Evaluation of WAEVE® Pre-Measurement Feature for Detection of Implanted Coronary Stents in Patients Undergoing Elective PCI (NIMBLE-II)

17. november 2026 oppdatert av: Nimble Diagnostics S.L.

Evaluation of WAEVE® Pre-Measurement Feature for Detection of Implanted Coronary Stents in Patients Undergoing Elective PCI: NIMBLE-II Study

Coronary stent implantation is a common treatment for coronary artery disease. However, current methods used to evaluate implanted stents and detect complications such as in-stent restenosis are generally invasive and may require ionizing radiation.

The WAEVE® device is an investigational, non-invasive, non-ionizing medical device that uses microwave interferometry to assess implanted coronary stents. The NIMBLE-II study is designed to evaluate the performance of a new pre-measurement feature incorporated into WAEVE® that detects heartbeat-synchronous microwave reflections and helps distinguish true stent-related signals from non-cardiac or artifactual signals. By confirming the cardiac origin of the detected reflection, this feature is intended to improve the robustness and reliability of WAEVE® measurements and reduce the risk of false detections arising from non-cardiac or artifactual reflections.

This prospective, multicenter, single-arm clinical investigation will enroll adults undergoing elective percutaneous coronary intervention (PCI). Participants will undergo WAEVE® measurements before PCI and, when a single coronary stent is implanted, after PCI. The study aims to generate technical and clinical evidence supporting the reliability and robustness of the WAEVE® pre-measurement feature under real clinical conditions.

Studieoversikt

Status

Rekruttering

Forhold

Intervensjon / Behandling

Detaljert beskrivelse

Coronary artery disease (CAD) remains one of the leading causes of morbidity and mortality worldwide. Percutaneous coronary intervention (PCI) with coronary stent implantation is a widely used treatment for restoring coronary blood flow in patients with obstructive coronary artery disease. Despite significant advances in stent technology, in-stent restenosis (ISR) and other stent-related complications remain clinically relevant and often require further diagnostic evaluation and treatment.

Current methods for evaluating implanted coronary stents, including invasive coronary angiography and intravascular imaging techniques, are associated with procedural risks, resource utilization, and patient burden. Furthermore, these methods are not suitable for frequent or routine monitoring. Consequently, there is an unmet clinical need for non-invasive, non-ionizing technologies capable of assessing implanted coronary stents and supporting longitudinal patient follow-up.

WAEVE® is an investigational active medical device developed by NIMBLE Diagnostics, S.L. that uses microwave interferometry to non-invasively assess implanted coronary stents. The technology is based on the interaction between low-power microwave signals and metallic coronary stents. Reflected signals are captured and analyzed to extract information related to the presence and behavior of the implanted device. Previous technical and clinical investigations have demonstrated the feasibility of detecting stent-related microwave reflections and have supported the continued development of the technology.

As part of the ongoing development of WAEVE®, a new pre-measurement feature has been incorporated into the system. This feature evaluates whether detected microwave reflections exhibit modulation synchronized with the cardiac cycle by using the electrocardiogram (ECG) signal as a temporal reference. The underlying rationale is that reflections originating from implanted coronary stents should demonstrate a heartbeat-synchronous pattern due to the cyclic motion and deformation of the coronary arteries during the cardiac cycle. In contrast, reflections originating from external structures, environmental sources, patient movement, or other non-cardiac artifacts are not expected to exhibit this physiological synchronization.

By confirming the cardiac origin of the detected reflection, the pre-measurement feature is intended to improve the robustness and reliability of WAEVE® measurements and reduce the risk of false detections arising from non-cardiac or artifactual reflections. The feature is designed to support operators in identifying measurements that are more likely to contain true stent-related information before the acquisition is accepted for further analysis.

The NIMBLE-II study is a prospective, multicenter, single-arm clinical investigation designed to evaluate the technical performance of this pre-measurement feature under real-world clinical conditions. The study will be conducted in Spain and will enroll adult patients undergoing elective PCI. Participants without previously implanted coronary stents will undergo WAEVE® measurements before PCI. Patients in whom a single coronary stent is successfully implanted during the procedure will undergo additional post-PCI measurements and will contribute to the primary analysis cohort.

The primary objective of the study is to determine the proportion of post-PCI pre-measurement acquisitions in which a heartbeat-synchronous microwave signal is detected. Secondary objectives include the characterization of false detections, assessment of heartbeat synchronism metrics, evaluation of signal repeatability and measurement stability, assessment of signal changes associated with stent implantation, and evaluation of operational performance indicators. Additional analyses will explore factors that may influence signal acquisition and detection performance.

An exploratory substudy conducted at Hospital Universitari Germans Trias i Pujol will investigate inflammatory biomarkers associated with PCI and evaluate potential relationships between biomarker levels and WAEVE® signal characteristics. These analyses are intended to generate hypotheses and support future research regarding the biological and physiological factors that may influence microwave signal behavior following stent implantation.

The results of the NIMBLE-II study are expected to provide technical and clinical validation evidence supporting the performance of the WAEVE® pre-measurement feature and contribute to the continued development of a non-invasive, non-ionizing technology for the assessment of implanted coronary stents.

Studietype

Intervensjonell

Registrering (Antatt)

120

Fase

  • Ikke aktuelt

Kontakter og plasseringer

Denne delen inneholder kontaktinformasjon for de som utfører studien, og informasjon om hvor denne studien blir utført.

Studiekontakt

Studiesteder

    • Catalonia
      • Badalona, Catalonia, Spania, 08916
        • Rekruttering
        • Hospital Universitari Germans Trias i Pujol
        • Ta kontakt med:
          • Oriol Rodriguez Leor, MD, PhD
          • Telefonnummer: (+34) 934 97 89 89
          • E-post: icor@icor.cat
      • L'Hospitalet de Llobregat, Catalonia, Spania, 08907
        • Har ikke rekruttert ennå
        • Hospital Universitari de Bellvitge

Deltakelseskriterier

Forskere ser etter personer som passer til en bestemt beskrivelse, kalt kvalifikasjonskriterier. Noen eksempler på disse kriteriene er en persons generelle helsetilstand eller tidligere behandlinger.

Kvalifikasjonskriterier

Alder som er kvalifisert for studier

18 år og eldre (Voksen, Eldre voksen)

Tar imot friske frivillige

Nei

Kjønn som er kvalifisert for studier

Alle

Beskrivelse

Inclusion Criteria:

  • Adults ≥18 years of age.
  • scheduled for a PCI procedure with potential stent implantation and without any previously implanted coronary stent.
  • Clinically stable condition allowing completion of pre-PCI WAEVE® measurements.
  • Ability to understand the study procedures and provide written informed consent.

Exclusion Criteria:

  • Hemodynamic instability, cardiogenic shock, or severe arrhythmias.
  • Active implantable electronic devices or metallic implants on the chest that may interfere with measurements (e.g., pacemakers, defibrillators, CRT devices, sternal wires, rib plates, metallic clips).
  • Presence of skin conditions, wounds, or lesions in the areas designated for transducer placement (including extensive thoracic skin lesions).
  • Inability to remain supine or still during WAEVE® measurements.
  • Pregnancy or absence of a negative pregnancy test in women of childbearing potential.

Studieplan

Denne delen gir detaljer om studieplanen, inkludert hvordan studien er utformet og hva studien måler.

Hvordan er studiet utformet?

Designdetaljer

  • Primært formål: Annen
  • Tildeling: N/A
  • Intervensjonsmodell: Enkeltgruppeoppdrag
  • Masking: Ingen (Open Label)

Våpen og intervensjoner

Deltakergruppe / Arm
Intervensjon / Behandling
Eksperimentell: WAEVE® Assessment

Participants undergoing elective percutaneous coronary intervention (PCI) will undergo WAEVE® measurements before PCI. Participants receiving a single coronary stent during PCI will undergo additional post-PCI WAEVE® measurements for evaluation of the study endpoints.

WAEVE® is an investigational, non-invasive, non-ionizing medical device that uses microwave interferometry to detect and analyze microwave reflections associated with implanted coronary stents. Participants undergo WAEVE® measurements before PCI and, when applicable, after coronary stent implantation. The study evaluates the performance of a pre-measurement feature designed to detect heartbeat-synchronous microwave reflections and distinguish true cardiac and stent-related signals from non-cardiac or artifactual reflections.

Hva måler studien?

Primære resultatmål

Resultatmål
Tiltaksbeskrivelse
Tidsramme
Heartbeat-Synchronous Signal Detection Rate in Post-PCI Acquisitions
Tidsramme: Day 1 (day of the PCI procedure)

Proportion of assessable post-PCI pre-measurement acquisitions in which a heartbeat-synchronous microwave signal is detected, defined by the presence of statistically significant periodic modulation of the microwave signal at the cardiac frequency using the electrocardiogram (ECG) signal as the temporal cardiac-cycle reference.

Day 1 (day of the PCI procedure)

Sekundære resultatmål

Resultatmål
Tiltaksbeskrivelse
Tidsramme
False Detection Rate
Tidsramme: Day 1 (day of the PCI procedure)

Proportion of assessable acquisitions exhibiting stent-compatible microwave patterns that do not demonstrate heartbeat-linked periodicity and are therefore classified as non-cardiac or artifactual reflections.

Day 1 (day of the PCI procedure)
Intra-Patient Measurement Repeatability
Tidsramme: Day 1 (day of the PCI procedure)

Repeatability of repeated WAEVE acquisitions performed under identical acquisition conditions within the same session, quantified using the Measurement Stability Index (MSI). The Measurement Stability Index ranges from 0 to 1, with higher values indicating greater agreement between repeated acquisitions

Day 1 (day of the PCI procedure)
Proportion of Technically Assessable Acquisitions
Tidsramme: Day 1 (day of the PCI procedure)

Proportion of acquisitions meeting predefined signal-quality criteria and considered suitable for analysis

Day 1 (day of the PCI procedure)
Signal Changes Associated With Stent Implantation
Tidsramme: Day 1 (day of the PCI procedure)

Within-subject differences between pre-PCI and post-PCI microwave reflection characteristics at matched anatomical sites, summarized using reflection morphology features and the Peak Prominence Difference Index (PPDI).

Day 1 (day of the PCI procedure)
Heart Rate Agreement Between Microwave-Derived and ECG-Derived Signals
Tidsramme: Day 1 (day of the PCI procedure)

Agreement between heart rate estimated from microwave reflections and heart rate derived from ECG, summarized using Heart Rate Error (HRE).

Day 1 (day of the PCI procedure)

Andre resultatmål

Resultatmål
Tiltaksbeskrivelse
Tidsramme
Association Between Inflammatory Biomarkers and WAEVE® Signal Parameters
Tidsramme: Day 1 (day of the PCI procedure)

Exploratory correlations between inflammatory biomarker levels and selected WAEVE® signal characteristics, including heartbeat-synchronous signal metrics and stent-related reflection features.

Day 1 (day of the PCI procedure)

Samarbeidspartnere og etterforskere

Det er her du vil finne personer og organisasjoner som er involvert i denne studien.

Sponsor

Samarbeidspartnere

Publikasjoner og nyttige lenker

Den som er ansvarlig for å legge inn informasjon om studien leverer frivillig disse publikasjonene. Disse kan handle om alt relatert til studiet.

Studierekorddatoer

Disse datoene sporer fremdriften for innsending av studieposter og sammendragsresultater til ClinicalTrials.gov. Studieposter og rapporterte resultater gjennomgås av National Library of Medicine (NLM) for å sikre at de oppfyller spesifikke kvalitetskontrollstandarder før de legges ut på det offentlige nettstedet.

Studer hoveddatoer

Studiestart (Faktiske)

8. juni 2026

Primær fullføring (Antatt)

31. desember 2027

Studiet fullført (Antatt)

31. desember 2027

Datoer for studieregistrering

Først innsendt

11. juni 2026

Først innsendt som oppfylte QC-kriteriene

17. november 2026

Først lagt ut (Faktiske)

24. november 2026

Oppdateringer av studieposter

Sist oppdatering lagt ut (Faktiske)

24. november 2026

Siste oppdatering sendt inn som oppfylte QC-kriteriene

17. november 2026

Sist bekreftet

1. november 2026

Mer informasjon

Begreper knyttet til denne studien

Andre studie-ID-numre

  • NDX-0002

Plan for individuelle deltakerdata (IPD)

Planlegger du å dele individuelle deltakerdata (IPD)?

NEI

IPD-planbeskrivelse

Individual participant data (IPD) will not be made publicly available. The study is an early-stage technical validation investigation of an investigational medical device and includes proprietary device-generated data and analyses. Any future data-sharing requests will be considered on a case-by-case basis in accordance with applicable data protection regulations, participant consent, and sponsor policies.

Legemiddel- og utstyrsinformasjon, studiedokumenter

Studerer et amerikansk FDA-regulert medikamentprodukt

Nei

Studerer et amerikansk FDA-regulert enhetsprodukt

Nei

Denne informasjonen ble hentet direkte fra nettstedet clinicaltrials.gov uten noen endringer. Hvis du har noen forespørsler om å endre, fjerne eller oppdatere studiedetaljene dine, vennligst kontakt register@clinicaltrials.gov. Så snart en endring er implementert på clinicaltrials.gov, vil denne også bli oppdatert automatisk på nettstedet vårt. .