Pacing Capture Threshold (>2V)
Periodo de tiempo: from insertion to Lead removal (Up to 7 days)
|
from insertion to Lead removal (Up to 7 days)
|
Medidas de resultado secundarias
Medidas de resultado secundarias
Medida de resultado |
Medida Descripción |
Periodo de tiempo |
|---|---|---|
|
Safety: Freedom from ADEs
Periodo de tiempo: from insertion up to 30 days post-lead removal
|
from insertion up to 30 days post-lead removal
|
|
|
Performance: Sensed R-wave Amplitude
Periodo de tiempo: from insertion to Lead removal (Up to 7 days)
|
Summary statistics for sensed R-wave amplitude for Lead from insertion through removal
|
from insertion to Lead removal (Up to 7 days)
|
|
Performance: Mean Pacing Impedance (Ohms)
Periodo de tiempo: from insertion to Lead removal (Up to 7 days)
|
Impedance measured while pacing the heart (in multiple postures)
|
from insertion to Lead removal (Up to 7 days)
|
Colaboradores e Investigadores
Patrocinador
Patrocinador
Fechas de registro del estudio
Fechas importantes del estudio
Inicio del estudio (Actual)
Inicio del estudio
Finalización primaria (Estimado)
Finalización primaria
Finalización del estudio (Estimado)
Finalización del estudio
Fechas de registro del estudio
Enviado por primera vez
Enviado por primera vez
Primero enviado que cumplió con los criterios de control de calidad
Primero enviado que cumplió con los criterios de control de calidad
Publicado por primera vez (Actual)
Publicado por primera vez
Actualizaciones de registros de estudio
Última actualización publicada (Actual)
Última actualización publicada
Última actualización enviada que cumplió con los criterios de control de calidad
Última actualización enviada que cumplió con los criterios de control de calidad
Última verificación
Última verificación
Más información
Términos relacionados con este estudio
Palabras clave
Términos MeSH relevantes adicionales
- Enfermedad del sistema de conducción cardíaca
- Enfermedades cardiovasculares
- Procesos Patológicos
- Enfermedades cardíacas
- Arritmias Cardiacas
- Bloqueo cardíaco
- Taquicardia
- Condiciones Patológicas, Signos y Síntomas
- Bloqueo auriculoventricular
- Paro cardíaco
- Taquicardia Ventricular
- Bradicardia
- Procedimientos quirúrgicos, operativo
- Eliminación de dispositivos
- Extracción de plomo del dispositivo
Otros números de identificación del estudio
Otros números de identificación del estudio
- IT-26-03-056934
Plan de datos de participantes individuales (IPD)
¿Planea compartir datos de participantes individuales (IPD)?
Información sobre medicamentos y dispositivos, documentos del estudio
Estudia un producto farmacéutico regulado por la FDA de EE. UU.
Estudia un producto de dispositivo regulado por la FDA de EE. UU.
Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .