Novel Non-Invasive Monitoring Parameter in Healthy Volunteers
Novel Cardio-Respiratory Parameter in Healthy Volunteers
Tutkimuksen yleiskatsaus
Tila
Yksityiskohtainen kuvaus
Opintotyyppi
Ilmoittautuminen (Todellinen)
Yhteystiedot ja paikat
Opiskelupaikat
-
-
Colorado
-
Boulder, Colorado, Yhdysvallat, 80301
- Covidien- RMS
-
-
Osallistumiskriteerit
Kelpoisuusvaatimukset
Opintokelpoiset iät
Hyväksyy terveitä vapaaehtoisia
Sukupuolet, jotka voivat opiskella
Näytteenottomenetelmä
Tutkimusväestö
Kuvaus
Inclusion Criteria:
- Subject is 18 or older
- Subject is willing and able to provide written consent
Exclusion Criteria:
- Severe contact allergies to standard adhesive materials
- Subjects with abnormalities that may prevent proper application of the device
- Women who are pregnant or lactating
- Subjects with significant Arrhythmias
Opintosuunnitelma
Miten tutkimus on suunniteltu?
Suunnittelun yksityiskohdat
Kohortit ja interventiot
Ryhmä/Kohortti |
|---|
|
Healthy Volunteers
Healthy volunteers ASA Class 1
|
Mitä tutkimuksessa mitataan?
Ensisijaiset tulostoimenpiteet
Tulosmittaus |
Toimenpiteen kuvaus |
Aikaikkuna |
|---|---|---|
|
The Covidien Nellcor Respiration Rate Software Shall Determine Respiration Rate Measured as Mean and Standard Deviation With Accuracy That is Non-inferior to Predicate Device.
Aikaikkuna: Participants were monitorerd on average of 30 minute period
|
Mean and standard deviations of respiration rates collected from healthy volunteers were compared between Covidien Respiration Rate Software, Transthoracic Impedance and Overscored End-Tidal Carbon Dioxide Waveforms.
Each volunteer served as its own control.
|
Participants were monitorerd on average of 30 minute period
|
Toissijaiset tulostoimenpiteet
Tulosmittaus |
Toimenpiteen kuvaus |
Aikaikkuna |
|---|---|---|
|
The Covidien Nellcor Respiration Rate Software Shall Calculate Respiration Rate With a Root Mean Square Difference (RMSD) of < 3 Breaths Per Minute Compared With a End-Tidal Carbon Dioxide Waveforms, With 95% Confidence.
Aikaikkuna: Participants were monitored on average of 30 minute periods
|
The data used for analysis consisted of multiple simultaneous measures of RR_TTI, RR_V1.0 and RR_EtCO2 for each subject. Given N sets of simultaneous RR values for each of P patients, the simultaneous RR values were used to calculate a pair of RMSD values for each subject. The two RMSD values, RMSDRR_V1.0_vs_EtCO2, and RMSDTTI_vs_EtCO2, quantify the mean absolute difference in simultaneous estimates of respiratory rate between RR_V1.0 compared with RR_EtCO2 and RR_TTI compared with RR_EtCO2, respectively for the subjects with 95% confidence of mean ± 3.38 BrPM. The Measure type is Number and represents the RMSD of Covidien Nellcor Respiration Rate Software |
Participants were monitored on average of 30 minute periods
|
Yhteistyökumppanit ja tutkijat
Sponsori
Tutkijat
- Päätutkija: Scott Kelley, MD, Medtronic - MITG
Opintojen ennätyspäivät
Opi tärkeimmät päivämäärät
Opiskelun aloitus
Ensisijainen valmistuminen (Todellinen)
Opintojen valmistuminen (Todellinen)
Opintoihin ilmoittautumispäivät
Ensimmäinen lähetetty
Ensimmäinen toimitettu, joka täytti QC-kriteerit
Ensimmäinen Lähetetty (Arvio)
Tutkimustietojen päivitykset
Viimeisin päivitys julkaistu (Arvio)
Viimeisin lähetetty päivitys, joka täytti QC-kriteerit
Viimeksi vahvistettu
Lisää tietoa
Tähän tutkimukseen liittyvät termit
Avainsanat
Muut tutkimustunnusnumerot
- COVMOPR0132
Nämä tiedot haettiin suoraan verkkosivustolta clinicaltrials.gov ilman muutoksia. Jos sinulla on pyyntöjä muuttaa, poistaa tai päivittää tutkimustietojasi, ota yhteyttä register@clinicaltrials.gov. Heti kun muutos on otettu käyttöön osoitteessa clinicaltrials.gov, se päivitetään automaattisesti myös verkkosivustollemme .