Novel Non-Invasive Monitoring Parameter in Healthy Volunteers
Novel Cardio-Respiratory Parameter in Healthy Volunteers
Descripción general del estudio
Estado
Condiciones
Descripción detallada
Tipo de estudio
Inscripción (Actual)
Contactos y Ubicaciones
Ubicaciones de estudio
-
-
Colorado
-
Boulder, Colorado, Estados Unidos, 80301
- Covidien- RMS
-
-
Criterios de participación
Criterio de elegibilidad
Edades elegibles para estudiar
Acepta Voluntarios Saludables
Géneros elegibles para el estudio
Método de muestreo
Población de estudio
Descripción
Inclusion Criteria:
- Subject is 18 or older
- Subject is willing and able to provide written consent
Exclusion Criteria:
- Severe contact allergies to standard adhesive materials
- Subjects with abnormalities that may prevent proper application of the device
- Women who are pregnant or lactating
- Subjects with significant Arrhythmias
Plan de estudios
¿Cómo está diseñado el estudio?
Detalles de diseño
Cohortes e Intervenciones
Grupo / Cohorte |
|---|
|
Healthy Volunteers
Healthy volunteers ASA Class 1
|
¿Qué mide el estudio?
Medidas de resultado primarias
Medida de resultado |
Medida Descripción |
Periodo de tiempo |
|---|---|---|
|
The Covidien Nellcor Respiration Rate Software Shall Determine Respiration Rate Measured as Mean and Standard Deviation With Accuracy That is Non-inferior to Predicate Device.
Periodo de tiempo: Participants were monitorerd on average of 30 minute period
|
Mean and standard deviations of respiration rates collected from healthy volunteers were compared between Covidien Respiration Rate Software, Transthoracic Impedance and Overscored End-Tidal Carbon Dioxide Waveforms.
Each volunteer served as its own control.
|
Participants were monitorerd on average of 30 minute period
|
Medidas de resultado secundarias
Medida de resultado |
Medida Descripción |
Periodo de tiempo |
|---|---|---|
|
The Covidien Nellcor Respiration Rate Software Shall Calculate Respiration Rate With a Root Mean Square Difference (RMSD) of < 3 Breaths Per Minute Compared With a End-Tidal Carbon Dioxide Waveforms, With 95% Confidence.
Periodo de tiempo: Participants were monitored on average of 30 minute periods
|
The data used for analysis consisted of multiple simultaneous measures of RR_TTI, RR_V1.0 and RR_EtCO2 for each subject. Given N sets of simultaneous RR values for each of P patients, the simultaneous RR values were used to calculate a pair of RMSD values for each subject. The two RMSD values, RMSDRR_V1.0_vs_EtCO2, and RMSDTTI_vs_EtCO2, quantify the mean absolute difference in simultaneous estimates of respiratory rate between RR_V1.0 compared with RR_EtCO2 and RR_TTI compared with RR_EtCO2, respectively for the subjects with 95% confidence of mean ± 3.38 BrPM. The Measure type is Number and represents the RMSD of Covidien Nellcor Respiration Rate Software |
Participants were monitored on average of 30 minute periods
|
Colaboradores e Investigadores
Patrocinador
Investigadores
- Investigador principal: Scott Kelley, MD, Medtronic - MITG
Fechas de registro del estudio
Fechas importantes del estudio
Inicio del estudio
Finalización primaria (Actual)
Finalización del estudio (Actual)
Fechas de registro del estudio
Enviado por primera vez
Primero enviado que cumplió con los criterios de control de calidad
Publicado por primera vez (Estimar)
Actualizaciones de registros de estudio
Última actualización publicada (Estimar)
Última actualización enviada que cumplió con los criterios de control de calidad
Última verificación
Más información
Términos relacionados con este estudio
Palabras clave
Otros números de identificación del estudio
- COVMOPR0132
Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .