このページは自動翻訳されたものであり、翻訳の正確性は保証されていません。を参照してください。 英語版 ソーステキスト用。

Real-World Validation of A Minimally Disruptive Ambulatory Blood Pressure Monitoring (ABPM) Huawei Smartwatch in Malaysia (Huawei-SunU02)

2026年10月28日 更新者:YookChin Chia、Sunway University

Background:

Ambulatory Blood Pressure Monitoring (ABPM) is the gold standard for diagnosing high blood pressure (hypertension) because it tracks blood pressure continuously across day and night. However, standard ABPM requires a cuff that inflates every 15 to 30 minutes during the day and every 30 to 60 minutes at night. These frequent inflations often cause discomfort, interrupt sleep, and cause patients to abandon testing.

Purpose:

This study aims to evaluate whether a new cuff-based Huawei smartwatch can provide a less-disturbing ABPM experience. Researchers will test whether a reduced-frequency measurement schedule (e.g., 7 readings: 4 awake, 3 asleep) can provide the same diagnostic accuracy as a conventional, full 24-hour ABPM protocol (typically 48 readings) while improving patient comfort.

Study Design & Procedures:

Study Population:

Approximately 500 adult participants will be recruited over a 24-month study period, with a targeted subgroup of 50 participants selected for clinical validation. The subgroup will represent normotensive, pre-hypertensive, and hypertensive individuals across diverse age groups and ethnicities.

Monitoring & Comparison: Participants in the validation group will undergo standard 24-hour ABPM at an accredited clinical facility alongside monitoring with the Huawei smartwatch. Data from the smartwatch will be used to compare a reduced-frequency reading schedule against full 24-hour ABPM recordings.

User Feedback: Participants will complete questionnaires assessing comfort, sleep quality, ease of use, and overall preference between the smartwatch and standard ABPM cuffs.

Goal:

The main goal is to determine if a reduced-frequency wearable monitoring protocol can maintain clinical diagnostic accuracy for blood pressure and sleep-dipping patterns while significantly reducing patient burden and sleep disruption.

調査の概要

状態

招待による登録

条件

研究の種類

観察的

入学 (推定)

500

連絡先と場所

このセクションには、調査を実施する担当者の連絡先の詳細と、この調査が実施されている場所に関する情報が記載されています。

研究場所

    • Selangor
      • Subang Jaya、Selangor、マレーシア、47500
        • Faculty of Medical and Life Sciences, Sunway University

参加基準

研究者は、適格基準と呼ばれる特定の説明に適合する人を探します。これらの基準のいくつかの例は、人の一般的な健康状態または以前の治療です。

適格基準

就学可能な年齢

18年歳以上 (大人、高齢者)

健康ボランティアの受け入れ

はい

受講資格のある性別

全て

サンプリング方法

非確率サンプル

調査対象母集団

The study population comprises adult community-dwelling individuals residing in Malaysia recruited via the Huawei Research Platform. From this general screening pool (target N = 500), a targeted clinical validation subgroup (n = 50) will be prospectively selected to undergo dual ABPM and smartwatch evaluation.

To ensure clinical applicability and statistical validity in alignment with ISO 81060-2 and AAMI validation standards, the validation subgroup will be stratified into three distinct cohorts:

  • Normotensive Cohort (n = 10): Participants with blood pressure < 135/85 mmHg.
  • Undiagnosed Elevated Blood Pressure Cohort (n = 20): Participants with untreated blood pressure > 135/85 mmHg.
  • Documented Hypertensive Cohort (n = 20): Participants with documented or self-reported history of history of hypertension.

Selection across all cohorts will maintain balanced demographic representation across age groups, biological sex, and major Malaysian ethnic populations (Malay, Chinese, Indian)

説明

Inclusion Criteria:

  • Possesses the specific Huawei wearable required for the study
  • Aged ≥18 years
  • Malaysian citizen/permanent resident
  • Understands English, Malay, or Chinese
  • Able to download the required application on a compatible smartphone
  • Willing to wear devices continuously
  • Able to provide informed consent

Exclusion Criteria:

  • Current pregnancy
  • Presence of implantable cardiac devices (e.g., permanent pacemakers, •implantable cardioverter-defibrillators [ICDs])
  • Uncontrolled cardiac arrhythmias (e.g., active atrial fibrillation, frequent premature ventricular contractions [PVCs])
  • Movement disorders or active tremors that interfere with blood pressure measurement or signal acquisition (e.g., severe Parkinson's disease, essential tremor)
  • Acute medical conditions (e.g., acute coronary syndromes, stroke/TIA within 3 months, acute decompensated heart failure, severe active infections, or hypertensive emergencies)
  • Resting blood pressure >=180/110 mmHg

研究計画

このセクションでは、研究がどのように設計され、研究が何を測定しているかなど、研究計画の詳細を提供します。

研究はどのように設計されていますか?

デザインの詳細

  • 観測モデル:ECOLOGIC_OR_COMMUNITY
  • 時間の展望:CROSS_SECTIONAL

コホートと介入

グループ/コホート
Normotensive Cohort

Adult participants with normal blood pressure (defined as blood pressure < 135/85 mmHg; n = 10) who undergo simultaneous standard 24-hour ABPM and Huawei smartwatch monitoring.

Undiagnosed Elevated Blood Pressure Cohort

Adult participants with undiagnosed elevated blood pressure (defined as blood pressure > 135/85 mmHg; n = 20) who undergo simultaneous standard 24-hour ABPM and Huawei smartwatch monitoring.

Documented Hypertensive Cohort

Adult participants with documented or self-reported hypertension (n = 20) who undergo simultaneous standard 24-hour ABPM and Huawei smartwatch monitoring.

Screening Population

General adult participants recruited across Malaysia who enroll via the Huawei Research Platform app to provide continuous wearable ambulatory blood pressure and physiological data and complete baseline health assessments over the study period.

この研究は何を測定していますか?

主要な結果の測定

結果測定
メジャーの説明
時間枠
Mean Difference in Blood Pressure Measured by Reduced-Frequency Smartwatch Schedule Versus Standard 24-Hour Ambulatory Blood Pressure Monitor (ABPM)
時間枠:Month 13 through Month 18 (during clinical ABPM validation testing).

Measure Description: Mean blood pressure difference (bias) for 24-hour, daytime, and night-time periods calculated by subtracting reference 24-hour ABPM values (SpaceLabs/Mobil-O-Graph cuff monitor) from reduced-frequency Huawei smartwatch ABPM values (7-measurement schedule).

Unit of Measure: mmHg

Month 13 through Month 18 (during clinical ABPM validation testing).
Mean Absolute Error (MAE) of Blood Pressure Readings Between Reduced-Frequency Smartwatch Schedule and Standard 24-Hour ABPM
時間枠:Month 13 through Month 18.

Measure Description: The average absolute difference between blood pressure readings obtained from the reduced-frequency Huawei smartwatch schedule and the standard 24-hour cuff-based ABPM monitor.

Unit of Measure: mmHg

Month 13 through Month 18.
Root Mean Square Error (RMSE) of Blood Pressure Readings Between Reduced-Frequency Smartwatch Schedule and Standard 24-Hour ABPM
時間枠:Month 13 through Month 18

Measure Description: The root mean square error evaluating agreement between blood pressure values recorded by the reduced-frequency Huawei smartwatch schedule and the standard 24-hour cuff-based ABPM monitor.

Unit of Measure: mmHg

Month 13 through Month 18
Intraclass Correlation Coefficient (ICC) of Blood Pressure Measurements Between Reduced-Frequency Smartwatch Schedule and Standard 24-Hour ABPM
時間枠:Month 13 through Month 18

Measure Description: Agreement reliability calculated using absolute-agreement two-way random-effects ICC comparing reduced-frequency Huawei smartwatch blood pressure estimates against full 24-hour ABPM readings.

Unit of Measure: Intraclass Correlation Coefficient (score ranging from -1.0 to +1.0)

Month 13 through Month 18

二次結果の測定

結果測定
メジャーの説明
時間枠
Limits of Agreement (Bland-Altman) for Blood Pressure Readings Between Huawei Smartwatch and Conventional Cuff ABPM
時間枠:Month 13 through Month 18

Measure Description: Upper and lower 95% limits of agreement (mean difference +/- 1.96 x standard deviation) calculated via Bland-Altman analysis comparing simultaneous Huawei smartwatch and conventional cuff-based ABPM measurements.

Unit of Measure: mmHg

Month 13 through Month 18
Participant Comfort Score Evaluated via Post-Monitoring User Experience Survey
時間枠:Immediately following completion of 24-hour monitoring (Month 13 through Month 18)

Measure Description: Mean total score for participant-reported physical comfort derived from a standardized 5-point Likert scale questionnaire administered immediately after 24-hour monitoring.

Unit of Measure: Score on a scale of 1 to 5 (where 1 = Extreme Discomfort and 5 = Complete Comfort)

Immediately following completion of 24-hour monitoring (Month 13 through Month 18)
Number of Participants Indicating Preference for Smartwatch Monitoring Over Conventional Cuff ABPM
時間枠:Immediately following completion of 24-hour monitoring (Month 13 through Month 18)

Measure Description: Count of participants who express a preference for smartwatch-based monitoring over traditional arm-cuff ABPM on the post-monitoring user experience survey.

Unit of Measure: Participant count

Immediately following completion of 24-hour monitoring (Month 13 through Month 18)

その他の成果指標

結果測定
メジャーの説明
時間枠
Cohen's Kappa Coefficient for Nocturnal Dipping Pattern Classification Concordance
時間枠:Month 16 through Month 20.

Measure Description: Inter-rater agreement (Cohen's Kappa) for categorizing participants into nocturnal dipping status categories (dipper vs. non-dipper) comparing reduced-frequency smartwatch schedules against standard full-day ABPM.

Unit of Measure: Kappa Statistic (score ranging from -1.0 to +1.0)

Month 16 through Month 20.
Minimum Number of Measurements Required to Maintain Diagnostic Accurac
時間枠:Month 16 through Month 20.

Measure Description: The minimum threshold count of active blood pressure inflations required per 24-hour period to maintain an ICC >= 0.80 relative to standard 24-hour ABPM.

Unit of Measure: Number of blood pressure inflations

Month 16 through Month 20.

協力者と研究者

ここでは、この調査に関係する人々や組織を見つけることができます。

スポンサー

協力者

捜査官

  • 主任研究者:Yook Chin Chia、Sunway University
  • 主任研究者:Yee How Say, BSc PhD、Sunway University

出版物と役立つリンク

研究に関する情報を入力する責任者は、自発的にこれらの出版物を提供します。これらは、研究に関連するあらゆるものに関するものである可能性があります。

一般刊行物

研究記録日

これらの日付は、ClinicalTrials.gov への研究記録と要約結果の提出の進捗状況を追跡します。研究記録と報告された結果は、国立医学図書館 (NLM) によって審査され、公開 Web サイトに掲載される前に、特定の品質管理基準を満たしていることが確認されます。

主要日程の研究

研究開始 (推定)

2026年12月15日

一次修了 (推定)

2028年11月30日

研究の完了 (推定)

2028年12月31日

試験登録日

最初に提出

2026年10月28日

QC基準を満たした最初の提出物

2026年10月28日

最初の投稿 (実際)

2026年11月3日

学習記録の更新

投稿された最後の更新 (実際)

2026年11月3日

QC基準を満たした最後の更新が送信されました

2026年10月28日

最終確認日

2026年10月1日

詳しくは

本研究に関する用語

追加の関連 MeSH 用語

その他の研究ID番号

  • Huawei-SunU-MinDisABPMStudy

個々の参加者データ (IPD) の計画

個々の参加者データ (IPD) を共有する予定はありますか?

はい

IPD プランの説明

De-identified individual participant data (IPD) collected during this study will be shared with qualified external researchers. Specific shared data elements include:

  • Demographics & Baseline Characteristics: De-identified participant age, sex, body mass index, and baseline clinical blood pressure measurements.
  • Wearable Sensor Data: Continuous physiological streams captured via the Huawei Research Platform, including raw photoplethysmography (PPG) signals, heart rate, heart rate variability, physical activity metrics, and sleep tracking parameters.
  • Reference ABPM Data: Matched 24-hour, daytime, and night-time blood pressure and heart rate values, along with derived nocturnal dipping indices, obtained from clinical cuff-based ABPM and reduced-frequency Huawei smartwatch schedules.
  • User Experience Data: De-identified survey scores on device comfort, ease of use, sleep disturbance, and monitoring preference.

IPD 共有時間枠

Access Start Date: 6 months after the publication of the primary study results in a peer-reviewed journal.

Access End Date: 5 years after initial publication.

IPD 共有アクセス基準

Method of Access: Data will be made available via a secure, managed data repository (or directly from the Principal Investigator upon formal request).

Qualified Requesters: Scientific researchers affiliated with recognized academic, clinical, or research institutions.

Request Requirements: Requesters must submit a methodologically sound research proposal and a clear scientific rationale for secondary analysis.

Data Use Agreement (DUA): Execution of a formal Data Use Agreement (DUA) is required to ensure data protection, mandate non-re-identification, and enforce secure handling.

Compliance: All shared files will be fully de-identified in compliance with the Personal Data Protection Act (PDPA) Malaysia and institutional Ethics Committee guidelines.

IPD 共有サポート情報タイプ

  • STUDY_PROTOCOL
  • SAP
  • ICF

医薬品およびデバイス情報、研究文書

米国FDA規制医薬品の研究

いいえ

米国FDA規制機器製品の研究

いいえ

この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。