Novel Non-Invasive Monitoring Parameter in Healthy Volunteers
Novel Cardio-Respiratory Parameter in Healthy Volunteers
Обзор исследования
Статус
Условия
Подробное описание
Тип исследования
Регистрация (Действительный)
Контакты и местонахождение
Места учебы
-
-
Colorado
-
Boulder, Colorado, Соединенные Штаты, 80301
- Covidien- RMS
-
-
Критерии участия
Критерии приемлемости
Возраст, подходящий для обучения
Принимает здоровых добровольцев
Полы, имеющие право на обучение
Метод выборки
Исследуемая популяция
Описание
Inclusion Criteria:
- Subject is 18 or older
- Subject is willing and able to provide written consent
Exclusion Criteria:
- Severe contact allergies to standard adhesive materials
- Subjects with abnormalities that may prevent proper application of the device
- Women who are pregnant or lactating
- Subjects with significant Arrhythmias
Учебный план
Как устроено исследование?
Детали дизайна
Когорты и вмешательства
Группа / когорта |
|---|
|
Healthy Volunteers
Healthy volunteers ASA Class 1
|
Что измеряет исследование?
Первичные показатели результатов
Мера результата |
Мера Описание |
Временное ограничение |
|---|---|---|
|
The Covidien Nellcor Respiration Rate Software Shall Determine Respiration Rate Measured as Mean and Standard Deviation With Accuracy That is Non-inferior to Predicate Device.
Временное ограничение: Participants were monitorerd on average of 30 minute period
|
Mean and standard deviations of respiration rates collected from healthy volunteers were compared between Covidien Respiration Rate Software, Transthoracic Impedance and Overscored End-Tidal Carbon Dioxide Waveforms.
Each volunteer served as its own control.
|
Participants were monitorerd on average of 30 minute period
|
Вторичные показатели результатов
Мера результата |
Мера Описание |
Временное ограничение |
|---|---|---|
|
The Covidien Nellcor Respiration Rate Software Shall Calculate Respiration Rate With a Root Mean Square Difference (RMSD) of < 3 Breaths Per Minute Compared With a End-Tidal Carbon Dioxide Waveforms, With 95% Confidence.
Временное ограничение: Participants were monitored on average of 30 minute periods
|
The data used for analysis consisted of multiple simultaneous measures of RR_TTI, RR_V1.0 and RR_EtCO2 for each subject. Given N sets of simultaneous RR values for each of P patients, the simultaneous RR values were used to calculate a pair of RMSD values for each subject. The two RMSD values, RMSDRR_V1.0_vs_EtCO2, and RMSDTTI_vs_EtCO2, quantify the mean absolute difference in simultaneous estimates of respiratory rate between RR_V1.0 compared with RR_EtCO2 and RR_TTI compared with RR_EtCO2, respectively for the subjects with 95% confidence of mean ± 3.38 BrPM. The Measure type is Number and represents the RMSD of Covidien Nellcor Respiration Rate Software |
Participants were monitored on average of 30 minute periods
|
Соавторы и исследователи
Спонсор
Следователи
- Главный следователь: Scott Kelley, MD, Medtronic - MITG
Даты записи исследования
Изучение основных дат
Начало исследования
Первичное завершение (Действительный)
Завершение исследования (Действительный)
Даты регистрации исследования
Первый отправленный
Впервые представлено, что соответствует критериям контроля качества
Первый опубликованный (Оценивать)
Обновления учебных записей
Последнее опубликованное обновление (Оценивать)
Последнее отправленное обновление, отвечающее критериям контроля качества
Последняя проверка
Дополнительная информация
Термины, связанные с этим исследованием
Ключевые слова
Другие идентификационные номера исследования
- COVMOPR0132
Эта информация была получена непосредственно с веб-сайта clinicaltrials.gov без каких-либо изменений. Если у вас есть запросы на изменение, удаление или обновление сведений об исследовании, обращайтесь по адресу register@clinicaltrials.gov. Как только изменение будет реализовано на clinicaltrials.gov, оно будет автоматически обновлено и на нашем веб-сайте. .