Esta página foi traduzida automaticamente e a precisão da tradução não é garantida. Por favor, consulte o versão em inglês para um texto fonte.

Blood'up Device Versus Manual Palpation for Arterial Puncture in Healthy Volunteers: A Pilot Randomized Study Under Normal and Simulated Hypotension

10 de agosto de 2026 atualizado por: Centre Hospitalier Régional d'Orléans

Arterial puncture is a common procedure in the management of hospitalized patients. It is most often performed at the level of the radial artery and allows arterial blood gas analysis.

The success rate of arterial puncture varies widely in the literature. This procedure is moderately painful and may cause significant discomfort, particularly in cases of repeated attempts. First-attempt success is therefore important to reduce both pain and the risk of local complications associated with multiple punctures.

In routine practice, arterial puncture is performed using manual palpation of the radial artery, typically by a trained nurse. In case of failure, a physician may attempt the procedure, often using ultrasound guidance. However, ultrasound is not routinely used and requires specific training and organizational protocols for nurse use.

In hypotensive conditions, such as shock, the radial pulse may become difficult to palpate, making arterial puncture more challenging when using conventional techniques.

The Blood'up device (Arterya, Hérouville, France) is a non-invasive device currently under development that assists in arterial localization. It is applied to the wrist and projects a light signal indicating the position of the artery. This localization is based on ultrasound technology using piezoelectric sensors integrated into the device.

The aim of this study is to evaluate the effectiveness of the Blood'up device for radial artery localization in healthy volunteers, both under physiological conditions (normal blood pressure) and under simulated hypotensive conditions.

Visão geral do estudo

Status

Ainda não está recrutando

Condições

Intervenção / Tratamento

Tipo de estudo

Intervencional

Inscrição (Estimado)

32

Estágio

  • Não aplicável

Contactos e Locais

Esta seção fornece os detalhes de contato para aqueles que conduzem o estudo e informações sobre onde este estudo está sendo realizado.

Contato de estudo

Locais de estudo

      • Orléans, França, 45067
        • Chu Orleans
        • Contato:
          • Mai-Anh NAY, MD
        • Investigador principal:
          • May Anh NAY, MD

Critérios de participação

Os pesquisadores procuram pessoas que se encaixem em uma determinada descrição, chamada de critérios de elegibilidade. Alguns exemplos desses critérios são a condição geral de saúde de uma pessoa ou tratamentos anteriores.

Critérios de elegibilidade

Idades elegíveis para estudo

  • Adulto
  • Adulto mais velho

Aceita Voluntários Saudáveis

Sim

Descrição

Inclusion Criteria:

  • Participant who has agreed to take part in the study
  • No history of cardiovascular disease and a positive Allen test (the Allen test confirms the presence of collateral circulation between the radial and ulnar arteries)
  • Age ≥ 18 years
  • Body mass index (BMI) < 30 kg/m²

Exclusion Criteria:

  • Raynaud's syndrome/disease
  • Damaged or injured skin at the puncture site or at the Blood'Up device contact area
  • Upper limb amputation
  • Contraindication to arterial puncture in one or both upper limbs (e.g., skin lesion, lymph node dissection, etc.)
  • Contraindication to extrinsic arterial compression
  • Therapeutic anticoagulation
  • Known allergy to thermoplastic (ABS)
  • Known allergy to EMLA patch (lidocaine or prilocaine)
  • Pregnant or breastfeeding women

Plano de estudo

Esta seção fornece detalhes do plano de estudo, incluindo como o estudo é projetado e o que o estudo está medindo.

Como o estudo é projetado?

Detalhes do projeto

  • Finalidade Principal: Tratamento
  • Alocação: Randomizado
  • Modelo Intervencional: Atribuição Paralela
  • Mascaramento: Nenhum (rótulo aberto)

Armas e Intervenções

Grupo de Participantes / Braço
Intervenção / Tratamento
Comparador Ativo: Arterial puncture using manual palpation

The traditional method for performing radial arterial puncture by manual palpation is as follows:

  • Identification of the radial artery pulsation by palpation by the nurse
  • Puncture performed at the location where the pulse amplitude is perceived as strongest.
Each healthy volunteer will undergo arterial blood gas sampling on both wrists using each randomized technique (manual palpation versus the Blood'Up device).

Only for the second cohort with simulated hypotension, which will include 16 healthy volunteers:

Hypotension will be simulated by extrinsic compression using a blood pressure cuff adapted to the volunteer's arm circumference. The (manual) blood pressure cuff will be inflated and maintained to achieve a mean arterial pressure of 60 mmHg (with hypotension defined as a mean arterial pressure below 65 mmHg). Blood pressure will be measured using a non-invasive hemodynamic monitoring device (ClearSight®, Edwards Lifesciences, Mississauga, Canada).

Experimental: Arterial puncture using Blood'Up device
Blood'Up is a portable, reusable, non-sterile Class I medical device that uses ultrasound to detect the radial artery beneath the skin. It projects two puncture points connected by a line onto the skin surface directly above the artery. Unlike ultrasound-guided or palpation-based techniques, the device allows the healthcare professional to keep both hands free, enabling the puncture to be performed under optimal conditions.
Each healthy volunteer will undergo arterial blood gas sampling on both wrists using each randomized technique (manual palpation versus the Blood'Up device).

Only for the second cohort with simulated hypotension, which will include 16 healthy volunteers:

Hypotension will be simulated by extrinsic compression using a blood pressure cuff adapted to the volunteer's arm circumference. The (manual) blood pressure cuff will be inflated and maintained to achieve a mean arterial pressure of 60 mmHg (with hypotension defined as a mean arterial pressure below 65 mmHg). Blood pressure will be measured using a non-invasive hemodynamic monitoring device (ClearSight®, Edwards Lifesciences, Mississauga, Canada).

Blood'Up will be used to detect the radial artery beneath the skin.

O que o estudo está medindo?

Medidas de resultados primários

Medida de resultado
Descrição da medida
Prazo
Success rate of radial arterial puncture performed with and without the Blood'Up device in healthy volunteers with no cardiovascular history
Prazo: 5 minutes

Success is defined as successful puncture of the radial artery with confirmation of the arterial nature of the sample by blood gas analysis. Only one puncture (defined as a single skin puncture) will be performed on each wrist.

If the needle is withdrawn without obtaining blood return, this will be considered a failed arterial puncture. Similarly, the absence of blood return within one minute after skin puncture will also be considered a failure.

5 minutes

Medidas de resultados secundários

Medida de resultado
Descrição da medida
Prazo
Procedure time: from device placement to blood return in the syringe when using the Blood'Up device; and from the start of palpation to blood return in the syringe for manual palpation.
Prazo: 30 minutes

Procedure time is defined as the time from the start of device placement to the onset of blood return in the arterial syringe (expected volume: 1 mL).

For the Blood'Up device, the start is defined as the moment the device is placed on the volunteer's skin.

For manual palpation, the start of palpation defines the start of the procedure.

In both cases (Blood'Up and manual palpation), the maximum allowed procedure time is 1 minute. Similarly, the absence of blood return within one minute after skin puncture will be considered a failure.

30 minutes
Pain with and without the Blood'Up device
Prazo: 30 minutes
Pain will be assessed immediately after each puncture by asking the participant to rate the maximum pain experienced during the procedure using a visual analog scale (VAS) ranging from 0 (no pain) to 10 (worst imaginable pain).
30 minutes
Number of needle redirections with and without the Blood'Up device.
Prazo: 30 minutes
Number of needle redirections during each puncture, defined as any change in needle angulation (self-reported by the nurse performing the procedure). A puncture is defined as a single skin breach by the needle; as long as the needle is not completely withdrawn from the skin, all manipulations are considered part of the same puncture, regardless of the number of redirections.
30 minutes
Immediate complications with and without the Blood'Up device.
Prazo: 30 minutes
Complications related to arterial puncture occurring within 5 minutes after the procedure: hematoma at the puncture site, suspected nerve injury (paresthesia), or vasovagal response (pallor, sweating, bradycardia, syncope).
30 minutes

Colaboradores e Investigadores

É aqui que você encontrará pessoas e organizações envolvidas com este estudo.

Patrocinador

Publicações e links úteis

A pessoa responsável por inserir informações sobre o estudo fornece voluntariamente essas publicações. Estes podem ser sobre qualquer coisa relacionada ao estudo.

Datas de registro do estudo

Essas datas acompanham o progresso do registro do estudo e os envios de resumo dos resultados para ClinicalTrials.gov. Os registros do estudo e os resultados relatados são revisados ​​pela National Library of Medicine (NLM) para garantir que atendam aos padrões específicos de controle de qualidade antes de serem publicados no site público.

Datas Principais do Estudo

Início do estudo (Estimado)

1 de setembro de 2026

Conclusão Primária (Estimado)

1 de setembro de 2027

Conclusão do estudo (Estimado)

1 de setembro de 2027

Datas de inscrição no estudo

Enviado pela primeira vez

26 de junho de 2026

Enviado pela primeira vez que atendeu aos critérios de CQ

10 de agosto de 2026

Primeira postagem (Real)

14 de agosto de 2026

Atualizações de registro de estudo

Última Atualização Postada (Real)

14 de agosto de 2026

Última atualização enviada que atendeu aos critérios de controle de qualidade

10 de agosto de 2026

Última verificação

1 de agosto de 2026

Mais Informações

Termos relacionados a este estudo

Outros números de identificação do estudo

  • CHUO-2025-04

Plano para dados de participantes individuais (IPD)

Planeja compartilhar dados de participantes individuais (IPD)?

NÃO

Informações sobre medicamentos e dispositivos, documentos de estudo

Estuda um medicamento regulamentado pela FDA dos EUA

Não

Estuda um produto de dispositivo regulamentado pela FDA dos EUA

Não

Essas informações foram obtidas diretamente do site clinicaltrials.gov sem nenhuma alteração. Se você tiver alguma solicitação para alterar, remover ou atualizar os detalhes do seu estudo, entre em contato com register@clinicaltrials.gov. Assim que uma alteração for implementada em clinicaltrials.gov, ela também será atualizada automaticamente em nosso site .