Smartphone Versus Back Range of Motion: Validity and Reliability in Lumbar Range of Motion Assessment

29 juni 2026 uppdaterad av: Loay Akmal Ali, Cairo University

Clinimetric Properties of Smartphone Application as a Suitable Alternative to Back Range of Motion Device in Assessing the Back Range of Motion

This study aims to investigate the clinimetric properties of the iPhone Level application as a suitable alternative to the Back Range of Motion device in assessing lumbar range of motion, including flexion, extension, and lateral flexion.

Studieöversikt

Detaljerad beskrivning

Range of motion measurement is vital in physiotherapy for assessing deficits, establishing diagnoses, and planning interventions. Digital advancements now allow Range of motion assessment through devices and applications, such as the back range of motion, which is a gold standard for measuring lateral flexion due to its validity. Furthermore, studies indicate that the iPhone Level app effectively measures Range of motion in various joints, demonstrating good reliability and validity. This study aims to evaluate the reliability and concurrent validity of the iPhone Level application specifically for measuring lumbar flexion, extension, and side bending Range of motion.

Studietyp

Observationell

Inskrivning (Faktisk)

120

Kontakter och platser

Det här avsnittet innehåller kontaktuppgifter för dem som genomför studien och information om var denna studie genomförs.

Studieorter

      • Giza, Egypten, 12613
        • Cairo University

Deltagandekriterier

Forskare letar efter personer som passar en viss beskrivning, så kallade behörighetskriterier. Några exempel på dessa kriterier är en persons allmänna hälsotillstånd eller tidigare behandlingar.

Urvalskriterier

Åldrar som är berättigade till studier

  • Vuxen

Tar emot friska volontärer

Ja

Testmetod

Sannolikhetsprov

Studera befolkning

Healthy University Students with normal BMI Healthy Male university students

Beskrivning

Inclusion Criteria:

  1. Age 18-25 years old
  2. Healthy Male university students
  3. All participants have normal BMI

Exclusion Criteria:

  1. Recent back, pelvic, or abdominal surgeries, traumatic injury to the back .
  2. Lower limb deformities, e.g., leg length discrepancy.
  3. Adolescent idiopathic scoliosis
  4. Any overweight or obese participants

Studieplan

Det här avsnittet ger detaljer om studieplanen, inklusive hur studien är utformad och vad studien mäter.

Hur är studien utformad?

Designdetaljer

Vad mäter studien?

Primära resultatmått

Resultatmått
Åtgärdsbeskrivning
Tidsram
assessment of flexion and extension range of motion using inclinometer
Tidsram: at baseline
During lumbar flexion and extension measurements using the Back Range of Motion (BROM) device, patients stand upright with extended knees and feet approximately 17 centimeters apart. For flexion, they bend forward to touch the floor, while for extension, they bend backward with crossed arms. An inclinometer is secured over the sacrum, and its position is centered to record readings in degrees. An L-shaped slide arm measures the distance between the T12 and S1 spinous processes. The device isolates lumbar motion from pelvic motion, and the stationary flexion/extension unit minimizes errors, enhancing measurement accuracy.
at baseline
assessment of lateral flexion range of motion using inclinometer
Tidsram: at baseline
For measuring lumbar lateral flexion, the subject's sides are parallel to a wall to prevent compensatory forward trunk flexion. An inclinometer is secured over the sacral region at S1, using a belt for stability. A gravity goniometer is placed on the upper thoracic region to capture trunk angular displacement during side bending. The inclinometer is levelled and zeroed before movement, allowing for precise measurement of lateral lumbar movement while reducing thoracic and pelvic compensations. Accurate placement of the device is essential for reliable results. The subject slides their hand down the thigh, keeping weight on the opposite leg to ensure the measurement reflects true lateral flexion.measures]
at baseline

Sekundära resultatmått

Resultatmått
Åtgärdsbeskrivning
Tidsram
assessment of flexion and extension range of motion using smartphone
Tidsram: at baseline
To measure lumbar flexion range of motion (ROM) using the iPhone Level app, participants are positioned standing upright with feet shoulder-width apart. T12-L1 and S1-S2 spinal levels are identified through palpation and marked on the skin. The iPhone is placed at these levels to record angles during maximum lumbar flexion. Total lumbar flexion ROM is calculated by subtracting the angle at S1-S2 from T12-L1. Subjects perform maximum lumbar flexion and extension with straight knees, and measurements are taken across three trials.
at baseline
assessment of lateral flexion range of motion using smartphone
Tidsram: at baseline
During the measurement of lumbar side bending using the iPhone Level application, the subject will stand in a neutral upright position with feet shoulder-width apart and hands placed on the hips to prevent arm movement from influencing the measurement. The spinous processes of T12 and S2 are identified and marked on the skin. The iPhone is first placed and zeroed at the T12 level, and the subject is instructed to bend sideways (toward one side) as far as possible without lifting the heels or rotating the trunk. The device remains aligned at T12 to capture the angular displacement. The same procedure is repeated with the device positioned at S2, and the final lumbar side bending range of motion is calculated by subtracting the angle at S2 (pelvic movement) from that at T12 (trunk movement). This method isolates lumbar lateral flexion and has shown good reliability and validity in clinical settings
at baseline

Samarbetspartners och utredare

Det är här du hittar personer och organisationer som är involverade i denna studie.

Studieavstämningsdatum

Dessa datum spårar framstegen för inlämningar av studieposter och sammanfattande resultat till ClinicalTrials.gov. Studieposter och rapporterade resultat granskas av National Library of Medicine (NLM) för att säkerställa att de uppfyller specifika kvalitetskontrollstandarder innan de publiceras på den offentliga webbplatsen.

Studera stora datum

Studiestart (Faktisk)

1 december 2025

Primärt slutförande (Faktisk)

1 mars 2026

Avslutad studie (Faktisk)

12 maj 2026

Studieregistreringsdatum

Först inskickad

14 juni 2026

Först inskickad som uppfyllde QC-kriterierna

29 juni 2026

Första postat (Faktisk)

2 juli 2026

Uppdateringar av studier

Senaste uppdatering publicerad (Faktisk)

2 juli 2026

Senaste inskickade uppdateringen som uppfyllde QC-kriterierna

29 juni 2026

Senast verifierad

1 juni 2026

Mer information

Termer relaterade till denna studie

Ytterligare relevanta MeSH-villkor

Andra studie-ID-nummer

  • P.T.REC/012/006199

Plan för individuella deltagardata (IPD)

Planerar du att dela individuella deltagardata (IPD)?

NEJ

Läkemedels- och apparatinformation, studiedokument

Studerar en amerikansk FDA-reglerad läkemedelsprodukt

Nej

Studerar en amerikansk FDA-reglerad produktprodukt

Nej

Denna information hämtades direkt från webbplatsen clinicaltrials.gov utan några ändringar. Om du har några önskemål om att ändra, ta bort eller uppdatera dina studieuppgifter, vänligen kontakta register@clinicaltrials.gov. Så snart en ändring har implementerats på clinicaltrials.gov, kommer denna att uppdateras automatiskt även på vår webbplats .