Determination of Specificity and Sensitivity in Detecting Myocardial Infarction Using the EDE CARDIO BIO-MAP Machine (CARDIO BIO-MAP)
Determination of Specificity and Sensitivity in Detecting Myocardial Infarction Using the EDE CARDIO BIO-MAP Machine, the EDE CARDIO BIO-MAP Study
The goal of this observational study is to learn whether a non-invasive electromagnetic wave scanner (EDE CARDIO BIO-MAP) can help diagnose myocardial infarction (heart attack) in adults (18 years and older) who come to the emergency department with chest pain and suspected myocardial infarction. The main questions it aims to answer are:
- How sensitive and specific is the EDE scanner in diagnosing myocardial infarction (MI) compared with standard blood tests (e.g., cardiac troponin)?
- How sensitive and specific are the immediate blood measurements from a portable lab device compared with conventional hospital laboratory measurements? Researchers will compare the results of the EDE scanner and the portable lab with standard hospital blood test results and the final clinical diagnosis to see how accurately each detects MI.
Participants will:
- Have a research visit (30 minutes to 1 hour) during their emergency care, including a quick scan with the Exponential Deep Examination (EDE) device (attached to a phone or tablet) and a small research blood sample tested on a portable lab device
- Repeat the scan and blood sample 4 to 6 hours later, only if a second blood draw is already needed for their clinical care
- Answer questions about their medical and social history, and allow the study team to collect relevant results from their medical record Scanner results will not be shared with participants or used for their clinical care.
Panoramica dello studio
Stato
Stato
Condizioni
Condizioni
Descrizione dettagliata
Myocardial infarction is a leading cause of death and disability worldwide. Current diagnosis relies on clinical assessment, electrocardiogram, and blood measurement of cardiac biomarkers such as troponin. Faster, non-invasive diagnostic tools could be especially valuable in the emergency setting, where timely intervention is critical.
The Exponential Deep Examination Cardio Bio-Map is a non-invasive device, classified as Class 1 under the European Conformity marking system, that analyzes passive electromagnetic signals (0.1 to 300 gigahertz) using artificial intelligence to measure cardiac biomarkers without a blood draw. Results are available within 50 milliseconds. The device does not collect or store patient data. It produces an immediate printed report, and all data processing occurs locally within Qatar. During the COVID-19 pandemic, the same technology detected the SARS-CoV-2 virus with a specificity of 95.62% and a sensitivity of 93.71% compared with polymerase chain reaction testing and antibody testing. This study will evaluate whether the device can screen for myocardial infarction through measurement of cardiac biomarkers, both in biological samples (in vitro, using a portable laboratory device) and directly in patients (in vivo, using the scanner).
Tipo di studio
Tipo di studio
Iscrizione (Stimato)
Iscrizione
Contatti e Sedi
Contatto studio
Contatto studio
- Nome: Charbel Abi Khalil, MD,PhD
- Numero di telefono: +974 4492 8484
- Email: cha2022@qatar-med.cornell.edu
Criteri di partecipazione
Criteri di ammissibilità
Criteri di ammissibilità
Età idonea allo studio
Accetta volontari sani
Sessi ammissibili allo studio
Metodo di campionamento
Popolazione di studio
Participants will be recruited at Hamad Medical Corporation
Descrizione
Inclusion Criteria:
- Age ≥18 years.
- Adults presenting to the emergency room for acute chest pain.
- English or Arabic speakers.
Exclusion Criteria:
- Pregnancy (self-reported).
- Adults with cognitive disorders.
- Unconscious patients.
- Citizens from countries where the General Data Protection Regulation (GDPR) applies, due to regulatory constraints.
Piano di studio
Come è strutturato lo studio?
Dettagli di progettazione
- Modelli osservazionali: COHORT
- Prospettive temporali: PROSPECTIVE
Cosa sta misurando lo studio?
Misure di risultato primarie
Misure di risultato primarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Sensitivity and specificity of the EDE CARDIO BIO-MAP scanner(in vivo) for detecting myocardial infarction
Lasso di tempo: Baseline
|
Sensitivity (true positives / [true positives + false negatives]) and specificity (true negatives / [true negatives + false positives]) of the scanner in emergency room patients with suspected myocardial infarction, compared with standard hospital laboratory results (cardiac troponin, creatine kinase-myocardial band, myoglobin, B-type natriuretic peptide, N-terminal pro-B-type natriuretic peptide) and the final clinical diagnosis in the medical record. |
Baseline
|
Misure di risultato secondarie
Misure di risultato secondarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Sensitivity and specificity of the portable lab's immediate blood measurements (in vitro)
Lasso di tempo: Baseline
|
Sensitivity and specificity of the portable lab's immediate blood measurements for myocardial infarction, compared with conventional hospital laboratory measurements, using a confusion matrix. |
Baseline
|
|
Accuracy of the EDE scanner
Lasso di tempo: Baseline
|
Accuracy = (true positives + true negatives) / all results; |
Baseline
|
|
Misclassification rate of the EDE scanner
Lasso di tempo: Baseline
|
Misclassification rate = (false positives + false negatives) / all results; |
Baseline
|
|
Precision of the EDE scanner
Lasso di tempo: Baseline
|
precision = true positives / (true positives + false positives) |
Baseline
|
|
Accuracy of the portable lab
Lasso di tempo: Baseline
|
Accuracy = (true positives + true negatives) / all results; |
Baseline
|
|
Misclassification rate of the portable lab
Lasso di tempo: Baseline
|
Misclassification rate = (false positives + false negatives) / all results; |
Baseline
|
|
Precision of the portable lab
Lasso di tempo: Baseline
|
precision = true positives / (true positives + false positives) |
Baseline
|
|
Correlation Between EDE Scanner and Portable Laboratory Values for Cardiac Biomarkers
Lasso di tempo: Baseline
|
Pearson or Spearman correlation coefficient between scanner and portable lab values for cardiac biomarkers : Troponin T, Troponin I, Myoglobin, B-type Natriuretic Peptide, NT-proBNP, and Creatine Kinase-MB. |
Baseline
|
|
Predictive power of each cardiac biomarker for myocardial infarction
Lasso di tempo: Baseline
|
Logistic regression, area under the receiver operating characteristic curve, and optimal diagnostic cutoffs for each cardiac biomarker: Troponin T, Troponin I, Myoglobin, B-type Natriuretic Peptide, NT-proBNP, and Creatine Kinase-MB. |
Baseline
|
Altre misure di risultato
Altre misure di risultato
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Change in cardiac biomarker levels between initial assessment and 4-6 hour follow-up
Lasso di tempo: Baseline and 4-6 hours after initial assessment
|
Change in cardiac biomarker levels( Troponin T, Troponin I, Myoglobin, B-type Natriuretic Peptide, NT-proBNP, and Creatine Kinase-MB) between the initial assessment and the follow-up, in participants who have a second clinical blood draw. |
Baseline and 4-6 hours after initial assessment
|
|
Diagnostic performance of the scanner by sex
Lasso di tempo: Baseline
|
Subgroup analysis of scanner sensitivity and specificity by sex |
Baseline
|
|
Diagnostic performance of the scanner by age group
Lasso di tempo: Baseline
|
Subgroup analysis of scanner sensitivity and specificity by age group |
Baseline
|
|
Diagnostic performance of the scanner by symptom profile
Lasso di tempo: Baseline
|
Subgroup analysis of scanner sensitivity and specificity by symptom profile. |
Baseline
|
Collaboratori e investigatori
Sponsor
Sponsor
Collaboratori
Collaboratori
Investigatori
Investigatori
- Investigatore principale: Charbel Abi Khalil, MD,PhD, Weill Cornell Medicine-Qatar
Studiare le date dei record
Studia le date principali
Inizio studio (Stimato)
Inizio studio
Completamento primario (Stimato)
Completamento primario
Completamento dello studio (Stimato)
Completamento dello studio
Date di iscrizione allo studio
Primo inviato
Primo inviato
Primo inviato che soddisfa i criteri di controllo qualità
Primo inviato che soddisfa i criteri di controllo qualità
Primo Inserito (Effettivo)
Primo Inserito
Aggiornamenti dei record di studio
Ultimo aggiornamento pubblicato (Effettivo)
Ultimo aggiornamento pubblicato
Ultimo aggiornamento inviato che soddisfa i criteri QC
Ultimo aggiornamento inviato che soddisfa i criteri QC
Ultimo verificato
Ultimo verificato
Maggiori informazioni
Termini relativi a questo studio
Termini MeSH pertinenti aggiuntivi
Altri numeri di identificazione dello studio
Altri numeri di identificazione dello studio
- 23-00017
Piano per i dati dei singoli partecipanti (IPD)
Hai intenzione di condividere i dati dei singoli partecipanti (IPD)?
Descrizione del piano IPD
ongoing discussion with the sponsor
Informazioni su farmaci e dispositivi, documenti di studio
Studia un prodotto farmaceutico regolamentato dalla FDA degli Stati Uniti
Studia un dispositivo regolamentato dalla FDA degli Stati Uniti
Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .