Evaluación del Sistema de Presión Positiva en las Vías Respiratorias (PAP) en Voluntarios
Evaluación de un sistema novedoso de presión positiva de la vía aérea en voluntarios
This study will test the control and measurements of a new system for airway support in healthy volunteers. The goals of the study are to evaluate how accurately the system can stabilize airway pressure to a desired level, evaluate the accuracy of the system's reported volume of each breath (tidal volume), evaluate the accuracy of the system's reported respiratory rate, evaluate the accuracy of exhaled carbon dioxide monitoring with a standard carbon dioxide monitor when the novel PAP system is used, and evaluate general comfort of the volunteers while using the system.
Descripción general del estudio
Descripción detallada
Sedation procedures involve administering drugs to patients that reduce awareness and pain without complete loss of consciousness. A common side effect of these drugs is partial or complete obstruction of the airway by the airway tissue, inhibiting the flow of oxygen to the lungs. Research has shown that this risk can be effectively reduced by administering PAP therapy during sedation, which involves directing air flow into a tight-fitting mask over the patient's nostrils and/or mouth. However, non-invasive ventilators used for administering PAP therapy are expensive and large, and often not available in areas of the hospital where sedation procedures take place.
The investigators have developed a small and low-cost novel PAP system for use in sedation procedures that controls airway pressure to desired levels (a feature that is generally only available in larger and higher-cost devices), provides estimates for tidal volumes (generally only available with a non-invasive ventilator), provides respiratory rate estimates, and allows for more accurate monitoring of exhaled carbon dioxide with a standard carbon dioxide monitor using a novel approach that to the investigator's knowledge has not been published. The goal of investigator's study is to demonstrate the accuracy of these features in human subjects. The system is much smaller and less expensive than a non-invasive ventilator, provides enhanced respiratory monitoring, and has the potential to be a practical alternative that would make PAP therapy during sedation much more feasible.
The investigator's evaluation will primarily include evaluating the investigator's system's performance against an FDA cleared commercial respiratory monitor to verify accuracy. The investigators anticipate that the investigator's system will automatically adjust oxygen flow to maintain a desired airway pressure even in the presence of variables such as mask leak and patient inhalation flow. The investigators also anticipate that the investigator's system will provide tidal volume estimates within 15% of the respiratory monitor reference measurement, as well as respiratory rate estimates that closely match the respiratory monitor. The investigator's system employs a method that detects when the patient exhales and removes mask pressure during exhalation, and the investigators anticipate that this will result in improved comfort for the patient, as well as improved carbon dioxide monitoring accuracy, since removing flow during exhalation reduces dilution of the expired gas sample that is used for measuring exhaled carbon dioxide levels.
Tipo de estudio
Inscripción (Actual)
Fase
- No aplica
Contactos y Ubicaciones
Ubicaciones de estudio
-
-
Utah
-
Salt Lake City, Utah, Estados Unidos, 84108
- University of Utah Pain Management Center
-
-
Criterios de participación
Criterio de elegibilidad
Edades elegibles para estudiar
Acepta Voluntarios Saludables
Géneros elegibles para el estudio
Descripción
Inclusion Criteria:
- Healthy volunteers
Exclusion Criteria:
- Have any respiratory or cardiovascular conditions
- i.e., cold, allergies, congestion, pneumonia, lung cancer, COPD, acute respiratory distress syndrome, etc.
Plan de estudios
¿Cómo está diseñado el estudio?
Detalles de diseño
- Propósito principal: Viabilidad del dispositivo
- Asignación: N / A
- Modelo Intervencionista: Asignación de un solo grupo
- Enmascaramiento: Ninguno (etiqueta abierta)
Armas e Intervenciones
Grupo de participantes/brazo |
Intervención / Tratamiento |
|---|---|
|
Experimental: CPAP
Enrolled healthy volunteers to examine airway pressure support through the investigational device. |
Airway pressure support through the investigational device |
¿Qué mide el estudio?
Medidas de resultado primarias
Medida de resultado |
Medida Descripción |
Periodo de tiempo |
|---|---|---|
|
Tidal Volume Accuracy
Periodo de tiempo: During the intervention, up to 20 minutes
|
The mean and standard deviation of the percent difference in tidal volume measurement between our PAP system and the respiratory monitor |
During the intervention, up to 20 minutes
|
Medidas de resultado secundarias
Medida de resultado |
Medida Descripción |
Periodo de tiempo |
|---|---|---|
|
Respiratory Rate
Periodo de tiempo: During the intervention, up to 20 minutes
|
The average respiratory rate reported by the respiratory monitor and rate reported by the PAP system for each volunteer will be compared. The average difference for each volunteer will be calculated, and a 95% confidence interval will be constructed, with the ideal mean difference being zero with low variance. |
During the intervention, up to 20 minutes
|
|
EtCO2
Periodo de tiempo: During the intervention, up to 20 minutes
|
End-tidal carbon dioxide measurements at 10 cmH2O of mask pressure will be compared between when the system turned off flow during exhalation and when delivering flow throughout the entire breath. We anticipate that end-tidal CO2 levels will read lower when flow is delivered continuously during exhalation because of dilution of the expired gas, and this will be demonstrated by calculating the average difference in end-tidal CO2 between these two approaches. |
During the intervention, up to 20 minutes
|
|
Mask Pressure Control Accuracy
Periodo de tiempo: During the intervention, up to 20 minutes
|
The measured pressure in the mask compared to the set pressure in the 50 milliseconds immediately preceding inhalation will be assessed for evaluating pressure control. Average error will be calculated and a 95% confidence interval will be constructed. We anticipate the error to be less than 1 cmH2O. |
During the intervention, up to 20 minutes
|
Colaboradores e Investigadores
Patrocinador
Colaboradores
Investigadores
- Investigador principal: Kai Kuck, PhD, University of Utah
Publicaciones y enlaces útiles
Publicaciones Generales
- Burk KM, Sakata DJ, Kuck K, Orr JA. Comparing Nasal End-Tidal Carbon Dioxide Measurement Variation and Agreement While Delivering Pulsed and Continuous Flow Oxygen in Volunteers and Patients. Anesth Analg. 2020 Mar;130(3):715-724. doi: 10.1213/ANE.0000000000004004.
- Fogarty M, Orr JA, Sakata D, Brewer L, Johnson K, Fang JC, Kuck K. A comparison of ventilation with a non-invasive ventilator versus standard O2 with a nasal cannula for colonoscopy with moderate sedation using propofol. J Clin Monit Comput. 2020 Dec;34(6):1215-1221. doi: 10.1007/s10877-019-00426-5. Epub 2019 Nov 23.
Fechas de registro del estudio
Fechas importantes del estudio
Inicio del estudio (Actual)
Finalización primaria (Actual)
Finalización del estudio (Actual)
Fechas de registro del estudio
Enviado por primera vez
Primero enviado que cumplió con los criterios de control de calidad
Publicado por primera vez (Actual)
Actualizaciones de registros de estudio
Última actualización publicada (Actual)
Última actualización enviada que cumplió con los criterios de control de calidad
Última verificación
Más información
Términos relacionados con este estudio
Palabras clave
Términos MeSH relevantes adicionales
Otros números de identificación del estudio
- IRB_00190165
Plan de datos de participantes individuales (IPD)
¿Planea compartir datos de participantes individuales (IPD)?
Información sobre medicamentos y dispositivos, documentos del estudio
Estudia un producto farmacéutico regulado por la FDA de EE. UU.
Estudia un producto de dispositivo regulado por la FDA de EE. UU.
producto fabricado y exportado desde los EE. UU.
Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .