Novel Non-Invasive Monitoring Parameter in Healthy Volunteers
Novel Cardio-Respiratory Parameter in Healthy Volunteers
Studieoversikt
Status
Detaljert beskrivelse
Studietype
Registrering (Faktiske)
Kontakter og plasseringer
Studiesteder
-
-
Colorado
-
Boulder, Colorado, Forente stater, 80301
- Covidien- RMS
-
-
Deltakelseskriterier
Kvalifikasjonskriterier
Alder som er kvalifisert for studier
Tar imot friske frivillige
Kjønn som er kvalifisert for studier
Prøvetakingsmetode
Studiepopulasjon
Beskrivelse
Inclusion Criteria:
- Subject is 18 or older
- Subject is willing and able to provide written consent
Exclusion Criteria:
- Severe contact allergies to standard adhesive materials
- Subjects with abnormalities that may prevent proper application of the device
- Women who are pregnant or lactating
- Subjects with significant Arrhythmias
Studieplan
Hvordan er studiet utformet?
Designdetaljer
Kohorter og intervensjoner
Gruppe / Kohort |
|---|
|
Healthy Volunteers
Healthy volunteers ASA Class 1
|
Hva måler studien?
Primære resultatmål
Resultatmål |
Tiltaksbeskrivelse |
Tidsramme |
|---|---|---|
|
The Covidien Nellcor Respiration Rate Software Shall Determine Respiration Rate Measured as Mean and Standard Deviation With Accuracy That is Non-inferior to Predicate Device.
Tidsramme: Participants were monitorerd on average of 30 minute period
|
Mean and standard deviations of respiration rates collected from healthy volunteers were compared between Covidien Respiration Rate Software, Transthoracic Impedance and Overscored End-Tidal Carbon Dioxide Waveforms.
Each volunteer served as its own control.
|
Participants were monitorerd on average of 30 minute period
|
Sekundære resultatmål
Resultatmål |
Tiltaksbeskrivelse |
Tidsramme |
|---|---|---|
|
The Covidien Nellcor Respiration Rate Software Shall Calculate Respiration Rate With a Root Mean Square Difference (RMSD) of < 3 Breaths Per Minute Compared With a End-Tidal Carbon Dioxide Waveforms, With 95% Confidence.
Tidsramme: Participants were monitored on average of 30 minute periods
|
The data used for analysis consisted of multiple simultaneous measures of RR_TTI, RR_V1.0 and RR_EtCO2 for each subject. Given N sets of simultaneous RR values for each of P patients, the simultaneous RR values were used to calculate a pair of RMSD values for each subject. The two RMSD values, RMSDRR_V1.0_vs_EtCO2, and RMSDTTI_vs_EtCO2, quantify the mean absolute difference in simultaneous estimates of respiratory rate between RR_V1.0 compared with RR_EtCO2 and RR_TTI compared with RR_EtCO2, respectively for the subjects with 95% confidence of mean ± 3.38 BrPM. The Measure type is Number and represents the RMSD of Covidien Nellcor Respiration Rate Software |
Participants were monitored on average of 30 minute periods
|
Samarbeidspartnere og etterforskere
Sponsor
Etterforskere
- Hovedetterforsker: Scott Kelley, MD, Medtronic - MITG
Studierekorddatoer
Studer hoveddatoer
Studiestart
Primær fullføring (Faktiske)
Studiet fullført (Faktiske)
Datoer for studieregistrering
Først innsendt
Først innsendt som oppfylte QC-kriteriene
Først lagt ut (Anslag)
Oppdateringer av studieposter
Sist oppdatering lagt ut (Anslag)
Siste oppdatering sendt inn som oppfylte QC-kriteriene
Sist bekreftet
Mer informasjon
Begreper knyttet til denne studien
Nøkkelord
Andre studie-ID-numre
- COVMOPR0132
Denne informasjonen ble hentet direkte fra nettstedet clinicaltrials.gov uten noen endringer. Hvis du har noen forespørsler om å endre, fjerne eller oppdatere studiedetaljene dine, vennligst kontakt register@clinicaltrials.gov. Så snart en endring er implementert på clinicaltrials.gov, vil denne også bli oppdatert automatisk på nettstedet vårt. .