Optimized Camera-based Patient Positioning in CT
Optimized Camera-based Patient Positioning in CT: Impact on Radiation Exposure.
Panoramica dello studio
Stato
Condizioni
Descrizione dettagliata
The positioning of the patient at the beginning of every CT examination has an major influence on the image quality and especially on the respective effective patient dose in the CT scan. If the patient position is closer to the X-ray tube, the patient dose increases significantly and if the vertical patient position is chosen too far away from the X-ray tube, the applied dose decreases, but also the image quality.
In a retrospective observational study, patients were examined on the CT scanner Somatom X.Cite and divided into two groups, one group was with intelligent camera positioning and the other with manual positioning.
The patients received their CT scan at the Institute Radiology and Nuclear Medicine at the Nuremberg North Hospital strictly according to the clinical indication and without influence of the study.
Tipo di studio
Iscrizione (Effettivo)
Contatti e Sedi
Luoghi di studio
-
-
Bavaria
-
Nuremberg, Bavaria, Germania, 90419
- Institute of Radiology and Nuclear Medicine
-
-
Criteri di partecipazione
Criteri di ammissibilità
Età idonea allo studio
Accetta volontari sani
Sessi ammissibili allo studio
Metodo di campionamento
Popolazione di studio
Descrizione
Inclusion Criteria:
Only patients with a clinical indication for a CT Scan
Exclusion Criteria:
renal insufficiency, manifest hyperthyroidism and pregnancy
Piano di studio
Come è strutturato lo studio?
Dettagli di progettazione
- Modelli osservazionali: Caso di controllo
- Prospettive temporali: Retrospettiva
Coorti e interventi
Gruppo / Coorte |
---|
experimental group
patients positioned with 3D camera
|
control group
patients positioned by experienced radiology assistants
|
Cosa sta misurando lo studio?
Misure di risultato primarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
---|---|---|
Effective dose
Lasso di tempo: 10 months
|
the absorbed dose to different organs of the body
|
10 months
|
Collaboratori e investigatori
Sponsor
Investigatori
- Direttore dello studio: Michael M. Lell, Prof. Dr., Department of Radiology and Nuclear Medicine, Klinikum Nuernberg, Paracelsus Medical University, Germany
Pubblicazioni e link utili
Pubblicazioni generali
- Booij R, Budde RPJ, Dijkshoorn ML, van Straten M. Accuracy of automated patient positioning in CT using a 3D camera for body contour detection. Eur Radiol. 2019 Apr;29(4):2079-2088. doi: 10.1007/s00330-018-5745-z. Epub 2018 Oct 10.
- Kaasalainen T, Palmu K, Reijonen V, Kortesniemi M. Effect of patient centering on patient dose and image noise in chest CT. AJR Am J Roentgenol. 2014 Jul;203(1):123-30. doi: 10.2214/AJR.13.12028.
- Saltybaeva N, Schmidt B, Wimmer A, Flohr T, Alkadhi H. Precise and Automatic Patient Positioning in Computed Tomography: Avatar Modeling of the Patient Surface Using a 3-Dimensional Camera. Invest Radiol. 2018 Nov;53(11):641-646. doi: 10.1097/RLI.0000000000000482.
- Marsh RM, Silosky MS. The effects of patient positioning when interpreting CT dose metrics: A phantom study. Med Phys. 2017 Apr;44(4):1514-1524. doi: 10.1002/mp.12137. Epub 2017 Mar 17.
- Dane B, O'Donnell T, Liu S, Vega E, Mohammed S, Singh V, Kapoor A, Megibow A. Radiation dose reduction, improved isocenter accuracy and CT scan time savings with automatic patient positioning by a 3D camera. Eur J Radiol. 2021 Mar;136:109537. doi: 10.1016/j.ejrad.2021.109537. Epub 2021 Jan 12.
- Manava P, Galster M, Ammon J, Singer J, Lell MM, Rieger V. Optimized Camera-Based Patient Positioning in CT: Impact on Radiation Exposure. Invest Radiol. 2022 Aug 2. doi: 10.1097/RLI.0000000000000904. [Epub ahead of print]
Studiare le date dei record
Studia le date principali
Inizio studio (Effettivo)
Completamento primario (Effettivo)
Completamento dello studio (Effettivo)
Date di iscrizione allo studio
Primo inviato
Primo inviato che soddisfa i criteri di controllo qualità
Primo Inserito (Effettivo)
Aggiornamenti dei record di studio
Ultimo aggiornamento pubblicato (Effettivo)
Ultimo aggiornamento inviato che soddisfa i criteri QC
Ultimo verificato
Maggiori informazioni
Termini relativi a questo studio
Altri numeri di identificazione dello studio
- CT_2021
Piano per i dati dei singoli partecipanti (IPD)
Hai intenzione di condividere i dati dei singoli partecipanti (IPD)?
Informazioni su farmaci e dispositivi, documenti di studio
Studia un prodotto farmaceutico regolamentato dalla FDA degli Stati Uniti
Studia un dispositivo regolamentato dalla FDA degli Stati Uniti
Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .