Performance of an Investigational Blood Glucose Monitoring System
Performance of the Ninja 3 Plus Blood Glucose Monitoring System
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Study Type
Study Type
Enrollment (Actual)
Enrollment
Phase
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
New Jersey
-
Fairfield, New Jersey, United States, 07004
- Consumer Product Testing Co.
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Description
Inclusion Criteria:
- Males and females, 18 years of age or older.
- People with type 1 or type 2 diabetes.
- Able to speak, read and understand English.
- Willing to complete all study procedures.
Exclusion Criteria:
- Hemophilia or any other bleeding disorder.
- Pregnancy.
- Physical, visual or neurological impairments that would make the person unable to perform testing with the BGM.
- Previously participated in a blood glucose monitoring study using the Ninja 3 Plus BGMS.
- Working for a medical laboratory, hospital or other clinical setting that involves training on or clinical use of blood glucose meters.
- Working for a competitive medical device company, or having an immediate family member who works for such a company.
- A condition, which in the opinion of the investigator or designee, would put the person or study conduct at risk.
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Diagnostic
- Allocation: N/A
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Experimental: Intended Users of the Monitoring System
Untrained subjects with diabetes used the Ninja 3 PLUS Investigational BG Monitoring System. General enrollment criteria for the 'Intended Users' population:
|
Untrained subjects with diabetes performed self Blood Glucose (BG) tests with capillary fingerstick and palm blood using the Ninja 3 PLUS Investigational BG Monitoring System.
Study staff tested subject fingerstick blood.
All BG results were compared to a reference laboratory glucose method.
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Percent of Self Test Fingerstick Blood Glucose Results Within ±15 mg/dL (< 75 mg/dL) and Within ±20% (≥ 75 mg/dL) of the Laboratory Glucose Method
Time Frame: 1 hour
|
Untrained subjects with diabetes who self-tested fingerstick blood used the Ninja 3 PLUS investigational Blood Glucose Monitoring System (BGMS).
BGMS results were compared with capillary plasma BG results obtained with a reference laboratory glucose method - YSI Life Sciences (YSI) Analyzer.
BG meter results were used to calculate the number of BGMS results within +/- 15mg/dL (for reference BG results <75mg/dL) and within +/- 20% (for reference BG results >=75mg/dL) of the YSI reference method results.
|
1 hour
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Percent of Self Test Fingerstick Blood Glucose Results Within ±15 mg/dL (< 100 mg/dL) and Within ±15% (≥ 100 mg/dL) of the Laboratory Glucose Method
Time Frame: 1 hour
|
Untrained subjects with diabetes who self-tested fingerstick blood used the Ninja 3 PLUS investigational Blood Glucose Monitoring System (BGMS).
BGMS results were compared with capillary plasma BG results obtained with a reference laboratory glucose method - YSI Life Sciences (YSI) Analyzer.
BG meter results were used to calculate the number of BGMS results within +/- 15mg/dL (for reference BG results <100mg/dL) and within +/- 15% (for reference BG results >=100mg/dL) of the YSI reference method results.
|
1 hour
|
|
Percent of Self Test Fingerstick Blood Glucose Results Within ±12.5 mg/dL (< 100 mg/dL) and Within ±12.5% (≥ 100 mg/dL) of the Laboratory Glucose Method
Time Frame: 1 hour
|
Untrained subjects with diabetes who self-tested fingerstick blood used the Ninja 3 PLUS investigational Blood Glucose Monitoring System (BGMS).
BGMS results were compared with capillary plasma BG results obtained with a reference laboratory glucose method - YSI Life Sciences (YSI) Analyzer.
BG meter results were used to calculate the number of BGMS results within +/- 12.5mg/dL (for reference BG results <100mg/dL) and within +/- 12.5% (for reference BG results >=100mg/dL) of the YSI reference method results.
|
1 hour
|
|
Percent of Blood Glucose Results From Alternative Site Testing (AST) of the Palm Within ±15 mg/dL (< 75 mg/dL) and Within ±20% (≥ 75 mg/dL) of the Laboratory Method.
Time Frame: 1 hour
|
Untrained subjects with diabetes self-tested Alternative Site (AST) Palm blood using the Ninja 3 PLUS investigational Blood Glucose Monitoring System (BGMS).
BGMS AST results were compared with capillary plasma BG results obtained with a reference laboratory glucose method - YSI Life Sciences (YSI) Analyzer.
BG meter results were used to calculate the number of AST BGMS results within +/- 15mg/dL (<75mg/dL YSI capillary plasma) and within +/- 15% (>=75mg/dL YSI capillary plasma).
|
1 hour
|
|
Percent of Fingerstick Blood Glucose Results Within ±15 mg/dL (< 75 mg/dL) and Within ±20% (≥ 75 mg/dL) of the Laboratory Glucose Method When Tested by Study Staff
Time Frame: one hour
|
Study Staff tested subject fingerstick blood using the Ninja 3 PLUS investigational Blood Glucose Monitoring System (BGMS).
BGMS results were compared with capillary plasma BG results obtained with a reference laboratory glucose method - YSI Life Sciences (YSI) Analyzer.
BG meter results were used to calculate the number of BGMS results within +/- 15mg/dL (for reference BG results <75mg/dL) and within +/- 20% (for reference BG results >=75mg/dL) of the YSI reference method results.
|
one hour
|
|
Number of Subject Responses That 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements
Time Frame: 1 hour
|
Staff obtained subject responses using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS.
Subjects could respond 'Strongly Agree' 'Agree' 'Neutral' 'Disagree' or 'Strongly Disagree.'
|
1 hour
|
Collaborators and Investigators
Sponsor
Sponsor
Investigators
Investigators
- Principal Investigator: Michael Caswell, PhD, Consumer Product Testing Co.
Publications and helpful links
Study record dates
Study Major Dates
Study Start
Study Start
Primary Completion (Actual)
Primary Completion
Study Completion (Actual)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Estimate)
First Posted
Study Record Updates
Last Update Posted (Estimate)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Other Study ID Numbers
Other Study ID Numbers
- CTD-2012-010-01
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.