Study of Clinical Performance of Corneal Topography Measurement Using an Optical Biometer

May 13, 2014 updated by: Haag-Streit AG

Lenstar LS 900 Topography

The purpose of this study is to quantify in-vivo repeatability of a new corneal topography measurement mode of an optical biometer, and the clinical performance of measurement in comparison to reference devices.

Study Overview

Status

Completed

Conditions

Study Type

Observational

Enrollment (Actual)

50

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

      • Berne, Switzerland, 3010
        • University Hospital Inselspital

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years and older (Adult, Older Adult)

Accepts Healthy Volunteers

Yes

Genders Eligible for Study

All

Sampling Method

Non-Probability Sample

Study Population

Volunteers recruited at University Hospital

Description

Inclusion Criteria:

  • 18 years or older

Exclusion Criteria:

  • No corneal transplants
  • No corneal scar(s)
  • No acute infections of the eye
  • No weak fixation
  • No suspected lack of compliance
  • No vulnerable population

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

Cohorts and Interventions

Group / Cohort
Normal eye
Astigmatism smaller than 1.5 diopters
Large regular astigmatism
Astigmatism of > 1.5 diopters, regular astigmatism.
Large irregular astigmatism
Astigmatism of > 1.5 diopters, irregular astigmatism.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
In-vivo Repeatability of Corneal Topography Measurements With Lenstar LS 900 Topography: Sample Mean of Differences of Mean Power and Sample Mean of Std. Dev of Local Power Differences Between Two Consecutive Measurements
Time Frame: 1 day of examination

Corneal topography is a measurement of the shape of the anterior cornea. The shape of a cornea can be fully quantified by providing a map of local power. "Diopter" is the unit of refractive power of a lens. In case of the cornea, the power K [diopter] is related to the radius (curvature) R [mm] of the best fitting sphere by the relation K=337.5/R.

Here, corneal topography measurements are implemented by the Placido method, i.e. by analyzing the reflection image of a ring-shaped illumination.

According to International Standards Organization (ISO) 19980-2012, repeatability of corneal topography is assessed on the central cornea: area with diameter d<=3mm, and middle cornea: 3mm<d<=6mm.

"power difference (2 rep. meas.)": sample mean and 95% C.I. of differences of spatial mean of corneal power between two consecutive measurements.

"power difference 2 std.dev.": sample mean of twice the deviation of local power differences (two consecutive measurements).

1 day of examination

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Equivalence of Corneal Topography Measurement Between Lenstar LS 900 Topography and Atlas 9000: Sample Mean of Differences of Mean Power and Sample Mean of Std. Dev of Local Power Differences for One Measurement Per Device.
Time Frame: 1 day of examination

For definition of corneal topography, corneal power in "diopter", areas of evaluation and methods, see outcome 1.

This outcome measure aims at testing the equivalence of corneal topography measurement between Lenstar LS 900 Topography (Haag Streit) and Atlas 9000 (Zeiss) by analyzing differences in measurement results for the same eye between both devices.

"power difference (2 devices)" is the sample mean and 95% C.I. of differences of spatial mean of corneal power between two devices. This value quantifies systematic differences between devices (e.g. calibration), ignoring local variations of the power measurement.

"power difference (2 std.dev.)" is the sample mean of twice the standard deviation of local power differences between measurements with both devices. This value quantifies the spatially resolved agreement of corneal shape measurement of the two devices.

1 day of examination

Other Outcome Measures

Outcome Measure
Measure Description
Time Frame
Equivalence of Corneal Topography Measurement Between Lenstar LS 900 Topography and Atlas 9000: Sample Mean of Std. Dev. of Local Corneal Elevation Differences for One Measurement Per Device.
Time Frame: 1 day of examination

For definition of corneal topography, areas of evaluation and methods, see outcome 1 and 2.

Corneal elevation refers to the distance between the measured corneal surface and the best fitting sphere, and is given in µm.

This outcome measure aims at testing the equivalence of corneal topography measurement between Lenstar LS 900 Topography (Haag Streit) and Atlas 9000 (Zeiss) by analyzing differences in measurement results for the same eye between both devices.

"elevation difference (2 std. dev.)" is the sample mean of twice the standard deviation of local corneal elevation differences between measurements with both devices. This value quantifies the spatially resolved agreement of corneal shape measurement of the two devices.

1 day of examination
Equivalence of Keratometry Radius Measurement Between Lenstar LS 900 Topography and Lenstar LS 900: Population Mean of Differences of Keratometry Radius Measurement Between Both Devices.
Time Frame: 1 day of examination

Keratometry radius refers to the corneal curvature R (see primary measure outcome). In this context, the cornea is approximated by a toric surface which can be characterized by a flat meridian (radius R1) and a steep meridian (radius R2) with an angle of 90 degrees between these meridians.

This outcome measure aims at testing the equivalence of keratometry radius measurement between Lenstar LS 900 Topography and Lenstar LS 900 (both Haag Streit) by analyzing differences in measurement results for the same eye between both devices.

Reported are: population mean of difference and 95% confidence interval of mean difference for radius of flat meridian (R1) and radius of steep meridian (R2).

1 day of examination
Equivalence of Keratometry Axis Measurement Between Lenstar LS 900 Topography and Lenstar LS 900: Population Mean of Differences of Keratometry Axis Measurement Between Both Devices.
Time Frame: 1 day of examination

Keratometry axis refers to the axis of the flat meridian of the toric representation of the cornea. For additional information see outcome 4.

This outcome measure aims at testing the equivalence of keratometry axis measurement between Lenstar LS 900 Topography and Lenstar LS 900 (both Haag Streit) by analyzing differences in measurement results for the same eye between both devices.

Reported are: population mean of difference and 95% confidence interval of mean difference for axis of flat meridian.

1 day of examination

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Sponsor

Collaborators

Investigators

  • Principal Investigator: Beatrice Frueh, Prof. Fr., University Hospital Inselspital, Berne

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start

June 1, 2013

Primary Completion (Actual)

July 1, 2013

Study Completion (Actual)

July 1, 2013

Study Registration Dates

First Submitted

May 21, 2013

First Submitted That Met QC Criteria

May 23, 2013

First Posted (Estimate)

May 24, 2013

Study Record Updates

Last Update Posted (Estimate)

June 5, 2014

Last Update Submitted That Met QC Criteria

May 13, 2014

Last Verified

May 1, 2014

More Information

Terms related to this study

Other Study ID Numbers

  • HS-LS900-TOPO-01

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Corneal Topography

Search Similar Trials