검하 몸체 센서를 사용한 Belun Ring Gen3 딥 러닝 알고리즘 검증
검하 신체 센서를 사용한 Belun Ring Gen3 딥 러닝 알고리즘: 바이오마커 역학 분석을 통해 수면 장애 호흡에 대한 진단 기능 탐색
Hypothesis: BR's Gen3 DL algorithms, combined with its subxiphoid body sensor, can accurately diagnose OSA, categorize its severity, identify REM OSA and supine OSA, and detect central sleep apnea (CSA).
Primary Objective:
To rigorously evaluate the overall performance of the BR with Gen3 DL Algorithms and Subxiphoid Body Sensor in assessing SDB in individuals referred to the sleep labs with clinical suspicion of sleep apnea and a STOP-Bang score > 3, by comparing to the attended in-lab PSG, the gold standard.
Secondary Objectives:
To determine the accuracy of BR sleep stage parameters using the Gen3 DL algorithms by comparing to the in-lab PSG;
To assess the accuracy of the BR arrhythmia detection algorithm;
To assess the impact of CPAP on HRV (both time- and frequency-domain), delta HR, hypoxic burden, and PWADI during split night studies;
To assess if any of the baseline HRV parameters (both time- and frequency-domain), delta heart rate (referred to as Delta HR), hypoxic burden, and pulse wave amplitude drop index (PWADI) or the change of these parameters may predict CPAP compliance;
To evaluate the minimum duration of quality data necessary for BR to achieve OSA diagnosis;
To examine the performance of OSA screening tools using OSA predictive AI models formulated by National Taiwan University Hospital (NTUH) and Northeast Ohio Medical University (NEOMED).
연구 개요
연구 유형
등록 (추정된)
단계
- 해당 없음
연락처 및 위치
연구 장소
-
-
Ohio
-
Chardon, Ohio, 미국, 44024
- UH Geauga Health Center Services
-
Cleveland, Ohio, 미국, 44106
- University Hospitals Cleveland Medical Center
-
-
참여기준
자격 기준
공부할 수 있는 나이
건강한 자원 봉사자를 받아들입니다
연구 대상 성별
설명
Inclusion Criteria:
- Provision of signed informed consent form.
- Clinically assessed and suspicious for OSA with a STOP-Bang score ≥ 3.
Exclusion Criteria:
- Full night PAP titration study.
- On home O2, noninvasive ventilator, diaphragmatic pacing, or any form of a nerve stimulator.
- Having atrial fibrillation-flutter, pacemaker/defibrillator, left ventricular assist device (LVAD), or status post cardiac transplantation.
- Recent hospitalization or recent surgery in the past 30 days.
- Unstable cardiopulmonary status on the night of the study judged to be unsafe for sleep study by the sleep tech and/or the on-call sleep physician.
If a participant did not sleep for at least 4 hours of technically valid sleep based on the Belun Ring method for diagnostic assessments, or a minimum of 3 hours of technically valid sleep during the diagnostic phase of a split-night study, the patient will be excluded from statistical analysis.
공부 계획
연구는 어떻게 설계됩니까?
디자인 세부사항
- 주 목적: 특수 증상
- 할당: 해당 없음
- 중재 모델: 단일 그룹 할당
- 마스킹: 없음(오픈 라벨)
무기와 개입
참가자 그룹 / 팔 |
개입 / 치료 |
|---|---|
|
실험적: UH-Participant
Potential participants will be identified from patients scheduled for in-lab PSG at the two medical centers of University Hospitals |
The Belun Ring sensor should be placed on the palmar side of the proximal phalanx of the index finger and the sensor should be placed along the radial artery such that the accuracy of the device will be minimally affected by skin color. The Ring has 7 adjustable arms for different finger sizes. Each device is reusable after thorough cleaning with an alcohol swab. The "Belun Cor" body sensor accessory is composed of an accelerometer, a temperature sensor, and a lithium battery. It will be placed immediately below the xiphoid process in the upper abdomen with a medical adhesive to detect the body temperature, body posture, respiratory rate, and respiratory efforts. |
연구는 무엇을 측정합니까?
주요 결과 측정
결과 측정 |
측정값 설명 |
기간 |
|---|---|---|
|
Performance of diagnosing sleep disordered breathing by BR's Gen3 DL algorithms
기간: 2 years
|
To establish mean, standard deviation, and target clinical agreement limit values for the differences in sleep-disordered breathing parameters (including AHI3%, AHI4%, REM AHI 3%, REM AHI4%, Supine AHI3%, Supine AHI4%, and CAI) from the BR against PSG. To calculate the sensitivity and specificity values, along with 95% confidence intervals (CIs), for sleep-disordered breathing parameters from the BR against PSG at PSG cutoffs of 5, 15, and 30 events/h. Pearson correlation and regression analysis will be employed to assess the association between sleep-disordered breathing parameters obtained from the BR and PSG. |
2 years
|
2차 결과 측정
결과 측정 |
측정값 설명 |
기간 |
|---|---|---|
|
Performance of sleep stage classification by BR's Gen3 DL algorithms
기간: 2 years
|
Sensitivity and specificity of the Epoch-by-Epoch (EBE) 4-stage classification (wake, deep sleep, light sleep, REM) of the BR will be compared to EBE PSG data for each stage against the combination of the other three classifications. For instance, sensitivity for detecting wake will measure the proportion of correctly classified wake epochs against the combined classifications of deep sleep, light sleep, and REM sleep. Sensitivity and specificity values, accompanied by 95% CIs, of EBE body position (supine vs. non-supine) against PSG will also be computed. |
2 years
|
|
To evaluate the minimum duration of quality data necessary for OSA diagnosis
기간: 2 years
|
To explore the minimum number of hours needed by the BR for an accurate OSA diagnosis, sensitivity, specificity, and Matthews Correlation Coefficient (MCC) values, along with 95% CIs, will be computed for sleep-disordered breathing parameters of Belun total record time (bTRT) and belun total sleep time (bTST) with different numbers of hours (e.g., 2, 3, 4, 5, 6, 7, and 8 hours) against PSG data. In cases where varying numbers of subjects meet different bTRT and bTST criteria at different durations, resampling methods like oversampling, undersampling, bootstrapping, etc., will be employed to address data imbalances. |
2 years
|
|
Accuracy of the BR arrhythmia-detecting algorithm
기간: 3 years
|
To calculate the performance metrics, including sensitivity and specificity, along with 95% CIs, for the detection of arrhythmias by BR device in comparison to assessments made by a board-certified sleep physician. The physician will review the PSG ECG signals of the subjects suspected of having significant arrhythmias during the sleep study. |
3 years
|
|
Biomarker relationship analysis
기간: 3 years
|
Pearson correlation and regression analysis will be employed: (1) to explore the relationships between autonomic nervous system (ANS) metrics and delta HR, hypoxic burden, and pulse wave amplitude drop index (PWADI); (2) to evaluate the association between HRV, Delta HR, hypoxic burden, PWADI, CPAP adherence, sleep-disordered breathing parameters, and demographic data (including age, gender, and BMI). If data distribution assumptions are violated, the Spearman correlation will be used instead of the Pearson correlation. |
3 years
|
|
Accuracy of the NTUH and NEOMED models
기간: 3 years
|
The investigators will compute sensitivity and specificity values, accompanied by their respective 95% CIs, for AHI3% and AHI4% from the NTUH/NEOMED models in comparison to PSG. Furthermore, the investigators will determine the ROC curve at PSG-defined cutoffs of 5, 15, and 30 events/h for both AHI3% and AHI4%. |
3 years
|
공동 작업자 및 조사자
수사관
- 수석 연구원: Ambrose A. Chiang, MD, University Hospitals Cleveland Medical Center
- 수석 연구원: Susheel P. Patil, MD, University Hospitals Cleveland Medical Center
- 수석 연구원: Kingman P. Strohl, MD, University Hospitals Cleveland Medical Center
연구 기록 날짜
연구 주요 날짜
연구 시작 (실제)
기본 완료 (추정된)
연구 완료 (추정된)
연구 등록 날짜
최초 제출
QC 기준을 충족하는 최초 제출
처음 게시됨 (실제)
연구 기록 업데이트
마지막 업데이트 게시됨 (실제)
QC 기준을 충족하는 마지막 업데이트 제출
마지막으로 확인됨
추가 정보
이 연구와 관련된 용어
추가 관련 MeSH 약관
기타 연구 ID 번호
- STUDY20231436
개별 참가자 데이터(IPD) 계획
개별 참가자 데이터(IPD)를 공유할 계획입니까?
약물 및 장치 정보, 연구 문서
미국 FDA 규제 의약품 연구
미국 FDA 규제 기기 제품 연구
미국에서 제조되어 미국에서 수출되는 제품
이 정보는 변경 없이 clinicaltrials.gov 웹사이트에서 직접 가져온 것입니다. 귀하의 연구 세부 정보를 변경, 제거 또는 업데이트하도록 요청하는 경우 register@clinicaltrials.gov. 문의하십시오. 변경 사항이 clinicaltrials.gov에 구현되는 즉시 저희 웹사이트에도 자동으로 업데이트됩니다. .