This study will test the control and measurements of a new system for airway support in healthy volunteers. The goals of the study are to evaluate how accurately the system can stabilize airway pressure to a desired level, evaluate the accuracy of the system's reported volume of each breath (tidal volume), evaluate the accuracy of the system's reported respiratory rate, evaluate the accuracy of exhaled carbon dioxide monitoring with a standard carbon dioxide monitor when the novel PAP system is used, and evaluate general comfort of the volunteers while using the system.
調査の概要
詳細な説明
Sedation procedures involve administering drugs to patients that reduce awareness and pain without complete loss of consciousness. A common side effect of these drugs is partial or complete obstruction of the airway by the airway tissue, inhibiting the flow of oxygen to the lungs. Research has shown that this risk can be effectively reduced by administering PAP therapy during sedation, which involves directing air flow into a tight-fitting mask over the patient's nostrils and/or mouth. However, non-invasive ventilators used for administering PAP therapy are expensive and large, and often not available in areas of the hospital where sedation procedures take place.
The investigators have developed a small and low-cost novel PAP system for use in sedation procedures that controls airway pressure to desired levels (a feature that is generally only available in larger and higher-cost devices), provides estimates for tidal volumes (generally only available with a non-invasive ventilator), provides respiratory rate estimates, and allows for more accurate monitoring of exhaled carbon dioxide with a standard carbon dioxide monitor using a novel approach that to the investigator's knowledge has not been published. The goal of investigator's study is to demonstrate the accuracy of these features in human subjects. The system is much smaller and less expensive than a non-invasive ventilator, provides enhanced respiratory monitoring, and has the potential to be a practical alternative that would make PAP therapy during sedation much more feasible.
The investigator's evaluation will primarily include evaluating the investigator's system's performance against an FDA cleared commercial respiratory monitor to verify accuracy. The investigators anticipate that the investigator's system will automatically adjust oxygen flow to maintain a desired airway pressure even in the presence of variables such as mask leak and patient inhalation flow. The investigators also anticipate that the investigator's system will provide tidal volume estimates within 15% of the respiratory monitor reference measurement, as well as respiratory rate estimates that closely match the respiratory monitor. The investigator's system employs a method that detects when the patient exhales and removes mask pressure during exhalation, and the investigators anticipate that this will result in improved comfort for the patient, as well as improved carbon dioxide monitoring accuracy, since removing flow during exhalation reduces dilution of the expired gas sample that is used for measuring exhaled carbon dioxide levels.
研究の種類
入学 (実際)
段階
- 適用できない
連絡先と場所
研究場所
-
-
Utah
-
Salt Lake City、Utah、アメリカ、84108
- University of Utah Pain Management Center
-
-
参加基準
適格基準
就学可能な年齢
健康ボランティアの受け入れ
受講資格のある性別
説明
Inclusion Criteria:
- Healthy volunteers
Exclusion Criteria:
- Have any respiratory or cardiovascular conditions
- i.e., cold, allergies, congestion, pneumonia, lung cancer, COPD, acute respiratory distress syndrome, etc.
研究計画
研究はどのように設計されていますか?
デザインの詳細
- 主な目的:デバイスの実現可能性
- 割り当て:なし
- 介入モデル:単一グループの割り当て
- マスキング:なし(オープンラベル)
武器と介入
参加者グループ / アーム |
介入・治療 |
|---|---|
|
実験的:CPAP
Enrolled healthy volunteers to examine airway pressure support through the investigational device. |
Airway pressure support through the investigational device |
この研究は何を測定していますか?
主要な結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Tidal Volume Accuracy
時間枠:During the intervention, up to 20 minutes
|
The mean and standard deviation of the percent difference in tidal volume measurement between our PAP system and the respiratory monitor |
During the intervention, up to 20 minutes
|
二次結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Respiratory Rate
時間枠:During the intervention, up to 20 minutes
|
The average respiratory rate reported by the respiratory monitor and rate reported by the PAP system for each volunteer will be compared. The average difference for each volunteer will be calculated, and a 95% confidence interval will be constructed, with the ideal mean difference being zero with low variance. |
During the intervention, up to 20 minutes
|
|
EtCO2
時間枠:During the intervention, up to 20 minutes
|
End-tidal carbon dioxide measurements at 10 cmH2O of mask pressure will be compared between when the system turned off flow during exhalation and when delivering flow throughout the entire breath. We anticipate that end-tidal CO2 levels will read lower when flow is delivered continuously during exhalation because of dilution of the expired gas, and this will be demonstrated by calculating the average difference in end-tidal CO2 between these two approaches. |
During the intervention, up to 20 minutes
|
|
Mask Pressure Control Accuracy
時間枠:During the intervention, up to 20 minutes
|
The measured pressure in the mask compared to the set pressure in the 50 milliseconds immediately preceding inhalation will be assessed for evaluating pressure control. Average error will be calculated and a 95% confidence interval will be constructed. We anticipate the error to be less than 1 cmH2O. |
During the intervention, up to 20 minutes
|
協力者と研究者
出版物と役立つリンク
一般刊行物
- Burk KM, Sakata DJ, Kuck K, Orr JA. Comparing Nasal End-Tidal Carbon Dioxide Measurement Variation and Agreement While Delivering Pulsed and Continuous Flow Oxygen in Volunteers and Patients. Anesth Analg. 2020 Mar;130(3):715-724. doi: 10.1213/ANE.0000000000004004.
- Fogarty M, Orr JA, Sakata D, Brewer L, Johnson K, Fang JC, Kuck K. A comparison of ventilation with a non-invasive ventilator versus standard O2 with a nasal cannula for colonoscopy with moderate sedation using propofol. J Clin Monit Comput. 2020 Dec;34(6):1215-1221. doi: 10.1007/s10877-019-00426-5. Epub 2019 Nov 23.
研究記録日
主要日程の研究
研究開始 (実際)
一次修了 (実際)
研究の完了 (実際)
試験登録日
最初に提出
QC基準を満たした最初の提出物
最初の投稿 (実際)
学習記録の更新
投稿された最後の更新 (実際)
QC基準を満たした最後の更新が送信されました
最終確認日
詳しくは
本研究に関する用語
個々の参加者データ (IPD) の計画
個々の参加者データ (IPD) を共有する予定はありますか?
医薬品およびデバイス情報、研究文書
米国FDA規制医薬品の研究
米国FDA規制機器製品の研究
米国で製造され、米国から輸出された製品。
この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。