PRecision Integrated Saturation Monitor (PRISM)

30 de junio de 2026 actualizado por: Danielle Gottlieb Sen, Le Bonheur Children's Hospital

PRISM: PRecision Integrated Saturation Monitor A Novel Approach to Monitoring Oxygen Saturation in Children A Prospective, Observational, Comparative Validation Clinical Study to Evaluate the Accuracy of a Novel Abdominal Pulse Oximeter With Integrated Motion Detection and Skin Tone Calibration Against Arterial Blood Gas (SaO₂) Measurements and Conventional Fingertip Pulse Oximetry (SpO₂) in Pediatric Patients.

The purpose of this study is to find out if an abdominal pulse oximeter device is as accurate as a conventional fingertip pulse oximeter in pediatric patients. In this study, we will be collecting data from your medical record while you are being treated in the hospital for your clinical care. Your baby will be in the study for 3 days. There will be no additional study visits. We will collect information already being obtained while your baby is at the hospital.

Descripción general del estudio

Descripción detallada

Conventional pulse oximeters are prone to motion artifacts due to their placement on high-activity limbs. Traditionally, inpatient pulse oximetry relies on clinical staff to verify the accuracy of readings and distinguish between real and false desaturation events. However, in outpatient environments, caregivers are left to manage and interpret signals that may not provide reliable information about their child's health. PRISM aims to improve pulse oximetry monitoring for children. We will test an innovative pulse oximeter prototype that mitigates motion artifacts by shifting the monitoring location from a high-activity limb to the abdomen while incorporating motion detection capabilities. Previously, we have successfully developed an abdomen-worn prototype for collection of ECG signals for infant monitoring (Figure 1). We obtained high-quality signals with a robust correlation to the gold-standard ECG. Our preliminary results demonstrate the feasibility of collecting physiological signals from the abdomen, supporting its use as a novel, baby-friendly monitoring site. Figure 1. Comparison of ECG signals between gold-standard chest ECG and abdomen ECG. Protocol Version 2: 11 June 2026 IRB NUMBER: STUDY00000178 IRB APPROVAL DATE: 6/29/2026 IRB EXPIRATION DATE: 6/2/2027 We will incorporate a commercially available pulse oximetry sensor into a biocompatible, flexible, 3D printed housing. The child-focused design will be specific for application to the abdomen and will clip onto the diaper or pants of the baby; the tension from the diaper or pants will provide ample force to ensure sufficient contact between the device and the skin to achieve robust signals (Figure 2). The device will be removed with each diaper change. Figure 2. Schematic demonstrating the proposed pulse oximeter and its abdominal application As pulse oximetry uses optical sensors, external factors such as ambient light may interfere with data acquisition. Additionally, maintaining reliable contact with the skin is crucial to ensure the emitted light penetrates the underlying tissue, rather than reflecting off the skin's surface. Here, we will design the sensor housing to shield the pulse oximetry sensors from ambient light and provide a contact surface flexible enough for direct skin-to-sensor contact. The PRISM prototype is intended to minimize the impact of motion artifacts, a common phenomenon that leads to inaccurate readings and unnecessary alarms. By shifting the monitoring location from high-activity limbs to the more stable abdomen, we aim to enhance the accuracy and reliability of the device's measurements compared to traditional pulse oximeters. This design choice is particularly relevant for infants and young children, who often experience restlessness and movement during routine monitoring.

Successful accomplishment of this will result in a working prototype with motion artifact-resistant SpO2 reading error of less than 3.0% RMSE from SaO2. This development will allow the caregiver to continuously monitor accurate oxygen status Hypoxemic conditions, or medical problems stemming from low oxygen saturation, can result in half a million deaths annually. Oxygen saturation levels, which determine hypoxic conditions, can be measured directly from arterial blood. This is the gold standard of oxygen saturation measurement and is often called SaO2. Oxygen saturation measured from pulse oximetry, referred to as SpO2, uses light transmission and can allow for early detection of hypoxic conditions. It's a popular belief that melanin can absorb light in pulse oximetry, leading to the overestimation of SpO2 in darker skin. Although individuals with darker skin are observed to have a SpO2 overestimation, clinicians have yet to devise a method to accurately predict oxygen saturation given this overestimation[1]. Various clinical data has been curated to specifically address the issue of biases and disparities of pulse oximeter measurements in darker skin tones including the MIT Critical Datathon 2023 utilizing MIMIC IV and the blood-gas and oximetry linked dataset (BOLD)[2]. One issue with these datasets is race is often used as a proxy for skin tone, but there is great variability in skin color, even within a single race. Without an algorithm to account for how variation in skin tone biases in SpO2 measurement, a disproportionate amount of people of color may continue to receive incorrect diagnoses, leading to health inequities.

These incorrect diagnoses can lead to hidden hypoxemia, which is a condition where the gold standard oxygen saturation of a patient is less than 88% while the pulse oximeter measurement reads over 88%5. Studies show a significant disparity with black patients experiencing nearly three times the frequency of undetected hypoxemia compared to white patients6. They also emphasize the need to integrate pulse oximetry data with other clinical and patient-reported information to address this bias in medical technology.

These results are expected to generate evidence for proof-of-concept for this advanced next-generation pulse oximetry to be further tested and commercialized.

Tipo de estudio

De observación

Inscripción (Estimado)

100

Contactos y Ubicaciones

Esta sección proporciona los datos de contacto de quienes realizan el estudio e información sobre dónde se lleva a cabo este estudio.

Ubicaciones de estudio

    • Tennessee
      • Memphis, Tennessee, Estados Unidos, 38105
        • Reclutamiento
        • Le Bonheur Children's Hospital
        • Contacto:

Criterios de participación

Los investigadores buscan personas que se ajusten a una determinada descripción, denominada criterio de elegibilidad. Algunos ejemplos de estos criterios son el estado de salud general de una persona o tratamientos previos.

Criterio de elegibilidad

Edades elegibles para estudiar

  • Niño

Acepta Voluntarios Saludables

No

Método de muestreo

Muestra no probabilística

Población de estudio

Eligible infants hospitalized in the intensive care unit (PICU, PCVICU,NICU) who have already had an arterial line placed and on continuous pulse oximetry monitoring will be considered for enrollment and will be identified and approached for participation during their admission.

Descripción

Inclusion Criteria:

  • 0-12 months of age. (Must be less than 1 year of life for the pulse oximeter portion of the study) Weight greater than 2 Kg at time of study
  • Admitted to an intensive care unit. (Not applicable for outpatient cohort)
  • Arterial line in place as standard of care (Not applicable for outpatient cohort)
  • Informed consent provided by parent or LAR.

Exclusion Criteria:

  • Open Sternum
  • Skin abnormalities at the proposed sensor site (e.g., rash, eczema, wounds, stoma, or infection)
  • Known allergy or sensitivity to adhesives or materials used in the sensor
  • Excessive movement or irritability that precludes safe and effective placement of the sensor
  • Parent or legal guardian unable to provide informed consent

Plan de estudios

Esta sección proporciona detalles del plan de estudio, incluido cómo está diseñado el estudio y qué mide el estudio.

¿Cómo está diseñado el estudio?

Detalles de diseño

¿Qué mide el estudio?

Medidas de resultado primarias

Medida de resultado
Medida Descripción
Periodo de tiempo
Abdominal SpO2 versus SaO2
Periodo de tiempo: From enrollment to end of study at 72 hours.
The primary outcome is the accuracy of abdominal SpO₂ versus SaO₂, measured as root mean square error (RMSE); target ≤ 3.0% RMSE.
From enrollment to end of study at 72 hours.

Colaboradores e Investigadores

Aquí es donde encontrará personas y organizaciones involucradas en este estudio.

Fechas de registro del estudio

Estas fechas rastrean el progreso del registro del estudio y los envíos de resultados resumidos a ClinicalTrials.gov. Los registros del estudio y los resultados informados son revisados ​​por la Biblioteca Nacional de Medicina (NLM) para asegurarse de que cumplan con los estándares de control de calidad específicos antes de publicarlos en el sitio web público.

Fechas importantes del estudio

Inicio del estudio (Estimado)

1 de agosto de 2026

Finalización primaria (Estimado)

1 de agosto de 2028

Finalización del estudio (Estimado)

1 de diciembre de 2028

Fechas de registro del estudio

Enviado por primera vez

30 de junio de 2026

Primero enviado que cumplió con los criterios de control de calidad

30 de junio de 2026

Publicado por primera vez (Actual)

7 de julio de 2026

Actualizaciones de registros de estudio

Última actualización publicada (Actual)

7 de julio de 2026

Última actualización enviada que cumplió con los criterios de control de calidad

30 de junio de 2026

Última verificación

1 de junio de 2026

Más información

Términos relacionados con este estudio

Información sobre medicamentos y dispositivos, documentos del estudio

Estudia un producto farmacéutico regulado por la FDA de EE. UU.

No

Estudia un producto de dispositivo regulado por la FDA de EE. UU.

No

Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .