Evaluation of WAEVE® Pre-Measurement Feature for Detection of Implanted Coronary Stents in Patients Undergoing Elective PCI (NIMBLE-II)
Evaluation of WAEVE® Pre-Measurement Feature for Detection of Implanted Coronary Stents in Patients Undergoing Elective PCI: NIMBLE-II Study
Coronary stent implantation is a common treatment for coronary artery disease. However, current methods used to evaluate implanted stents and detect complications such as in-stent restenosis are generally invasive and may require ionizing radiation.
The WAEVE® device is an investigational, non-invasive, non-ionizing medical device that uses microwave interferometry to assess implanted coronary stents. The NIMBLE-II study is designed to evaluate the performance of a new pre-measurement feature incorporated into WAEVE® that detects heartbeat-synchronous microwave reflections and helps distinguish true stent-related signals from non-cardiac or artifactual signals. By confirming the cardiac origin of the detected reflection, this feature is intended to improve the robustness and reliability of WAEVE® measurements and reduce the risk of false detections arising from non-cardiac or artifactual reflections.
This prospective, multicenter, single-arm clinical investigation will enroll adults undergoing elective percutaneous coronary intervention (PCI). Participants will undergo WAEVE® measurements before PCI and, when a single coronary stent is implanted, after PCI. The study aims to generate technical and clinical evidence supporting the reliability and robustness of the WAEVE® pre-measurement feature under real clinical conditions.
Visão geral do estudo
Status
Status
Condições
Condições
Intervenção / Tratamento
Intervenção / Tratamento
Descrição detalhada
Coronary artery disease (CAD) remains one of the leading causes of morbidity and mortality worldwide. Percutaneous coronary intervention (PCI) with coronary stent implantation is a widely used treatment for restoring coronary blood flow in patients with obstructive coronary artery disease. Despite significant advances in stent technology, in-stent restenosis (ISR) and other stent-related complications remain clinically relevant and often require further diagnostic evaluation and treatment.
Current methods for evaluating implanted coronary stents, including invasive coronary angiography and intravascular imaging techniques, are associated with procedural risks, resource utilization, and patient burden. Furthermore, these methods are not suitable for frequent or routine monitoring. Consequently, there is an unmet clinical need for non-invasive, non-ionizing technologies capable of assessing implanted coronary stents and supporting longitudinal patient follow-up.
WAEVE® is an investigational active medical device developed by NIMBLE Diagnostics, S.L. that uses microwave interferometry to non-invasively assess implanted coronary stents. The technology is based on the interaction between low-power microwave signals and metallic coronary stents. Reflected signals are captured and analyzed to extract information related to the presence and behavior of the implanted device. Previous technical and clinical investigations have demonstrated the feasibility of detecting stent-related microwave reflections and have supported the continued development of the technology.
As part of the ongoing development of WAEVE®, a new pre-measurement feature has been incorporated into the system. This feature evaluates whether detected microwave reflections exhibit modulation synchronized with the cardiac cycle by using the electrocardiogram (ECG) signal as a temporal reference. The underlying rationale is that reflections originating from implanted coronary stents should demonstrate a heartbeat-synchronous pattern due to the cyclic motion and deformation of the coronary arteries during the cardiac cycle. In contrast, reflections originating from external structures, environmental sources, patient movement, or other non-cardiac artifacts are not expected to exhibit this physiological synchronization.
By confirming the cardiac origin of the detected reflection, the pre-measurement feature is intended to improve the robustness and reliability of WAEVE® measurements and reduce the risk of false detections arising from non-cardiac or artifactual reflections. The feature is designed to support operators in identifying measurements that are more likely to contain true stent-related information before the acquisition is accepted for further analysis.
The NIMBLE-II study is a prospective, multicenter, single-arm clinical investigation designed to evaluate the technical performance of this pre-measurement feature under real-world clinical conditions. The study will be conducted in Spain and will enroll adult patients undergoing elective PCI. Participants without previously implanted coronary stents will undergo WAEVE® measurements before PCI. Patients in whom a single coronary stent is successfully implanted during the procedure will undergo additional post-PCI measurements and will contribute to the primary analysis cohort.
The primary objective of the study is to determine the proportion of post-PCI pre-measurement acquisitions in which a heartbeat-synchronous microwave signal is detected. Secondary objectives include the characterization of false detections, assessment of heartbeat synchronism metrics, evaluation of signal repeatability and measurement stability, assessment of signal changes associated with stent implantation, and evaluation of operational performance indicators. Additional analyses will explore factors that may influence signal acquisition and detection performance.
An exploratory substudy conducted at Hospital Universitari Germans Trias i Pujol will investigate inflammatory biomarkers associated with PCI and evaluate potential relationships between biomarker levels and WAEVE® signal characteristics. These analyses are intended to generate hypotheses and support future research regarding the biological and physiological factors that may influence microwave signal behavior following stent implantation.
The results of the NIMBLE-II study are expected to provide technical and clinical validation evidence supporting the performance of the WAEVE® pre-measurement feature and contribute to the continued development of a non-invasive, non-ionizing technology for the assessment of implanted coronary stents.
Tipo de estudo
Tipo de estudo
Inscrição (Estimado)
Inscrição
Estágio
Estágio
- Não aplicável
Contactos e Locais
Contato de estudo
Contato de estudo
- Nome: Oriol Rodríguez Leor, MD, PhD
- Número de telefone: 93 465 12 00
- E-mail: orodriguez.germanstrias@gencat.cat
Locais de estudo
-
-
Catalonia
-
Badalona, Catalonia, Espanha, 08916
- Recrutamento
- Hospital Universitari Germans Trias i Pujol
-
Contato:
- Oriol Rodriguez Leor, MD, PhD
- Número de telefone: (+34) 934 97 89 89
- E-mail: icor@icor.cat
-
L'Hospitalet de Llobregat, Catalonia, Espanha, 08907
- Ainda não está recrutando
- Hospital Universitari de Bellvitge
-
-
Critérios de participação
Critérios de elegibilidade
Critérios de elegibilidade
Idades elegíveis para estudo
Aceita Voluntários Saudáveis
Gêneros Elegíveis para o Estudo
Descrição
Inclusion Criteria:
- Adults ≥18 years of age.
- scheduled for a PCI procedure with potential stent implantation and without any previously implanted coronary stent.
- Clinically stable condition allowing completion of pre-PCI WAEVE® measurements.
- Ability to understand the study procedures and provide written informed consent.
Exclusion Criteria:
- Hemodynamic instability, cardiogenic shock, or severe arrhythmias.
- Active implantable electronic devices or metallic implants on the chest that may interfere with measurements (e.g., pacemakers, defibrillators, CRT devices, sternal wires, rib plates, metallic clips).
- Presence of skin conditions, wounds, or lesions in the areas designated for transducer placement (including extensive thoracic skin lesions).
- Inability to remain supine or still during WAEVE® measurements.
- Pregnancy or absence of a negative pregnancy test in women of childbearing potential.
Plano de estudo
Como o estudo é projetado?
Detalhes do projeto
- Finalidade Principal: Outro
- Alocação: N / D
- Modelo Intervencional: Atribuição de grupo único
- Mascaramento: Nenhum (rótulo aberto)
Número de braços
Armas e Intervenções
Grupo de Participantes / BraçoGrupo de Participantes / Braço |
Intervenção / TratamentoIntervenção / Tratamento |
|---|---|
|
Experimental: WAEVE® Assessment
Participants undergoing elective percutaneous coronary intervention (PCI) will undergo WAEVE® measurements before PCI. Participants receiving a single coronary stent during PCI will undergo additional post-PCI WAEVE® measurements for evaluation of the study endpoints. |
WAEVE® is an investigational, non-invasive, non-ionizing medical device that uses microwave interferometry to detect and analyze microwave reflections associated with implanted coronary stents. Participants undergo WAEVE® measurements before PCI and, when applicable, after coronary stent implantation. The study evaluates the performance of a pre-measurement feature designed to detect heartbeat-synchronous microwave reflections and distinguish true cardiac and stent-related signals from non-cardiac or artifactual reflections. |
O que o estudo está medindo?
Medidas de resultados primários
Medidas de resultados primários
Medida de resultado |
Descrição da medida |
Prazo |
|---|---|---|
|
Heartbeat-Synchronous Signal Detection Rate in Post-PCI Acquisitions
Prazo: Day 1 (day of the PCI procedure)
|
Proportion of assessable post-PCI pre-measurement acquisitions in which a heartbeat-synchronous microwave signal is detected, defined by the presence of statistically significant periodic modulation of the microwave signal at the cardiac frequency using the electrocardiogram (ECG) signal as the temporal cardiac-cycle reference. |
Day 1 (day of the PCI procedure)
|
Medidas de resultados secundários
Medidas de resultados secundários
Medida de resultado |
Descrição da medida |
Prazo |
|---|---|---|
|
False Detection Rate
Prazo: Day 1 (day of the PCI procedure)
|
Proportion of assessable acquisitions exhibiting stent-compatible microwave patterns that do not demonstrate heartbeat-linked periodicity and are therefore classified as non-cardiac or artifactual reflections. |
Day 1 (day of the PCI procedure)
|
|
Intra-Patient Measurement Repeatability
Prazo: Day 1 (day of the PCI procedure)
|
Repeatability of repeated WAEVE acquisitions performed under identical acquisition conditions within the same session, quantified using the Measurement Stability Index (MSI). The Measurement Stability Index ranges from 0 to 1, with higher values indicating greater agreement between repeated acquisitions |
Day 1 (day of the PCI procedure)
|
|
Proportion of Technically Assessable Acquisitions
Prazo: Day 1 (day of the PCI procedure)
|
Proportion of acquisitions meeting predefined signal-quality criteria and considered suitable for analysis |
Day 1 (day of the PCI procedure)
|
|
Signal Changes Associated With Stent Implantation
Prazo: Day 1 (day of the PCI procedure)
|
Within-subject differences between pre-PCI and post-PCI microwave reflection characteristics at matched anatomical sites, summarized using reflection morphology features and the Peak Prominence Difference Index (PPDI). |
Day 1 (day of the PCI procedure)
|
|
Heart Rate Agreement Between Microwave-Derived and ECG-Derived Signals
Prazo: Day 1 (day of the PCI procedure)
|
Agreement between heart rate estimated from microwave reflections and heart rate derived from ECG, summarized using Heart Rate Error (HRE). |
Day 1 (day of the PCI procedure)
|
Outras medidas de resultado
Outras medidas de resultado
Medida de resultado |
Descrição da medida |
Prazo |
|---|---|---|
|
Association Between Inflammatory Biomarkers and WAEVE® Signal Parameters
Prazo: Day 1 (day of the PCI procedure)
|
Exploratory correlations between inflammatory biomarker levels and selected WAEVE® signal characteristics, including heartbeat-synchronous signal metrics and stent-related reflection features. |
Day 1 (day of the PCI procedure)
|
Colaboradores e Investigadores
Patrocinador
Patrocinador
Colaboradores
Colaboradores
Publicações e links úteis
Links úteis
Datas de registro do estudo
Datas Principais do Estudo
Início do estudo (Real)
Início do estudo
Conclusão Primária (Estimado)
Conclusão Primária
Conclusão do estudo (Estimado)
Conclusão do estudo
Datas de inscrição no estudo
Enviado pela primeira vez
Enviado pela primeira vez
Enviado pela primeira vez que atendeu aos critérios de CQ
Enviado pela primeira vez que atendeu aos critérios de CQ
Primeira postagem (Real)
Primeira postagem
Atualizações de registro de estudo
Última Atualização Postada (Real)
Última Atualização Postada
Última atualização enviada que atendeu aos critérios de controle de qualidade
Última atualização enviada que atendeu aos critérios de controle de qualidade
Última verificação
Última verificação
Mais Informações
Termos relacionados a este estudo
Palavras-chave
Termos MeSH relevantes adicionais
Outros números de identificação do estudo
Outros números de identificação do estudo
- NDX-0002
Plano para dados de participantes individuais (IPD)
Planeja compartilhar dados de participantes individuais (IPD)?
Descrição do plano IPD
Individual participant data (IPD) will not be made publicly available. The study is an early-stage technical validation investigation of an investigational medical device and includes proprietary device-generated data and analyses. Any future data-sharing requests will be considered on a case-by-case basis in accordance with applicable data protection regulations, participant consent, and sponsor policies.
Informações sobre medicamentos e dispositivos, documentos de estudo
Estuda um medicamento regulamentado pela FDA dos EUA
Estuda um produto de dispositivo regulamentado pela FDA dos EUA
Essas informações foram obtidas diretamente do site clinicaltrials.gov sem nenhuma alteração. Se você tiver alguma solicitação para alterar, remover ou atualizar os detalhes do seu estudo, entre em contato com register@clinicaltrials.gov. Assim que uma alteração for implementada em clinicaltrials.gov, ela também será atualizada automaticamente em nosso site .