Effect of Double-Tap Optimization on Valve Expansion and Hemodynamics in Balloon-expandable Transcatheter Aortic Valve Implantation: A Randomized Clinical Trial (DO-OPT TAVI Trial)
This study will compare two balloon inflation strategies during transcatheter aortic valve implantation (TAVI), a procedure used to replace a severely narrowed aortic valve through a catheter. A balloon-expandable replacement valve may not fully expand during its initial placement. An additional inflation using the same balloon and the same inflation volume, called "Double-Tap," may improve valve expansion and function. Its benefits and risks need to be evaluated in a randomized trial.
This prospective, multicenter study aims to include 250 adults with symptomatic severe aortic stenosis undergoing TAVI with an Edwards SAPIEN-family balloon-expandable valve. After successful valve implantation, participants for whom an additional inflation is considered technically feasible and clinically appropriate will be randomly assigned in equal numbers to Double-Tap or conventional single inflation. Patients requiring therapeutic post-dilatation for moderate or greater leakage around the implanted valve will not undergo randomization.
The main question is whether Double-Tap improves valve expansion immediately after TAVI, as measured using fluoroscopic X-ray images. The study will also compare blood flow through the valve, valve expansion and signs of valve thrombosis on cardiac computed tomography, procedure-related complications, and clinical outcomes. Participants will undergo echocardiography before discharge and at 3-6 and 12 months, and cardiac computed tomography at 3-6 months. Clinical follow-up will continue for at least 12 months, with annual follow-up thereafter until the last enrolled participant completes the 12-month visit.
Descripción general del estudio
Estado
Estado
Condiciones
Condiciones
Intervención / Tratamiento
Intervención / Tratamiento
Tipo de estudio
Tipo de estudio
Inscripción (Estimado)
Inscripción
Fase
Fase
- No aplica
Contactos y Ubicaciones
Estudio Contacto
Estudio Contacto
- Nombre: Young-Guk Ko, Professor
- Número de teléfono: +82-22228-8460
- Correo electrónico: ygko@yuhs.ac
Criterios de participación
Criterio de elegibilidad
Criterio de elegibilidad
Edades elegibles para estudiar
Acepta Voluntarios Saludables
Géneros elegibles para el estudio
Descripción
Inclusion Criteria:
- Patients ≥ 19 years old
- Patients undergoing TAVI using a Sapien family valve for severe AS
- Provision of informed consent
Exclusion criteria:
- Subannular calcification
- Bicuspid aortic valve with raphe calcification
- Valve-in-valve TAVI procedure
- Planned major surgery
- Cardiogenic shock or hemodynamic instability
- Chronic kidney disease stage 3b, 4, or 5 (eGFR <45mL/min)
- Thrombocytopenia (platelet count <100,000) or contraindication to antiplatelet therapy or anticoagulation despite clinical indications
Plan de estudios
¿Cómo está diseñado el estudio?
Detalles de diseño
- Propósito principal: Tratamiento
- Asignación: Aleatorizado
- Modelo Intervencionista: Asignación paralela
- Enmascaramiento: Ninguno (etiqueta abierta)
Número de brazos
Armas e Intervenciones
Grupo de participantes/brazoGrupo de participantes/brazo |
Intervención / TratamientoIntervención / Tratamiento |
|---|---|
|
Experimental: Double-Tap
Participants undergo TAVI with an Edwards SAPIEN-family balloon-expandable valve. After successful initial valve implantation and randomization, one additional balloon inflation is performed using the same delivery balloon and the same inflation volume used for initial deployment. The additional inflation is performed immediately after confirming appropriate valve position and procedural stability. |
Following successful implantation of an Edwards SAPIEN-family balloon-expandable transcatheter heart valve, one additional balloon inflation is performed using the same delivery balloon and the same inflation volume used for initial valve deployment. Balloon inflation volume is not increased for the Double-Tap inflation. The procedure is performed after randomization and confirmation of appropriate valve position and procedural stability. |
|
Comparador activo: No Double-Tap
Participants undergo standard TAVI with an Edwards SAPIEN-family balloon-expandable valve deployed using a single balloon inflation. After successful initial valve implantation and randomization, no planned additional balloon inflation for Double-Tap optimization is performed. |
An Edwards SAPIEN-family balloon-expandable transcatheter heart valve is implanted using a single balloon inflation according to standard practice and the manufacturer's instructions for use. No planned additional balloon inflation for Double-Tap optimization is performed after randomization. |
¿Qué mide el estudio?
Medidas de resultado primarias
Medidas de resultado primarias
Medida de resultado |
Medida Descripción |
Periodo de tiempo |
|---|---|---|
|
Percentage valve expansion measured by fluoroscopy
Periodo de tiempo: immediately after TAVI procedure
|
Expansion of the implanted transcatheter heart valve, expressed as a percentage of the manufacturer's nominal cross-sectional area for the implanted valve size. Valve diameters are measured at the inflow, midframe, and outflow levels in three-cusp and cusp-overlap fluoroscopic views. At each level, the diameters from the two views are averaged, and the cross-sectional area is calculated using the circular area formula. Fluoroscopic images are evaluated by independent imaging core laboratory assessors blinded to treatment assignment and clinical outcomes. |
immediately after TAVI procedure
|
Colaboradores e Investigadores
Patrocinador
Patrocinador
Fechas de registro del estudio
Fechas importantes del estudio
Inicio del estudio (Estimado)
Inicio del estudio
Finalización primaria (Estimado)
Finalización primaria
Finalización del estudio (Estimado)
Finalización del estudio
Fechas de registro del estudio
Enviado por primera vez
Enviado por primera vez
Primero enviado que cumplió con los criterios de control de calidad
Primero enviado que cumplió con los criterios de control de calidad
Publicado por primera vez (Actual)
Publicado por primera vez
Actualizaciones de registros de estudio
Última actualización publicada (Actual)
Última actualización publicada
Última actualización enviada que cumplió con los criterios de control de calidad
Última actualización enviada que cumplió con los criterios de control de calidad
Última verificación
Última verificación
Más información
Términos relacionados con este estudio
Términos MeSH relevantes adicionales
Otros números de identificación del estudio
Otros números de identificación del estudio
- 4-2026-1423
Plan de datos de participantes individuales (IPD)
¿Planea compartir datos de participantes individuales (IPD)?
Información sobre medicamentos y dispositivos, documentos del estudio
Estudia un producto farmacéutico regulado por la FDA de EE. UU.
Estudia un producto de dispositivo regulado por la FDA de EE. UU.
Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .