Novel Non-Invasive Monitoring Parameter in Healthy Volunteers
Novel Cardio-Respiratory Parameter in Healthy Volunteers
Studie Overzicht
Toestand
Gedetailleerde beschrijving
Studietype
Inschrijving (Werkelijk)
Contacten en locaties
Studie Locaties
-
-
Colorado
-
Boulder, Colorado, Verenigde Staten, 80301
- Covidien- RMS
-
-
Deelname Criteria
Geschiktheidscriteria
Leeftijden die in aanmerking komen voor studie
Accepteert gezonde vrijwilligers
Geslachten die in aanmerking komen voor studie
Bemonsteringsmethode
Studie Bevolking
Beschrijving
Inclusion Criteria:
- Subject is 18 or older
- Subject is willing and able to provide written consent
Exclusion Criteria:
- Severe contact allergies to standard adhesive materials
- Subjects with abnormalities that may prevent proper application of the device
- Women who are pregnant or lactating
- Subjects with significant Arrhythmias
Studie plan
Hoe is de studie opgezet?
Ontwerpdetails
Cohorten en interventies
Groep / Cohort |
|---|
|
Healthy Volunteers
Healthy volunteers ASA Class 1
|
Wat meet het onderzoek?
Primaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
|---|---|---|
|
The Covidien Nellcor Respiration Rate Software Shall Determine Respiration Rate Measured as Mean and Standard Deviation With Accuracy That is Non-inferior to Predicate Device.
Tijdsspanne: Participants were monitorerd on average of 30 minute period
|
Mean and standard deviations of respiration rates collected from healthy volunteers were compared between Covidien Respiration Rate Software, Transthoracic Impedance and Overscored End-Tidal Carbon Dioxide Waveforms.
Each volunteer served as its own control.
|
Participants were monitorerd on average of 30 minute period
|
Secundaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
|---|---|---|
|
The Covidien Nellcor Respiration Rate Software Shall Calculate Respiration Rate With a Root Mean Square Difference (RMSD) of < 3 Breaths Per Minute Compared With a End-Tidal Carbon Dioxide Waveforms, With 95% Confidence.
Tijdsspanne: Participants were monitored on average of 30 minute periods
|
The data used for analysis consisted of multiple simultaneous measures of RR_TTI, RR_V1.0 and RR_EtCO2 for each subject. Given N sets of simultaneous RR values for each of P patients, the simultaneous RR values were used to calculate a pair of RMSD values for each subject. The two RMSD values, RMSDRR_V1.0_vs_EtCO2, and RMSDTTI_vs_EtCO2, quantify the mean absolute difference in simultaneous estimates of respiratory rate between RR_V1.0 compared with RR_EtCO2 and RR_TTI compared with RR_EtCO2, respectively for the subjects with 95% confidence of mean ± 3.38 BrPM. The Measure type is Number and represents the RMSD of Covidien Nellcor Respiration Rate Software |
Participants were monitored on average of 30 minute periods
|
Medewerkers en onderzoekers
Sponsor
Onderzoekers
- Hoofdonderzoeker: Scott Kelley, MD, Medtronic - MITG
Studie record data
Bestudeer belangrijke data
Studie start
Primaire voltooiing (Werkelijk)
Studie voltooiing (Werkelijk)
Studieregistratiedata
Eerst ingediend
Eerst ingediend dat voldeed aan de QC-criteria
Eerst geplaatst (Schatting)
Updates van studierecords
Laatste update geplaatst (Schatting)
Laatste update ingediend die voldeed aan QC-criteria
Laatst geverifieerd
Meer informatie
Termen gerelateerd aan deze studie
Trefwoorden
Andere studie-ID-nummers
- COVMOPR0132
Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .