Questa pagina è stata tradotta automaticamente e l'accuratezza della traduzione non è garantita. Si prega di fare riferimento al Versione inglese per un testo di partenza.

A Clinical Study to Collect Real-wordt Data on Safety and Clinical Performance of the VCFix Spinal System for Treatment of Vertebral Compression Fractures of (Suspected) Osteoporotic Origin. (EXPAND-US)

4 novembre 2026 aggiornato da: Amber Implants B.V.

A Prospective, Multicenter, Post-Market Clinical Study Collecting Real-World Data of the VCFix Spinal System in Stand-Alone Configuration for Vertebral Augmentation (EXPAND-US)

This study is designed as a prospective, post-market, multi-center, single-arm study to collect real-world data of the VCFix Spinal System used in US clinical practice.

Approximately 5 centers in the US will enroll subjects with VCF in a non-randomized fashion, and VCFix will be used within its cleared purpose and indications for use.

The study has been designed to be methodologically compatible with the EU EXPAND (CIP-VS02-EU) study, thereby aiming for the pooling of data to the EU study. Data pooling will be pursued only as a sensitivity analysis if pre-specified similarity criteria are met. The US study provides meaningful real-world evidence regardless of pooling feasibility.

The targeted patient population, assessment schedule, and study endpoints are therefore similar to those of the EU EXPAND study.

Panoramica dello studio

Stato

Reclutamento

Condizioni

Intervento / Trattamento

Tipo di studio

Osservativo

Iscrizione (Stimato)

30

Contatti e Sedi

Questa sezione fornisce i recapiti di coloro che conducono lo studio e informazioni su dove viene condotto lo studio.

Contatto studio

Luoghi di studio

    • New York
      • New York, New York, Stati Uniti, 10467
        • Reclutamento
        • Montefiore Medical Center
        • Contatto:
          • Research Administrator
          • Numero di telefono: (718) 920-6202
    • Oklahoma
      • Oklahoma City, Oklahoma, Stati Uniti, 73104
        • Non ancora reclutamento
        • University of Oklahoma Health Campus
        • Contatto:
          • Research Project Manager
          • Numero di telefono: (405) 271-8001
    • Pennsylvania
      • Philadelphia, Pennsylvania, Stati Uniti, 19107
        • Non ancora reclutamento
        • Rothman Orthopaedics
        • Contatto:
          • Clinical Research Coordinator
          • Numero di telefono: (484) 335-2910

Criteri di partecipazione

I ricercatori cercano persone che corrispondano a una certa descrizione, chiamata criteri di ammissibilità. Alcuni esempi di questi criteri sono le condizioni generali di salute di una persona o trattamenti precedenti.

Criteri di ammissibilità

Età idonea allo studio

Da 21 anni a 85 anni (Adulto, Adulto più anziano)

Accetta volontari sani

No

Sessi ammissibili allo studio

Tutto

Metodo di campionamento

Campione non probabilistico

Popolazione di studio

Adult patients undergoing treatment with the VCFix Spinal System as part of standard clinical care for vertebral compression fractures. The study will enroll patients treated according to routine clinical practice and will collect real-world evidence on clinical outcomes, radiographic outcomes, and patient-reported outcomes following treatment.

Descrizione

Inclusion Criteria:

  • 1. Subject is ≥ 21 to ≤ 85 years old. 2.

Single vertebral fracture which meets all of the following criteria:

  • Type A fractures from AO spine classification (compression injuries);
  • Vertebral fracture located in the thoracic and lumbar regions of the spine
  • Fracture age < 6 weeks or fractures with anatomic signs of reducibility (based on radiographic evidence as well as patient history)
  • Vertebral fracture shows an estimated height loss in the anterior or mid third of the vertebral body (VB) of at least 10% but not more than 60% based on radiographic evidence
  • Target VB has appropriate anatomy, i.e. suitable pedicle diameter, pedicle length and chord length that would allow for selection of correct implant size, as described in the IFU and based on axial MRI/CT scan before surgery 3. Subject is diagnosed with, or has a suspected diagnosis of, severe osteoporosis, including low trabecular bone quality , and/or an oversized vertebral body as compared to implant size (e.g. relatively small pedicle requiring smaller implant size than desired for vertebral body size and/or fracture characteristics and location) 4. Subject has a NPRS back pain score of ≥5 5. Subject is a candidate for surgical intervention based on investigator opinion 6. Subject has a body Mass Index (BMI) < 35 7. Subject is mentally capable of complying with trial protocol requirements for the duration of the study 8. Subject can understand the risks and benefits of participating in the study and is able to provide written informed consent

Exclusion Criteria:

  1. Neoplasms with posterior involvement and/or presence of a mass within the spinal canal

  2. Non-mobile fractures (i.e., fracture is not recent (>6 weeks), bone marrow edema or fluid or empty cleft are not visible in radiographic imaging, or fracture mode does not allow for fracture reduction in craniocaudal direction).

  3. Spondylolisthesis > Grade 1 at target vertebral body(s)

  4. Local kyphotic angle > 30°

  5. Any pre-existing vertebral fracture prior to the index fracture

  6. Previous spine surgery, balloon kyphoplasty or vertebroplasty, for any vertebral fracture

  7. Subjects that require anterior stabilization of the index fracture

  8. Fracture to the pedicle based on radiographic evaluation

  9. Spinal cord compression or canal compromise requiring decompression

  10. Severe back pain due to causes other than acute fracture with NPRS score > 5

  11. The subject has pain based on clinical diagnosis of herniated nucleus pulposus or severe spinal stenosis (progressive weakness or paralysis)

  12. Pain due to any other condition that requires daily narcotic (opiates or opioids) medication

  13. Pre-existing neurological deficit, radiculopathy or myelopathy

  14. Pre-existing condition or significant co-morbidity:

    1. Uncontrolled diabetes (HbA1c >8%)
    2. Severe cardiopulmonary deficiencies
    3. Any other condition that would impact study outcome in the opinion of the investigator
  15. Irreversible coagulopathy or bleeding disorder

  16. Contraindications to both MRI and Computed Tomography (CT) imaging

  17. Concurrent participation in another clinical study which could potentially interfere with the outcome of this study

  18. A life expectancy less than the study duration or undergoing palliative care

  19. Subject non-ambulatory prior to fracture Allergy to any components of the device/instruments used during the procedure

  20. Allergy to any components of the device/instruments used during the procedure

  21. Active or incompletely treated infection of the vertebral column or active systemic infection, including unresolved urinary tract infection

  22. Any underlying systemic bone disease other than osteoporosis (e.g., osteomalacia, osteogenesis imperfecta, Paget's disease, etc.)

  23. Any evidence of alcohol or drug abuse

  24. Ongoing long-term steroid therapy (steroid dose ≥30 mg /day for >3 months)

  25. The subject is currently on anti-cancer therapy or anti-HIV therapy

  26. Pregnancy or subjects with child-bearing potential that are unwilling to use contraception throughout the study duration.

  27. Patient known to be involved in medical litigation including Workmen's Compensation prior to the surgery

Piano di studio

Questa sezione fornisce i dettagli del piano di studio, compreso il modo in cui lo studio è progettato e ciò che lo studio sta misurando.

Come è strutturato lo studio?

Dettagli di progettazione

  • Modelli osservazionali: COHORT
  • Prospettive temporali: PROSPECTIVE

Coorti e interventi

Gruppo / Coorte
Intervento / Trattamento
Implant of the VCFix device per IFU

Patients with a (suspected) diagnosis of severe osteoporosis and a single vertebral compression fracture

The distinguishing features of this inetrvention reside in: 1. In situ adjustment and maintenance of the height and angle of the device that allows expansion of the fractured VB in the cranio-caudal direction to realign the vertebral endplates and correct kyphosis. 2. The complex titanium perforated structure with optimized surface roughness that allows possible bone attachment, thereby permitting natural healing of the bone and potentially providing more stability to the device and the treated vertebra over time. 3. The ability to connect the posterior to the anterior column of the spine through the pedicle screw component of the VCFix implant, therefore improving the biomechanical load distribution and consequently the ability to stabilize more severe fractures.

Cosa sta misurando lo studio?

Misure di risultato primarie

Misura del risultato
Misura Descrizione
Lasso di tempo
Incidence of serious adverse events (SAEs) related to the device and/or investigational procedure (safety)
Lasso di tempo: 1 month

Incidence of serious adverse events (SAEs) related to the device and/or investigational procedure within 30 days post-procedure. This includes SAEs resulting from mechanical malfunction, structural compromise, device failure, or loss of intended function, as well as events representing deviations from the device's intended medical purpose.

1 month
Vertebral fracture-related pain as assessed by the Numeric Pain Rating Scale (NPRS)
Lasso di tempo: 6 months

Difference in vertebral fracture-related pain from baseline 6 months post procedure measured by the 11-point (0 = no pain, 10 is worst pain) Numerical Pain Rating Scale (NPRS).

6 months
Functional disability, measured by the Oswestry disability index (ODI)
Lasso di tempo: 6 months

Difference in functional disability, measured by the Oswestry disability index (ODI) (0-100 scale, where 0 is no disability and 100 is bed bound), from baseline to 6 months post procedure.

6 months

Misure di risultato secondarie

Misura del risultato
Misura Descrizione
Lasso di tempo
Vertebral height difference (anterior, midline, and posterior aspects)
Lasso di tempo: Post-Operative, 1Month FU, 6 Months FU, 12 Months FU

Vertebral height difference (anterior, midline, and posterior aspects) at PO (Post-OP), 1M FU, 6M FU and 12M FU as compared to baseline.

Post-Operative, 1Month FU, 6 Months FU, 12 Months FU
Kyphotic angle and Cobb angle improvement at Post-Operative, 1 Month FU, 6 Months FU and 12 Months FU.
Lasso di tempo: Post-Operative, 1 Month FU, 6 Months FU and 12 Months FU.

Kyphotic angle and Cobb angle improvement

Post-Operative, 1 Month FU, 6 Months FU and 12 Months FU.
Vertebral fracture-related pain as assessed by the Numeric Pain Rating Scale (NPRS)
Lasso di tempo: Discharge, 1, 3, 6 and 12 Months FU compared to baseline.

Difference in vertebral fracture-related pain from baseline to Discharge, 1, 3, 6 and 12 Months post procedure measured by the 11-point (0 = no pain, 10 is worst pain) Numerical Pain Rating Scale (NPRS).

Discharge, 1, 3, 6 and 12 Months FU compared to baseline.
Functional disability, measured by the Oswestry disability index (ODI)
Lasso di tempo: Discharge, 1, 3, 6 and 12 months FU

Improvement in function (measured by the 100-point Oswestry Disability Index (ODI)) at discharge, 1, 3, 6 and 12 months FU compared to baseline.

Discharge, 1, 3, 6 and 12 months FU
Health-related quality of life (HRQoL) measured by EuroQol 5-Dimension Scale (EQ-5D-5L) at 1 Month, 3 Months, 6 Months and 12 Months post-procedure compared to baseline
Lasso di tempo: 1 Month, 3 Months, 6 Months and 12 Months post-procedure compared to baseline

Improvement of Health-related quality of life (HRQoL)

1 Month, 3 Months, 6 Months and 12 Months post-procedure compared to baseline
Device-related serious adverse events assessed at 12- and 24- months post-procedure.
Lasso di tempo: 12 months FU, 24 months FU

Device-related serious adverse events

12 months FU, 24 months FU
All (per protocol reportable) adverse events assessed at 12- and 24- months post-procedure.
Lasso di tempo: 12 Months FU, 24 Months FU

All (per protocol reportable) adverse events assessed at 12- and 24- months post-procedure.

12 Months FU, 24 Months FU

Collaboratori e investigatori

Qui è dove troverai le persone e le organizzazioni coinvolte in questo studio.

Sponsor

Studiare le date dei record

Queste date tengono traccia dell'avanzamento della registrazione dello studio e dell'invio dei risultati di sintesi a ClinicalTrials.gov. I record degli studi e i risultati riportati vengono esaminati dalla National Library of Medicine (NLM) per assicurarsi che soddisfino specifici standard di controllo della qualità prima di essere pubblicati sul sito Web pubblico.

Studia le date principali

Inizio studio (Effettivo)

4 ottobre 2026

Completamento primario (Stimato)

21 luglio 2027

Completamento dello studio (Stimato)

31 dicembre 2028

Date di iscrizione allo studio

Primo inviato

31 luglio 2026

Primo inviato che soddisfa i criteri di controllo qualità

4 novembre 2026

Primo Inserito (Effettivo)

5 novembre 2026

Aggiornamenti dei record di studio

Ultimo aggiornamento pubblicato (Effettivo)

5 novembre 2026

Ultimo aggiornamento inviato che soddisfa i criteri QC

4 novembre 2026

Ultimo verificato

1 novembre 2026

Maggiori informazioni

Termini relativi a questo studio

Altri numeri di identificazione dello studio

  • CIP-VS03-US

Piano per i dati dei singoli partecipanti (IPD)

Hai intenzione di condividere i dati dei singoli partecipanti (IPD)?

INDECISO

Informazioni su farmaci e dispositivi, documenti di studio

Studia un prodotto farmaceutico regolamentato dalla FDA degli Stati Uniti

No

Studia un dispositivo regolamentato dalla FDA degli Stati Uniti

Sì

prodotto fabbricato ed esportato dagli Stati Uniti

No

Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .