Assessing Blood Pressure Variability in a Healthy Adult Population During Transcutaneous Hypoglossal Nerve Stimulation With the Zeus Sleep Device (Zeus-BPV)

3 november 2026 bijgewerkt door: Royal Brompton & Harefield NHS Foundation Trust

Transcutaneous hypoglossal nerve stimulation is an emerging treatment option for Obstructive Sleep Apnoea (OSA), a sleep condition where breathing is repeatedly limited. The standard treatment for OSA is not always well tolerated therefore, more alternatives are needed.

Transcutaneous hypoglossal nerve stimulation uses a device, such as Zeus, which sticks underneath the chin to deliver electrical pulses during sleep. These pulses stimulate the hypoglossal nerve to prevent the tongue from relaxing and blocking the airway.

Device placement is important for specific nerve targeting; however, its proximity is close to receptors involved in blood pressure regulation. Whilst current research indicates that transcutaneous stimulation is unlikely to increase blood pressure, OSA patients have an increased risk of developing high blood pressure, therefore treatment options should not make this worse. Further research to understand the impact of transcutaneous hypoglossal nerve stimulation on blood pressure is therefore needed. In this study, the blood pressure of participants will be measured whilst they wear the Zeus sleep device. It is expected that blood pressure will either decrease or remain constant in line with current research. Findings could facilitate improved confidence in device use for OSA treatment and provide patients with alternative treatment options.

Studie Overzicht

Toestand

Werving

Conditie

Interventie / Behandeling

Gedetailleerde beschrijving

During sleep the upper-airway muscles relax, in obstructive sleep apnoea (OSA) this causes repeated narrowing (hypopnoea) and obstruction (apnoea) (NICE, 2021). These periods of reduced breathing cause arousals fragmenting sleep. Symptoms include snoring, impaired performance, waking headaches, and excessive daytime sleepiness (Arnardottir et al., 2016, Spicuzza et al., 2015). OSA additionally impacts quality of life and cognition (Vaessen et al., 2015) and increases risks of hypertension and cardiovascular morbidity (Peppard et al., 2000, Nieto, 2000, Marin et al., 2012). Diagnosis of OSA uses sleep studies, requiring an apnoea hypopnoea index (AHI) of ≥5/hr (NICE, 2021). Peppard et al. (2013) estimate the prevalence of sleep disordered breathing, of which OSA is the most common, to be 26% of 30-70-year-olds and rising.

The gold standard treatment of OSA is continuous positive airway pressure (CPAP) (NICE, 2021), requiring >4hrs/night usage for effective treatment and risk reduction; however, CPAP is not always well tolerated. A review by Weaver et al. (2008) found CPAP nonadherence rates of 29-83%. Alternative treatments include positional modifiers, mandibular advancement devices, and surgery (NICE, 2021, Spicuzza et al., 2015) whose usage is limited by pathophysiology. Further CPAP alternatives are needed.

Hypoglossal nerve stimulation (HNS) is an emerging OSA therapy which uses electrical stimulation to contract the genioglossus muscle, preventing prolapse and maintaining airway patency (NICE, 2017). Strollo et al. (2014) conducted a prospective, single-group, cohort study where non-CPAP compliant OSA patients had a subcutaneous HNS device implanted. A clinically significant decrease in median AHI of 68% (29.3 to 7.4/hr, P<0.001) was seen 12 months post-implantation alongside improved sleepiness and quality of life. A separate study of 113 participants by Woodson et al. (2025) also showed significant improvements in mean AHI (-18.3 ± 11.8/hour, P<0.001) following HNS implantation. Whilst HNS has shown promising efficacy and acceptable safety profile (NICE, 2017, Strollo et al., 2014, Woodson et al. 2025, Friedman et al. 2016), HNS requires a general anaesthesia and carries all the associated risks and costs of an invasive procedure. Interest has therefore turned to transcutaneous HNS as a non-invasive, cost-efficient alternative.

According to Steier et al. (2011), transcutaneous electrical stimulation (TESLA) of the hypoglossal nerve via bilateral hydrogel patches located in the submental area causes contraction of the genioglossus muscle similarly to subcutaneous HNS. A recent randomised controlled trial in OSA patients by Ratneswaran et al., (2023) compared TESLA to usual care (CPAP). An adjusted reduction in AHI of 7/hr with TESLA was seen at 3 months, however this was not significant (P=0.12) potentially due to small sample size or absent additional interrogation. Indeed, Pengo et al. (2016) indicated that only some participants are responders to TESLA, in their randomised, sham-controlled crossover trial a significant reduction in AHI was only seen in responders (9.1/hr, p=0.004). Meta-analysis by Wang et al. (2025) and Young Jae Byun et al. (2020) both concluded that TESLA results in a meaningful AHI reduction. TESLA appears to be well tolerated by patients with minimal adverse events (Pengo et al., 2016, Ratneswaran et al., 2023, Reed et al., 2016).

Despite the promising research, it is possible that submental electrical stimulation could affect blood pressure due to its proximity to carotid sinus baroreceptors. A similar device to TESLA, TENS machines, are contraindicated for use in the neck region (NHS, 2017). However, a sham-controlled, randomized crossover trial of HNS in 60 participants by Dedhia et al., (2023) concluded there was not a significant difference in mean 24-systolic blood pressure between sham and active groups. Furthermore, a study by Alsharifi et al. (2023) found submental TESLA in 13 healthy participants caused a significant decrease in blood pressure (systolic: p=4.93E-06, diastolic: p=3.48E-09, mean: p=3.88E-08); supported by a meta-analysis that concluded electrical stimulation of the upper airways reduced both systolic and diastolic blood pressure (Khaled et al., 2025). Whilst current evidence indicates that TESLA is unlikely to elicit hypertension, in a patient cohort already at increased risk of hypertension and cardiovascular morbidity, it is vital risk is not compounded by treatment.

This study aims to evaluate the short-term impact of hypoglossal nerve TESLA on blood pressure. We hypothesise that submental TESLA will not increase blood pressure or short-term blood pressure variability and may lead to modest reductions, consistent with previous studies.

Studietype

Ingrijpend

Inschrijving (Geschat)

30

Fase

  • Niet toepasbaar

Contacten en locaties

In dit gedeelte vindt u de contactgegevens van degenen die het onderzoek uitvoeren en informatie over waar dit onderzoek wordt uitgevoerd.

Studiecontact

Studie Contact Back-up

Studie Locaties

Deelname Criteria

Onderzoekers zoeken naar mensen die aan een bepaalde beschrijving voldoen, de zogenaamde geschiktheidscriteria. Enkele voorbeelden van deze criteria zijn iemands algemene gezondheidstoestand of eerdere behandelingen.

Geschiktheidscriteria

Leeftijden die in aanmerking komen voor studie

18 jaar en ouder (Volwassen, Oudere volwassene)

Accepteert gezonde vrijwilligers

Ja

Geslachten die in aanmerking komen voor studie

Allemaal

Beschrijving

Inclusion Criteria:

  1. Adults aged 18 or older
  2. Either biological sex
  3. Healthy volunteers
  4. Able to provide informed consent
  5. No contraindications to hypoglossal nerve stimulation using the Zeus device
  6. BMI >18.5 and <30kg/m2

Exclusion Criteria:

  1. Skin allergy, including eczema or allergic contact dermatitis in device placement area
  2. Diagnosed heart condition or cardiovascular disease
  3. Cardiovascular event in the previous 2 months
  4. Implanted medical devices including cardioverter defibrillator and pacemaker
  5. Previous or current cardiac arrhythmia
  6. Autonomic neuropathy
  7. Neuromuscular disease
  8. Diagnosed epilepsy or recurrent seizures
  9. Hypertension (>140/90 mmHg)
  10. Hypotension (<95/65 mmHg)
  11. Metal work in the head or neck, excluding dental fillings
  12. Current migraine
  13. Pregnancy
  14. Diagnosed or suspected deep vein thrombosis within previous 2 months
  15. Recent cancer diagnosis or cancerous legions in proximity to device placement area
  16. Alcohol dependence
  17. Recent infection requiring hospital admission or medical treatment within the previous month, including leprosy
  18. Swollen, inflamed, infected, or broken skin in device placement area
  19. Recent surgery within 10 months in, or near, device placement area
  20. Currently taking medications, excluding contraceptive medication
  21. Long facial hair in device placement area

Studie plan

Dit gedeelte bevat details van het studieplan, inclusief hoe de studie is opgezet en wat de studie meet.

Hoe is de studie opgezet?

Ontwerpdetails

  • Primair doel: Fundamentele wetenschap
  • Toewijzing: Gerandomiseerd
  • Interventioneel model: Crossover-opdracht
  • Masker: Geen (open label)

Wapens en interventies

Deelnemersgroep / Arm
Interventie / Behandeling
Experimenteel: Group 1- Active stimulation followed by sham stimulation

Transcutaneous hypoglossal nerve stimulation using the Zeus Sleep device

Experimenteel: Group 2- Sham stimulation followed by active stimulation

Transcutaneous hypoglossal nerve stimulation using the Zeus Sleep device

Wat meet het onderzoek?

Primaire uitkomstmaten

Uitkomstmaat
Tijdsspanne
Change from baseline in systolic and diastolic blood pressure during Zeus sleep device usage
Tijdsspanne: baseline, intervention
baseline, intervention

Secundaire uitkomstmaten

Uitkomstmaat
Tijdsspanne
Change in pulse rate variability from baseline during active stimulation
Tijdsspanne: baseline, intervention
baseline, intervention

Medewerkers en onderzoekers

Hier vindt u mensen en organisaties die betrokken zijn bij dit onderzoek.

Sponsor

Studie record data

Deze datums volgen de voortgang van het onderzoeksdossier en de samenvatting van de ingediende resultaten bij ClinicalTrials.gov. Studieverslagen en gerapporteerde resultaten worden beoordeeld door de National Library of Medicine (NLM) om er zeker van te zijn dat ze voldoen aan specifieke kwaliteitscontrolenormen voordat ze op de openbare website worden geplaatst.

Bestudeer belangrijke data

Studie start (Werkelijk)

29 september 2026

Primaire voltooiing (Geschat)

19 februari 2027

Studie voltooiing (Geschat)

29 maart 2027

Studieregistratiedata

Eerst ingediend

25 september 2026

Eerst ingediend dat voldeed aan de QC-criteria

3 november 2026

Eerst geplaatst (Werkelijk)

4 november 2026

Updates van studierecords

Laatste update geplaatst (Werkelijk)

4 november 2026

Laatste update ingediend die voldeed aan QC-criteria

3 november 2026

Laatst geverifieerd

1 november 2026

Meer informatie

Termen gerelateerd aan deze studie

Andere studie-ID-nummers

  • 368532

Plan Individuele Deelnemersgegevens (IPD)

Bent u van plan om gegevens van individuele deelnemers (IPD) te delen?

ONBESLIST

Informatie over medicijnen en apparaten, studiedocumenten

Bestudeert een door de Amerikaanse FDA gereguleerd geneesmiddel

Nee

Bestudeert een door de Amerikaanse FDA gereguleerd apparaatproduct

Nee

Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .