- ICH GCP
- US Clinical Trials Registry
- Klinisk utprøving NCT07680231
Smartphone Versus Back Range of Motion: Validity and Reliability in Lumbar Range of Motion Assessment
29. juni 2026 oppdatert av: Loay Akmal Ali, Cairo University
Clinimetric Properties of Smartphone Application as a Suitable Alternative to Back Range of Motion Device in Assessing the Back Range of Motion
This study aims to investigate the clinimetric properties of the iPhone Level application as a suitable alternative to the Back Range of Motion device in assessing lumbar range of motion, including flexion, extension, and lateral flexion.
Studieoversikt
Status
Fullført
Forhold
Intervensjon / Behandling
Detaljert beskrivelse
Range of motion measurement is vital in physiotherapy for assessing deficits, establishing diagnoses, and planning interventions.
Digital advancements now allow Range of motion assessment through devices and applications, such as the back range of motion, which is a gold standard for measuring lateral flexion due to its validity.
Furthermore, studies indicate that the iPhone Level app effectively measures Range of motion in various joints, demonstrating good reliability and validity.
This study aims to evaluate the reliability and concurrent validity of the iPhone Level application specifically for measuring lumbar flexion, extension, and side bending Range of motion.
Studietype
Observasjonsmessig
Registrering (Faktiske)
120
Kontakter og plasseringer
Denne delen inneholder kontaktinformasjon for de som utfører studien, og informasjon om hvor denne studien blir utført.
Studiesteder
-
-
-
Giza, Egypt, 12613
- Cairo University
-
-
Deltakelseskriterier
Forskere ser etter personer som passer til en bestemt beskrivelse, kalt kvalifikasjonskriterier. Noen eksempler på disse kriteriene er en persons generelle helsetilstand eller tidligere behandlinger.
Kvalifikasjonskriterier
Alder som er kvalifisert for studier
- Voksen
Tar imot friske frivillige
Ja
Prøvetakingsmetode
Sannsynlighetsprøve
Studiepopulasjon
Healthy University Students with normal BMI Healthy Male university students
Beskrivelse
Inclusion Criteria:
- Age 18-25 years old
- Healthy Male university students
- All participants have normal BMI
Exclusion Criteria:
- Recent back, pelvic, or abdominal surgeries, traumatic injury to the back .
- Lower limb deformities, e.g., leg length discrepancy.
- Adolescent idiopathic scoliosis
- Any overweight or obese participants
Studieplan
Denne delen gir detaljer om studieplanen, inkludert hvordan studien er utformet og hva studien måler.
Hvordan er studiet utformet?
Designdetaljer
Hva måler studien?
Primære resultatmål
Resultatmål |
Tiltaksbeskrivelse |
Tidsramme |
|---|---|---|
|
assessment of flexion and extension range of motion using inclinometer
Tidsramme: at baseline
|
During lumbar flexion and extension measurements using the Back Range of Motion (BROM) device, patients stand upright with extended knees and feet approximately 17 centimeters apart.
For flexion, they bend forward to touch the floor, while for extension, they bend backward with crossed arms.
An inclinometer is secured over the sacrum, and its position is centered to record readings in degrees.
An L-shaped slide arm measures the distance between the T12 and S1 spinous processes.
The device isolates lumbar motion from pelvic motion, and the stationary flexion/extension unit minimizes errors, enhancing measurement accuracy.
|
at baseline
|
|
assessment of lateral flexion range of motion using inclinometer
Tidsramme: at baseline
|
For measuring lumbar lateral flexion, the subject's sides are parallel to a wall to prevent compensatory forward trunk flexion.
An inclinometer is secured over the sacral region at S1, using a belt for stability.
A gravity goniometer is placed on the upper thoracic region to capture trunk angular displacement during side bending.
The inclinometer is levelled and zeroed before movement, allowing for precise measurement of lateral lumbar movement while reducing thoracic and pelvic compensations.
Accurate placement of the device is essential for reliable results.
The subject slides their hand down the thigh, keeping weight on the opposite leg to ensure the measurement reflects true lateral flexion.measures]
|
at baseline
|
Sekundære resultatmål
Resultatmål |
Tiltaksbeskrivelse |
Tidsramme |
|---|---|---|
|
assessment of flexion and extension range of motion using smartphone
Tidsramme: at baseline
|
To measure lumbar flexion range of motion (ROM) using the iPhone Level app, participants are positioned standing upright with feet shoulder-width apart.
T12-L1 and S1-S2 spinal levels are identified through palpation and marked on the skin.
The iPhone is placed at these levels to record angles during maximum lumbar flexion.
Total lumbar flexion ROM is calculated by subtracting the angle at S1-S2 from T12-L1.
Subjects perform maximum lumbar flexion and extension with straight knees, and measurements are taken across three trials.
|
at baseline
|
|
assessment of lateral flexion range of motion using smartphone
Tidsramme: at baseline
|
During the measurement of lumbar side bending using the iPhone Level application, the subject will stand in a neutral upright position with feet shoulder-width apart and hands placed on the hips to prevent arm movement from influencing the measurement.
The spinous processes of T12 and S2 are identified and marked on the skin.
The iPhone is first placed and zeroed at the T12 level, and the subject is instructed to bend sideways (toward one side) as far as possible without lifting the heels or rotating the trunk.
The device remains aligned at T12 to capture the angular displacement.
The same procedure is repeated with the device positioned at S2, and the final lumbar side bending range of motion is calculated by subtracting the angle at S2 (pelvic movement) from that at T12 (trunk movement).
This method isolates lumbar lateral flexion and has shown good reliability and validity in clinical settings
|
at baseline
|
Samarbeidspartnere og etterforskere
Det er her du vil finne personer og organisasjoner som er involvert i denne studien.
Sponsor
Samarbeidspartnere
Studierekorddatoer
Disse datoene sporer fremdriften for innsending av studieposter og sammendragsresultater til ClinicalTrials.gov. Studieposter og rapporterte resultater gjennomgås av National Library of Medicine (NLM) for å sikre at de oppfyller spesifikke kvalitetskontrollstandarder før de legges ut på det offentlige nettstedet.
Studer hoveddatoer
Studiestart (Faktiske)
1. desember 2025
Primær fullføring (Faktiske)
1. mars 2026
Studiet fullført (Faktiske)
12. mai 2026
Datoer for studieregistrering
Først innsendt
14. juni 2026
Først innsendt som oppfylte QC-kriteriene
29. juni 2026
Først lagt ut (Faktiske)
2. juli 2026
Oppdateringer av studieposter
Sist oppdatering lagt ut (Faktiske)
2. juli 2026
Siste oppdatering sendt inn som oppfylte QC-kriteriene
29. juni 2026
Sist bekreftet
1. juni 2026
Mer informasjon
Begreper knyttet til denne studien
Ytterligere relevante MeSH-vilkår
Andre studie-ID-numre
- P.T.REC/012/006199
Plan for individuelle deltakerdata (IPD)
Planlegger du å dele individuelle deltakerdata (IPD)?
NEI
Legemiddel- og utstyrsinformasjon, studiedokumenter
Studerer et amerikansk FDA-regulert medikamentprodukt
Nei
Studerer et amerikansk FDA-regulert enhetsprodukt
Nei
Denne informasjonen ble hentet direkte fra nettstedet clinicaltrials.gov uten noen endringer. Hvis du har noen forespørsler om å endre, fjerne eller oppdatere studiedetaljene dine, vennligst kontakt register@clinicaltrials.gov. Så snart en endring er implementert på clinicaltrials.gov, vil denne også bli oppdatert automatisk på nettstedet vårt. .