Continuous Glucose Sensing at the Site of Subcutaneous Insulin Administration
Continuous Glucose Sensing at the Site of Subcutaneous Insulin Administration: Evaluation of a Novel Single-Port Treatment Approach in Type 1 Diabetic Patients During a 1-Day Stay in a Clinical Research Center and a 6-Day Period at Home
Przegląd badań
Status
Status
Warunki
Warunki
Interwencja / Leczenie
Interwencja / Leczenie
Szczegółowy opis
The treatment of type 1 diabetes usually comprises the measurement of glucose in blood obtained by finger pricking and the administration of insulin via subcutaneous bolus injection or continuous subcutaneous infusion.
This study seeks to test a new treatment approach where glucose sensing and insulin administration are combined and performed at a single subcutaneous tissue site (single-port treatment approach). A single-port device consisting of a continuous glucose sensor and an insulin infusion cannula is inserted into the subcutaneous tissue of 10 type-1 diabetes patients and used for insulin infusion and simultaneous glucose sensing over a 7-day period. The performance of the single-port device is assessed by comparing the obtained glucose readings with those of a blood glucose meter and an additionally worn control sensor.
Typ studiów
Typ studiów
Zapisy (Rzeczywisty)
Zapisy
Faza
Faza
- Nie dotyczy
Kontakty i lokalizacje
Lokalizacje studiów
-
-
Styria
-
Graz, Styria, Austria, A-8036
- Medical University of Graz
-
-
Kryteria uczestnictwa
Kryteria kwalifikacji
Kryteria kwalifikacji
Wiek uprawniający do nauki
Akceptuje zdrowych ochotników
Płeć kwalifikująca się do nauki
Opis
Inclusion Criteria:
- 18 to 65 years of age, both inclusive
- Type 1 diabetes treated with Continuous Subcutaneous Insulin Infusion (CSII)
- HbA1C < 10%
- Signed informed consent before any study-related activities
Exclusion Criteria:
- Severe acute diseases
- Clinically overt diabetic complications
- Mental incapacity, unwillingness or language barriers precluding adequate understanding or co-operation
- Taking of any vasoactive substances or anticoagulation medication
- Diseases of the skin which could interfere with application of the catheters and Sensors as judged by the investigator
- Pregnancy, breastfeeding, intention of becoming pregnant or not using adequate contraception.
- Increased tendency towards development of hypoglycaemia
- Any disease or condition which the investigator or treating physician feels would interfere with the trial or the safety of the subject
- Concurrent participation in another study
Plan studiów
Jak projektuje się badanie?
Szczegóły projektu
- Główny cel: Podstawowa nauka
- Przydział: Nie dotyczy
- Model interwencyjny: Zadanie dla jednej grupy
- Maskowanie: Brak (otwarta etykieta)
Liczba ramion
Broń i interwencje
Grupa uczestników / ArmGrupa uczestników / Arm |
Interwencja / LeczenieInterwencja / Leczenie |
|---|---|
|
Eksperymentalny: glucose sensing and insulin delivery
Subcutaneous insulin delivery with an insulin pump plus glucose sensing with a continuous glucose sensor placed at the site of subcutaneous insulin delivery. In parallel, capillary glucose determinations using a conventional glucose meter as well as glucose sensing in subcutaneous, insulin-unexposed tissue using a continuous glucose sensor. |
Subcutaneous insulin (100 U/ml, NovoRapid®; Novo Nordisk, Bagsvaerd, Denmark) delivery with an insulin pump (Vibe-TM; Animas Corp., West Chester, PA, USA) and an insulin infusion cannula (SOF-SET MICRO QR 6mm; Medtronic MiniMed, USA) placed at the same site as the continuous glucose sensor.
Glucose sensing with a continuous glucose sensor (Dexcom G4 Platinum Sensor; Dexcom Inc., San Diego, USA) placed at the site of subcutaneous insulin delivery.
|
Co mierzy badanie?
Podstawowe miary wyniku
Podstawowe miary wyniku
Miara wyniku |
Opis środka |
Ramy czasowe |
|---|---|---|
|
Mean absolute relative difference between sensor readings and capillary glucose readings
Ramy czasowe: 7 days
|
Mean absolute relative differences calculated by determining the mean of the absolute value of the percentage difference between paired sensor and capillary glucose concentration
|
7 days
|
Współpracownicy i badacze
Sponsor
Sponsor
Współpracownicy
Współpracownicy
Śledczy
Śledczy
- Główny śledczy: Thomeas R Pieber, MD, Medical University of Graz, Internal Medicine, Endocrinology and Metabolism
Daty zapisu na studia
Główne daty studiów
Rozpoczęcie studiów
Rozpoczęcie studiów
Zakończenie podstawowe (Rzeczywisty)
Zakończenie podstawowe
Ukończenie studiów (Rzeczywisty)
Ukończenie studiów
Daty rejestracji na studia
Pierwszy przesłany
Pierwszy przesłany
Pierwszy przesłany, który spełnia kryteria kontroli jakości
Pierwszy przesłany, który spełnia kryteria kontroli jakości
Pierwszy wysłany (Oszacować)
Pierwszy wysłany
Aktualizacje rekordów badań
Ostatnia wysłana aktualizacja (Oszacować)
Ostatnia wysłana aktualizacja
Ostatnia przesłana aktualizacja, która spełniała kryteria kontroli jakości
Ostatnia przesłana aktualizacja, która spełniała kryteria kontroli jakości
Ostatnia weryfikacja
Ostatnia weryfikacja
Więcej informacji
Terminy związane z tym badaniem
Słowa kluczowe
Dodatkowe istotne warunki MeSH
Inne numery identyfikacyjne badania
Inne numery identyfikacyjne badania
- Single-Port@Home-01
Te informacje zostały pobrane bezpośrednio ze strony internetowej clinicaltrials.gov bez żadnych zmian. Jeśli chcesz zmienić, usunąć lub zaktualizować dane swojego badania, skontaktuj się z register@clinicaltrials.gov. Gdy tylko zmiana zostanie wprowadzona na stronie clinicaltrials.gov, zostanie ona automatycznie zaktualizowana również na naszej stronie internetowej .