A Comparison of Two Urinary Catheters of Different Lengths for Female Use, in Intermittent Catheterization
A Multi Centre, Single-blind, Randomized, Cross-over Study Comparing Two Female Catheters With Different Length in Terms of Residual Urine After Intermittent Catheterization
A tanulmány áttekintése
Állapot
Körülmények
Beavatkozás / kezelés
Részletes leírás
The study personnel will conduct three bladder scans after each catheterization in order to obtain the mean residual urine volume. Each patient will catheterize herself two times with each product, i.e. a total of four catheterizations.
Diagnosis and BMI will be collected at visit 1 to assess whether the residual urine volume differs between these subgroups.
The hypothesis that the mean residual volumes were equal when using the 7 and the 10 centimeter catheters was tested by means of the Wilcoxon signed rank test.
Tanulmány típusa
Beiratkozás (Tényleges)
Fázis
- Nem alkalmazható
Kapcsolatok és helyek
Tanulmányi helyek
-
-
-
Borås, Svédország, SE-501 82
- Rehabiliteringskliniken, Södra Älvsborgs Sjukhus
-
Gothenburg, Svédország, SE-413 45
- Verksamhet Urologi, Sahlgrenska Universitetssjukhuset
-
Lund, Svédország, SE-221 85
- Clinical Research and Trial Centre, Lund University Hospital
-
-
Részvételi kritériumok
Jogosultsági kritériumok
Tanulmányozható életkorok
Egészséges önkénteseket fogad
Tanulmányozható nemek
Leírás
Inclusion Criteria:
- Provision of signed informed consent
- Female subjects aged 18 years and over
- Experienced users of CIC with a minimum of one month of use
- Practice CIC a minimum of 1 time per day
- Practice CIC with female catheters with a maximum length of 20 cm
- Perform self-catheterization
- Sufficient hand function to be able to open and handle catheter packages
- Use catheters with Ch 10, 12 or 14
Exclusion Criteria:
- Pregnancy
- Ongoing symptomatic UTI
- Involvement in the planning and conduct of the study (applies to both Astra Tech staff or staff at the study site)
- Previous enrolment or randomisation of treatment in the present study
- Participation in another clinical study during the study period
- Severe non-compliance to protocol as judged by the investigator and/or Astra Tech
Tanulási terv
Hogyan készül a tanulmány?
Tervezési részletek
- Elsődleges cél: Egyéb
- Kiosztás: Véletlenszerűsített
- Beavatkozó modell: Crossover kiosztás
- Maszkolás: Kettős
Fegyverek és beavatkozások
Résztvevő csoport / kar |
Beavatkozás / kezelés |
---|---|
Kísérleti: AABB
A=Coloplast SpeediCath Compact catheter, 7 cm B=Astra Tech POBE catheter, 10 cm 7 cm then 7 cm then 10 cm then 10 cm Four catheterizations in total, two catheterizations with each study product in a randomized order. |
Catheterization should resemble the daily use as far as possible and at least 2 hours had to pass between 2 catheterizations.The study centre decided whether the subject performed all four catheterizations on one or several days.
Más nevek:
|
Kísérleti: ABAB
A=Coloplast SpeediCath Compact catheter, 7 cm B=Astra Tech POBE catheter, 10 cm 7 cm then 10 cm then 7 cm then 10 cm Four catheterizations in total, two catheterizations with each study product in a randomized order. |
Catheterization should resemble the daily use as far as possible and at least 2 hours had to pass between 2 catheterizations.The study centre decided whether the subject performed all four catheterizations on one or several days.
Más nevek:
|
Kísérleti: ABBA
A=Coloplast SpeediCath Compact catheter, 7 cm B=Astra Tech POBE catheter, 10 cm 7 cm then 10 cm then 10 cm then 7 cm Four catheterizations in total, two catheterizations with each study product in a randomized order. |
Catheterization should resemble the daily use as far as possible and at least 2 hours had to pass between 2 catheterizations.The study centre decided whether the subject performed all four catheterizations on one or several days.
Más nevek:
|
Kísérleti: BAAB
A=Coloplast SpeediCath Compact catheter, 7 cm B=Astra Tech POBE catheter, 10 cm 10 cm then 7 cm then 7 cm then 10 cm Four catheterizations in total, two catheterizations with each study product in a randomized order. |
Catheterization should resemble the daily use as far as possible and at least 2 hours had to pass between 2 catheterizations.The study centre decided whether the subject performed all four catheterizations on one or several days.
Más nevek:
|
Kísérleti: BABA
A=Coloplast SpeediCath Compact catheter, 7 cm B=Astra Tech POBE catheter, 10 cm 10 cm then then 7 cm then 10 cm then 7 cm Four catheterizations in total, two catheterizations with each study product in a randomized order. |
Catheterization should resemble the daily use as far as possible and at least 2 hours had to pass between 2 catheterizations.The study centre decided whether the subject performed all four catheterizations on one or several days.
Más nevek:
|
Kísérleti: BBAA
A=Coloplast SpeediCath Compact catheter, 7 cm B=Astra Tech POBE catheter, 10 cm 10 cm then 10 cm then 7 cm then 7 cm Four catheterizations in total, two catheterizations with each study product in a randomized order. |
Catheterization should resemble the daily use as far as possible and at least 2 hours had to pass between 2 catheterizations.The study centre decided whether the subject performed all four catheterizations on one or several days.
Más nevek:
|
Mit mér a tanulmány?
Elsődleges eredményintézkedések
Eredménymérő |
Intézkedés leírása |
Időkeret |
---|---|---|
Bladder Scan Measurements of Residual Urine Volume
Időkeret: Three bladder scan measurements were performed immediately after each catheterization and a mean value was calculated. A maximum of 5 minutes were allowed to elapse between catheterization and bladder scan.
|
Measurements were performed with the bladder scan BVI 3000. The primary outcome variable was average residual urine volume (mL) calculated from repetitive bladder scans performed after each catheterization. Three bladder scan measurements were performed immediately after each catheterization and a mean value was calculated. |
Three bladder scan measurements were performed immediately after each catheterization and a mean value was calculated. A maximum of 5 minutes were allowed to elapse between catheterization and bladder scan.
|
Másodlagos eredményintézkedések
Eredménymérő |
Intézkedés leírása |
Időkeret |
---|---|---|
Number of Subjects With a Residual Volume >5mL or ≤5mL
Időkeret: Three bladder scan measurements were performed immediately after each catheterization and a mean value was calculated. At least 2 hours had to pass between 2 catheterizations.
|
Three bladder scan measurements were performed immediately after each catheterization and a mean value was calculated. At least 2 hours had to pass between 2 catheterizations.
|
|
Number of Subjects With a Residual Volume >10mL or ≤10mL
Időkeret: Three bladder scan measurements were performed immediately after each catheterization and a mean value was calculated. A maximum of 5 minutes were allowed to elapse between catheterization and bladder scan.
|
Three bladder scan measurements were performed immediately after each catheterization and a mean value was calculated. A maximum of 5 minutes were allowed to elapse between catheterization and bladder scan.
|
|
Number of Subjects With a Residual Volume >30mL or ≤30mL
Időkeret: Three bladder scan measurements were performed immediately after each catheterization and a mean value was calculated. A maximum of 5 minutes were allowed to elapse between catheterization and bladder scan.
|
Three bladder scan measurements were performed immediately after each catheterization and a mean value was calculated. A maximum of 5 minutes were allowed to elapse between catheterization and bladder scan.
|
|
Number of Subjects With a Residual Volume >50mL or ≤50mL
Időkeret: Three bladder scan measurements were performed immediately after each catheterization and a mean value was calculated. A maximum of 5 minutes were allowed to elapse between catheterization and bladder scan.
|
Three bladder scan measurements were performed immediately after each catheterization and a mean value was calculated. A maximum of 5 minutes were allowed to elapse between catheterization and bladder scan.
|
|
Does the Subject's Bladder Feel Completely Empty After Catheterisation (All Subjects)
Időkeret: Question asked minutes after catheterization.
|
After each catheterization the study nurse will ask the subject whether it feels as if the bladder is completely emptied or not. 0=No and 1=Yes. |
Question asked minutes after catheterization.
|
Együttműködők és nyomozók
Szponzor
Nyomozók
- Kutatásvezető: Ulf Malmqvist, MD, PhD, Assoc.Prof., Clinical Research and Trial Centre, Lund University Hospital
Tanulmányi rekorddátumok
Tanulmány főbb dátumok
Tanulmány kezdete (Tényleges)
Elsődleges befejezés (Tényleges)
A tanulmány befejezése (Tényleges)
Tanulmányi regisztráció dátumai
Először benyújtva
Először nyújtották be, amely megfelel a minőségbiztosítási kritériumoknak
Első közzététel (Becslés)
Tanulmányi rekordok frissítései
Utolsó frissítés közzétéve (Tényleges)
Az utolsó frissítés elküldve, amely megfelel a minőségbiztosítási kritériumoknak
Utolsó ellenőrzés
Több információ
A tanulmányhoz kapcsolódó kifejezések
Egyéb vizsgálati azonosító számok
- YA-LOF-0015
Gyógyszer- és eszközinformációk, tanulmányi dokumentumok
Egy amerikai FDA által szabályozott gyógyszerkészítményt tanulmányoz
Egy amerikai FDA által szabályozott eszközterméket tanulmányoz
az Egyesült Államokban gyártott és onnan exportált termék
Ezt az információt közvetlenül a clinicaltrials.gov webhelyről szereztük be, változtatás nélkül. Ha bármilyen kérése van vizsgálati adatainak módosítására, eltávolítására vagy frissítésére, kérjük, írjon a következő címre: register@clinicaltrials.gov. Amint a változás bevezetésre kerül a clinicaltrials.gov oldalon, ez a webhelyünkön is automatikusan frissül. .